| Literature DB >> 24701570 |
Marcelo P de Castro1, Marco Meucci2, Denise P Soares3, Pedro Fonseca4, Márcio Borgonovo-Santos3, Filipa Sousa3, Leandro Machado3, João Paulo Vilas-Boas3.
Abstract
WalkinSense is a new device designed to monitor walking. The aim of this study was to measure the accuracy and repeatability of the gait analysis performed by the WalkinSense system. Descriptions of values recorded by WalkinSense depicting typical gait in adults are also presented. A bench experiment using the Trublu calibration device was conducted to statically test the WalkinSense. Following this, a dynamic test was carried out overlapping the WalkinSense and the Pedar insoles in 40 healthy participants during walking. Pressure peak, pressure peak time, pressure-time integral, and mean pressure at eight-foot regions were calculated. In the bench experiments, the repeatability (i) among the WalkinSense sensors (within), (ii) between two WalkinSense devices, and (iii) between the WalkinSense and the Trublu devices was excellent. In the dynamic tests, the repeatability of the WalkinSense (i) between stances in the same trial (within-trial) and (ii) between trials was also excellent (ICC > 0.90). When the eight-foot regions were analyzed separately, the within-trial and between-trials repeatability was good-to-excellent in 88% (ICC > 0.80) of the data and fair in 11%. In short, the data suggest that the WalkinSense has good-to-excellent levels of accuracy and repeatability for plantar pressure variables.Entities:
Mesh:
Year: 2014 PMID: 24701570 PMCID: PMC3950554 DOI: 10.1155/2014/348659
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1(a) WalkinSense and Pedar attached on one of the participants during data collection. (b) Experimental setup of the bench experiment: (&) position of the eight WalkinSense sensors on (∗) the Pedar insoles prior to insertion in (#) the Trublu calibration device.
Figure 2Position of the WalkinSense sensors at the Pedar insole.
Figure 3Bench experiment: relation between the applied loads and WalkinSense records at the 10 levels of load. ICC: intraclass correlation coefficient. At the second and third column of the table, positive values indicate greater values in the WalkinSense system and negative values indicate lower values in the WalkinSense.
WalkinSense overall (all regions together) within- and between-trials intraclass correlation coefficients (ICC) for all measurements during gait.
| Variables | Within-trial | Between-trial | ||||
|---|---|---|---|---|---|---|
| ICC | CI95% | ICC | CI95% | |||
|
| 0.972 | 0.969 | 0.975 | 0.979 | 0.977 | 0.981 |
|
| 0.987 | 0.986 | 0.988 | 0.915 | 0.904 | 0.925 |
|
| 0.940 | 0.933 | 0.946 | 0.993 | 0.992 | 0.994 |
|
| 0.938 | 0.931 | 0.944 | 0.965 | 0.960 | 0.969 |
P Peak: peak pressure; P Time: pressure time; P Mean: mean pressure; P Integral: pressure-time integral.
WalkinSense regional within- and between-trial intraclass correlation coefficients (ICC) during gait.
| Variable | Region | Within-trial | Between-trial | ||||
|---|---|---|---|---|---|---|---|
| ICC | CI95% | ICC | CI95% | ||||
| Peak pressure | RFLat | 0.978 | 0.971 | 0.983 | 0.971 | 0.961 | 0.979 |
| RFMed | 0.964 | 0.953 | 0.973 | 0.961 | 0.948 | 0.972 | |
| MFLat | 0.763 | 0.688 | 0.821 | 0.794 | 0.713 | 0.855 | |
| MFMed | & | & | & | & | & | & | |
| FFLat | 0.868 | 0.826 | 0.900 | 0.896 | 0.858 | 0.926 | |
| FFCt | 0.949 | 0.932 | 0.961 | 0.956 | 0.940 | 0.968 | |
| FFMed | 0.834 | 0.779 | 0.875 | 0.858 | 0.802 | 0.900 | |
| GToe | 0.928 | 0.905 | 0.946 | 0.966 | 0.954 | 0.976 | |
|
| |||||||
| Peak pressure time | RFLat | 0.718 | 0.630 | 0.785 | 0.859 | 0.808 | 0.897 |
| RFMed | 0.763 | 0.684 | 0.823 | 0.826 | 0.758 | 0.877 | |
| MFLat | 0.665 | 0.554 | 0.749 | 0.763 | 0.669 | 0.834 | |
| MFMed | & | & | & | & | & | & | |
| FFLat | 0.873 | 0.831 | 0.904 | 0.857 | 0.801 | 0.900 | |
| FFCt | 0.786 | 0.717 | 0.838 | 0.808 | 0.736 | 0.863 | |
| FFMed | 0.748 | 0.664 | 0.811 | 0.843 | 0.780 | 0.891 | |
| GToe | 0.900 | 0.867 | 0.924 | 0.946 | 0.926 | 0.962 | |
|
| |||||||
| Pressure-time integral | RFLat | 0.959 | 0.946 | 0.969 | 0.960 | 0.946 | 0.971 |
| RFMed | 0.908 | 0.879 | 0.930 | 0.928 | 0.901 | 0.948 | |
| MFLat | 0.848 | 0.798 | 0.885 | 0.873 | 0.825 | 0.910 | |
| MFMed | & | & | & | & | & | & | |
| FFLat | 0.890 | 0.855 | 0.917 | 0.912 | 0.879 | 0.937 | |
| FFCt | 0.960 | 0.948 | 0.970 | 0.965 | 0.953 | 0.975 | |
| FFMed | 0.810 | 0.746 | 0.857 | 0.835 | 0.769 | 0.885 | |
| GToe | 0.882 | 0.843 | 0.911 | 0.915 | 0.882 | 0.939 | |
|
| |||||||
| Mean pressure | RFLat | 0.965 | 0.955 | 0.974 | 0.961 | 0.947 | 0.972 |
| RFMed | 0.903 | 0.871 | 0.926 | 0.928 | 0.902 | 0.949 | |
| MFLat | 0.817 | 0.758 | 0.862 | 0.867 | 0.814 | 0.906 | |
| MFMed | & | & | & | & | & | & | |
| FFLat | 0.884 | 0.847 | 0.912 | 0.915 | 0.883 | 0.939 | |
| FFCt | 0.959 | 0.947 | 0.969 | 0.962 | 0.948 | 0.973 | |
| FFMed | 0.816 | 0.755 | 0.862 | 0.840 | 0.774 | 0.889 | |
| GToe | 0.890 | 0.853 | 0.917 | 0.916 | 0.885 | 0.941 | |
RFLat: lateral rearfoot; RFMed: medial rearfoot; MFLat: lateral midfoot; MFMed: medial midfoot; FFLat: lateral forefoot; FFCt: central forefoot; MFMed: medial forefoot; GToe: great toe; CI95%: 95% confidence interval. &: as the midfoot region was loaded only in ≅5% of the trials, the percentage difference and ICC were not calculated for this region.
Percentage difference, intraclass correlation coefficient (ICC), and Pearson correlation coefficient (CC) between Pedar and WalkinSense during gait.
| Percentage differences (%) | ICC | Pearson CC | ||||||
|---|---|---|---|---|---|---|---|---|
| Average difference | CI95% | ICC | CI95% |
|
| |||
|
| −6.2 | −6.8 | −5.6 | 0.973 | 0.971 | 0.975 | 0.869 | <0.001 |
|
| 3.4 | 3.1 | 3.6 | 0.997 | 0.997 | 0.997 | 0.973 | <0.001 |
|
| −14.1 | −14.9 | −13.2 | 0.955 | 0.952 | 0.958 | 0.872 | <0.001 |
|
| −13.2 | −14.1 | −12.4 | 0.956 | 0.953 | 0.959 | 0.872 | <0.001 |
P Peak: peak pressure; P Time: pressure time; P Mean: mean pressure; P Integral: pressure-time integral.
Percentage differences and intraclass correlation coefficient (ICC) between Pedar and WalkinSense for each foot region during gait.
| Variable | Region | Percentage differences (%) | ICC | CI95% | |||
|---|---|---|---|---|---|---|---|
| Average difference | CI95% | ||||||
| Peak pressure | RFLat | −10.8 | −12.4 | −9.1 | 0.859 | 0.827 | 0.885 |
| RFMed | −3.6 | −5.0 | −2.2 | 0.958 | 0.949 | 0.965 | |
| MFLat | −16.4 | −18.1 | −14.6 | 0.963 | 0.954 | 0.971 | |
| MFMed | & | & | & | & | & | & | |
| FFLat | −15.2 | −16.7 | −13.7 | 0.936 | 0.922 | 0.947 | |
| FFCt | −5.4 | −6.9 | −3.9 | 0.953 | 0.942 | 0.961 | |
| FFMed | 10.4 | 8.5 | 12.4 | 0.945 | 0.932 | 0.956 | |
| GToe | −7.9 | −9.8 | −6.1 | 0.953 | 0.942 | 0.961 | |
|
| |||||||
| Peak pressure time | RFLat | 4.3 | 3.1 | 5.5 | 0.974 | 0.967 | 0.980 |
| RFMed | 3.4 | 2.2 | 4.5 | 0.982 | 0.977 | 0.985 | |
| MFLat | 3.6 | 2.9 | 4.3 | 0.983 | 0.980 | 0.986 | |
| MFMed | & | & | & | & | & | & | |
| FFLat | 3.7 | 3.3 | 4.1 | 0.964 | 0.956 | 0.971 | |
| FFCt | 2.3 | 1.9 | 2.7 | 0.839 | 0.803 | 0.868 | |
| FFMed | 3.0 | 2.6 | 3.5 | 0.944 | 0.931 | 0.954 | |
| GToe | 3.5 | 3.1 | 3.8 | 0.938 | 0.923 | 0.949 | |
|
| |||||||
| Pressure-time integral | RFLat | −18.4 | −20.7 | −16.2 | 0.907 | 0.888 | 0.923 |
| RFMed | −7.7 | −9.6 | −5.8 | 0.949 | 0.937 | 0.958 | |
| MFLat | −31.2 | −33.7 | −28.8 | 0.937 | 0.923 | 0.948 | |
| MFMed | & | & | & | & | & | & | |
| FFLat | −24.7 | −26.6 | −22.8 | 0.915 | 0.897 | 0.929 | |
| FFCt | 8.4 | 6.4 | 10.4 | 0.923 | 0.907 | 0.937 | |
| FFMed | −12.2 | −13.8 | −10.6 | 0.961 | 0.952 | 0.968 | |
| GToe | −14.0 | −15.8 | −12.2 | 0.959 | 0.950 | 0.966 | |
|
| |||||||
| Mean pressure | RFLat | −17.5 | −19.8 | −15.2 | 0.915 | 0.897 | 0.930 |
| RFMed | −6.3 | −8.1 | −4.4 | 0.947 | 0.936 | 0.957 | |
| MFLat | −30.1 | −32.6 | −27.7 | 0.942 | 0.927 | 0.953 | |
| MFMed | & | & | & | & | & | & | |
| FFLat | −23.7 | −25.6 | −21.8 | 0.913 | 0.895 | 0.929 | |
| FFCt | 9.8 | 7.8 | 11.8 | 0.949 | 0.937 | 0.959 | |
| FFMed | −11.0 | −12.6 | −9.4 | 0.959 | 0.950 | 0.967 | |
| GToe | −13.3 | −15.1 | −11.4 | 0.970 | 0.963 | 0.975 | |
RFLat: lateral rearfoot; RFMed: medial rearfoot; MFLat: lateral midfoot; MFMed: medial midfoot; FFLat: lateral forefoot; FFCt: central forefoot; MFMed: medial forefoot; GToe: great toe; CI95%: 95% confidence interval. &: as the midfoot region was loaded only in ≅5% of the trials, the percentage difference and ICC were not calculated for this region.
Description of values for the WalkinSense and Pedar during gait.
| Variable | Region | Pedar | WalkinSense | ||
|---|---|---|---|---|---|
| Mean | SD | Mean | SD | ||
| Peak pressure (kPa) | RFLat | 291.0 | 73.0 | 257.2 | 71.6 |
| RFMed | 285.3 | 95.8 | 274.7 | 96.3 | |
| MFLat | 80.8 | 35.0 | 67.8 | 34.2 | |
| MFMed | 0.8 | 6.4 | 0.3 | 2.7 | |
| FFLat | 250.0 | 86.7 | 211.9 | 80.8 | |
| FFCt | 302.9 | 104.1 | 330.7 | 110.8 | |
| FFMed | 257.0 | 88.8 | 242.4 | 86.9 | |
| GToe | 247.0 | 121.7 | 224.2 | 111.9 | |
|
| |||||
| Peak pressure time (%Stance) | RFLat | 16.2 | 5.1 | 16.9 | 5.4 |
| RFMed | 17.8 | 6.5 | 18.4 | 6.8 | |
| MFLat | 48.2 | 11.5 | 49.6 | 11.3 | |
| MFMed | 26.7 | 8.7 | 33.5 | 13.3 | |
| FFLat | 71.0 | 7.4 | 73.6 | 7.4 | |
| FFCt | 76.9 | 4.1 | 78.6 | 3.6 | |
| FFMed | 72.3 | 6.7 | 74.3 | 6.4 | |
| GToe | 80.6 | 6.0 | 83.3 | 5.6 | |
|
| |||||
| Pressure-time integral (kPa·s) | RFLat | 70.5 | 26.6 | 58.1 | 27.8 |
| RFMed | 70.7 | 27.0 | 65.6 | 28.9 | |
| MFLat | 25.8 | 10.6 | 18.3 | 11.0 | |
| MFMed | 0.4 | 1.6 | 0.0 | 0.4 | |
| FFLat | 82.5 | 26.0 | 63.5 | 27.9 | |
| FFCt | 84.1 | 30.7 | 90.7 | 35.2 | |
| FFMed | 78.4 | 28.3 | 69.9 | 30.3 | |
| GToe | 57.7 | 32.9 | 50.4 | 32.0 | |
|
| |||||
| Mean pressure (kPa) | RFLat | 94.5 | 33.0 | 79.1 | 36.3 |
| RFMed | 96.2 | 37.1 | 90.4 | 39.5 | |
| MFLat | 34.5 | 14.4 | 24.7 | 14.3 | |
| MFMed | 0.5 | 2.3 | 0.1 | 0.7 | |
| FFLat | 111.5 | 33.2 | 86.5 | 36.1 | |
| FFCt | 112.9 | 39.4 | 123.2 | 44.9 | |
| FFMed | 106.1 | 37.4 | 95.7 | 40.4 | |
| GToe | 78.3 | 43.5 | 69.0 | 42.8 | |
RFLat: lateral rearfoot; RFMed: medial rearfoot; MFLat: lateral midfoot; MFMed: medial midfoot; FFLat: lateral forefoot; FFCt: central forefoot; MFMed: medial forefoot; GToe: great toe; CI95%: 95% confidence interval; SD: standard deviation.