Abstract

A three-dimensional (3D) intensity-modulated radiotherapy (IMRT) pretreatment verification procedure has been developed based on the measurement of two-dimensional (2D) primary fluence profiles using an amorphous silicon flat-panel electronic portal imaging device (EPID). As described in our previous work, fluence profiles are extracted from EPID images by deconvolution with kernels that represent signal spread in the EPID due to radiation and optical scattering. The deconvolution kernels are derived using Monte Carlo simulations of dose deposition in the EPID and empirical fitting methods, for both 6 and 15 MV photon energies. In our new 3D verification technique, 2D fluence modulation profiles for each IMRT field in a treatment are used as input to a treatment planning system (TPS), which then generates 3D doses. Verification is accomplished by comparing this new EPID-based 3D dose distribution to the planned dose distribution calculated by the TPS. Thermoluminescent dosimeter (TLD) point dose measurements for an IMRT treatment of an anthropomorphic phantom were in good agreement with the EPID-based 3D doses; in contrast, the planned dose under-predicts the TLD measurement in a high-gradient region by approximately 16%. Similarly, large discrepancies between EPID-based and TPS doses were also evident in dose profiles of small fields incident on a water phantom. These results suggest that our 3D EPID-based method is effective in quantifying relevant uncertainties in the dose calculations of our TPS for IMRT treatments. For three clinical head and neck cancer IMRT treatment plans, our TPS was found to underestimate the mean EPID-based doses in the critical structures of the spinal cord and the parotids by approximately 4 Gy (11%-14%). According to radiobiological modeling calculations that were performed, such underestimates can potentially lead to clinically significant underpredictions of normal tissue complication rates.

Entities:  

Mesh:

Year:  2005        PMID: 15789607     DOI: 10.1118/1.1843471

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

1.  Practical guidelines for routine intensity-modulated radiotherapy verification: pre-treatment verification with portal dosimetry and treatment verification with in vivo dosimetry.

Authors:  A J Vinall; A J Williams; V E Currie; A Van Esch; D Huyskens
Journal:  Br J Radiol       Date:  2010-11       Impact factor: 3.039

2.  Monte Carlo-based adaptive EPID dose kernel accounting for different field size responses of imagers.

Authors:  Song Wang; Joseph K Gardner; John J Gordon; Weidong Li; Luke Clews; Peter B Greer; Jeffrey V Siebers
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

3.  Investigating the Electronic Portal Imaging Device for Small Radiation Field Measurements.

Authors:  Arpita Agarwal; Nikhil Rastogi; K J Maria Das; S A Yoganathan; D Udayakumar; Shaleen Kumar
Journal:  J Med Phys       Date:  2017 Apr-Jun

4.  Development of multi-planar dose verification by use of a flat panel EPID for intensity-modulated radiation therapy.

Authors:  Yuji Nakaguchi; Fujio Araki; Tomohiro Kouno; Takeshi Ono; Kazunari Hioki
Journal:  Radiol Phys Technol       Date:  2012-12-11

5.  Camera selection for real-time in vivo radiation treatment verification systems using Cherenkov imaging.

Authors:  Jacqueline M Andreozzi; Rongxiao Zhang; Adam K Glaser; Lesley A Jarvis; Brian W Pogue; David J Gladstone
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

6.  Transit dosimetry in dynamic IMRT with an a-Si EPID.

Authors:  Mahsheed Sabet; Pejman Rowshanfarzad; Fred W Menk; Peter B Greer
Journal:  Med Biol Eng Comput       Date:  2014-05-31       Impact factor: 2.602

7.  Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software.

Authors:  Mareike Held; Joey Cheung; Angelica Perez Andujar; François Husson; Olivier Morin
Journal:  Cureus       Date:  2018-02-02

8.  EPID in vivo dosimetry in RapidArc technique.

Authors:  Krzysztof Slosarek; Marta Szlag; Barbara Bekman; Aleksandra Grzadziel
Journal:  Rep Pract Oncol Radiother       Date:  2010-02-20

9.  Fast 3D dosimetric verifications based on an electronic portal imaging device using a GPU calculation engine.

Authors:  Jinhan Zhu; Lixin Chen; Along Chen; Guangwen Luo; Xiaowu Deng; Xiaowei Liu
Journal:  Radiat Oncol       Date:  2015-04-11       Impact factor: 3.481

10.  The use of extended dose range film for dosimetric calibration of a scanning liquid-filled ionization chamber electronic portal imaging device.

Authors:  Mohammad Mohammadi; Eva Bezak; Paul Reich
Journal:  J Appl Clin Med Phys       Date:  2006-05-15       Impact factor: 2.102

View more