Influence of partial pressure of oxygen in blood samples on measurement performance in glucose-oxidase-based systems for self-monitoring of blood glucose.

Annette Baumstark1, Christina Schmid, Stefan Pleus, Cornelia Haug, Guido Freckmann.   

Abstract

BACKGROUND: Partial pressure of oxygen (pO2) in blood samples can affect blood glucose (BG) measurements, particularly in systems that employ the glucose oxidase (GOx) enzyme reaction on test strips. In this study, we assessed the impact of different pO2 values on the performance of five GOx systems and one glucose dehydrogenase (GDH) system. Two of the GOx systems are labeled by the manufacturers to be sensitive to increased blood oxygen content, while the other three GOx systems are not.
METHODS: Aliquots of 20 venous samples were adjusted to the following pO2 values: <45, ~70, and ≥150 mmHg. For each system, five consecutive measurements on each sample aliquot were performed using the same test strip lot. Relative differences between the mean BG results at pO2 ~70 mmHg, which is considered to be similar to pO2 in capillary blood samples, and the mean BG result at pO2 <45 and ≥150 mmHg were calculated.
RESULTS: For all tested GOx systems, mean relative differences in the BG measurement results were between 6.1% and 22.6% at pO2 <45 mmHg and between -7.9% and -14.9% at pO2 ≥150 mmHg. For both pO2 levels, relative differences of all tested GOx systems were significant (p < .0001). The GDH system showed mean relative differences of -1.0% and -0.4% at pO2 values <45 and ≥150 mmHg, respectively, which were not significant.
CONCLUSIONS: These data suggest that capillary blood pO2 variations lead to clinically relevant BG measurement deviations in GOx systems, even in GOx systems that are not labeled as being oxygen sensitive.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24351177      PMCID: PMC3876329          DOI: 10.1177/193229681300700611

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  17 in total

Review 1.  Review of glucose oxidases and glucose dehydrogenases: a bird's eye view of glucose sensing enzymes.

Authors:  Stefano Ferri; Katsuhiro Kojima; Koji Sode
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

Review 2.  Electrochemical glucose biosensors.

Authors:  Joseph Wang
Journal:  Chem Rev       Date:  2007-12-23       Impact factor: 60.622

3.  Partial pressure of oxygen in capillary blood samples from the fingertip.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Stefan Pleus; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

4.  Impact of partial pressure of oxygen in blood samples on the performance of systems for self-monitoring of blood glucose.

Authors:  Christina Schmid; Annette Baumstark; Stefan Pleus; Cornelia Haug; Martina Tesar; Guido Freckmann
Journal:  Diabetes Technol Ther       Date:  2013-11-08       Impact factor: 6.118

5.  Oxygen effects on glucose measurements with a reference analyzer and three handheld meters.

Authors:  Z Tang; R F Louie; M Payes; K C Chang; G J Kost
Journal:  Diabetes Technol Ther       Date:  2000       Impact factor: 6.118

6.  Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing.

Authors:  Z Tang; R F Louie; J H Lee; D M Lee; E E Miller; G J Kost
Journal:  Crit Care Med       Date:  2001-05       Impact factor: 7.598

7.  Comparison of arterial and venous blood gases and the effects of analysis delay and air contamination on arterial samples in patients with chronic obstructive pulmonary disease and healthy controls.

Authors:  T M O'Connor; P J Barry; A Jahangir; C Finn; B M Buckley; A El-Gammal
Journal:  Respiration       Date:  2010-02-04       Impact factor: 3.580

Review 8.  [Air travel and respiratory diseases].

Authors:  Francisco García Río; Luis Borderías Clau; Ciro Casanova Macario; Bartolomé R Celli; Joan Escarrabill Sanglás; Nicolás González Mangado; Josep Roca Torrent; Fernando Uresandi Romero
Journal:  Arch Bronconeumol       Date:  2007-02       Impact factor: 4.872

Review 9.  Glucose biosensors: an overview of use in clinical practice.

Authors:  Eun-Hyung Yoo; Soo-Youn Lee
Journal:  Sensors (Basel)       Date:  2010-05-04       Impact factor: 3.576

Review 10.  Chronic obstructive pulmonary disease and glucose metabolism: a bitter sweet symphony.

Authors:  Aibek E Mirrakhimov
Journal:  Cardiovasc Diabetol       Date:  2012-10-27       Impact factor: 9.951

View more
  8 in total

1.  Pilots and Diabetes Technology.

Authors:  Johan Jendle; Kurt Rinnert; Anton Westman; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2016-12-05

2.  Proof of Concept Study to Assess the Influence of Oxygen Partial Pressure in Capillary Blood on SMBG Measurements.

Authors:  Annette Baumstark; Stefan Pleus; Nina Jendrike; Christina Liebing; Rolf Hinzmann; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2019-03-06

Review 3.  Glucose homeostasis during short-term and prolonged exposure to high altitudes.

Authors:  Orison O Woolcott; Marilyn Ader; Richard N Bergman
Journal:  Endocr Rev       Date:  2015-02-12       Impact factor: 19.871

4.  In Vitro Evaluation of Miniaturized Amperometric Enzyme Sensor Based on the Direct Electron Transfer Principle for Continuous Glucose Monitoring.

Authors:  Yutaro Inoue; Yasuhide Kusaka; Kotaro Shinozaki; Inyoung Lee; Koji Sode
Journal:  J Diabetes Sci Technol       Date:  2022-01-05

5.  Evaluation of Hematocrit Influence on Self-Monitoring of Blood Glucose Based on ISO 15197:2013: Comparison of a Novel System With Five Systems With Different Hematocrit Ranges.

Authors:  Andrew Hattemer; Sami Wardat
Journal:  J Diabetes Sci Technol       Date:  2018-01-29

Review 6.  Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge.

Authors:  Michael Erbach; Guido Freckmann; Rolf Hinzmann; Bernhard Kulzer; Ralph Ziegler; Lutz Heinemann; Oliver Schnell
Journal:  J Diabetes Sci Technol       Date:  2016-08-22

7.  Structural analysis of fungus-derived FAD glucose dehydrogenase.

Authors:  Hiromi Yoshida; Genki Sakai; Kazushige Mori; Katsuhiro Kojima; Shigehiro Kamitori; Koji Sode
Journal:  Sci Rep       Date:  2015-08-27       Impact factor: 4.379

Review 8.  Monitoring of Pediatric Type 1 Diabetes.

Authors:  Brynn E Marks; Joseph I Wolfsdorf
Journal:  Front Endocrinol (Lausanne)       Date:  2020-03-17       Impact factor: 5.555

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.