EPID in vivo dosimetry in RapidArc technique.

Krzysztof Slosarek1, Marta Szlag1, Barbara Bekman1, Aleksandra Grzadziel1.   

Abstract

AIM: The aim of the study was to estimate the dose at the reference point applying an aSi-EPID device in the course of patient treatment.
MATERIALS AND METHODS: The method assumes direct proportionality between EPID signal and dose delivered to the patient reference point during the treatment session. The procedure consists of treatment plan calculation for the actual patient in the arc technique. The plan was realized with an elliptic water-equivalent phantom. An ionization chamber inside the phantom measured the dose delivered to the reference point. Simultaneously, the EPID matrix measured the CU distribution. EPID signal was also registered during patient irradiation with the same treatment plan. The formula for in vivo dose calculation was based on the CU(g) function, EPID signal registered during therapy and the relation between the dose and EPID signal level measured for the phantom. In vivo dose was compared with dose planned with the treatment planning system. Irradiation was performed with a Clinac accelerator by Varian Medical Systems in the RapidArc technique. The Clinac was equipped with an EPID matrix (electronic portal image device) of aSi-1000. Treatment plans were calculated with the Eclipse/Helios system. The phantom was a Scanditronix/Wellhöfer Slab phantom, and the ionization chamber was a 0.6 ccm PTW chamber.
RESULTS: In vivo dose calculations were performed for five patients. Planned dose at the reference point was 2 Gy for each treatment plan. Mean in vivo dose was in the range of 1.96-2.09.
CONCLUSIONS: Our method was shown to be appropriate for in vivo dose evaluation in the RapidArc technique.

Entities:  

Keywords:  Arc technique; Electronic portal imaging device; IMRT; Portal dosimetry; Verification

Year:  2010        PMID: 24376916      PMCID: PMC3863289          DOI: 10.1016/j.rpor.2010.01.003

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  18 in total

Review 1.  Intensity-modulated radiotherapy: current status and issues of interest.

Authors: 
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-11-15       Impact factor: 7.038

2.  Pre-treatment verification of intensity modulated photon beams with films and electronic portal imaging--two years of clinical experience.

Authors:  Luca Cozzi; Antonella Fogliata; Giorgia Nicolini
Journal:  Z Med Phys       Date:  2004       Impact factor: 4.820

3.  Three-dimensional IMRT verification with a flat-panel EPID.

Authors:  S Steciw; B Warkentin; S Rathee; B G Fallone
Journal:  Med Phys       Date:  2005-02       Impact factor: 4.071

4.  Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry.

Authors:  Peter B Greer
Journal:  Med Phys       Date:  2005-12       Impact factor: 4.071

5.  Volumetric modulated arc therapy: IMRT in a single gantry arc.

Authors:  Karl Otto
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

6.  Volumetric modulated Arc therapy and conventional intensity-modulated radiotherapy for simultaneous maximal intraprostatic boost: a planning comparison study.

Authors:  R Shaffer; W J Morris; V Moiseenko; M Welsh; C Crumley; S Nakano; M Schmuland; T Pickles; K Otto
Journal:  Clin Oncol (R Coll Radiol)       Date:  2009-03-05       Impact factor: 4.126

7.  Relative dosimetry using active matrix flat-panel imager (AMFPI) technology.

Authors:  Y El-Mohri; L E Antonuk; J Yorkston; K W Jee; M Maolinbay; K L Lam; J H Siewerdsen
Journal:  Med Phys       Date:  1999-08       Impact factor: 4.071

8.  Intensity modulation with photons for benign intracranial tumours: a planning comparison of volumetric single arc, helical arc and fixed gantry techniques.

Authors:  Antonella Fogliata; Alessandro Clivio; Giorgia Nicolini; Eugenio Vanetti; Luca Cozzi
Journal:  Radiother Oncol       Date:  2008-08-28       Impact factor: 6.280

9.  Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study.

Authors:  Wilko F A R Verbakel; Johan P Cuijpers; Daan Hoffmans; Michael Bieker; Ben J Slotman; Suresh Senan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-05-01       Impact factor: 7.038

10.  The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy.

Authors:  Giorgia Nicolini; Eugenio Vanetti; Alessandro Clivio; Antonella Fogliata; Stine Korreman; Jiri Bocanek; Luca Cozzi
Journal:  Radiat Oncol       Date:  2008-09-09       Impact factor: 3.481

View more
  12 in total

1.  The importance of accurate treatment planning, delivery, and dose verification.

Authors:  Julian Malicki
Journal:  Rep Pract Oncol Radiother       Date:  2012-03-06

Review 2.  Organizational, technical, physical and clinical quality standards for radiotherapy.

Authors:  Marta Bogusz-Czerniewicz; Daniel Kaźmierczak
Journal:  Rep Pract Oncol Radiother       Date:  2012-06-03

3.  Comparison of individual and composite field analysis using array detector for Intensity Modulated Radiotherapy dose verification.

Authors:  Sathiyan Saminathan; Varatharaj Chandraraj; C H Sridhar; Ravikumar Manickam
Journal:  Rep Pract Oncol Radiother       Date:  2012-05-19

4.  Commissioning and Acceptance Testing of the existing linear accelerator upgraded to volumetric modulated arc therapy.

Authors:  Ekambaram Varadharajan; Velayudham Ramasubramanian
Journal:  Rep Pract Oncol Radiother       Date:  2013-08-06

5.  Comparison of dosimetric variation between prostate IMRT and VMAT due to patient's weight loss: Patient and phantom study.

Authors:  James C L Chow; Runqing Jiang
Journal:  Rep Pract Oncol Radiother       Date:  2013-06-25

6.  Clinical examples of 3D dose distribution reconstruction, based on the actual MLC leaves movement, for dynamic treatment techniques.

Authors:  Wojciech Osewski; Lukasz Dolla; Michał Radwan; Marta Szlag; Roman Rutkowski; Barbara Smolińska; Krzysztof Slosarek
Journal:  Rep Pract Oncol Radiother       Date:  2014-07-17

7.  A study on conventional IMRT and RapidArc treatment planning techniques for head and neck cancers.

Authors:  S A Syam Kumar; Nagarajan Vivekanandan; Padmanaban Sriram
Journal:  Rep Pract Oncol Radiother       Date:  2012-03-03

8.  EPID-based daily verification of reproducibility of patients' irradiation with IMRT plans.

Authors:  Marta Kruszyna-Mochalska
Journal:  Rep Pract Oncol Radiother       Date:  2018-08-08

9.  Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance.

Authors:  Hosang Jin; Vance P Keeling; Daniel A Johnson; Salahuddin Ahmad
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

10.  Diode-based transmission detector for IMRT delivery monitoring: a validation study.

Authors:  Taoran Li; Q Jackie Wu; Thomas Matzen; Fang-Fang Yin; Jennifer C O'Daniel
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.