Concomitant field terms for asymmetric gradient coils: consequences for diffusion, flow, and echo-planar imaging.

C Meier1, M Zwanger, T Feiweier, D Porter.   

Abstract

As a consequence of the Maxwell equations, linear field gradients are accompanied by additional spatially dependent field components. A description of the Maxwell field terms is presented which explicitly takes into account the asymmetry of the gradient coil. It is shown both theoretically and experimentally that, in contrast to symmetric coils, an asymmetric coil generates concomitant field terms of zeroth and first order in space. Artifacts induced by concomitant fields can be much more pronounced for asymmetric coil designs than for symmetric ones. For the strong gradient amplitudes available on modern MR systems the effect of these concomitant magnetic fields on the evolution of magnetization has to be taken into consideration in a variety of NMR acquisition techniques. The formalism is used experimentally to compensate for artifacts observed in three different imaging methods: an image shift in standard echo planar imaging (EPI), an echo shift in diffusion-weighted EPI, and a phase shift in a flow quantification technique based on phase contrast images. (c) 2008 Wiley-Liss, Inc.

Mesh:

Year:  2008        PMID: 18581353     DOI: 10.1002/mrm.21615

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  29 in total

1.  Optimized Diffusion-Weighting Gradient Waveform Design (ODGD) formulation for motion compensation and concomitant gradient nulling.

Authors:  Óscar Peña-Nogales; Yuxin Zhang; Xiaoke Wang; Rodrigo de Luis-Garcia; Santiago Aja-Fernández; James H Holmes; Diego Hernando
Journal:  Magn Reson Med       Date:  2018-11-05       Impact factor: 4.668

2.  B0 concomitant field compensation for MRI systems employing asymmetric transverse gradient coils.

Authors:  Paul T Weavers; Shengzhen Tao; Joshua D Trzasko; Louis M Frigo; Yunhong Shu; Matthew A Frick; Seung-Kyun Lee; Thomas K-F Foo; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2017-06-21       Impact factor: 4.668

3.  Pushing the limits of in vivo diffusion MRI for the Human Connectome Project.

Authors:  K Setsompop; R Kimmlingen; E Eberlein; T Witzel; J Cohen-Adad; J A McNab; B Keil; M D Tisdall; P Hoecht; P Dietz; S F Cauley; V Tountcheva; V Matschl; V H Lenz; K Heberlein; A Potthast; H Thein; J Van Horn; A Toga; F Schmitt; D Lehne; B R Rosen; V Wedeen; L L Wald
Journal:  Neuroimage       Date:  2013-05-24       Impact factor: 6.556

4.  Addressing a systematic vibration artifact in diffusion-weighted MRI.

Authors:  Daniel Gallichan; Jan Scholz; Andreas Bartsch; Timothy E Behrens; Matthew D Robson; Karla L Miller
Journal:  Hum Brain Mapp       Date:  2010-02       Impact factor: 5.038

5.  High slew-rate head-only gradient for improving distortion in echo planar imaging: Preliminary experience.

Authors:  Ek T Tan; Seung-Kyun Lee; Paul T Weavers; Dominic Graziani; Joseph E Piel; Yunhong Shu; John Huston; Matt A Bernstein; Thomas K F Foo
Journal:  J Magn Reson Imaging       Date:  2016-02-26       Impact factor: 4.813

6.  Technical Note: Compact three-tesla magnetic resonance imager with high-performance gradients passes ACR image quality and acoustic noise tests.

Authors:  Paul T Weavers; Yunhong Shu; Shengzhen Tao; John Huston; Seung-Kyun Lee; Dominic Graziani; Jean-Baptiste Mathieu; Joshua D Trzasko; Thomas K-F Foo; Matt A Bernstein
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

7.  ADC measurements on the Unity MR-linac - A recommendation on behalf of the Elekta Unity MR-linac consortium.

Authors:  Ernst S Kooreman; Petra J van Houdt; Rick Keesman; Floris J Pos; Vivian W J van Pelt; Marlies E Nowee; Andreas Wetscherek; Rob H N Tijssen; Marielle E P Philippens; Daniela Thorwarth; Jihong Wang; Amita Shukla-Dave; William A Hall; Eric S Paulson; Uulke A van der Heide
Journal:  Radiother Oncol       Date:  2020-10-02       Impact factor: 6.280

8.  Multi-system repeatability and reproducibility of apparent diffusion coefficient measurement using an ice-water phantom.

Authors:  Dariya Malyarenko; Craig J Galbán; Frank J Londy; Charles R Meyer; Timothy D Johnson; Alnawaz Rehemtulla; Brian D Ross; Thomas L Chenevert
Journal:  J Magn Reson Imaging       Date:  2012-09-28       Impact factor: 4.813

9.  Optimization of MRI Gradient Coils With Explicit Peripheral Nerve Stimulation Constraints.

Authors:  Mathias Davids; Bastien Guerin; Valerie Klein; Lawrence L Wald
Journal:  IEEE Trans Med Imaging       Date:  2020-12-29       Impact factor: 10.048

10.  The effects of concomitant gradients on chemical shift encoded MRI.

Authors:  Timothy J Colgan; Diego Hernando; Samir D Sharma; Scott B Reeder
Journal:  Magn Reson Med       Date:  2016-09-21       Impact factor: 4.668