Improved visual delineation of the intimal flap in Stanford type A and B dissections at 3rd generation dual-source high-pitch CT angiography.

Martin Beeres1, Andreas M Bucher2, Julian L Wichmann2, Claudia Frellesen2, Jan E Scholtz2, Moritz Albrecht2, Boris Bodelle2, Nour-Eldin A Nour-Eldin2,3, Clara Lee2, Moritz Kaup2, Thomas J Vogl2, Tatjana Gruber-Rouh2.   

Abstract

OBJECTIVE: Evaluation of the intimal flap visibility comparing 2nd and 3rd generation dual-source high-pitch CT.
METHODS: Twenty-five consecutive patients with aortic dissection underwent CT angiography on a second and third generation dual-source CT scanner using prospective ECG-gated high-pitch dual-source CT acquisition mode. Contrast material, saline flush and flow rate were kept equal for optimum comparability. The visibility of the intimal flap as well as the delineation of the different vascular structures was evaluated.
RESULTS: In 3rd generation dual-source high-pitch CT we could show a significant improvement of intimal flap visibility in aortic dissection. Especially, the far end of the dissection membrane could be better evaluated in 3rd generation high-pitch CT, reaching statistical significance (P < 0.01).
CONCLUSION: 3rd Generation high-pitch CT angiography shows a better delineation of the aortic intimal flap in a small patient cohort, especially in the far ends of the dissection membrane. This might be due to higher tube power in this CT generation. However, to generalise these findings larger trials are needed.

Entities:  

Keywords:  Aorta; CT angiography; Dual-source CT; High-pitch CT

Mesh:

Substances:

Year:  2016        PMID: 27100720     DOI: 10.1007/s11547-016-0634-5

Source DB:  PubMed          Journal:  Radiol Med        ISSN: 0033-8362            Impact factor:   3.469


  22 in total

Review 1.  Diagnosis and management of aortic dissection.

Authors:  R Erbel; F Alfonso; C Boileau; O Dirsch; B Eber; A Haverich; H Rakowski; J Struyven; K Radegran; U Sechtem; J Taylor; C Zollikofer; W W Klein; B Mulder; L A Providencia
Journal:  Eur Heart J       Date:  2001-09       Impact factor: 29.983

2.  Intravenous contrast material administration at high-pitch dual-source CT pulmonary angiography: test bolus versus bolus-tracking technique.

Authors:  J Matthias Kerl; Thomas Lehnert; Boris Schell; Boris Bodelle; Martin Beeres; Volkmar Jacobi; Thomas J Vogl; Ralf W Bauer
Journal:  Eur J Radiol       Date:  2011-10-22       Impact factor: 3.528

Review 3.  Acute type-B aortic dissection: the treatment strategy.

Authors:  Efstratios Apostolakis; Nikolaos G Baikoussis; Miltiadis Georgiopoulos
Journal:  Hellenic J Cardiol       Date:  2010 Jul-Aug

Review 4.  Intravenous contrast medium administration and scan timing at CT: considerations and approaches.

Authors:  Kyongtae T Bae
Journal:  Radiology       Date:  2010-07       Impact factor: 11.105

Review 5.  Cardiothoracic CT angiography: current contrast medium delivery strategies.

Authors:  Markus Weininger; J Michael Barraza; Corey A Kemper; John F Kalafut; Philip Costello; U Joseph Schoepf
Journal:  AJR Am J Roentgenol       Date:  2011-03       Impact factor: 3.959

6.  Dual source multidetector CT-angiography before Transcatheter Aortic Valve Implantation (TAVI) using a high-pitch spiral acquisition mode.

Authors:  W Wuest; K Anders; A Schuhbaeck; M S May; S Gauss; M Marwan; M Arnold; S Ensminger; G Muschiol; W G Daniel; M Uder; S Achenbach
Journal:  Eur Radiol       Date:  2011-08-17       Impact factor: 5.315

7.  Comparison of image quality and radiation dose of different pulmonary CTA protocols on a 128-slice CT: high-pitch dual source CT, dual energy CT and conventional spiral CT.

Authors:  Tobias De Zordo; Klemens von Lutterotti; Christian Dejaco; Peter F Soegner; Renate Frank; Friedrich Aigner; Andrea S Klauser; Christoph Pechlaner; U Joseph Schoepf; Werner R Jaschke; Gudrun M Feuchtner
Journal:  Eur Radiol       Date:  2011-08-28       Impact factor: 5.315

8.  High-pitch dual-source CT coronary angiography: systolic data acquisition at high heart rates.

Authors:  Robert Goetti; Gudrun Feuchtner; Paul Stolzmann; Lotus Desbiolles; Michael Alexander Fischer; Christoph Karlo; Stephan Baumueller; Hans Scheffel; Hatem Alkadhi; Sebastian Leschka
Journal:  Eur Radiol       Date:  2010-06-29       Impact factor: 5.315

9.  Closing in on the K edge: coronary CT angiography at 100, 80, and 70 kV-initial comparison of a second- versus a third-generation dual-source CT system.

Authors:  Mathias Meyer; Holger Haubenreisser; U Joseph Schoepf; Rozemarijn Vliegenthart; Christianne Leidecker; Thomas Allmendinger; Ralf Lehmann; Sonja Sudarski; Martin Borggrefe; Stefan O Schoenberg; Thomas Henzler
Journal:  Radiology       Date:  2014-05-31       Impact factor: 11.105

10.  Energy Limits in Second Generation High-pitch Dual Source CT - Comparison in an Upper Abdominal Phantom.

Authors:  Martin Beeres; Ralf W Bauer; Josef M Kerl; Thomas J Vogl; Clara Lee
Journal:  J Clin Imaging Sci       Date:  2015-01-30
View more
  2 in total

1.  Optimal phase analysis of electrocardiogram-gated computed tomography angiography in patients with Stanford type A acute aortic dissection.

Authors:  Kenji Nishida; Yuki Yokoi; Ayumi Yamada; Nobuhiro Takaya; Ken Yamagiwa; Shuichi Kawada; Koichi Mori; Susumu Manabe; Eiichiro Kanda; Tomoyuki Fujioka; Mitsuhiro Kishino; Ukihide Tateishi
Journal:  Eur J Radiol Open       Date:  2020-12-09

2.  The ultrafast, high-pitch turbo FLASH mode of third-generation dual-source CT: Effect of different pitch and corresponding SFOV on image quality in a phantom study.

Authors:  Yang Zhou; Lei Hu; Silin Du; Rui Jin; Wangjia Li; Fajin Lv; Zhiwei Zhang
Journal:  J Appl Clin Med Phys       Date:  2021-11-09       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.