Abstract

PURPOSE: Quantification of positron emission tomography/magnetic resonance imaging (PET/MRI) studies is hampered by inaccurate MR-based attenuation correction (MR-AC). To date, most studies on MR-AC have been performed using PET/MR systems without time of flight (TOF). Maximum likelihood reconstruction of attenuation and activity (MLAA), however, has the potential to improve MR-AC by exploiting TOF. The purpose of this study is to assess the impact of MR-AC on PET image quantification for TOF-PET/MR systems and to evaluate PET accuracy when using TOF in combination with MLAA (TOF-MLAA). PROCEDURES: Simulations were designed to evaluate (1) the impact of MR-AC on PET quantification for different TOF windows (667, 500, 333 and 167 ps) and (2) use of TOF-MLAA for improving PET quantification. TOF-ordered subset expectation maximisation (OSEM) and TOF-MLAA reconstructions using MR-AC were compared with those obtained using TOF-OSEM with computed tomography-based AC (CT-AC).
RESULTS: OSEM reconstructions without TOF showed a negative MR-AC-induced bias of -50 % in the bone. TOF-OSEM was able to reduce this bias down to -15 %, with more accurate results for better TOF. TOF-MLAA was able to reduce the bias to within 5 % but at the cost of a ∼40 % increase in image variance.
CONCLUSIONS: TOF-MLAA can improve quantitative PET accuracy of PET/MR studies. Further improvements are anticipated with improving TOF performance.

Entities:  

Mesh:

Year:  2014        PMID: 24430291     DOI: 10.1007/s11307-013-0716-x

Source DB:  PubMed          Journal:  Mol Imaging Biol        ISSN: 1536-1632            Impact factor:   3.488


  35 in total

1.  Simultaneous reconstruction of activity and attenuation for PET/MR.

Authors:  André Salomon; Andreas Goedicke; Bernd Schweizer; Til Aach; Volkmar Schulz
Journal:  IEEE Trans Med Imaging       Date:  2010-11-29       Impact factor: 10.048

Review 2.  Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques.

Authors:  Matthias Hofmann; Bernd Pichler; Bernhard Schölkopf; Thomas Beyer
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-03       Impact factor: 9.236

3.  A decade of combined imaging: from a PET attached to a CT to a PET inside an MR.

Authors:  Thomas Beyer; Bernd Pichler
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-03       Impact factor: 9.236

4.  Combined PET/MR imaging--technology and applications.

Authors:  H F Wehrl; A W Sauter; M S Judenhofer; B J Pichler
Journal:  Technol Cancer Res Treat       Date:  2010-02

5.  MR-Based PET attenuation correction for PET/MR imaging.

Authors:  Ilja Bezrukov; Frédéric Mantlik; Holger Schmidt; Bernhard Schölkopf; Bernd J Pichler
Journal:  Semin Nucl Med       Date:  2013-01       Impact factor: 4.446

6.  MRI-guided attenuation correction in whole-body PET/MR: assessment of the effect of bone attenuation.

Authors:  A Akbarzadeh; M R Ay; A Ahmadian; N Riahi Alam; H Zaidi
Journal:  Ann Nucl Med       Date:  2012-12-21       Impact factor: 2.668

7.  MRI-based attenuation correction for PET/MRI using ultrashort echo time sequences.

Authors:  Vincent Keereman; Yves Fierens; Tom Broux; Yves De Deene; Max Lonneux; Stefaan Vandenberghe
Journal:  J Nucl Med       Date:  2010-05       Impact factor: 10.057

8.  Toward implementing an MRI-based PET attenuation-correction method for neurologic studies on the MR-PET brain prototype.

Authors:  Ciprian Catana; Andre van der Kouwe; Thomas Benner; Christian J Michel; Michael Hamm; Matthias Fenchel; Bruce Fischl; Bruce Rosen; Matthias Schmand; A Gregory Sorensen
Journal:  J Nucl Med       Date:  2010-09       Impact factor: 10.057

9.  Performance measurements of the Siemens mMR integrated whole-body PET/MR scanner.

Authors:  Gaspar Delso; Sebastian Fürst; Björn Jakoby; Ralf Ladebeck; Carl Ganter; Stephan G Nekolla; Markus Schwaiger; Sibylle I Ziegler
Journal:  J Nucl Med       Date:  2011-11-11       Impact factor: 10.057

10.  Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study.

Authors:  Ronald Boellaard; Nanda C Krak; Otto S Hoekstra; Adriaan A Lammertsma
Journal:  J Nucl Med       Date:  2004-09       Impact factor: 10.057

View more
  23 in total

1.  Improvements in PET Image Quality in Time of Flight (TOF) Simultaneous PET/MRI.

Authors:  Ryogo Minamimoto; Craig Levin; Mehran Jamali; Dawn Holley; Amir Barkhodari; Greg Zaharchuk; Andrei Iagaru
Journal:  Mol Imaging Biol       Date:  2016-10       Impact factor: 3.488

2.  Attenuation correction in emission tomography using the emission data--A review.

Authors:  Yannick Berker; Yusheng Li
Journal:  Med Phys       Date:  2016-02       Impact factor: 4.071

3.  Clinical evaluation of TOF versus non-TOF on PET artifacts in simultaneous PET/MR: a dual centre experience.

Authors:  Edwin E G W Ter Voert; Patrick Veit-Haibach; Sangtae Ahn; Florian Wiesinger; M Mehdi Khalighi; Craig S Levin; Andrei H Iagaru; Greg Zaharchuk; Martin Huellner; Gaspar Delso
Journal:  Eur J Nucl Med Mol Imaging       Date:  2017-01-26       Impact factor: 9.236

4.  The Effect of Defective PET Detectors in Clinical Simultaneous [18F]FDG Time-of-Flight PET/MR Imaging.

Authors:  Edwin E G W Ter Voert; Gaspar Delso; Felipe de Galiza Barbosa; Martin Huellner; Patrick Veit-Haibach
Journal:  Mol Imaging Biol       Date:  2017-08       Impact factor: 3.488

5.  Joint Reconstruction of Activity and Attenuation in Time-of-Flight PET: A Quantitative Analysis.

Authors:  Ahmadreza Rezaei; Christophe M Deroose; Thomas Vahle; Fernando Boada; Johan Nuyts
Journal:  J Nucl Med       Date:  2018-03-01       Impact factor: 10.057

Review 6.  Attenuation Correction of PET/MR Imaging.

Authors:  Yasheng Chen; Hongyu An
Journal:  Magn Reson Imaging Clin N Am       Date:  2017-01-18       Impact factor: 2.266

Review 7.  MR Imaging-Guided Attenuation Correction of PET Data in PET/MR Imaging.

Authors:  David Izquierdo-Garcia; Ciprian Catana
Journal:  PET Clin       Date:  2016-01-26

Review 8.  Morphology supporting function: attenuation correction for SPECT/CT, PET/CT, and PET/MR imaging.

Authors:  Tzu C Lee; Adam M Alessio; Robert M Miyaoka; Paul E Kinahan
Journal:  Q J Nucl Med Mol Imaging       Date:  2015-11-17       Impact factor: 2.346

9.  Investigating the state-of-the-art in whole-body MR-based attenuation correction: an intra-individual, inter-system, inventory study on three clinical PET/MR systems.

Authors:  Thomas Beyer; Martin L Lassen; Ronald Boellaard; Gaspar Delso; Maqsood Yaqub; Bernhard Sattler; Harald H Quick
Journal:  MAGMA       Date:  2016-01-06       Impact factor: 2.310

10.  Generation of PET Attenuation Map for Whole-Body Time-of-Flight 18F-FDG PET/MRI Using a Deep Neural Network Trained with Simultaneously Reconstructed Activity and Attenuation Maps.

Authors:  Donghwi Hwang; Seung Kwan Kang; Kyeong Yun Kim; Seongho Seo; Jin Chul Paeng; Dong Soo Lee; Jae Sung Lee
Journal:  J Nucl Med       Date:  2019-01-25       Impact factor: 10.057

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.