A quality assurance framework for the fully automated and objective evaluation of image quality in cone-beam computed tomography.

Christian Steiding1, Daniel Kolditz1, Willi A Kalender1.   

Abstract

PURPOSE: Thousands of cone-beam computed tomography (CBCT) scanners for vascular, maxillofacial, neurological, and body imaging are in clinical use today, but there is no consensus on uniform acceptance and constancy testing for image quality (IQ) and dose yet. The authors developed a quality assurance (QA) framework for fully automated and time-efficient performance evaluation of these systems. In addition, the dependence of objective Fourier-based IQ metrics on direction and position in 3D volumes was investigated for CBCT.
METHODS: The authors designed a dedicated QA phantom 10 cm in length consisting of five compartments, each with a diameter of 10 cm, and an optional extension ring 16 cm in diameter. A homogeneous section of water-equivalent material allows measuring CT value accuracy, image noise and uniformity, and multidimensional global and local noise power spectra (NPS). For the quantitative determination of 3D high-contrast spatial resolution, the modulation transfer function (MTF) of centrally and peripherally positioned aluminum spheres was computed from edge profiles. Additional in-plane and axial resolution patterns were used to assess resolution qualitatively. The characterization of low-contrast detectability as well as CT value linearity and artifact behavior was tested by utilizing sections with soft-tissue-equivalent and metallic inserts. For an automated QA procedure, a phantom detection algorithm was implemented. All tests used in the dedicated QA program were initially verified in simulation studies and experimentally confirmed on a clinical dental CBCT system.
RESULTS: The automated IQ evaluation of volume data sets of the dental CBCT system was achieved with the proposed phantom requiring only one scan for the determination of all desired parameters. Typically, less than 5 min were needed for phantom set-up, scanning, and data analysis. Quantitative evaluation of system performance over time by comparison to previous examinations was also verified. The maximum percentage interscan variation of repeated measurements was less than 4% and 1.7% on average for all investigated quality criteria. The NPS-based image noise differed by less than 5% from the conventional standard deviation approach and spatially selective 10% MTF values were well comparable to subjective results obtained with 3D resolution pattern. Determining only transverse spatial resolution and global noise behavior in the central field of measurement turned out to be insufficient.
CONCLUSIONS: The proposed framework transfers QA routines employed in conventional CT in an advanced version to CBCT for fully automated and time-efficient evaluation of technical equipment. With the modular phantom design, a routine as well as an expert version for assessing IQ is provided. The QA program can be used for arbitrary CT units to evaluate 3D imaging characteristics automatically.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24593719     DOI: 10.1118/1.4863507

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

Review 1.  Quality assurance phantoms for cone beam computed tomography: a systematic literature review.

Authors:  Marcus V L de Oliveira; Ann Wenzel; Paulo S F Campos; Rubens Spin-Neto
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

2.  Long-term Inter-protocol kV CBCT image quality assessment for a ring-gantry linac via automated QA approach.

Authors:  Jiayuan Peng; Hua Li; Eric Laugeman; Thomas Mazur; Dao Lam; Taoran Li; Baozhou Sun; Weigang Hu; Lei Dong; Geoffrey D Hugo; Sasa Mutic; Bin Cai
Journal:  Biomed Phys Eng Express       Date:  2020-01-20

3.  Standardized method to quantify the variation in voxel value distribution in patient-simulated CBCT data sets.

Authors:  R Spin-Neto; E Gotfredsen; A Wenzel
Journal:  Dentomaxillofac Radiol       Date:  2014-10-29       Impact factor: 2.419

Review 4.  Spatial resolution in CBCT machines for dental/maxillofacial applications-what do we know today?

Authors:  D Brüllmann; R K W Schulze
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

5.  Quantitative performance characterization of image quality and radiation dose for a CS 9300 dental cone beam computed tomography machine.

Authors:  Elham Abouei; Sierra Lee; Nancy L Ford
Journal:  J Med Imaging (Bellingham)       Date:  2015-11-18

6.  Factors affecting modulation transfer function measurements in cone-beam computed tomographic images.

Authors:  Jin-Woo Choi
Journal:  Imaging Sci Dent       Date:  2019-06-24

7.  Simple computation of the approximated modulation transfer function (MTF) using spreadsheet-software: method and evaluation in five maxillofacial CBCT-devices.

Authors:  Ralf Kurt Willy Schulze; Clara I Doering
Journal:  Dentomaxillofac Radiol       Date:  2019-02-28       Impact factor: 2.419

8.  Application of a newly developed software program for image quality assessment in cone-beam computed tomography.

Authors:  Marcus Vinicius Linhares de Oliveira; António Carvalho Santos; Graciano Paulo; Paulo Sergio Flores Campos; Joana Santos
Journal:  Imaging Sci Dent       Date:  2017-06-22

9.  Quantitative analysis of image quality for acceptance and commissioning of an MRI simulator with a semiautomatic method.

Authors:  Xinyuan Chen; Jianrong Dai
Journal:  J Appl Clin Med Phys       Date:  2018-03-24       Impact factor: 2.102

10.  Evaluation of metal artefacts for two CBCT devices with a new dental arch phantom.

Authors:  Luciano Augusto Cano Martins; Polyane Mazucatto Queiroz; Yuri Nejaim; Karla de Faria Vasconcelos; Francisco Carlos Groppo; Francisco Haiter-Neto
Journal:  Dentomaxillofac Radiol       Date:  2020-03-16       Impact factor: 2.419

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.