Dietrich Beitzke1, Richard Nolz1, Sylvia Unterhumer1, Christina Plank1, Michael Weber1, Rüdiger Schernthaner1, Veronika Schöpf2, Florian Wolf1, Christian Loewe1.
Abstract
OBJECTIVE: To prospectively evaluate image quality and radiation dose using a low-dose computed tomography angiography protocol and iterative image reconstruction for high-pitch dual-source CT-angiography (DSCTA) of the supraaortic arteries.Entities:
Mesh:
Year: 2014 PMID: 24919195 PMCID: PMC4053523 DOI: 10.1371/journal.pone.0099832
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
CT study protocols and patient characteristics. Scan Range/Patients data presented as means±standard deviations (range).
| Standard | Low-dose | p-value | |
|
| |||
| CT scanner | Somatom Definition Flash | ||
| Slice collimation | 2×64×0.625 mm | ||
| Kilovoltage (kV) | 120 | 100 | |
| Tube current (mAs) ref. | 120 | 100 | |
| Tube current modulation | CareDose4D | ||
| Pitch | 2.4 | 3.2 | |
| Contrast/flow | 70 ml Optiray 350 at 5 ml/sec | ||
| Image reconstruction | FBP | SAFIRE (Level 3) | |
| Scan Range (cm) | 42±3.5 (35.8–48.5) | 42.1±2.9 (37.4–48.6) | 0.83 (n.s.) |
|
| |||
| Sex (female/male) | 9/12 | 7/14 | |
| Age (years) | 66.5±12 (43–86) | 64.1±11.8 (39–90) | 0.69 (n.s.) |
| Body Mass Index | 27.3±4.1 (20,7–36,2) | 27.1±2.8(22.1–33.5) | 0.49 (n.s.) |
Figure 1Comparison of image quality in two patients with suspected stenosis of the internal carotid artery.
Curved planar reconstructions of (A) the standard protocol and (B) the low-dose protocol demonstrate similar diagnostic image quality in both protocols with a more smoothed appearance and reduced image noise at the level of the common carotid arteries using SAFIRE. Significantly lower radiation dose at B.
Figure 2Comparison of image quality at the thoracic outlet.
Axial images with similar window setting (W900/L250) using the standard protocol and FBP (A) and at low-dose, using SAFIRE for image reconstruction (B). Both studies were rated having excellent diagnostic image quality. Higher vessel attenuation and reduced image noise is evident at low-dose and SAFIRE (B).
Group comparison of objective image quality (mean vessel attenuation, muscle attenuation, background noise, CNR, SNR).
| Standard Protocol mean±SD | Low-dose protocol mean±SD | p-value | |||
|
| |||||
| Aorta | 362.3±91.48 | 420.2±75.97 | |||
| Right | Left | Right | Left | ||
| Common carotid artery | 358.1±93.3 | 370.4±115.1 | 419.1±79.5 | 423.5±81.1 | |
| Bifurcation | 405.3±93.9 | 417.3±114.2 | 477.8±91.7 | 480.7±95.7 | |
| Internal carotid artery | 354.2±87.1 | 370.4±104.9 | 433.8±78.4 | 436.3±81.2 | |
| Carotid sinus | 362,6±81,5 | 378,3±103,6 | 475,4±87,6 | 462,3±84,8 | |
| Medial cerebral artery | 345,5±75,3 | 349,0±93,8 | 428,9±81,0 | 444,9±89,3 | |
|
|
| ||||
|
| |||||
| Common carotid artery | 6.32±1.9 | 6.08±1.7 | |||
| Bifurcation | 5.53±1.5 | 5.34±0.7 | |||
| Internal carotid artery | 6.13±1.4 | 6.12±1.2 | |||
|
|
| ||||
|
| |||||
| Aorta | 46.7±12.71 | 50±13.33 | |||
| Right | Left | Right | Left | ||
| Common carotid artery | 58.3±14.7 | 56.3±17.3 | 70.8±17.8 | 67.7±21.9 | |
| Bifurcation | 58.4±10.1 | 59.6±11.7 | 64.8±7.9 | 61±9.1 | |
| Internal carotid artery | 49.5±13.9 | 50.2±10.7 | 54.3±16.2 | 53.3±11.8 | |
|
|
| ||||
|
| |||||
| Aorta | 55.78±28.40 | 65.81±23.10 | |||
| Right | Left | Right | Left | ||
| Common carotid artery | 53.4±29.6 | 55.5±36.7 | 62.2±23.1 | 63.6±24.3 | |
| Bifurcation | 68.1±27.9 | 70.6±32.7 | 77.7±19.9 | 80.5±23.8 | |
| Internal carotid artery | 52.6±21.5 | 58.6±31.9 | 64.5±19.8 | 63.6±18.2 | |
|
|
| ||||
|
| |||||
| Aorta | 58.41±27.9 | 73.11±26.9 | |||
| Right | Left | Right | Left | ||
| Common carotid artery | 63.4±32.1 | 66.1±38.1 | 74.5±24.9 | 75.3±26 | |
| Bifurcation | 79.3±30.3 | 82.2±35.2 | 89.8±20.9 | 91.9±23.9 | |
| Internal carotid artery | 61.1±22.7 | 67.5±33.6 | 73.4±20 | 72.9±20.3 | |
|
|
| ||||
|
| |||||
| Sagital Sinus | 238.5±74.76 | 256.1±60.2 | |||
| Δ to M1 (mean) | 108.7 | 224 | |||
| Right | Left | Right | Left | ||
| Transverse Sinus | 194.0±44.15 | 192.2±47.5 | 230.8±60.5 | 221±44.1 | |
| Jugular Vein | 126.5±129.7 | 106.4±34.4 | 108.2±23.9 | 102.2±22 | |
| Δto Bifurcation (mean) | 278.8 | 310.9 | 369.6 | 378.5 | |
|
|
| ||||
Mixed model ANOVA was used for group comparison of objective image quality. Data are means±standard deviations.
HU = Houndsfield units, CNR = contrast-to-noise ratio, SNR = signal-to-noise ratio. BN = Standard deviation HU.
*BN at the level of the aorta = BN ACC.
= Significance level <0.05.
Image reading parameters including interrater agreement measures for both protocols.
| Image Quality | Standard Protocol | Low-dose Protocol | ||
| 0 | 0.0% | n = 0 | 0.0% | n = 0 |
| 1 | 0.3% | n = 2 | 0.3% | n = 2 |
| 2 | 10.6% | n = 76 | 9.1% | n = 65 |
| 3 | 86.4% | n = 617 | 86.1% | n = 615 |
| Occluded Segments | 2.7% | n = 23 | 4.5% | n = 36 |
|
|
| |||
|
|
| |||
|
|
| |||
|
|
|
| ||
| 0 | 77.6% | n = 554 | 75.9% | n = 542 |
| 1 | 19.5% | n = 139 | 19.3% | n = 138 |
| 2 | 0.3% | n = 2 | 0.3% | n = 2 |
| 3 | 0.0% | n = 0 | 0.0% | n = 0 |
| Occluded Segments | 2.6% | n = 23 | 4.5% | n = 36 |
|
|
| |||
|
|
| |||
|
|
| |||
CI indicates confidence interval.
Results from group comparison of radiation dose (unpaired t-test).
| Standard protocol mean±SD | Low-dose protocol mean±SD | p-value | |
|
| |||
| CTDI (mGy) | 5.3±1.1 | 2.4±0.8 | <0.01 |
| DLP (mGy*cm) | 225.4±55.3 | 99.2±15.9 | <0.01 |
| ED (mSv) | 2.8±0.7 | 1.2±0.2 | <0.01 |
Data are means±standard deviations.
CDTI = Computed Tomography Dose Index (mGy); DLP = Dose Length Product (mGy*cm); ED = estimated effective dose (mSv).
Figure 3Box-Whisker plots of the estimated mean effective dose (mSv) for the study protocols.