Pulse pileup analysis for a double-sided silicon strip detector using variable pulse shapes.

Jinghui Wang1, Linchuan Chen2, Mats Persson3, Paurakh L Rajbhandary4, Praneeth Kandlakunta5, Gabriella Carini6, Rebecca Fahrig7.   

Abstract

Due to pulse pileup, photon counting detectors (PCDs) suffer from count loss and energy distortion when operating in high count rate environments. In this paper, we studied the pulse pileup of a double-sided silicon strip detector (DSSSD) to evaluate its potential application in a mammography system. We analyzed the pulse pileup using pulses of varied shapes, where the shape of the pulse depends on the location of photon interaction within the detector. To obtain the shaped pulses, first, transient currents for photons interacting at different locations were simulated using a Technology Computer-Aided Design (TCAD) software. Next, the currents were shaped by a CR-RC2 shaping circuit, calculated using Simulink. After obtaining these pulses, both the different orders of pileup and the energy spectrum were calculated by taking into account the following two factors: 1) spatial distribution of photon interactions within the detector, and 2) time interval distribution between successive photons under a given photon flux. We found that for a DSSSD with thickness of 300 μm, pitch of 25 μm and strip length of 1 cm, under a bias voltage of 50 V, the variable pulse shape model predicts the fraction free of pileup can be > 90 % under a photon flux of 3.75 Mcps/mm2. The double-sided silicon-strip detector is a promising candidate for digital mammography applications.

Entities:  

Keywords:  double-sided silicon strip detector; paralyzable detection model; photon counting detector; pulse pileup

Year:  2019        PMID: 31327872      PMCID: PMC6640861          DOI: 10.1109/TNS.2019.2917144

Source DB:  PubMed          Journal:  IEEE Trans Nucl Sci        ISSN: 0018-9499            Impact factor:   1.679


  12 in total

1.  An analytical model of the effects of pulse pileup on the energy spectrum recorded by energy resolved photon counting x-ray detectors.

Authors:  Katsuyuki Taguchi; Eric C Frey; Xiaolan Wang; Jan S Iwanczyk; William C Barber
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

2.  Dual-energy approach to contrast-enhanced mammography using the balanced filter method: spectral optimization and preliminary phantom measurement.

Authors:  Masatoshi Saito
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

Review 3.  Vision 20/20: Single photon counting x-ray detectors in medical imaging.

Authors:  Katsuyuki Taguchi; Jan S Iwanczyk
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

4.  High-rate x-ray spectroscopy in mammography with a CdTe detector: a digital pulse processing approach.

Authors:  L Abbene; G Gerardi; F Principato; S Del Sordo; R Ienzi; G Raso
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

5.  A cascaded model of spectral distortions due to spectral response effects and pulse pileup effects in a photon-counting x-ray detector for CT.

Authors:  Jochen Cammin; Jennifer Xu; William C Barber; Jan S Iwanczyk; Neal E Hartsough; Katsuyuki Taguchi
Journal:  Med Phys       Date:  2014-04       Impact factor: 4.071

6.  Improving pulse detection in multibin photon-counting detectors.

Authors:  Scott S Hsieh; Norbert J Pelc
Journal:  J Med Imaging (Bellingham)       Date:  2016-06-01

7.  Evaluation of models of spectral distortions in photon-counting detectors for computed tomography.

Authors:  Jochen Cammin; Steffen Kappler; Thomas Weidinger; Katsuyuki Taguchi
Journal:  J Med Imaging (Bellingham)       Date:  2016-05-06

Review 8.  Energy-sensitive photon counting detector-based X-ray computed tomography.

Authors:  Katsuyuki Taguchi
Journal:  Radiol Phys Technol       Date:  2017-01-30

9.  Modeling the performance of a photon counting x-ray detector for CT: energy response and pulse pileup effects.

Authors:  Katsuyuki Taguchi; Mengxi Zhang; Eric C Frey; Xiaolan Wang; Jan S Iwanczyk; Einar Nygard; Neal E Hartsough; Benjamin M W Tsui; William C Barber
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

10.  Upper limits of the photon fluence rate on CT detectors: Case study on a commercial scanner.

Authors:  Mats Persson; Robert Bujila; Patrik Nowik; Henrik Andersson; Love Kull; Jonas Andersson; Hans Bornefalk; Mats Danielsson
Journal:  Med Phys       Date:  2016-07       Impact factor: 4.071

View more
  1 in total

1.  Anomalous edge response of cadmium telluride-based photon counting detectors jointly caused by high-flux radiation and inter-pixel communication.

Authors:  Xu Ji; Kevin Treb; Ke Li
Journal:  Phys Med Biol       Date:  2021-04-14       Impact factor: 3.609

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.