BACKGROUND: The most commonly used form of reporting doses in multislice computed tomography involves a CT dose index per slice and dose-length product for the whole series. The purpose of this study was to analyze the actual dose distribution in routine chest CT examination protocols using an antropomorphic phantom. MATERIAL/Entities:
Keywords: Chest CT; Computed Tomography (CT); Low Dose CT; Lung CT; Radiation Protection; Radiation Safety
Year: 2014 PMID: 24744819 PMCID: PMC3990109 DOI: 10.12659/PJR.889951
Source DB: PubMed Journal: Pol J Radiol ISSN: 1733-134X
Figure 1A computed tomography scan of a phantom.
Descriptive statistics concerning doses measured by detectors during scanning of the phantom and statistical significance of differences between mean dose values.
| Protocol | Low-dose | Helical | Angio-CT | ||||||
|---|---|---|---|---|---|---|---|---|---|
|
| |||||||||
| Section part | Section part | Section part | |||||||
|
| |||||||||
| Part of phantom’s section | Anterior | Middle | Posterior | Anterior | Middle | Posterior | Anterior | Middle | Posterior |
| Dose (mSv) | |||||||||
|
| |||||||||
| Mean | 3.74 | 3.18 | 3.01 | 16.95 | 14.15 | 12.47 | 30.17 | 26.71 | 24.98 |
|
| |||||||||
| Min./max. | 3.32 | 2.60 | 2.63 | 14.64 | 12.04 | 10.38 | 26.47 | 21.67 | 21.47 |
| 4.50 | 3.61 | 3.29 | 22.14 | 16.09 | 14.18 | 34.35 | 30.17 | 27.43 | |
|
| |||||||||
| Median | 3.66 | 3.24 | 3.04 | 15.99 | 14.53 | 12.57 | 30.26 | 27.04 | 25.33 |
|
| |||||||||
| Standard deviation | 0.34 | 0.31 | 0.19 | 2.34 | 1.34 | 1.14 | 2.18 | 2.53 | 1.73 |
|
| |||||||||
| Statistical significance of differences between means (p) | 0.0001 | 0.0429 | 0.0003 | 0.0007 | 0.0003 | 0.0222 | |||
Figure 2Dose distribution in the short axis of the phantom with a division into anterior, medial and posterior parts of phantom sections.
Figure 3Dose distribution in the long axis of the phantom.
Correlations between mean doses acquired from anterior, medial and posterior parts of the phantom in the course of three scanning protocols. Significant correlations with p<0.05 are denoted with*.
| Correlation coefficient | Low-dose/helical | Low-dose/angio-CT | Helical/angio-CT |
|---|---|---|---|
| Anterior | 0.49 | 0.63* | 0.36 |
| Middle | 0.73* | 0.66* | 0.83* |
| Posterior | 0.06 | 0.21 | 0.57* |