Coronary CT angiography in obese patients using 3(rd) generation dual-source CT: effect of body mass index on image quality.

Stefanie Mangold1,2, Julian L Wichmann1,3, U Joseph Schoepf4,5, Sheldon E Litwin6, Christian Canstein7, Akos Varga-Szemes1, Giuseppe Muscogiuri1,8, Stephen R Fuller1, Andrew C Stubenrauch1, Konstantin Nikolaou2, Carlo N De Cecco1.   

Abstract

OBJECTIVES: To evaluate the image quality of coronary CT angiography (CCTA) in obese patients using a 3(rd) generation, dual-source CT scanner.
METHODS: We retrospectively evaluated 102 overweight and obese patients who had undergone CCTA. Studies were performed with 3(rd) generation dual-source CT, prospectively ECG-triggered acquisition at 120 kV, and automated tube current modulation. Advanced modeled iterative reconstruction was used. Patients were divided into three BMI groups: 1)25-29.9 kg/m(2); 2)30-39.9 kg/m(2); 3) ≥ 40 kg/m(2). Vascular attenuation in the coronary arteries was measured. Contrast-to-noise ratio (CNR) was calculated. Image quality was subjectively evaluated using five-point scales.
RESULTS: Image quality was considered diagnostic in 97.6 % of examinations. CNR was consistently adequate in all groups but decreased for groups 2 and 3 in comparison to group 1 as well as for group 3 compared to group 2 (p = 0.001, respectively). Subjective image quality was significantly higher in group 1 compared to group 3 (attenuation proximal: 4.8 ± 0.4 vs. 4.4 ± 0.6, p = 0.011; attenuation distal: 4.5 ± 0.7 vs. 4.0 ± 0.8, p = 0.019; noise: 4.7 ± 0.6 vs. 3.8 ± 0.7, p < 0.001). The mean effective dose was 9.5 ± 3.9 mSv for group 1, 11.4 ± 4.7 mSv for group 2 and 14.0 ± 6.4 mSv for group 3.
CONCLUSION: Diagnostic image quality can be routinely obtained at CCTA in obese patients with 3(rd) generation DSCT at 120 kV. KEY POINTS: • Diagnostic CCTA can be routinely performed in obese patients with 3 (rd) generation DSCT. • 120-kV tube voltage allows diagnostic image quality in patients with BMI > 40 kg/m (2) . • 80-ml contrast medium can be administered without significant decline in vascular attenuation.

Entities:  

Keywords:  Body mass index; Coronary computed tomography angiography; Dual-source; Obesity; Tube current

Mesh:

Year:  2015        PMID: 26712517     DOI: 10.1007/s00330-015-4161-x

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  30 in total

Review 1.  Intravenous contrast medium administration and scan timing at CT: considerations and approaches.

Authors:  Kyongtae T Bae
Journal:  Radiology       Date:  2010-07       Impact factor: 11.105

2.  Effect of increasing body mass index on image quality and positive predictive value of 100-kV coronary computed tomographic angiography.

Authors:  John S Ho; Carolyn E Barlow; Dale B Reinhardt; Wendy A Wade; John J Cannaday
Journal:  Am J Cardiol       Date:  2010-10-15       Impact factor: 2.778

3.  Image quality and radiation dose of low dose coronary CT angiography in obese patients: sinogram affirmed iterative reconstruction versus filtered back projection.

Authors:  Rui Wang; U Joseph Schoepf; Runze Wu; Ryan P Reddy; Chuanchen Zhang; Wei Yu; Yi Liu; Zhaoqi Zhang
Journal:  Eur J Radiol       Date:  2012-05-10       Impact factor: 3.528

4.  Prospectively ECG-triggered high-pitch coronary angiography with third-generation dual-source CT at 70 kVp tube voltage: feasibility, image quality, radiation dose, and effect of iterative reconstruction.

Authors:  Michaela M Hell; Daniel Bittner; Annika Schuhbaeck; Gerd Muschiol; Michael Brand; Michael Lell; Michael Uder; Stephan Achenbach; Mohamed Marwan
Journal:  J Cardiovasc Comput Tomogr       Date:  2014-09-16

5.  Diagnostic accuracy of low-dose 256-slice multi-detector coronary CT angiography using iterative reconstruction in patients with suspected coronary artery disease.

Authors:  Yang Hou; Yue Ma; Weipeng Fan; Yuke Wang; Mei Yu; Mani Vembar; Qiyong Guo
Journal:  Eur Radiol       Date:  2013-07-26       Impact factor: 5.315

6.  Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study.

Authors:  Felix G Meinel; Christian Canstein; U Joseph Schoepf; Martin Sedlmaier; Bernhard Schmidt; Brett S Harris; Thomas G Flohr; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2014-05-10       Impact factor: 5.315

7.  Accuracy of prospectively ECG-triggered very low-dose coronary dual-source CT angiography using iterative reconstruction for the detection of coronary artery stenosis: comparison with invasive catheterization.

Authors:  Christian Layritz; Jasmin Schmid; Stephan Achenbach; Stefan Ulzheimer; Wolfgang Wuest; Matthias May; Dieter Ropers; Lutz Klinghammer; Werner G Daniel; Tobias Pflederer; Michael Lell
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2014-06-17       Impact factor: 6.875

Review 8.  Multidetector computed tomography for the diagnosis of coronary artery disease: a systematic review.

Authors:  Paul D Stein; Afzal Beemath; Fadi Kayali; Elias Skaf; Julia Sanchez; Ronald E Olson
Journal:  Am J Med       Date:  2006-03       Impact factor: 4.965

9.  Iterative model reconstruction: improved image quality of low-tube-voltage prospective ECG-gated coronary CT angiography images at 256-slice CT.

Authors:  Seitaro Oda; Gaby Weissman; Mani Vembar; Wm Guy Weigold
Journal:  Eur J Radiol       Date:  2014-05-02       Impact factor: 3.528

10.  Iterative reconstruction to preserve image quality and diagnostic accuracy at reduced radiation dose in coronary CT angiography: an intraindividual comparison.

Authors:  Wei-Hua Yin; Bin Lu; Nan Li; Lei Han; Zhi-Hui Hou; Run-Ze Wu; Yong-Jian Wu; Hong-Xia Niu; Shi-Liang Jiang; Aleksander W Krazinski; Ullrich Ebersberger; Felix G Meinel; U Joseph Schoepf
Journal:  JACC Cardiovasc Imaging       Date:  2013-10-23
View more
  9 in total

1.  The associated factors for radiation dose variation in cardiac CT angiography.

Authors:  Ali B Alhailiy; Ernest U Ekpo; Peter L Kench; Elaine A Ryan; Patrick C Brennan; Mark McEntee
Journal:  Br J Radiol       Date:  2019-01-31       Impact factor: 3.039

Review 2.  Recent advances in cardiac computed tomography dose reduction strategies: a review of scientific evidence and technical developments.

Authors:  Sandeep S Hedgire; Vinit Baliyan; Brian B Ghoshhajra; Mannudeep K Kalra
Journal:  J Med Imaging (Bellingham)       Date:  2017-08-24

3.  Clinical implementation of an emergency department coronary computed tomographic angiography protocol for triage of patients with suspected acute coronary syndrome.

Authors:  Brian B Ghoshhajra; Richard A P Takx; Pedro V Staziaki; Harshna Vadvala; Phillip Kim; Tomas G Neilan; Nandini M Meyersohn; Daniel Bittner; Sumbal A Janjua; Thomas Mayrhofer; Jeffrey L Greenwald; Quyhn A Truong; Suhny Abbara; David F M Brown; James L Januzzi; Sanjeev Francis; John T Nagurney; Udo Hoffmann
Journal:  Eur Radiol       Date:  2016-11-24       Impact factor: 5.315

4.  Contrast agent volume in coronary computer tomography angiography-where are the limits?

Authors:  Grigorios Korosoglou; Sorin Giusca
Journal:  Quant Imaging Med Surg       Date:  2021-10

5.  Contrast dose reduction with shortened injection durations in coronary CT angiography on 16-cm Wide-detector CT scanner.

Authors:  Ting-Ting Qu; Jian-Ying Li; Xi-Jun Jiao; Xiang-Li Zhang; Zhe-Fan Song; Yin-Xia Guo; Jian-Xin Guo
Journal:  Br J Radiol       Date:  2018-09-04       Impact factor: 3.039

6.  The effect of four-phasic versus three-phasic contrast media injection protocols on extravasation rate in coronary CT angiography: a randomized controlled trial.

Authors:  Júlia Karády; Alexisz Panajotu; Márton Kolossváry; Bálint Szilveszter; Ádám L Jermendy; Andrea Bartykowszki; Mihály Károlyi; Csilla Celeng; Béla Merkely; Pál Maurovich-Horvat
Journal:  Eur Radiol       Date:  2017-05-24       Impact factor: 5.315

7.  Effect of Body Mass Index in Coronary CT Angiography Performed on a 256-Slice Multi-Detector CT Scanner.

Authors:  Wei-Yip Law; Guan-Lin Huang; Ching-Ching Yang
Journal:  Diagnostics (Basel)       Date:  2022-01-27

8.  Evaluation of image quality and radiation dose in low tube voltage coronary computed tomography angiography.

Authors:  Seyedeh Shokoofeh Mousavi-Gazafroudi; Amirreza Sajjadieh-Khajouei; Maryam Moradi; Seyedeh Shabnam Mousavi-Gazafroudi; Ghasem Yadegarfar; Mohammad Bagher Tavakoli
Journal:  ARYA Atheroscler       Date:  2019-09

9.  Coronary computed tomography angiography (CCTA): effect of bolus-tracking ROI positioning on image quality.

Authors:  Heiner Nebelung; Thomas Brauer; Danilo Seppelt; Ralf-Thorsten Hoffmann; Ivan Platzek
Journal:  Eur Radiol       Date:  2020-08-18       Impact factor: 5.315

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.