| Literature DB >> 24484613 |
Keith A Gillis, Christie McComb, John E Foster, Alison H M Taylor, Rajan K Patel, Scott T W Morris, Alan G Jardine, Markus P Schneider, Giles H Roditi, Christian Delles, Patrick B Mark1.
Abstract
BACKGROUND: Measurement of renal perfusion is a crucial part of measuring kidney function. Arterial spin labelling magnetic resonance imaging (ASL MRI) is a non-invasive method of measuring renal perfusion using magnetised blood as endogenous contrast. We studied the reproducibility of ASL MRI in normal volunteers.Entities:
Mesh:
Substances:
Year: 2014 PMID: 24484613 PMCID: PMC3909760 DOI: 10.1186/1471-2369-15-23
Source DB: PubMed Journal: BMC Nephrol ISSN: 1471-2369 Impact factor: 2.388
Demographics and biochemistry of participating volunteers
| Age | 44.1 | 14.6 |
| Male | 5 | |
| BMI | 26.5 | 6.6 |
| Systolic BP ( | 136 | 23 |
| Diastolic BP | 82 | 9 |
| Urea | 5.0 | 1.1 |
| Creatinine | 72.3 | 10.6 |
| CKD EPI eGFR | 98.3 | 15.1 |
| Total Cholesterol | 4.8 | 0.7 |
| HDL cholesterol | 1.8 | 1.2 |
| Glucose | 5.2 | 1.0 |
| Haemoglobin | 144 | 12 |
| Protein to creatinine ratio | 0.58 | 2.02 |
| Albumin to creatinine ratio | 0.17 | 0.32 |
| Adjusted serum calcium | 2.47 | 0.05 |
| Serum phosphate | 0.96 | 0.35 |
| Calcium phosphate product | 2.38 | 0.89 |
Figure 1Renal arterial spin labelling MRI images. a. Single contrast ASL image. b. M0 image demonstrating magnetisation equilibrium c. T1 map. d. Composite image of 5 ASL MRI contrast images registered with post processing software.
Kidney morphological, T1 and perfusion measurements made by ASL MRI
| Kidney length | 10.6 | 0.8 | 10.8 | 0.8 | 0.02 |
| Kidney volume Ellipsoid formula | 120.5 | 26.1 | 126.4 | 24.9 | 0.02 |
| Kidney volume voxel count | 155.7 | 29.2 | 157.7 | 28.6 | 0.39 |
| Whole kidney T1 | 1491 | 61 | 1499 | 52 | 0.52 |
| Cortical T1 | 1376 | 104 | 1406 | 96 | 0.07 |
| Medullary T1 | 1651 | 86 | 1639 | 80 | 0.38 |
| Whole kidney perfusion | 228 | 40 | 230 | 41 | 0.66 |
| Cortical perfusion | 321 | 63 | 334 | 63 | 0.18 |
| Absolute perfusion | 367 | 66 | 379 | 86 | 0.33 |
Test of significance is Student’s t-test.
Figure 2Correlation plot of voxel count and ellipsoid formula methods of renal volume measurement using MRI. R is the Pearson correlation coefficient.
Figure 3Correlation of kidney perfusion measured in left and right kidneys. R is the Pearson correlation coefficient.
Figure 4Correlation between whole and cortical kidney perfusion measurements made at MRI 1 and 2. R is the Pearson correlation coefficient.
Figure 5Bland Altman plot of cortical perfusion measurements made at MRI 1 and 2. Solid line and adjacent number indicates mean difference, whilst dashed line and number indicates limits of agreement.
Figure 6Bland Altman plot of whole kidney perfusion measurements at MRI 1 and 2. Solid line and adjacent number indicates mean difference, whilst dashed line and number indicates limits of agreement.