Dose reduction in abdominal computed tomography: intraindividual comparison of image quality of full-dose standard and half-dose iterative reconstructions with dual-source computed tomography.

Matthias S May1, Wolfgang Wüst, Michael Brand, Christian Stahl, Thomas Allmendinger, Bernhard Schmidt, Michael Uder, Michael M Lell.   

Abstract

OBJECTIVES: We sought to evaluate the image quality of iterative reconstruction in image space (IRIS) in half-dose (HD) datasets compared with full-dose (FD) and HD filtered back projection (FBP) reconstruction in abdominal computed tomography (CT).
MATERIALS AND METHODS: To acquire data with FD and HD simultaneously, contrast-enhanced abdominal CT was performed with a dual-source CT system, both tubes operating at 120 kV, 100 ref.mAs, and pitch 0.8. Three different image datasets were reconstructed from the raw data: Standard FD images applying FBP which served as reference, HD images applying FBP and HD images applying IRIS. For the HD data sets, only data from 1 tube detector-system was used. Quantitative image quality analysis was performed by measuring image noise in tissue and air. Qualitative image quality was evaluated according to the European Guidelines on Quality criteria for CT. Additional assessment of artifacts, lesion conspicuity, and edge sharpness was performed.
RESULTS: : Image noise in soft tissue was substantially decreased in HD-IRIS (-3.4 HU, -22%) and increased in HD-FBP (+6.2 HU, +39%) images when compared with the reference (mean noise, 15.9 HU). No significant differences between the FD-FBP and HD-IRIS images were found for the visually sharp anatomic reproduction, overall diagnostic acceptability (P = 0.923), lesion conspicuity (P = 0.592), and edge sharpness (P = 0.589), while HD-FBP was rated inferior. Streak artifacts and beam hardening was significantly more prominent in HD-FBP while HD-IRIS images exhibited a slightly different noise pattern.
CONCLUSIONS: Direct intrapatient comparison of standard FD body protocols and HD-IRIS reconstruction suggest that the latest iterative reconstruction algorithms allow for approximately 50% dose reduction without deterioration of the high image quality necessary for confident diagnosis.

Entities:  

Mesh:

Year:  2011        PMID: 21467948     DOI: 10.1097/RLI.0b013e31821690a1

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  38 in total

1.  Iterative reconstruction of dual-source coronary CT angiography: assessment of image quality and radiation dose.

Authors:  Eun-Ah Park; Whal Lee; Kwang Woo Kim; Kwang Gi Kim; Allmendinger Thomas; Jin Wook Chung; Jae Hyung Park
Journal:  Int J Cardiovasc Imaging       Date:  2011-12-21       Impact factor: 2.357

2.  Impact of iterative reconstruction on image quality and radiation dose in multidetector CT of large body size adults.

Authors:  Gaurav S Desai; Raul N Uppot; Elaine W Yu; Avinash R Kambadakone; Dushyant V Sahani
Journal:  Eur Radiol       Date:  2012-04-19       Impact factor: 5.315

3.  Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT.

Authors:  Cynthia H McCollough; Guang Hong Chen; Willi Kalender; Shuai Leng; Ehsan Samei; Katsuyuki Taguchi; Ge Wang; Lifeng Yu; Roderic I Pettigrew
Journal:  Radiology       Date:  2012-06-12       Impact factor: 11.105

4.  The adaptive statistical iterative reconstruction-V technique for radiation dose reduction in abdominal CT: comparison with the adaptive statistical iterative reconstruction technique.

Authors:  Heejin Kwon; Jinhan Cho; Jongyeong Oh; Dongwon Kim; Junghyun Cho; Sanghyun Kim; Sangyun Lee; Jihyun Lee
Journal:  Br J Radiol       Date:  2015-08-03       Impact factor: 3.039

5.  Dose reduction with adaptive statistical iterative reconstruction for paediatric CT: phantom study and clinical experience on chest and abdomen CT.

Authors:  F Gay; Y Pavia; N Pierrat; S Lasalle; S Neuenschwander; H J Brisse
Journal:  Eur Radiol       Date:  2013-08-31       Impact factor: 5.315

6.  Combining automated attenuation-based tube voltage selection and iterative reconstruction: a liver phantom study.

Authors:  Daniela B Husarik; Sebastian T Schindera; Fabian Morsbach; Natalie Chuck; Burkhardt Seifert; Zsolt Szucs-Farkas; Hatem Alkadhi
Journal:  Eur Radiol       Date:  2013-10-24       Impact factor: 5.315

Review 7.  Task-based measures of image quality and their relation to radiation dose and patient risk.

Authors:  Harrison H Barrett; Kyle J Myers; Christoph Hoeschen; Matthew A Kupinski; Mark P Little
Journal:  Phys Med Biol       Date:  2015-01-07       Impact factor: 3.609

8.  90-kVp low-tube-voltage CT pulmonary angiography in combination with advanced modeled iterative reconstruction algorithm: effects on radiation dose, image quality and diagnostic accuracy for the detection of pulmonary embolism.

Authors:  Doris Leithner; Tatjana Gruber-Rouh; Martin Beeres; Julian L Wichmann; Scherwin Mahmoudi; Simon S Martin; Lukas Lenga; Moritz H Albrecht; Christian Booz; Thomas J Vogl; Jan-Erik Scholtz
Journal:  Br J Radiol       Date:  2018-06-05       Impact factor: 3.039

9.  Image quality of adaptive iterative dose reduction 3D of coronary CT angiography of 640-slice CT: comparison with filtered back-projection.

Authors:  Roh-Eul Yoo; Eun-Ah Park; Whal Lee; Hackjoon Shim; Yeo Koon Kim; Jin Wook Chung; Jae Hyung Park
Journal:  Int J Cardiovasc Imaging       Date:  2012-08-26       Impact factor: 2.357

10.  Quantitative syndesmophyte measurement in ankylosing spondylitis using CT: longitudinal validity and sensitivity to change over 2 years.

Authors:  Sovira Tan; Jianhua Yao; John A Flynn; Lawrence Yao; Michael M Ward
Journal:  Ann Rheum Dis       Date:  2013-12-02       Impact factor: 19.103

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.