Marcus Vinicius Linhares de Oliveira1,2, António Carvalho Santos3, Graciano Paulo4, Paulo Sergio Flores Campos2, Joana Santos4.
Abstract
PURPOSE: The purpose of this study was to apply a newly developed free software program, at low cost and with minimal time, to evaluate the quality of dental and maxillofacial cone-beam computed tomography (CBCT) images.Entities:
Keywords: Cone-Beam Computed Tomography; Image Quality; Quality Assurance; Quality Control
Year: 2017 PMID: 28680843 PMCID: PMC5489672 DOI: 10.5624/isd.2017.47.2.75
Source DB: PubMed Journal: Imaging Sci Dent ISSN: 2233-7822
Fig. 1Structure of study
Image quality parameters and structure of phantom CQP-IFBA
PMMA: polymethyl methacrylate, PTFE: polytetrafluoroethylene, PP: polypropylene, PVC: polyvinyl chloride
Fig. 2Design of the phantom CQP-IFBA. A. Seven discs and location landmarks, B. docking slot inside the regular discs
Fig. 3Rectangular ROIs (vertical and horizontal) are drawn to identify highest standard deviation on disc 3.
Methods automatically defined to select the best image per disc
Cone beam computed tomography (CBCT) phantom acquisition exposure parameters per scanner
Reg: regular time of exposure; Hi Reg: high resolution with regular time of exposure, Eco: low time of exposure, Mini: low dose, Std: standard
Fig. 4Images obtained in CBCT scanners. A. Disc 1 for geometric accuracy, B. Disc 2 with ROI's positioned to evaluate the image uniformity, C. Disc 2 with a circular ROI (Ø 30 mm) in the center, D. Disc 3 with central hole containing air, E. Disc 4 used to evaluate the SNR and CNR in centered and periphery elements, F. Disc 05 with pattern bars, G. Disc 06 to evaluate the low contrast, and H. Disc 7 with three metallic rods and subtracted PMMA background
Image analysis results for Cone beam computed tomography scanners and protocols
SNRC: signal to noise ratio in the center, CNRC: contrast to noise ratio in the center, SNRP: signal to noise ratio in the periphery, CNRP: contrast to noise ratio in the periphery, Ver: vertical, Hor: horizontal, PMMA: Polymethyl methacrylate, PTFE: polytetrafluoroethylene, PVC: Polyvinyl chloride, PP: polypropylene
Fig. 5Low contrast evaluation according to the protocol and contrast-to-noise ratio (CNR) in seven rods with diverse diameters
Fig. 6Spatial resolution of CBCT units: A. Cranex 3Dx_ C1, B. Cranex 3Dx_ C2, C. Scanora 3D_ S1, D. Scanora 3D_ S2, E. NewTom 5G_ N1, F. NewTom 5G_ N2, and G. NewTom 5G_ N3
Fig. 7Data storage system.