Literature DB >> 24694149

Tungsten anode spectral model using interpolating cubic splines: unfiltered x-ray spectra from 20 kV to 640 kV.

Andrew M Hernandez1, John M Boone2.   

Abstract

PURPOSE: Monte Carlo methods were used to generate lightly filtered high resolution x-ray spectra spanning from 20 kV to 640 kV.
METHODS: X-ray spectra were simulated for a conventional tungsten anode. The Monte Carlo N-Particle eXtended radiation transport code (MCNPX 2.6.0) was used to produce 35 spectra over the tube potential range from 20 kV to 640 kV, and cubic spline interpolation procedures were used to create piecewise polynomials characterizing the photon fluence per energy bin as a function of x-ray tube potential. Using these basis spectra and the cubic spline interpolation, 621 spectra were generated at 1 kV intervals from 20 to 640 kV. The tungsten anode spectral model using interpolating cubic splines (TASMICS) produces minimally filtered (0.8 mm Be) x-ray spectra with 1 keV energy resolution. The TASMICS spectra were compared mathematically with other, previously reported spectra.
RESULTS: Using pairedt-test analyses, no statistically significant difference (i.e., p > 0.05) was observed between compared spectra over energy bins above 1% of peak bremsstrahlung fluence. For all energy bins, the correlation of determination (R(2)) demonstrated good correlation for all spectral comparisons. The mean overall difference (MOD) and mean absolute difference (MAD) were computed over energy bins (above 1% of peak bremsstrahlung fluence) and over all the kV permutations compared. MOD and MAD comparisons with previously reported spectra were 2.7% and 9.7%, respectively (TASMIP), 0.1% and 12.0%, respectively [R. Birch and M. Marshall, "Computation of bremsstrahlung x-ray spectra and comparison with spectra measured with a Ge(Li) detector," Phys. Med. Biol. 24, 505-517 (1979)], 0.4% and 8.1%, respectively (Poludniowski), and 0.4% and 8.1%, respectively (AAPM TG 195). The effective energy of TASMICS spectra with 2.5 mm of added Al filtration ranged from 17 keV (at 20 kV) to 138 keV (at 640 kV); with 0.2 mm of added Cu filtration the effective energy was 9 keV at 20 kV and 169 keV at 640 kV.
CONCLUSIONS: Ranging from 20 kV to 640 kV, 621 x-ray spectra were produced and are available at 1 kV tube potential intervals. The spectra are tabulated at 1 keV intervals. TASMICS spectra were shown to be largely equivalent to published spectral models and are available in spreadsheet format for interested users by emailing the corresponding author (JMB).
© 2014 American Association of Physicists in Medicine.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24694149      PMCID: PMC3985923          DOI: 10.1118/1.4866216

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Identification of heavy metal origins related to chemical and morphological soil properties using several non-destructive X-ray analytical methods.

Authors:  Songul Akbulut; Renevan Grieken; Mehmet A Kılıc; Ugur Cevik; Giuliana G Rotondo
Journal:  Environ Monit Assess       Date:  2012-06-22       Impact factor: 2.513

2.  Simulation of image detectors in radiology for determination of scatter-to-primary ratios using Monte Carlo radiation transport code MCNP/MCNPX.

Authors:  Kristien Smans; Johannes Zoetelief; Beatrijs Verbrugge; Wim Haeck; Lara Struelens; Filip Vanhavere; Hilde Bosmans
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

3.  Methodology for digital radiography simulation using the Monte Carlo code MCNPX for industrial applications.

Authors:  E M Souza; S C A Correa; A X Silva; R T Lopes; D F Oliveira
Journal:  Appl Radiat Isot       Date:  2007-11-21       Impact factor: 1.513

4.  Multi-energy X-ray imaging as a quantitative method for materials characterization.

Authors:  M Firsching; F Nachtrab; N Uhlmann; R Hanke
Journal:  Adv Mater       Date:  2011-04-12       Impact factor: 30.849

5.  Inconsistency in 9 mm bullets measured with non-destructive X-ray computed tomography.

Authors:  Jagadeesha Kumar; Dirk Landheer; Jane Barnes-Warden; Paul Fenne; Alex Attridge; Mark A Williams
Journal:  Forensic Sci Int       Date:  2011-08-03       Impact factor: 2.395

6.  An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1997-11       Impact factor: 4.071

7.  Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography.

Authors:  J M Boone; T R Fewell; R J Jennings
Journal:  Med Phys       Date:  1997-12       Impact factor: 4.071

8.  Kilovoltage rotational external beam radiotherapy on a breast computed tomography platform: a feasibility study.

Authors:  Nicolas D Prionas; Sarah E McKenney; Robin L Stern; John M Boone
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-03-19       Impact factor: 7.038

9.  Computation of bremsstrahlung X-ray spectra and comparison with spectra measured with a Ge(Li) detector.

Authors:  R Birch; M Marshall
Journal:  Phys Med Biol       Date:  1979-05       Impact factor: 3.609

  9 in total
  43 in total

1.  A new technique to characterize CT scanner bow-tie filter attenuation and applications in human cadaver dosimetry simulations.

Authors:  Xinhua Li; Jim Q Shi; Da Zhang; Sarabjeet Singh; Atul Padole; Alexi Otrakji; Mannudeep K Kalra; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

2.  Breast dose in mammography is about 30% lower when realistic heterogeneous glandular distributions are considered.

Authors:  Andrew M Hernandez; J Anthony Seibert; John M Boone
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

3.  Detective quantum efficiency of photon-counting CdTe and Si detectors for computed tomography: a simulation study.

Authors:  Mats Persson; Adam Wang; Norbert J Pelc
Journal:  J Med Imaging (Bellingham)       Date:  2020-07-17

4.  Organ dose variability and trends in tomosynthesis and radiography.

Authors:  Jocelyn Hoye; Yakun Zhang; Greeshma Agasthya; Greg Sturgeon; Anuj Kapadia; W Paul Segars; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2017-08-01

5.  Segmentation-free x-ray energy spectrum estimation for computed tomography using dual-energy material decomposition.

Authors:  Wei Zhao; Lei Xing; Qiude Zhang; Qingguo Xie; Tianye Niu
Journal:  J Med Imaging (Bellingham)       Date:  2017-06-30

6.  Monte Carlo Basics for Radiation Dose Assessment in Diagnostic Radiology.

Authors:  John M Boone; Michael F McNitt-Gray; Andrew M Hernandez
Journal:  J Am Coll Radiol       Date:  2017-04-29       Impact factor: 5.532

7.  The Effect of Iodine-based Contrast Material on Radiation Dose at CT: It's Complicated.

Authors:  John M Boone; Andrew M Hernandez
Journal:  Radiology       Date:  2017-06       Impact factor: 11.105

8.  A Monte Carlo model for mean glandular dose evaluation in spot compression mammography.

Authors:  Antonio Sarno; David R Dance; Ruben E van Engen; Kenneth C Young; Paolo Russo; Francesca Di Lillo; Giovanni Mettivier; Kristina Bliznakova; Baowei Fei; Ioannis Sechopoulos
Journal:  Med Phys       Date:  2017-06-13       Impact factor: 4.071

9.  A framework for performance characterization of energy-resolving photon-counting detectors.

Authors:  Mats Persson; Paurakh L Rajbhandary; Norbert J Pelc
Journal:  Med Phys       Date:  2018-10-12       Impact factor: 4.071

10.  The Influence of X-Ray Computed Tomography Acquisition Parameters on Image Quality and Probability of Detection of Additive Manufacturing Defects.

Authors:  Felix H Kim; Adam L Pintar; Shawn P Moylan; Edward J Garboczi
Journal:  J Manuf Sci Eng       Date:  2019-11       Impact factor: 3.033

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.