Impact of partial pressure of oxygen in blood samples on the performance of systems for self-monitoring of blood glucose.

Christina Schmid1, Annette Baumstark, Stefan Pleus, Cornelia Haug, Martina Tesar, Guido Freckmann.   

Abstract

BACKGROUND: The partial pressure of oxygen (pO2) in blood samples can affect glucose measurements with oxygen-sensitive systems. In this study, we assessed the influence of different pO2 levels on blood glucose (BG) measurements with five glucose oxidase (GOD) systems and one glucose dehydrogenase (GDH) system. All selected GOD systems were indicated by the manufacturers to be sensitive to increased oxygen content of the blood sample.
MATERIALS AND METHODS: Venous blood samples of 16 subjects (eight women, eight men; mean age, 52 years; three with type 1 diabetes, four with type 2 diabetes, and nine without diabetes) were collected. Aliquots of each sample were adjusted to the following pO2 values: ≤45 mm Hg, approximately 70 mm Hg, and ≥150 mm Hg. For each system, five consecutive measurements on each sample were performed using the same test strip lot. Relative differences between the mean BG value at a pO2 level of approximately 70 mm Hg, which was considered to be similar to pO2 values in capillary blood samples, and the mean BG value at pO2 levels ≤45 mm Hg and ≥150 mm Hg were calculated.
RESULTS: The GOD systems showed mean relative differences between 11.8% and 44.5% at pO2 values ≤45 mm Hg and between -14.6% and -21.2% at pO2 values ≥150 mm Hg. For the GDH system, the mean relative differences were -0.3% and -0.2% at pO2 values ≤45 mm Hg and ≥150 mm Hg, respectively.
CONCLUSIONS: The magnitude of the pO2 impact on BG measurements seems to vary among the tested oxygen-sensitive GOD systems. The pO2 range in which oxygen-sensitive systems operate well should be provided in the product information.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24205977      PMCID: PMC3934604          DOI: 10.1089/dia.2013.0184

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  25 in total

Review 1.  Mediated biosensors.

Authors:  Asha Chaubey; B D Malhotra
Journal:  Biosens Bioelectron       Date:  2002-06       Impact factor: 10.618

2.  Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design.

Authors:  R F Louie; Z Tang; D V Sutton; J H Lee; G J Kost
Journal:  Arch Pathol Lab Med       Date:  2000-02       Impact factor: 5.534

3.  What happens to patients with respiratory disease when they fly?

Authors:  R K Coker; M R Partridge
Journal:  Thorax       Date:  2004-11       Impact factor: 9.139

4.  Influence of blood sample oxygen tension on blood glucose concentration measured using an enzyme-electrode method.

Authors:  K Kurahashi; H Maruta; Y Usuda; M Ohtsuka
Journal:  Crit Care Med       Date:  1997-02       Impact factor: 7.598

5.  Time required for partial pressure of arterial oxygen equilibration during mechanical ventilation after a step change in fractional inspired oxygen concentration.

Authors:  N Cakar; M Tuŏrul; A Demirarslan; A Nahum; A Adams; O Akýncý; F Esen; L Telci
Journal:  Intensive Care Med       Date:  2001-04       Impact factor: 17.440

6.  An arterial blood gas diagram for clinical use.

Authors:  E E Mays
Journal:  Chest       Date:  1973-05       Impact factor: 9.410

7.  Oxygen effects on glucose measurements with a reference analyzer and three handheld meters.

Authors:  Z Tang; R F Louie; M Payes; K C Chang; G J Kost
Journal:  Diabetes Technol Ther       Date:  2000       Impact factor: 6.118

8.  Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing.

Authors:  Z Tang; R F Louie; J H Lee; D M Lee; E E Miller; G J Kost
Journal:  Crit Care Med       Date:  2001-05       Impact factor: 7.598

9.  Effect of high altitude on blood glucose meter performance.

Authors:  Kenneth S Fink; Dale B Christensen; Allan Ellsworth
Journal:  Diabetes Technol Ther       Date:  2002       Impact factor: 6.118

10.  The effect of the partial pressure of oxygen on blood glucose concentration examined using glucose oxidase with ferricyan ion.

Authors:  T Y Chun; M Hirose; T Sawa; M Harada; T Hosokawa; Y Tanaka; M Miyazaki
Journal:  Anesth Analg       Date:  1994-11       Impact factor: 5.108

View more
  6 in total

1.  Partial pressure of oxygen in capillary blood samples from the fingertip.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Stefan Pleus; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

2.  Pilots and Diabetes Technology.

Authors:  Johan Jendle; Kurt Rinnert; Anton Westman; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2016-12-05

3.  Self-monitoring of blood glucose: impact of a time delay between capillary blood sampling and glucose measurement.

Authors:  Guido Freckmann; Stefan Pleus; Annette Baumstark; Christina Schmid; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2014-08-21

4.  Influence of partial pressure of oxygen in blood samples on measurement performance in glucose-oxidase-based systems for self-monitoring of blood glucose.

Authors:  Annette Baumstark; Christina Schmid; Stefan Pleus; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

5.  Proof of Concept Study to Assess the Influence of Oxygen Partial Pressure in Capillary Blood on SMBG Measurements.

Authors:  Annette Baumstark; Stefan Pleus; Nina Jendrike; Christina Liebing; Rolf Hinzmann; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2019-03-06

6.  Performance evaluation of three i-SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer.

Authors:  Ha Nui Kim; Kyung Chul Moon; Soo-Young Yoon
Journal:  J Clin Lab Anal       Date:  2020-05-19       Impact factor: 2.352

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.