Performance measurements of the Siemens mMR integrated whole-body PET/MR scanner.

Gaspar Delso1, Sebastian Fürst, Björn Jakoby, Ralf Ladebeck, Carl Ganter, Stephan G Nekolla, Markus Schwaiger, Sibylle I Ziegler.   

Abstract

UNLABELLED: The recently released Biograph mMR is the first commercially available integrated whole-body PET/MR scanner. There are considerable advantages to integrating both modalities in a single scanner that enables truly simultaneous acquisition. However, there are also concerns about the possible degradation of both PET and MR performance in an integrated system. This paper evaluates the performance of the Biograph mMR during independent and simultaneous acquisition of PET and morphologic MR data.
METHODS: The NEMA NU 2-2007 protocol was followed for studying the PET performance. The following measurements were performed: spatial resolution; scatter fraction, count losses, and randoms; sensitivity; accuracy of the correction for count losses and randoms; and image quality. The quality control manual of the American College of Radiology was followed for studying the MR performance. The following measurements were performed: geometric accuracy, spatial resolution, low-contrast detectability, signal-to-noise ratio, static field (B(0)) homogeneity, radiofrequency field (B(1)) homogeneity, and radiofrequency noise.
RESULTS: An average spatial resolution of 4.3 mm in full width at half maximum was measured at 1 cm offset from the center of the field of view. The system sensitivity was 15.0 kcps/MBq along the center of the scanner. The scatter fraction was 37.9%, and the peak noise-equivalent count rate was 184 kcps at 23.1 kBq/mL. The maximum absolute value of the relative count rate error due to dead-time losses and randoms was 5.5%. The average residual error in scatter and attenuation correction was 12.1%. All MR parameters were within the tolerances defined by the American College of Radiology. B(0) inhomogeneities below 1 ppm were measured in a 120-mm radius. B(1) homogeneity and signal-to-noise ratio were equivalent to those of a standard MR scanner. No radiofrequency interference was detected.
CONCLUSION: These results compare favorably with other state-of-the-art PET/CT and PET/MR scanners, indicating that the integration of the PET detectors in the MR scanner and their operation within the magnetic field do not have a perceptible impact on the overall performance. The MR subsystem performs essentially like a standalone system. However, further work is necessary to evaluate the more advanced MR applications, such as functional imaging and spectroscopy.

Entities:  

Mesh:

Year:  2011        PMID: 22080447     DOI: 10.2967/jnumed.111.092726

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  239 in total

Review 1.  Molecular brain imaging in the multimodality era.

Authors:  Julie C Price
Journal:  J Cereb Blood Flow Metab       Date:  2012-03-21       Impact factor: 6.200

2.  High-risk plaque features can be detected in non-stenotic carotid plaques of patients with ischaemic stroke classified as cryptogenic using combined (18)F-FDG PET/MR imaging.

Authors:  Fabien Hyafil; Andreas Schindler; Dominik Sepp; Tilman Obenhuber; Anna Bayer-Karpinska; Tobias Boeckh-Behrens; Sabine Höhn; Marcus Hacker; Stephan G Nekolla; Axel Rominger; Martin Dichgans; Markus Schwaiger; Tobias Saam; Holger Poppert
Journal:  Eur J Nucl Med Mol Imaging       Date:  2015-10-03       Impact factor: 9.236

Review 3.  [Multiparametric imaging with simultaneous MRI/PET: Methodological aspects and possible clinical applications].

Authors:  S Gatidis; H Schmidt; C D Claussen; N F Schwenzer
Journal:  Z Rheumatol       Date:  2015-12       Impact factor: 1.372

4.  Sensitivity estimation in time-of-flight list-mode positron emission tomography.

Authors:  J L Herraiz; A Sitek
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

5.  Feasibility of (18)F-Fluorodeoxyglucose radiotracer dose reduction in simultaneous carotid PET/MR imaging.

Authors:  Mootaz Eldib; Jason Bini; Olivier Lairez; David D Faul; Niels Oesingmann; Zahi A Fayad; Venkatesh Mani
Journal:  Am J Nucl Med Mol Imaging       Date:  2015-06-15

6.  Imaging of differentiated thyroid carcinoma: (124)I-PET/MRI may not be superior to (124)I-PET/CT.

Authors:  A Vrachimis; M Weckesser; M Schäfers; L Stegger
Journal:  Eur J Nucl Med Mol Imaging       Date:  2016-01-26       Impact factor: 9.236

7.  Treatment-related changes in neuroendocrine tumors as assessed by textural features derived from 68Ga-DOTATOC PET/MRI with simultaneous acquisition of apparent diffusion coefficient.

Authors:  Manuel Weber; Lukas Kessler; Benedikt Schaarschmidt; Wolfgang Peter Fendler; Harald Lahner; Gerald Antoch; Lale Umutlu; Ken Herrmann; Christoph Rischpler
Journal:  BMC Cancer       Date:  2020-04-16       Impact factor: 4.430

8.  Improvement of attenuation correction in time-of-flight PET/MR imaging with a positron-emitting source.

Authors:  Pieter Mollet; Vincent Keereman; Jason Bini; David Izquierdo-Garcia; Zahi A Fayad; Stefaan Vandenberghe
Journal:  J Nucl Med       Date:  2014-01-16       Impact factor: 10.057

9.  Accurate PET/MR quantification using time of flight MLAA image reconstruction.

Authors:  R Boellaard; M B M Hofman; O S Hoekstra; A A Lammertsma
Journal:  Mol Imaging Biol       Date:  2014-08       Impact factor: 3.488

Review 10.  Positron emission tomography-magnetic resonance imaging: technical review.

Authors:  Raymond F Muzic; Frank P DiFilippo
Journal:  Semin Roentgenol       Date:  2014-10-18       Impact factor: 0.800

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.