Evaluation of metal artefacts for two CBCT devices with a new dental arch phantom.

Luciano Augusto Cano Martins1, Polyane Mazucatto Queiroz1, Yuri Nejaim1, Karla de Faria Vasconcelos2, Francisco Carlos Groppo3, Francisco Haiter-Neto1.   

Abstract

OBJECTIVES: To create a new phantom design to evaluate the real impact of artefacts caused by titanium on bone structures in cone beam CT images considering different positions and quantity of metals in the dental arch, with and without metal artefact reduction (MAR).
METHODS: A three cylindrical polymethyl methacrylate (PMMA) plate phantom was designed containing eight perforations arranged to simulate the lower dental arch in the intermediate plate. Three titanium cylinders were positioned in different locations and quantities to test different clinical conditions and to quantify the impact of the metal artefact around five bone cylinders. Scans were carried out in seven different protocols (Control, A-F) in two cone beam CT devices (OP300 Maxio and Picasso Trio). Eight regions of interest around each cortical and trabecular bone were used to measure the grey value standard deviation corresponding the artefact expression in the Image J software. Both the artefact expression and the MAR effect were assessed using the Wilcoxon, Friedman (Dunn) and Kruskal-Wallis tests (significance level of 5%).
RESULTS: For both devices, MAR was statistically efficient only for the protocols E, and F. Protocol F (three metals on the adjacent area of the analysis region) showed higher artefact expression when compared to the others.
CONCLUSION: In conclusion, the new phantom design allowed the quantification of the metal artefact expression caused by titanium. The metal artefact expression is higher when more metal objects are positioned in the adjacent bone structures. MAR may not be effective to reduce artefact expression on the adjacencies of those objects for the devices studied.

Entities:  

Keywords:  Artefacts; Cone-Beam Computed Tomography; Dental Implants

Mesh:

Substances:

Year:  2020        PMID: 32155087      PMCID: PMC7333471          DOI: 10.1259/dmfr.20190385

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  29 in total

1.  Prospects and challenges of rendering tissue density in Hounsfield units for cone beam computed tomography.

Authors:  Roberto Molteni
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2013-07

2.  Assessment of linear measurements of bone for implant sites in the presence of metallic artefacts using cone beam computed tomography and multislice computed tomography.

Authors:  C C Cremonini; M Dumas; C M Pannuti; J B C Neto; M G P Cavalcanti; L A Lima
Journal:  Int J Oral Maxillofac Surg       Date:  2011-05-31       Impact factor: 2.789

3.  The performance of metal artifact reduction algorithms in cone beam computed tomography images considering the effects of materials, metal positions, and fields of view.

Authors:  Karla de Faria Vasconcelos; Marina Codari; Polyane Mazucatto Queiroz; Laura Ferreira Pinheiro Nicolielo; Deborah Queiroz Freitas; Chiarella Sforza; Reinhilde Jacobs; Francisco Haiter-Neto
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2018-09-26

4.  Quantitative evaluation of metal artifacts using different CBCT devices, high-density materials and field of views.

Authors:  Marina Codari; Karla de Faria Vasconcelos; Laura Ferreira Pinheiro Nicolielo; Francisco Haiter Neto; Reinhilde Jacobs
Journal:  Clin Oral Implants Res       Date:  2017-04-22       Impact factor: 5.977

5.  Quantification of bone quality using different cone beam computed tomography devices: Accuracy assessment for edentulous human mandibles.

Authors:  Jeroen Van Dessel; Laura Ferreira Pinheiro Nicolielo; Yan Huang; Pieter Slagmolen; Constantinus Politis; Ivo Lambrichts; Reinhilde Jacobs
Journal:  Eur J Oral Implantol       Date:  2016       Impact factor: 3.123

6.  Evaluation of metal artefact reduction in cone-beam computed tomography images of different dental materials.

Authors:  Polyane Mazucatto Queiroz; Matheus Lima Oliveira; Francisco Carlos Groppo; Francisco Haiter-Neto; Deborah Queiroz Freitas
Journal:  Clin Oral Investig       Date:  2017-05-23       Impact factor: 3.573

7.  Metal artefact reduction with cone beam CT: an in vitro study.

Authors:  B B Bechara; W S Moore; C A McMahan; M Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2012-01-12       Impact factor: 2.419

8.  Accuracy of linear measurements on CBCT images related to presurgical implant treatment planning: A systematic review.

Authors:  George Fokas; Vida M Vaughn; William C Scarfe; Michael M Bornstein
Journal:  Clin Oral Implants Res       Date:  2018-10       Impact factor: 5.977

9.  Presurgical evaluation of bony implant sites using panoramic radiography and cone beam computed tomography-influence of medical education.

Authors:  Michael Dau; Amir Edalatpour; Ralf Schulze; Bilal Al-Nawas; Abdulmonem Alshihri; Peer W Kämmerer
Journal:  Dentomaxillofac Radiol       Date:  2016-10-19       Impact factor: 2.419

Review 10.  Cone beam computed tomography in implant dentistry: recommendations for clinical use.

Authors:  Reinhilde Jacobs; Benjamin Salmon; Marina Codari; Bassam Hassan; Michael M Bornstein
Journal:  BMC Oral Health       Date:  2018-05-15       Impact factor: 2.757

View more
  6 in total

1.  Does the size of an object containing dental implant affect the expression of artifacts in cone beam computed tomography imaging?

Authors:  Mahkameh Moshfeghi; Yaser Safi; Ingrid Różyło-Kalinowska; Shiva Gandomi
Journal:  Head Face Med       Date:  2022-06-29       Impact factor: 2.246

2.  Comparison of the expression of the volumetric alteration artifact in cylindrical and triangular fields of view in two cone-beam computed tomography devices.

Authors:  Deivi Cascante-Sequeira; Fernanda Coelho-Silva; Lucas P Lopes Rosado; Deborah Queiroz Freitas; Sergio Lins de-Azevedo-Vaz; Francisco Haiter-Neto
Journal:  Clin Oral Investig       Date:  2021-07-29       Impact factor: 3.606

3.  Effect of the Field of View Size on CBCT Artifacts Caused by the Presence of Metal Objects in the Exomass.

Authors:  Yaser Safi; Mitra Ghazizadeh Ahsaie; Maede Jafarian Amiri
Journal:  Int J Dent       Date:  2022-09-09

4.  Influence of windowing and metal artefact reduction algorithms on the volumetric dimensions of five different high-density materials: a cone-beam CT study.

Authors:  Fernanda Coelho-Silva; Luciano Augusto Cano Martins; Daniela Azeredo Braga; Eliana Zandonade; Francisco Haiter-Neto; Sergio Lins de-Azevedo-Vaz
Journal:  Dentomaxillofac Radiol       Date:  2020-06-25       Impact factor: 2.419

5.  Distortion or magnification? An in vitro cone-beam CT study of dimensional changes of objects with different compositions.

Authors:  Fernanda Coelho-Silva; Hugo Gaêta-Araujo; Lucas P Lopes Rosado; Deborah Queiroz Freitas; Francisco Haiter-Neto; Sergio Lins de-Azevedo-Vaz
Journal:  Dentomaxillofac Radiol       Date:  2021-06-16       Impact factor: 2.419

6.  Cone-beam computed tomography artifacts in the presence of dental implants and associated factors: an integrative review.

Authors:  Bianca Rodrigues Terrabuio; Caroline Gomes Carvalho; Mariela Peralta-Mamani; Paulo Sérgio da Silva Santos; Izabel Regina Fischer Rubira-Bullen; Cássia Maria Fischer Rubira
Journal:  Imaging Sci Dent       Date:  2021-03-11
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.