Abstract

PURPOSE: This study was designed to investigate the effects of voxel size, the oversampling technique, and the direction and area of measurement on modulation transfer function (MTF) values to identify the optimal method of MTF measurement.
MATERIALS AND METHODS: Images of the wire inserts of the SedentexCT IQ phantom were acquired, and MTF values were calculated under different conditions (voxel size of 0.1, 0.2, and 0.3 mm; 5 oversampling techniques; simulated pixel location errors; and different directions and areas of measurement). The differences in the MTF values across various conditions were evaluated.
RESULTS: The MTF 10 values showed smaller standard deviations than the MTF 50 values. Stable and accurate MTF values were obtained in the 0.1-mm voxel images. In the 0.3-mm voxel images, oversampling techniques of 11 lines or more did not show significant differences in MTF values depending on the presence of simulated location errors. MTF 10 values showed significant differences according to the direction and area of the measurement.
CONCLUSION: To measure more accurate and stable MTF values, it is better to measure MTF 10 values in small-voxel images. In large-voxel images, the proper oversampling technique is required. MTF values from the radial and tangential directions may be different, and MTF values vary depending on the measured area.

Entities:  

Keywords:  Cone-Beam Computed Tomography; Quality Control; Radiography, Dental

Year:  2019        PMID: 31281790      PMCID: PMC6597368          DOI: 10.5624/isd.2019.49.2.131

Source DB:  PubMed          Journal:  Imaging Sci Dent        ISSN: 2233-7822


  20 in total

1.  Characteristics of a newly developed dentomaxillofacial X-ray cone beam CT scanner (CB MercuRay): system configuration and physical properties.

Authors:  K Araki; K Maki; K Seki; K Sakamaki; Y Harata; R Sakaino; T Okano; K Seo
Journal:  Dentomaxillofac Radiol       Date:  2004-01       Impact factor: 2.419

2.  Managing patient dose in digital radiology. A report of the International Commission on Radiological Protection.

Authors: 
Journal:  Ann ICRP       Date:  2004

3.  Comparison of spatial and contrast resolution for cone-beam computed tomography scanners.

Authors:  Ruben Pauwels; Jilke Beinsberger; Harry Stamatakis; Kostas Tsiklakis; Adrian Walker; Hilde Bosmans; Ria Bogaerts; Reinhilde Jacobs; Keith Horner
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2012-05-15

4.  Evaluation of the spatial resolution characteristics of a cone-beam breast CT scanner.

Authors:  Alexander L C Kwan; John M Boone; Kai Yang; Shih-Ying Huang
Journal:  Med Phys       Date:  2007-01       Impact factor: 4.071

5.  Influence of voxel size in the diagnostic ability of cone beam tomography to evaluate simulated external root resorption.

Authors:  Gabriela Salatino Liedke; Heloísa Emília Dias da Silveira; Heraldo Luis Dias da Silveira; Vinícius Dutra; José Antônio Poli de Figueiredo
Journal:  J Endod       Date:  2009-02       Impact factor: 4.171

6.  Image quality assessment of three cone beam CT machines using the SEDENTEXCT CT phantom.

Authors:  J Bamba; K Araki; A Endo; T Okano
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

7.  Modulation transfer function evaluation of cone beam computed tomography for dental use with the oversampling method.

Authors:  H Watanabe; E Honda; T Kurabayashi
Journal:  Dentomaxillofac Radiol       Date:  2010-01       Impact factor: 2.419

8.  Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners.

Authors:  A Suomalainen; T Kiljunen; Y Käser; J Peltola; M Kortesniemi
Journal:  Dentomaxillofac Radiol       Date:  2009-09       Impact factor: 2.419

9.  A comparative study for spatial resolution and subjective image characteristics of a multi-slice CT and a cone-beam CT for dental use.

Authors:  Hiroshi Watanabe; Eiichi Honda; Akemi Tetsumura; Tohru Kurabayashi
Journal:  Eur J Radiol       Date:  2009-10-09       Impact factor: 3.528

10.  Detection of vertical root fractures by using cone beam computed tomography with variable voxel sizes in an in vitro model.

Authors:  Senem Yiğit Özer
Journal:  J Endod       Date:  2010-11-05       Impact factor: 4.171

View more
  1 in total

1.  How image-processing parameters can influence the assessment of dental materials using micro-CT.

Authors:  Fernanda Ferrari Esteves Torres; Reinhilde Jacobs; Mostafa EzEldeen; Karla de Faria-Vasconcelos; Juliane Maria Guerreiro-Tanomaru; Bernardo Camargo Dos Santos; Mário Tanomaru-Filho
Journal:  Imaging Sci Dent       Date:  2020-06-18
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.