Rapid Monte Carlo simulation of detector DQE(f).

Josh Star-Lack1, Mingshan Sun1, Andre Meyer2, Daniel Morf2, Dragos Constantin3, Rebecca Fahrig3, Eric Abel1.   

Abstract

PURPOSE: Performance optimization of indirect x-ray detectors requires proper characterization of both ionizing (gamma) and optical photon transport in a heterogeneous medium. As the tool of choice for modeling detector physics, Monte Carlo methods have failed to gain traction as a design utility, due mostly to excessive simulation times and a lack of convenient simulation packages. The most important figure-of-merit in assessing detector performance is the detective quantum efficiency (DQE), for which most of the computational burden has traditionally been associated with the determination of the noise power spectrum (NPS) from an ensemble of flood images, each conventionally having 10(7) - 10(9) detected gamma photons. In this work, the authors show that the idealized conditions inherent in a numerical simulation allow for a dramatic reduction in the number of gamma and optical photons required to accurately predict the NPS.
METHODS: The authors derived an expression for the mean squared error (MSE) of a simulated NPS when computed using the International Electrotechnical Commission-recommended technique based on taking the 2D Fourier transform of flood images. It is shown that the MSE is inversely proportional to the number of flood images, and is independent of the input fluence provided that the input fluence is above a minimal value that avoids biasing the estimate. The authors then propose to further lower the input fluence so that each event creates a point-spread function rather than a flood field. The authors use this finding as the foundation for a novel algorithm in which the characteristic MTF(f), NPS(f), and DQE(f) curves are simultaneously generated from the results of a single run. The authors also investigate lowering the number of optical photons used in a scintillator simulation to further increase efficiency. Simulation results are compared with measurements performed on a Varian AS1000 portal imager, and with a previously published simulation performed using clinical fluence levels.
RESULTS: On the order of only 10-100 gamma photons per flood image were required to be detected to avoid biasing the NPS estimate. This allowed for a factor of 10(7) reduction in fluence compared to clinical levels with no loss of accuracy. An optimal signal-to-noise ratio (SNR) was achieved by increasing the number of flood images from a typical value of 100 up to 500, thereby illustrating the importance of flood image quantity over the number of gammas per flood. For the point-spread ensemble technique, an additional 2× reduction in the number of incident gammas was realized. As a result, when modeling gamma transport in a thick pixelated array, the simulation time was reduced from 2.5 × 10(6) CPU min if using clinical fluence levels to 3.1 CPU min if using optimized fluence levels while also producing a higher SNR. The AS1000 DQE(f) simulation entailing both optical and radiative transport matched experimental results to within 11%, and required 14.5 min to complete on a single CPU.
CONCLUSIONS: The authors demonstrate the feasibility of accurately modeling x-ray detector DQE(f) with completion times on the order of several minutes using a single CPU. Convenience of simulation can be achieved using GEANT4 which offers both gamma and optical photon transport capabilities.

Entities:  

Mesh:

Year:  2014        PMID: 24593734      PMCID: PMC3978362          DOI: 10.1118/1.4865761

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  Monte Carlo simulations of the imaging performance of metal plate/phosphor screens used in radiotherapy.

Authors:  C Kausch; B Schreiber; F Kreuder; R Schmidt; O Dössel
Journal:  Med Phys       Date:  1999-10       Impact factor: 4.071

Review 2.  Inorganic scintillators in medical imaging.

Authors:  Carel W E van Eijk
Journal:  Phys Med Biol       Date:  2002-04-21       Impact factor: 3.609

3.  Segmented crystalline scintillators: an initial investigation of high quantum efficiency detectors for megavoltage x-ray imaging.

Authors:  Amit Sawant; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Yixin Li; Zhong Su; Yi Wang; Jin Yamamoto; Hong Du; Ian Cunningham; Misha Klugerman; Kanai Shah
Journal:  Med Phys       Date:  2005-10       Impact factor: 4.071

4.  Thick, segmented CdWO4-photodiode detector for cone beam megavoltage CT: a Monte Carlo study of system design parameters.

Authors:  T T Monajemi; B G Fallone; S Rathee
Journal:  Med Phys       Date:  2006-12       Impact factor: 4.071

5.  A simple method for determining the modulation transfer function in digital radiography.

Authors:  H Fujita; D Y Tsai; T Itoh; K Doi; J Morishita; K Ueda; A Ohtsuka
Journal:  IEEE Trans Med Imaging       Date:  1992       Impact factor: 10.048

6.  Signal, noise power spectrum, and detective quantum efficiency of indirect-detection flat-panel imagers for diagnostic radiology.

Authors:  J H Siewerdsen; L E Antonuk; Y el-Mohri; J Yorkston; W Huang; I A Cunningham
Journal:  Med Phys       Date:  1998-05       Impact factor: 4.071

7.  X-ray quantum limited portal imaging using amorphous silicon flat-panel arrays.

Authors:  P Munro; D C Bouius
Journal:  Med Phys       Date:  1998-05       Impact factor: 4.071

8.  Optical photon transport in powdered-phosphor scintillators. Part II. Calculation of single-scattering transport parameters.

Authors:  Gavin G Poludniowski; Philip M Evans
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

9.  Monte Carlo simulation of the scattered radiation distribution in diagnostic radiology.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1988 Sep-Oct       Impact factor: 4.071

10.  Countering beam divergence effects with focused segmented scintillators for high DQE megavoltage active matrix imagers.

Authors:  Langechuan Liu; Larry E Antonuk; Qihua Zhao; Youcef El-Mohri; Hao Jiang
Journal:  Phys Med Biol       Date:  2012-08-01       Impact factor: 3.609

View more
  12 in total

1.  Physics considerations in MV-CBCT multi-layer imager design.

Authors:  Yue-Houng Hu; Rony Fueglistaller; Marios Myronakis; Joerg Rottmann; Adam Wang; Daniel Shedlock; Daniel Morf; Paul Baturin; Pascal Huber; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-06-20       Impact factor: 3.609

2.  Comparison of CsI:Tl and Gd2 O2 S:Tb indirect flat panel detector x-ray imaging performance in front- and back-irradiation geometries.

Authors:  Adrian Howansky; Anastasiia Mishchenko; A R Lubinsky; Wei Zhao
Journal:  Med Phys       Date:  2019-09-23       Impact factor: 4.071

3.  An apparatus and method for directly measuring the depth-dependent gain and spatial resolution of turbid scintillators.

Authors:  Adrian Howansky; A R Lubinsky; Katsuhiko Suzuki; S Ghose; Wei Zhao
Journal:  Med Phys       Date:  2018-10-01       Impact factor: 4.071

4.  A novel EPID design for enhanced contrast and detective quantum efficiency.

Authors:  Joerg Rottmann; Daniel Morf; Rony Fueglistaller; George Zentai; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2016-08-05       Impact factor: 3.609

5.  Evaluation of a hybrid direct-indirect active matrix flat-panel imager using Monte Carlo simulation.

Authors:  Scott Dow; Adrian Howansky; Anthony R Lubinsky; Wei Zhao
Journal:  J Med Imaging (Bellingham)       Date:  2020-05-12

6.  A Monte Carlo study of the impact of phosphor optical properties on EPID imaging performance.

Authors:  Mengying Shi; Marios Myronakis; Yue-Houng Hu; Daniel Morf; Joerg Rottmann; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-08-20       Impact factor: 3.609

7.  A novel multilayer MV imager computational model for component optimization.

Authors:  Marios Myronakis; Josh Star-Lack; Paul Baturin; Joerg Rottmann; Daniel Morf; Adam Wang; Yue-Houng Hu; Daniel Shedlock; Ross I Berbeco
Journal:  Med Phys       Date:  2017-06-28       Impact factor: 4.071

8.  A piecewise-focused high DQE detector for MV imaging.

Authors:  Josh Star-Lack; Daniel Shedlock; Dennis Swahn; Dave Humber; Adam Wang; Hayley Hirsh; George Zentai; Daren Sawkey; Isaac Kruger; Mingshan Sun; Eric Abel; Gary Virshup; Mihye Shin; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

Review 9.  Monte Carlo methods for device simulations in radiation therapy.

Authors:  Hyojun Park; Harald Paganetti; Jan Schuemann; Xun Jia; Chul Hee Min
Journal:  Phys Med Biol       Date:  2021-09-14       Impact factor: 4.174

10.  Frequency-dependent optimal weighting approach for megavoltage multilayer imagers.

Authors:  Ingrid Valencia Lozano; Mengying Shi; Marios Myronakis; Paul Baturin; Rony Fueglistaller; Pascal Huber; Mathias Lehmann; Daniel Morf; Dianne Ferguson; Matthew W Jacobson; Thomas Harris; Ross I Berbeco; Christopher L Williams
Journal:  Phys Med Biol       Date:  2021-04-16       Impact factor: 4.174

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.