Comparison of arterial and venous blood gases and the effects of analysis delay and air contamination on arterial samples in patients with chronic obstructive pulmonary disease and healthy controls.

T M O'Connor1, P J Barry, A Jahangir, C Finn, B M Buckley, A El-Gammal.   

Abstract

BACKGROUND: Arterial blood gases (ABGs) are often sampled incorrectly, leading to a 'mixed' or venous sample. Delays in analysis and air contamination are common.
OBJECTIVES: We measured the effects of these errors in patients with chronic obstructive pulmonary disease (COPD) exacerbations and controls.
METHODS: Arterial and venous samples were analyzed from 30 patients with COPD exacerbation and 30 controls. Venous samples were analysed immediately and arterial samples separated into non-air-contaminated and air-contaminated specimens and analysed at 0, 30, 60, 90 and 180 min.
RESULTS: Mean venous pH was 7.371 and arterial pH was 7.407 (p < 0.0001). There was a correlation between venous and arterial pH (r = 0.5347, p < 0.0001). The regression equation to predict arterial pH was: arterial pH = 4.2289 + 0.43113 · venous pH. There were no clinically significant differences in arterial PO₂ associated with analysis delay. A statistically significant decline in pH was detected at 30 min in patients with COPD exacerbation (p = 0.0042) and 90 min in controls (p < 0.0001). A clinically significant decline in pH emerged at 73 min in patients with COPD exacerbation and 87 min in controls. Air contamination was associated with a clinically significant increase in PO₂ in all samples, including those that were immediately analyzed.
CONCLUSIONS: Arterial and venous pH differ significantly. Venous pH cannot accurately replace arterial pH. Temporal delays in ABG analysis result in a significant decline in measured pH. ABGs should be analysed within 30 min. Air contamination leads to an immediate increase in measured PO₂, indicating that air-contaminated ABGs should be discarded.
Copyright © 2010 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2010        PMID: 20134147     DOI: 10.1159/000281879

Source DB:  PubMed          Journal:  Respiration        ISSN: 0025-7931            Impact factor:   3.580


  7 in total

1.  Partial pressure of oxygen in capillary blood samples from the fingertip.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Stefan Pleus; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

2.  Impact of partial pressure of oxygen in blood samples on the performance of systems for self-monitoring of blood glucose.

Authors:  Christina Schmid; Annette Baumstark; Stefan Pleus; Cornelia Haug; Martina Tesar; Guido Freckmann
Journal:  Diabetes Technol Ther       Date:  2013-11-08       Impact factor: 6.118

3.  Influence of partial pressure of oxygen in blood samples on measurement performance in glucose-oxidase-based systems for self-monitoring of blood glucose.

Authors:  Annette Baumstark; Christina Schmid; Stefan Pleus; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

4.  Proof of Concept Study to Assess the Influence of Oxygen Partial Pressure in Capillary Blood on SMBG Measurements.

Authors:  Annette Baumstark; Stefan Pleus; Nina Jendrike; Christina Liebing; Rolf Hinzmann; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2019-03-06

5.  Detection of preanalytical errors in arterial blood gas analysis.

Authors:  Serap Çuhadar; Hayat Özkanay-Yörük; Mehmet Köseoğlu; Kaan Katırcıoğlu
Journal:  Biochem Med (Zagreb)       Date:  2022-06-15       Impact factor: 2.515

6.  Agreement and Correlation between Arterial and Central Venous Blood Gas Following Coronary Artery Bypass Graft Surgery.

Authors:  Masoumeh Esmaeilivand; Alireza Khatony; Gholamreza Moradi; Farid Najafi; Alireza Abdi
Journal:  J Clin Diagn Res       Date:  2017-03-01

7.  Time-dependent variation of ionized calcium in serum samples.

Authors:  Antonija Perović; Marina Njire Bratičević
Journal:  Biochem Med (Zagreb)       Date:  2019-10-15       Impact factor: 2.313

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.