MR scanner systems should be adequately characterized in diffusion-MRI of the breast.

Marco Giannelli1, Roberto Sghedoni2, Chiara Iacconi3, Mauro Iori2, Antonio Claudio Traino1, Maria Guerrisi4, Mario Mascalchi5, Nicola Toschi6, Stefano Diciotti7.   

Abstract

Breast imaging represents a relatively recent and promising field of application of quantitative diffusion-MRI techniques. In view of the importance of guaranteeing and assessing its reliability in clinical as well as research settings, the aim of this study was to specifically characterize how the main MR scanner system-related factors affect quantitative measurements in diffusion-MRI of the breast. In particular, phantom acquisitions were performed on three 1.5 T MR scanner systems by different manufacturers, all equipped with a dedicated multi-channel breast coil as well as acquisition sequences for diffusion-MRI of the breast. We assessed the accuracy, inter-scan and inter-scanner reproducibility of the mean apparent diffusion coefficient measured along the main orthogonal directions (<ADC>) as well as of diffusion-tensor imaging (DTI)-derived mean diffusivity (MD) measurements. Additionally, we estimated spatial non-uniformity of <ADC> (NU<ADC>) and MD (NUMD) maps. We showed that the signal-to-noise ratio as well as overall calibration of high strength diffusion gradients system in typical acquisition sequences for diffusion-MRI of the breast varied across MR scanner systems, introducing systematic bias in the measurements of diffusion indices. While <ADC> and MD values were not appreciably different from each other, they substantially varied across MR scanner systems. The mean of the accuracies of measured <ADC> and MD was in the range [-2.3%,11.9%], and the mean of the coefficients of variation for <ADC> and MD measurements across MR scanner systems was 6.8%. The coefficient of variation for repeated measurements of both <ADC> and MD was < 1%, while NU<ADC> and NUMD values were <4%. Our results highlight that MR scanner system-related factors can substantially affect quantitative diffusion-MRI of the breast. Therefore, a specific quality control program for assessing and monitoring the performance of MR scanner systems for diffusion-MRI of the breast is highly recommended at every site, especially in multicenter and longitudinal studies.

Entities:  

Mesh:

Year:  2014        PMID: 24489711      PMCID: PMC3904912          DOI: 10.1371/journal.pone.0086280

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

In magnetic resonance imaging (MRI), “diffusion” (i.e. the random, thermally-induced displacements of water molecules over time) [1] represents an extraordinarily sensitive contrast mechanism, and the exquisite structural detail it affords has proven useful in a vast number of clinical as well as research applications, especially in neuroimaging [2]. Currently, diffusion-MRI is a promising and potentially useful MRI technique for improving the diagnostic accuracy of breast imaging without administering contrast agents [3]–[8]. Indeed, previous studies have shown a potential role of quantitative diffusion-MRI in differentiating between benign and malignant breast lesions [9]–[12], with the majority of malignant lesions showing reduced diffusion when compared to benign lesions, and diffusion-MRI may aid in identifying patients with low grade ductal carcinoma in situ (DCIS) as compared to high grade DCIS, hence contributing to risk-stratification in DCIS [13], [14]. Some studies have revealed an inverse correlation between the cellularity of breast cancer and diffusion indices [15], [16], and diffusion indices have been seen to vary significantly according to various histopathological and immunohistochemical tumour features [17]. In locally advanced breast cancer, another potential application of diffusion-MRI is in the evaluation and assessment of the early response of cancer to neoadjuvant chemotherapy [18]–[20]. Previous studies [21], [22] have reported a detectable increase of diffusion which manifested itself before quantifiable decrease in tumour size, and the diffusion change was observed as early as right upon completion of the first cycle of neoadjuvant chemotherapy. Moreover, preliminary studies have suggested that diffusion may be used as a biomarker for pre-treatment prediction of response to neoadjuvant chemotherapy in patients with locally advanced breast cancer [5], [23], although this hypothesis needs further validation [24]. Diffusion-MRI has also shown potential for evaluating residual breast cancer following neoadjuvant chemotherapy [25]. Several factors, both in data acquisition and processing, can influence the accuracy and precision of quantitative diffusion-MRI measurements [26]–[31]. In particular, it should be noted that the signal-to-noise ratio (SNR) as well as the overall degree of calibration of the high strength diffusion gradients system (which are intrinsically linked to all stages of the diffusion-MRI pipeline, from sequence design through data analysis) can directly and systematically bias the measurement of diffusion indices [32]. Accordingly, some studies have emphasized the importance of implementing specific diffusion-MRI related quality control protocols as well as correction methods [33]–[44], which should be put into practice in addition to standard quality assurance routines in order to guarantee the reliability of quantitative diffusion-MRI measurements. Furthermore, in diffusion-MRI studies, a time- and site-dependency of MR scanner system performance can introduce bias in diffusion-MRI measurements, increase the variance of measured diffusion indices and substantially reduce the power of statistical inference for detecting group differences [30]. In this context, a number of in vivo studies have analyzed intra-scanner variability of diffusion-MRI measurements of the brain [27], [28], [45]–[53]. Moreover, given that the integration of multicenter data would greatly improve the sensitivity of diffusion-MRI studies, recent clinical investigations have specifically evaluated the inter-scanner reproducibility of measurements of different diffusion-tensor imaging (DTI)-derived indices in the human brain [54]–[60]. In diffusion-MRI of the body, some in vivo studies have evaluated the inter-scan reproducibility of measurements of diffusion indices of the abdomen [61]–[64], liver [65]–[67], prostate [68], anal canal [69] and kidney [70]. However, so far, only a few clinical studies [71]–[74] have specifically investigated the reliability of diffusion-MRI measurements in the breast in terms of inter-scan reproducibility as well as intra- and inter-observer reproducibility. In view of the fact that breast imaging represents a relatively recent field of application of quantitative diffusion-MRI techniques, and based on the importance of guaranteeing and assessing its reliability in clinical as well as research investigations, the aim of this study was to specifically characterize how the main MR scanner system-related factors affect quantitative measurements in diffusion-MRI of the breast. In particular, we evaluated the accuracy, inter-scan and inter-scanner reproducibility of measurements of phantom diffusion indices performed on 1.5 T MR scanner systems by different manufacturers, all equipped with a dedicated multi-channel breast coil as well as acquisition sequences for diffusion-MRI of the breast.

Materials and Methods

2.1. MR scanner systems and phantom

All diffusion-MRI acquisitions were performed on three commercial 1.5 T MR scanner systems by three different manufacturers, operating in three distinct centers: scanner-A [GE Signa HDx TwinSpeed (GE Medical Systems - Milwaukee, WI, USA) with 50 mT/m maximum gradient strength and 150 T/m/s slew rate], scanner-B [Philips Achieva (Philips Medical Systems - Eindhoven, the Nederlands) with 66 mT/m maximum gradient strength and 90 T/m/s slew rate] and scanner-C [Siemens Avanto (Siemens Healthcare - Erlangen, Germany) with 45 mT/m maximum gradient strength and 200 T/m/s slew rate]. All MR scanner systems were equipped with a dedicated multi-channel breast coil with 8, 7 and 4 elements for scanner-A, scanner-B and scanner-C, respectively. For each MR scanner system, standard maintenance and quality assurance procedures were routinely performed. The same doped (per 1000 g H2O distilled: 1.25 g NiSO4 × 6H2O + 5 g NaCl) isotropic water phantom (i.e. two identical cylindrical bottles with diameter 11.5 cm and length 20 cm) was employed in all acquisitions.

2.2. Data acquisition

Images from different MR scanner systems were obtained using pulse sequences provided by the manufacturers. For diffusion-weighted image acquisition, we used a 2D axial spin echo - echo planar imaging sequence, sensitized to diffusion (DWI-SE-EPI) through strong magnetic field gradient pulses. The acquisition protocols and parameters are reported in Table 1.
Table 1

Axial 2D diffusion-weighted spin echo - echo planar imaging (DWI-SE-EPI) sequence: acquisition parameters for scanner-A, scanner-B and scanner-C.

Scanner-AScanner-BScanner-C
TR (ms)800080008000
TE (ms)818183
BW (Hz/pixel)390629052170
b-value (s/mm2)850850850
FOV (mm×mm)350×350350×350350×350
Matrix128×128128×128128×128
Slice thickness (mm)555
Interslice gap (mm)000
Number of slices212121
K space sampling5/85/85/8
Parallel imagingASSETSENSEmSENSE
Acceleration factor222
Phase encoding directionanterior/posterioranterior/posterioranterior/posterior
For each MR scanner system, all acquisitions were performed on the same day in order to avoid any mid- and long-term changes in scanner performance as well as any potential variability induced by phantom repositioning. The phantom (i.e. two cylindrical bottles as described above) was stored in the scanner room for at least 24 hours prior to scanning and was positioned in the gantry 1 hour before acquisition. The centre of each of the two cylindrical bottles was placed in the centre of each of the two sides of the breast coil and secured using foam padding. The central slice of the acquisition slab (21 slices) was placed at the centre of the two bottles (Figure 1). The temperature of the scanner bore was recorded during data acquisition.
Figure 1

Schematic description of a) breast coil, phantom and acquisition slab (21 slices), as well as b) ROIref positioning.

2.2.1. SNR and calibration of diffusion gradients system

In order to evaluate SNR as well as the calibration of the high strength diffusion gradients system in each MR scanner, the axial DWI-SE-EPI sequence (Table 1) was acquired both without (b-value = 0 s/mm2) and with (b-value = 850 s/mm2) sensitization to diffusion along each of the main orthogonal directions (readout/left-right, phase-encoding/anterior-posterior, slice-selection/head-foot). In order to improve SNR, we employed a number of excitations (NEX) equal to 14. The above acquisitions were repeated obtaining a total of 5 measurements.

2.2.2. Accuracy, inter-scan and inter-scanner reproducibility of diffusion-MRI measurements

In order to assess the accuracy, inter-scan and inter-scanner reproducibility of measurements of conventional mean apparent diffusion coefficient (ADC) measured along the main orthogonal directions () as well as DTI-derived indices (which a few breast studies have preliminarily incorporated in their scanning protocols [72], [73], [75]–[77]) additional diffusion-weighted images along each of the main orthogonal directions and DTI data sets were acquired using the DWI-SE-EPI sequence (Table 1) with b-value = 850 s/mm2. For each MR scanner system, DTI acquisitions of the DWI-SE-EPI sequence with sensitization to diffusion along 6 non-collinear and non-coplanar directions [72], [75], [76] were performed. For as well as DTI measurements, an additional acquisition of the DWI-SE-EPI sequence without sensitization to diffusion (b-value = 0 s/mm2) was carried out. In order to guarantee a constant acquisition time, the NEX value was 7 and 4 for and DTI data sets, respectively. The entire set of acquisitions was repeated obtaining a total of 5 measurements for both and DTI data.

2.3. Image processing and analysis

Except for diffusion tensor estimation, all processing and analysis of diffusion-MRI data was performed using custom scripts developed in MATLAB 7.1 (MathWorks, Natick, MA, USA). In order to better evaluate inter-scanner variability of diffusion-MRI measurements independently of temperature (Ta) during data acquisition, ADC measured along each of the main orthogonal directions, and DTI-derived mean diffusivity (MD) values were normalized to a reference value corresponding to a temperature of 22°C (at which the phantom diffusion coefficient, D0, is equal to 2.14±0.03×10−3 mm2/s) [78]. In particular, for each MR scanner system, we used an analytical equation derived by fitting experimental water diffusion coefficients measured at different temperature values with the Arrhenius activation law to obtain the true phantom diffusion coefficient at Ta (Da) [78]. Given that for an isotropic phantom the ratio (R) between the value of a diffusion index measured at Ta and Da depends exclusively on the ratios (Rb i = 1-n, n = 1, 3 and 6 for ADC measured along each of the main orthogonal directions, and MD, respectively) between the nominal and the effective b-value applied along the diffusion sensitized directions (which can reasonably be considered independent of temperature), the normalized values of diffusion indices were calculated as RD0. All analyses were carried out in the central slice of the acquired phantom volume, within a reference region of interest (ROIref) that consisted of two rectangles (size 29×41 voxels), each placed in the centre of the image of the bottle on each of two sides of the breast coil (Figure 1).

2.3.1. SNR and maps of ADC along each of the main orthogonal directions

The SNR was calculated using non-diffusion-weighted (b0) images. Conventional approaches to evaluate SNR are based on the signal statistics in one or two separate large regions of interest of a single image or the signal statistics in a large region of interest of a difference image of two repeated acquisitions [79], [80]. In order to take into account spatial variations in SNR (which can be substantial in acquisitions performed using multi-channel coils and parallel imaging techniques) [81], maps of SNR in small adjacent ROIs of 8×8 voxels (SNRROI) were computed as previously described [81], [82]:where Sb0(r,k) is the signal of the voxel at position r within the selected ROI for the kth repetition of the b0 image. The overall SNR was computed as the mean value of SNRROI within ROIref. For each repetition (k = 1–5), ADC maps along each of the main orthogonal directions [ADCi,k(r) - i = 1, readout/left-right; i = 2, phase-encoding/anterior-posterior; i = 3, slice-selection/head-foot] were computed. For the ith direction, the mean, (ADCi)mean, and standard deviation, (ADCi)SD, images across repetitions were calculated. Then, the overall ADC along the ith diffusion weighting direction (ADCi) was obtained as the average (ADCi)mean within ROIref. Furthermore, the overall percent coefficient of variation for repeated measurements of ADC along the ith diffusion weighting direction was computed as follows: The spatial non-uniformity levels of maps of ADC along each of the main orthogonal directions were evaluated by adapting a method proposed by Magnusson and Olsson [83]. The ADCi,k maps were smoothed using a low-pass spatial filter with a 3×3 kernel which reduces noise by computing the mean value of a voxel and its 8 neighbours, and replacing the value of the voxel with this mean. Then, the mean value (C) within ROIref was estimated. For each voxel, the deviation from this value was calculated as the absolute value of [100×(voxel value - C)/C], obtaining a new image which represents the absolute value of the percentage deviation from C. The mean value of this new image within ROIref was recorded, obtaining the non-uniformity value of ADCi,k maps () for each diffusion weighting direction (i = 1–3) and repetition (k = 1–5). Finally, for each diffusion weighting direction, the overall non-uniformity degree () was estimated as the mean of across repetitions.

2.3.2. Maps of , MD and FA

For each repetition (k = 1–5), the mean ADC along the main orthogonal directions [k(r)] was calculated voxel-wise. The overall mean ADC () and its coefficient of variation for repeated measurements (CV) were calculated as described above for ADCi and , respectively. Moreover, the overall spatial non-uniformity degree of maps (NU) was estimated using the same method employed for calculating . In order to estimate the diffusion tensor, we adopted a method similar to that described in previous breast DTI studies [72], [76]. In particular, we performed the standard steps implemented in the diffusion toolbox (FDT) of FSL 4.1.4 (Oxford Centre for Functional Magnetic Resonance Imaging of the Brain (FMRIB) software library) [84], [85] using the weighted linear least square approach. For each repetition (k = 1–5), the mean diffusivity [MDk(r)] and fractional anisotropy [FAk(r)] were computed voxel-wise. Then, the overall mean diffusivity (MD) and fractional anisotropy (FA) were calculated as described for ADCi. The coefficient of variation for repeated measurements of MD (CVMD) and spatial non-uniformity of MD maps (NUMD) were then obtained using the same procedure adopted for and calculation.

2.3.3. Statistical analysis

Any significant difference in quality control data and measured diffusion metrics, both across the main orthogonal directions within a single MR scanner system and across MR scanner systems, was assessed through a one-way analysis of variance (ANOVA). When the ANOVA revealed a significant difference (p<0.05), a post-hoc analysis was performed using the two sample t-test, with Bonferroni correction for multiple comparisons. For each MR scanner system, any significant difference between and MD maps was assessed similarly. The one-sample t-test, with Bonferroni correction for multiple comparisons, was used to evaluate any significant difference between the true diffusion indices and estimated diffusion indices.

Results

3.1. SNR and calibration of diffusion gradients system

The SNR (mean value ± standard deviation within ROIref) was 242±55, 184±26 and 309±60 for scanner-A, scanner-B and scanner-C, respectively. The SNR values varied significantly across MR scanner systems (ANOVA: p<0.0001 – post-hoc analysis: p<0.001 for scanner-A vs scanner-B, scanner-A vs scanner-C and scanner-B vs scanner-C). For each of the main orthogonal directions (i = 1–3), the ADCi, and results are reported in Figure 2, Figure 3 and Figure 4, respectively. ADCi (i = 1–3) values varied significantly with diffusion gradient direction for scanner-A (ANOVA: p<0.0001 – post-hoc analysis: p<0.001 for i = 1 vs i = 2, i = 1 vs i = 3, i = 2 vs i = 3), scanner-B (ANOVA: p<0.0001 – post-hoc analysis: p<0.001 for i = 1 vs i = 2, i = 1 vs i = 3, i = 2 vs i = 3) and scanner-C (ANOVA: p<0.0001 – post-hoc analysis: p<0.01 and p<0.001 for i = 1 vs i = 2 and i = 1 vs i = 3, i = 2 vs i = 3, respectively). Moreover, except for ADC1 for scanner-C (p>0.05), all measured ADCi values were significantly (p<0.05) different from the known phantom diffusion coefficient. For each MR scanner system, and values varied significantly with diffusion gradient direction (ANOVA: p<0.0001 – post-hoc analysis: p<0.001 for i = 1 vs i = 2, i = 1 vs i = 3, i = 2 vs i = 3).
Figure 2

Phantom ADC along each of the main orthogonal directions (ADCi - i = 1, readout/left-right; i = 2, phase-encoding/anterior-posterior; i = 3, slice-selection/head-foot) for scanner-A, scanner-B and scanner-C.

The bar charts depict the mean of the average value within ROIref ± standard deviation across five repetitions. The dashed line represents the known phantom diffusion coefficient (2.14±0.03×10−3 mm2/s) at the reference temperature value of 22°C.

Figure 3

Coefficient of variation for repeated measurements of phantom ADC along each of the main orthogonal directions ( - i = 1, readout/left-right; i = 2, phase-encoding/anterior-posterior; i = 3, slice-selection/head-foot) for scanner-A, scanner-B and scanner-C.

The bar charts depict the mean value ± standard deviation within ROIref.

Figure 4

Non-uniformity of maps of phantom ADC along each of the main orthogonal directions ( - i = 1, readout/left-right; i = 2, phase-encoding/anterior-posterior; i = 3, slice-selection/head-foot) for scanner-A, scanner-B and scanner-C.

The bar charts depict the mean value ± standard deviation across five repetitions.

Phantom ADC along each of the main orthogonal directions (ADCi - i = 1, readout/left-right; i = 2, phase-encoding/anterior-posterior; i = 3, slice-selection/head-foot) for scanner-A, scanner-B and scanner-C.

The bar charts depict the mean of the average value within ROIref ± standard deviation across five repetitions. The dashed line represents the known phantom diffusion coefficient (2.14±0.03×10−3 mm2/s) at the reference temperature value of 22°C.

Coefficient of variation for repeated measurements of phantom ADC along each of the main orthogonal directions ( - i = 1, readout/left-right; i = 2, phase-encoding/anterior-posterior; i = 3, slice-selection/head-foot) for scanner-A, scanner-B and scanner-C.

The bar charts depict the mean value ± standard deviation within ROIref.

Non-uniformity of maps of phantom ADC along each of the main orthogonal directions ( - i = 1, readout/left-right; i = 2, phase-encoding/anterior-posterior; i = 3, slice-selection/head-foot) for scanner-A, scanner-B and scanner-C.

The bar charts depict the mean value ± standard deviation across five repetitions.

3.2. Accuracy, inter-scan and inter-scanner reproducibility of diffusion-MRI measurements

Both and MD data are reported in Figure 5. Within each MR scanner system, and MD values were not significantly different (p>0.05) from each other. On the other hand, and MD values varied significantly across MR scanner systems (ANOVA: p<0.0001 – post-hoc analysis: p<0.001 for scanner-A vs scanner-B, scanner-A vs scanner-C and scanner-B vs scanner-C). Both and MD values were significantly (p<0.01) different from their true value, with an accuracy (percentage difference between the measured and known diffusion value) ranging from -2.6% to 12.0%.
Figure 5

Mean ADC along the main orthogonal directions () as well as mean diffusivity (MD) of the phantom for scanner-A, scanner-B and scanner-C.

The bar charts depict the mean of the average value within ROIref ± standard deviation across five repetitions. The dashed line represents the known phantom diffusion coefficient (2.14±0.03×10−3 mm2/s) at the reference temperature value of 22°C.

Mean ADC along the main orthogonal directions () as well as mean diffusivity (MD) of the phantom for scanner-A, scanner-B and scanner-C.

The bar charts depict the mean of the average value within ROIref ± standard deviation across five repetitions. The dashed line represents the known phantom diffusion coefficient (2.14±0.03×10−3 mm2/s) at the reference temperature value of 22°C. The CV and CVMD results as well as NU and NUMD results are reported in Table 2. For each MR scanner system, CV and CVMD differed significantly (p<0.001) with values which were less than 1%. Moreover, CV and CVMD varied significantly across MR scanner systems (ANOVA: p<0.0001 – post-hoc analysis: p<0.001 for scanner-A vs scanner-B, scanner-A vs scanner-C and scanner-B vs scanner-C). For scanner-A and scanner-B, NU and NUMD were not significantly (p>0.05) different, whereas, for scanner-C, NU was significantly (p<0.01) lower than NUMD. Furthermore, both NU (ANOVA: p<0.0001 – post-hoc analysis: p<0.001 and p>0.05 for scanner-A vs scanner-B, scanner-A vs scanner-C and scanner-B vs scanner-C, respectively) and NUMD (ANOVA: p<0.0001 – post-hoc analysis: p<0.001 for scanner-A vs scanner-B, scanner-A vs scanner-C and scanner-B vs scanner-C) varied significantly across MR scanner systems, with values which were less than 4%.
Table 2

CV and CVMD results [mean value (standard deviation within ROIref)] as well as NU and NUMD results [mean value (standard deviation across five repetitions)] for scanner-A, scanner-B and scanner-C.

Scanner-AScanner-BScanner-C
CV<ADC> (%)0.69 (0.34)0.64 (0.31)0.81 (0.38)
CVMD (%)0.53 (0.28)0.67 (0.32)0.92 (0.58)
NU<ADC> (%)3.70 (0.04)1.70 (0.08)1.83 (0.14)
NUMD (%)3.65 (0.09)1.67 (0.08)2.52 (0.09)
For each MR scanner system, the overall FA of the phantom was significantly (p<0.001) greater than 0. Moreover, FA values varied significantly across MR scanner systems (ANOVA: p<0.0001 – post-hoc analysis: p<0.001 for scanner-A vs scanner-B, scanner-A vs scanner-C and scanner-B vs scanner-C). In particular, FA values (mean of the average value within ROIref ± standard deviation across five repetitions) were 0.086±0.001, 0.050±0.001 and 0.076±0.001 for scanner-A, scanner-B and scanner-C, respectively.

Discussion

A number of in vivo studies have evaluated the reliability of diffusion-MRI measurements in the brain as well as body [27], [28], [45]–[70], [86]. However, a more specific and careful evaluation of the reliability of diffusion-MRI measurements of the breast would be of practical interest. Recently, O'Flynn et al. [71] have assessed the mid-term reproducibility and inter-observer variability of ADC measurements of fibroglandular tissue at 3 T, obtaining a within-subject coefficient of variation of 22–25% and a kappa value (κ) of 0.83. Partridge et al. [72] have evaluated the reproducibility of DTI-derived parameter measurements in normal breast tissue at 1.5 T after repositioning and rescanning, reporting a between-scan coefficient of variation of 4.5% and 11.4% for MD and FA, respectively. Tagliafico et al. [73] have reported a between-scan coefficient of variation of 15% and 30% for MD and FA measurements in normal breast tissue at 3 T, respectively; moreover, when looking at intra-/inter-observer variability, the κ values were 0.82–0.89/0.73–0.83 and 0.60–0.84/0.64–0.80 for MD and FA, respectively. Additionally, when measuring ADC at 1.5 T in breast carcinomas, Petralia et al. [74] estimated an intra- and inter-observer variability of 1.1% and 2%, respectively. It should be noted that assessing and guaranteeing reliability of quantitative diffusion-MRI measurements, which is a prerequisite for successful clinical as well as research studies, necessarily includes a characterization of the MR scanner system. Indeed, although in vivo studies can evaluate repeatability and reproducibility of diffusion-MRI measurements in a clinical setting (which are fundamental elements toward carrying out longitudinal as well as multicenter studies), such studies do not allow to address measurement accuracy as well as some of the main characteristics of MR scanner system (e.g. the SNR or the calibration of high strength diffusion gradients system) that can systematically bias quantitative diffusion-MRI measurements [30], [32]. A limited number of studies have reported phantom data specific to the characterization of MR scanner systems for diffusion-MRI of the brain as well as the body [34]–[41], [43], [57], [58], [87], [88]. However, in diffusion-MRI of the breast, only a few clinical studies have incorporated a basic verification of the calibration of diffusion gradients [11], [22], [72], [89], [90]. To the best of our knowledge, this is the first phantom study which carries out multiple and specific quality controls in order to characterize in detail different 1.5 T MR scanner systems by three different manufacturers, all equipped with a dedicated multi-channel breast coil as well as acquisition sequence for quantitative diffusion-MRI of the breast. In particular, for each MR scanner system, we evaluated the calibration of high strength diffusion gradients for the three main orthogonal axes along which diffusion-sensitizing gradients can be applied. Then, we assessed how the MR scanner system-related factors affect the accuracy, inter-scan and inter-scanner reproducibility of diffusion measurements of as well as of DTI measurements of MD. We used acquisition protocols and parameters typically employed in diffusion-MRI of the breast, which, except for small differences in readout bandwidth (BW) values, were similar for all MR scanner systems. As suggested by Bogner et al. [89], we employed a b-value of 850 s/mm2. For all acquisitions, we used the same homogeneous and isotropic phantom with known diffusion coefficient, allowing a proper evaluation of the accuracy of estimated diffusion indices as well as non-uniformity of maps of diffusion indices. The diffusion coefficient of the phantom at room temperature (∼2×10−3 mm2/s) is similar to water diffusion coefficient in normal breast tissue (1.8–2.1×10−3 mm2/s), while it is slightly higher than water diffusion coefficient in malign as well benign breast tissue (0.9–1.7×10−3 mm2/s) [7], [22], [72], [89], [91]. In this context, as previously described by Delakis et al. [35], the use of a phantom with a relatively high diffusion coefficient is recommended in order to improve the sensitivity to any discrepancies in measured diffusion indices induced by differences between the nominal and the effective b-value applied along the diffusion sensitized directions. Diffusion-MRI measurements are affected by an inherently low SNR. In particular, both precision and accuracy of diffusion indices can depend on SNR [27]–[29], [82], [92]–[94]. Therefore, we began by characterizing each MR scanner system in terms of SNR. In particular, the overall SNR of scanner-B was 24% and 40% lower than that of scanner-A and scanner-C, respectively. Interestingly, based on the BW values of the acquisition protocols (Table 1), scanner-A (highest BW value) was expected to show the lowest SNR across MR scanner systems. Therefore, SNR results cannot be ascribed to differences in BW values only, and are likely to also reflect different overall sensitivities of the breast coils. All MR scanner systems showed a high short term stability of the performance of diffusion gradients. For each MR scanner system, the overall coefficient of variation for repeated measurements of ADC along each of the main orthogonal directions was less than 1.1% (Figure 3). Nonetheless, for each MR scanner system and direction (readout/left-right, phase-encoding/anterior-posterior, slice-selection/head-foot) except for ADC measurements along the readout/left-right direction for scanner-C, we revealed a significant difference between the measured ADC and the true diffusion coefficient. Moreover, for each MR scanner system, the entity of this difference varied significantly with diffusion direction (Figure 2). This effect, when quantified in terms of the coefficient of variation of ADC measurements across the main orthogonal directions, was more relevant for scanner-A (7.9%) as compared to scanner-B (1.7%) and scanner-C (2.7%). As a whole, these results indicate a mismatch between the theoretically assumed and the effective b-value. This could originate from errors in diffusion gradients amplitude, eddy current fields, concomitant field terms and cross terms between diffusion gradients and imaging gradients [2], [32], [87], [88], [95]. These factors are direction-dependent and can have deleterious effects that are more prominent at the high gradient strengths usually employed in diffusion-MRI [2], [32]. In addition, any diffusion gradient non-uniformity is expected to yield a spatial variation in measured diffusion indices. For each MR scanner system, we observed that the spatial non-uniformity values of maps of ADC along each of the main orthogonal directions depended significantly on the diffusion weighting direction. Scanner-A showed a relatively high spatial non-uniformity value (7.3%) of ADC along the phase-encoding/anterior-posterior direction, while for both scanner-B and scanner-C the degree of non-uniformity of ADC along each diffusion weighting direction was less than 4.5% (Figure 4). In general, when DWI-SE-EPI sequences (Table 1) are acquired, the high strength diffusion gradients system belonging to each MR scanner system presented an overall mis-calibration (not documented by standard maintenance procedures and quality assurance routines), which can affect diffusion indices measurement. Therefore, in order to improve the reliability of quantitative diffusion-MRI of the breast, suitable correction methods could be employed [34], [36], [37], [40]. For each MR scanner system, the coefficient of variation for short term repeated measurements of both and MD was less than 1% (Table 2), while previous clinical studies [71]–[73] which measured breast diffusion indices have reported a between-scan coefficient of variation in the range 5–15%. The greater experimental variability of in vivo diffusion-MRI measurements when compared to our phantom study is likely due to patient repositioning, manual ROI positioning and motion induced effects. For every MR scanner system, the spatial non-uniformities of and MD maps were less than 4% (Table 2). For scanner-A and scanner-B, non-uniformities of and MD maps were not significantly different. Conversely, for scanner-C, the non-uniformity of the map (1.8%) was significantly lower than the non-uniformity of the MD map (2.5%). For each MR scanner system, we did not reveal any significant difference between estimated and MD values, with an absolute percentage difference between and MD of less than 0.6%. This indicates a correct pulse timing when using multiple oblique diffusion gradients as employed in DTI, and may suggest the negligibility of cross-term effects between diffusion and imaging gradients along different directions [26], [95], [96]. However, the accuracy of and MD measurements varied significantly with the MR scanner system (Figure 5). In particular, the mean value of and MD accuracies was 11.9%, 6.0% and -2.3% for scanner-A, scanner-B and scanner-C, respectively, while the mean value of the coefficients of variation for and MD measurements across MR scanner systems was 6.8%. Previous phantom studies of diffusion-MRI both using a head coil [35], [36], [38]–[40], [43], [58], [87] and a body coil [34], [43] have reported accuracy values of estimated diffusion indices ranging from −15% to 30%. Other in vivo studies of the brain [45], [46], [54]–[59] have reported a coefficient of variation in MD and FA across different MR scanner systems in the ranges 4–15% and 5–29%, respectively. In this context, it is important to note that the differences in diffusion indices reported in previous clinical diffusion-MRI studies of the breast range from 5% to 45% [9]–[13], [17], [21]–[23], [25], [76]. Therefore, a comparison of breast diffusion-MRI data from different centers should be performed with great caution. Moreover, during the planning of a multicenter study, the accuracy of diffusion-MRI measurements should be carefully assessed in every participating center. Additionally, in longitudinal studies, a periodic monitoring of the accuracy of measured diffusion indices is highly recommended. In a meta-analysis of 13 studies dealing with quantitative diffusion-MRI in the differential diagnosis of breast lesions, Chen et al. [91] have shown that a) the ADC values of benign lesions ranged from 1×10−3 mm2/s to 1.82×10−3 mm2/s, b) the cutoff values for differentiating malignant from benign lesions ranged from 0.9×10−3 mm2/s to 1.76×10−3 mm2/s, and that c) the sensitivity and specificity ranged from 63% to 100% and 46% to 97%, respectively. This heterogeneity could be due to differences in patient characteristics and diagnostic criteria, as well as to different diffusion-MRI acquisition and analysis methods. However, we hypothesize that potential differences in MR scanner system-related factors between different MR scanner systems, which can systematically bias accuracy and precision of diffusion-MRI measurements, may contribute to explaining the results heterogeneity reported by Chen et al. [91]. Besides DTI-based measurements of MD, we also performed diffusion anisotropy estimation, and the overall FA value of the isotropic phantom (true FA = 0) was found to be significantly greater than 0 for every MR scanner system. This could reflect effects of relatively low SNR (high SNR has been shown to reduce the brain anisotropy overestimation due to noise at a b-value typically used in clinical DTI examinations, ∼1000 s/mm2) [93], [94], as well as errors in diffusion gradients amplitude (which can result in mimicking anisotropy). While for each MR scanner system FA values were less than 0.09, they were significantly different among MR scanner systems. In particular, data acquired on scanner-B resulted in the lowest FA estimate (∼0.05).

Conclusions

Although breast imaging is an appealing and promising application field of diffusion-MRI, only few in vivo studies have recently evaluated the inter-scan reproducibility as well as intra- and inter-observer reproducibility of diffusion measurements of the breast [71]–[74]. In this phantom study, we characterized in detail three 1.5 T MR scanner systems by three different manufacturers, all equipped with a dedicated multi-channel breast coil as well as acquisition sequences for quantitative diffusion-MRI of the breast. The SNR as well as overall calibration of high strength diffusion gradients system varied substantially across MR scanner systems, introducing systematic bias in measurements of diffusion indices. We note that in vivo diffusion-MRI measurements of the breast can also depend on other non-MR scanner system-related factors – such as subject-related artifacts (e.g. motion and cardiac pulsation, physiological noise), perfusion and non-Gaussian processes – that could further increase the variability in diffusion measurements. Nonetheless, in order to improve the reliability of quantitative breast diffusion-MRI and, hence, the sensitivity of clinical studies, a specific and periodic quality control program for characterizing and monitoring the performance of breast coil and high strength diffusion gradients of MR scanner system is highly recommended at every site, especially before multicenter studies are tackled as well as during longitudinal studies. In this context, we agree with Jones [30] and De Santis et al. [41] who have recently emphasized that the quality control culture in diffusion-MRI remains limited. Therefore, we feel that in diffusion-MRI, which is a truly quantitative technique, enabling a suitable and dedicated quality assurance program at every site would represent a major step toward the effective use of every MR scanner system as a “measurement tool”, hence further improving and strengthening the capabilities of this powerful diagnostic modality. Maps of phantom ADC along each of the main orthogonal directions (  = 1, readout/left-right;  = 2, phase-encoding/anterior-posterior;  = 3, slice-selection/head-foot), calculated using the first (  = 1) of 5 repetitions (ADC In order to facilitate visual assessment, the figure depicts a zoomed region (located on one side of the breast coil) of the phantom containing one of the two rectangular ROIs (highlighted in red) which make up ROIref. (TIF) Click here for additional data file.
  92 in total

1.  Acceptance testing of magnetic resonance imaging systems: report of AAPM Nuclear Magnetic Resonance Task Group No. 6.

Authors:  J G Och; G D Clarke; W T Sobol; C W Rosen; S K Mun
Journal:  Med Phys       Date:  1992 Jan-Feb       Impact factor: 4.071

Review 2.  Comparison of diffusion tensor imaging measurements at 3.0 T versus 1.5 T with and without parallel imaging.

Authors:  Andrew L Alexander; Jee Eun Lee; Yu-Chien Wu; Aaron S Field
Journal:  Neuroimaging Clin N Am       Date:  2006-05       Impact factor: 2.264

Review 3.  Advances in functional and structural MR image analysis and implementation as FSL.

Authors:  Stephen M Smith; Mark Jenkinson; Mark W Woolrich; Christian F Beckmann; Timothy E J Behrens; Heidi Johansen-Berg; Peter R Bannister; Marilena De Luca; Ivana Drobnjak; David E Flitney; Rami K Niazy; James Saunders; John Vickers; Yongyue Zhang; Nicola De Stefano; J Michael Brady; Paul M Matthews
Journal:  Neuroimage       Date:  2004       Impact factor: 6.556

4.  The b matrix in diffusion tensor echo-planar imaging.

Authors:  J Mattiello; P J Basser; D Le Bihan
Journal:  Magn Reson Med       Date:  1997-02       Impact factor: 4.668

5.  Toward a quantitative assessment of diffusion anisotropy.

Authors:  C Pierpaoli; P J Basser
Journal:  Magn Reson Med       Date:  1996-12       Impact factor: 4.668

6.  Quality assurance methods and phantoms for magnetic resonance imaging: report of AAPM nuclear magnetic resonance Task Group No. 1.

Authors:  R R Price; L Axel; T Morgan; R Newman; W Perman; N Schneiders; M Selikson; M Wood; S R Thomas
Journal:  Med Phys       Date:  1990 Mar-Apr       Impact factor: 4.071

7.  Changes in water mobility measured by diffusion MRI predict response of metastatic breast cancer to chemotherapy.

Authors:  Rebecca J Theilmann; Rebecca Borders; Theodore P Trouard; Guowei Xia; Eric Outwater; James Ranger-Moore; Robert J Gillies; Alison Stopeck
Journal:  Neoplasia       Date:  2004 Nov-Dec       Impact factor: 5.715

8.  Diffusion-weighted imaging of malignant breast tumors: the usefulness of apparent diffusion coefficient (ADC) value and ADC map for the detection of malignant breast tumors and evaluation of cancer extension.

Authors:  Reiko Woodhams; Keiji Matsunaga; Keiichi Iwabuchi; Shinichi Kan; Hirofumi Hata; Masaru Kuranami; Masahiko Watanabe; Kazushige Hayakawa
Journal:  J Comput Assist Tomogr       Date:  2005 Sep-Oct       Impact factor: 1.826

9.  "Squashing peanuts and smashing pumpkins": how noise distorts diffusion-weighted MR data.

Authors:  Derek K Jones; Peter J Basser
Journal:  Magn Reson Med       Date:  2004-11       Impact factor: 4.668

10.  Neoadjuvant chemotherapy in breast cancer: early response prediction with quantitative MR imaging and spectroscopy.

Authors:  D J Manton; A Chaturvedi; A Hubbard; M J Lind; M Lowry; A Maraveyas; M D Pickles; D J Tozer; L W Turnbull
Journal:  Br J Cancer       Date:  2006-02-13       Impact factor: 7.640

View more
  8 in total

1.  Diffusional kurtosis imaging for differentiation of additional suspicious lesions on preoperative breast MRI of patients with known breast cancer.

Authors:  Vivian Youngjean Park; Sungheon G Kim; Eun-Kyung Kim; Hee Jung Moon; Jung Hyun Yoon; Min Jung Kim
Journal:  Magn Reson Imaging       Date:  2019-07-16       Impact factor: 2.546

2.  Multi-site harmonization of diffusion MRI data in a registration framework.

Authors:  Hengameh Mirzaalian; Lipeng Ning; Peter Savadjiev; Ofer Pasternak; Sylvain Bouix; Oleg Michailovich; Sarina Karmacharya; Gerald Grant; Christine E Marx; Rajendra A Morey; Laura A Flashman; Mark S George; Thomas W McAllister; Norberto Andaluz; Lori Shutter; Raul Coimbra; Ross D Zafonte; Mike J Coleman; Marek Kubicki; Carl-Fredrik Westin; Murray B Stein; Martha E Shenton; Yogesh Rathi
Journal:  Brain Imaging Behav       Date:  2018-02       Impact factor: 3.978

3.  Inter-site and inter-scanner diffusion MRI data harmonization.

Authors:  H Mirzaalian; L Ning; P Savadjiev; O Pasternak; S Bouix; O Michailovich; G Grant; C E Marx; R A Morey; L A Flashman; M S George; T W McAllister; N Andaluz; L Shutter; R Coimbra; R D Zafonte; M J Coleman; M Kubicki; C F Westin; M B Stein; M E Shenton; Y Rathi
Journal:  Neuroimage       Date:  2016-04-30       Impact factor: 6.556

Review 4.  Diffusion-weighted magnetic resonance imaging in cancer: Reported apparent diffusion coefficients, in-vitro and in-vivo reproducibility.

Authors:  Maysam M Jafar; Arman Parsai; Marc E Miquel
Journal:  World J Radiol       Date:  2016-01-28

5.  Apparent diffusion coefficient normalization of normal liver: Will it improve the reproducibility of diffusion-weighted imaging at different MR scanners as a new biomarker?

Authors:  Jie Zhu; Jie Zhang; Jia-Yin Gao; Jin-Ning Li; Da-Wei Yang; Min Chen; Cheng Zhou; Zheng-Han Yang
Journal:  Medicine (Baltimore)       Date:  2017-01       Impact factor: 1.889

6.  Temperature and concentration calibration of aqueous polyvinylpyrrolidone (PVP) solutions for isotropic diffusion MRI phantoms.

Authors:  Friedrich Wagner; Frederik B Laun; Tristan A Kuder; Anna Mlynarska; Florian Maier; Jonas Faust; Kerstin Demberg; Linus Lindemann; Boris Rivkin; Armin M Nagel; Mark E Ladd; Klaus Maier-Hein; Sebastian Bickelhaupt; Michael Bach
Journal:  PLoS One       Date:  2017-06-19       Impact factor: 3.240

Review 7.  Factors affecting the value of diffusion-weighted imaging for identifying breast cancer patients with pathological complete response on neoadjuvant systemic therapy: a systematic review.

Authors:  Kay J J van der Hoogt; Robert J Schipper; Gonneke A Winter-Warnars; Leon C Ter Beek; Claudette E Loo; Ritse M Mann; Regina G H Beets-Tan
Journal:  Insights Imaging       Date:  2021-12-18

8.  ADC histograms from routine DWI for longitudinal studies in cerebral small vessel disease: a field study in CADASIL.

Authors:  Bence Gunda; Raphael Porcher; Marco Duering; Jean-Pierre Guichard; Jerome Mawet; Eric Jouvent; Martin Dichgans; Hugues Chabriat
Journal:  PLoS One       Date:  2014-05-12       Impact factor: 3.240

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.