High-pitch coronary CT angiography with third generation dual-source CT: limits of heart rate.

Sonja Gordic1, Daniela B Husarik, Lotus Desbiolles, Sebastian Leschka, Thomas Frauenfelder, Hatem Alkadhi.   

Abstract

To determine the average heart rate (HR) and heart rate variability (HRV) required for diagnostic imaging of the coronary arteries in patients undergoing high-pitch CT-angiography (CTA) with third-generation dual-source CT. Fifty consecutive patients underwent CTA of the thoracic (n = 8) and thoracoabdominal (n = 42) aorta with third-generation dual-source 192-slice CT with prospective electrocardiography (ECG)-gating at a pitch of 3.2. No β-blockers were administered. Motion artifacts of coronary arteries were graded on a 4-point scale. Average HR and HRV were noted. The average HR was 66 ± 11 beats per minute (bpm) (range 45-96 bpm); the HRV was 7.3 ± 4.4 bpm (range 3-20 bpm). Interobserver agreement on grade of image quality for the 642 coronary segments evaluated by both observers was good (κ = 0.71). Diagnostic image quality was found for 608 of the 642 segments (95%) in 43 of 50 patients (86%). In 14% of the patients, image quality was nondiagnostic for at least one segment. HR (p = 0.001) was significantly higher in patients with at least one non-diagnostic segment compared to those without. There was no significant difference (p > 0.05) in HRV between patients with nondiagnostic segments and those with diagnostic images of all segments. All patients with a HR < 70 bpm had diagnostic image quality in all coronary segments. The effective radiation dose and scan time for the heart were 0.4 ± 0.1 mSv and 0.17 ± 0.02 s, respectively. Third-generation dual-source 192-slice CT allows for coronary angiography in the prospectively ECG-gated high-pitch mode with diagnostic image quality at HR up to 70 bpm. HRV is not significantly related to image quality of coronary CTA.

Entities:  

Mesh:

Year:  2014        PMID: 24816910     DOI: 10.1007/s10554-014-0445-5

Source DB:  PubMed          Journal:  Int J Cardiovasc Imaging        ISSN: 1569-5794            Impact factor:   2.357


  21 in total

1.  A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association.

Authors:  W G Austen; J E Edwards; R L Frye; G G Gensini; V L Gott; L S Griffith; D C McGoon; M L Murphy; B B Roe
Journal:  Circulation       Date:  1975-04       Impact factor: 29.690

2.  Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography.

Authors:  Jiang Hsieh; John Londt; Melissa Vass; Jay Li; Xiangyang Tang; Darin Okerlund
Journal:  Med Phys       Date:  2006-11       Impact factor: 4.071

3.  Detection of coronary artery stenoses by low-dose, prospectively ECG-triggered, high-pitch spiral coronary CT angiography.

Authors:  Stephan Achenbach; Tobias Goroll; Martin Seltmann; Tobias Pflederer; Katharina Anders; Dieter Ropers; Werner G Daniel; Michael Uder; Michael Lell; Mohamed Marwan
Journal:  JACC Cardiovasc Imaging       Date:  2011-04

4.  High-pitch electrocardiogram-triggered computed tomography of the chest: initial results.

Authors:  Michael Lell; Fabian Hinkmann; Katharina Anders; Paul Deak; Willi A Kalender; Michael Uder; Stephan Achenbach
Journal:  Invest Radiol       Date:  2009-11       Impact factor: 6.016

5.  High-pitch thoracic CT with simultaneous assessment of coronary arteries: effect of heart rate and heart rate variability on image quality and diagnostic accuracy.

Authors:  Michael Scharf; Ricarda Bink; Matthias S May; Christian Hentschke; Stephan Achenbach; Michael Uder; Michael M Lell
Journal:  JACC Cardiovasc Imaging       Date:  2011-06

6.  Predictors of image quality in high-pitch coronary CT angiography.

Authors:  Paul Stolzmann; Robert P Goetti; Pal Maurovich-Horvat; Udo Hoffmann; Thomas G Flohr; Sebastian Leschka; Hatem Alkadhi
Journal:  AJR Am J Roentgenol       Date:  2011-10       Impact factor: 3.959

7.  High-pitch dual-source CT angiography of the thoracic and abdominal aorta: is simultaneous coronary artery assessment possible?

Authors:  Robert Goetti; Stephan Baumüller; Gudrun Feuchtner; Paul Stolzmann; Christoph Karlo; Hatem Alkadhi; Sebastian Leschka
Journal:  AJR Am J Roentgenol       Date:  2010-04       Impact factor: 3.959

8.  Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experience.

Authors:  Sebastian Leschka; Paul Stolzmann; Lotus Desbiolles; Stephan Baumueller; Robert Goetti; Thomas Schertler; Hans Scheffel; Andre Plass; Volkmar Falk; Gudrun Feuchtner; Borut Marincek; Hatem Alkadhi
Journal:  Eur Radiol       Date:  2009-09-16       Impact factor: 5.315

9.  Submillisievert median radiation dose for coronary angiography with a second-generation 320-detector row CT scanner in 107 consecutive patients.

Authors:  Marcus Y Chen; Sujata M Shanbhag; Andrew E Arai
Journal:  Radiology       Date:  2013-01-22       Impact factor: 11.105

10.  Optimizing radiation dose levels in prospectively electrocardiogram-triggered coronary computed tomography angiography using iterative reconstruction techniques: a phantom and patient study.

Authors:  Yang Hou; Jiahe Zheng; Yuke Wang; Mei Yu; Mani Vembar; Qiyong Guo
Journal:  PLoS One       Date:  2013-02-20       Impact factor: 3.240

View more
  16 in total

1.  Application of intelligent optimal kV scanning technology (CARE kV) in dual-source computed tomography (DSCT) coronary angiography.

Authors:  Jun Zhang; Shaolei Kang; Dan Han; Xiaojie Xie; Yaming Deng
Journal:  Int J Clin Exp Med       Date:  2015-10-15

2.  Improved visual delineation of the intimal flap in Stanford type A and B dissections at 3rd generation dual-source high-pitch CT angiography.

Authors:  Martin Beeres; Andreas M Bucher; Julian L Wichmann; Claudia Frellesen; Jan E Scholtz; Moritz Albrecht; Boris Bodelle; Nour-Eldin A Nour-Eldin; Clara Lee; Moritz Kaup; Thomas J Vogl; Tatjana Gruber-Rouh
Journal:  Radiol Med       Date:  2016-04-21       Impact factor: 3.469

3.  Optimizing radiation dose by using advanced modelled iterative reconstruction in high-pitch coronary CT angiography.

Authors:  Sonja Gordic; Lotus Desbiolles; Martin Sedlmair; Robert Manka; André Plass; Bernhard Schmidt; Daniela B Husarik; Francesco Maisano; Simon Wildermuth; Hatem Alkadhi; Sebastian Leschka
Journal:  Eur Radiol       Date:  2015-06-03       Impact factor: 5.315

Review 4.  Cardiovascular imaging 2014 in the International Journal of Cardiovascular Imaging.

Authors: 
Journal:  Int J Cardiovasc Imaging       Date:  2015-03       Impact factor: 2.357

Review 5.  New Applications of Cardiac Computed Tomography: Dual-Energy, Spectral, and Molecular CT Imaging.

Authors:  Ibrahim Danad; Zahi A Fayad; Martin J Willemink; James K Min
Journal:  JACC Cardiovasc Imaging       Date:  2015-06

6.  Coronary CT angiography in obese patients using 3(rd) generation dual-source CT: effect of body mass index on image quality.

Authors:  Stefanie Mangold; Julian L Wichmann; U Joseph Schoepf; Sheldon E Litwin; Christian Canstein; Akos Varga-Szemes; Giuseppe Muscogiuri; Stephen R Fuller; Andrew C Stubenrauch; Konstantin Nikolaou; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2015-12-28       Impact factor: 5.315

7.  Advanced image-based virtual monoenergetic dual-energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast.

Authors:  Moritz H Albrecht; Jan-Erik Scholtz; Kristina Hüsers; Martin Beeres; Andreas M Bucher; Moritz Kaup; Simon S Martin; Sebastian Fischer; Boris Bodelle; Ralf W Bauer; Thomas Lehnert; Thomas J Vogl; Julian L Wichmann
Journal:  Eur Radiol       Date:  2015-09-03       Impact factor: 5.315

Review 8.  Dual-energy computed tomography for detection of coronary artery disease.

Authors:  Ibrahim Danad; Bríain Ó Hartaigh; James K Min
Journal:  Expert Rev Cardiovasc Ther       Date:  2015-11-07

9.  Low dose contrast media in step-and-shoot coronary angiography with third-generation dual-source computed tomography: feasibility of using 30 mL of contrast media in patients with body surface area <1.7 m2.

Authors:  Liang Jin; Bing Jie; Yiyi Gao; An'qi Jiang; Tingwen Weng; Ming Li
Journal:  Quant Imaging Med Surg       Date:  2021-06

10.  Influence of trigger type, tube voltage and heart rate on calcified plaque imaging in dual source cardiac computed tomography: phantom study.

Authors:  Tobias Penzkofer; Eva Donandt; Peter Isfort; Thomas Allmendinger; Christiane K Kuhl; Andreas H Mahnken; Philipp Bruners
Journal:  BMC Med Imaging       Date:  2014-09-01       Impact factor: 1.930

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.