Proof of Concept Study to Assess the Influence of Oxygen Partial Pressure in Capillary Blood on SMBG Measurements.

Annette Baumstark1, Stefan Pleus1, Nina Jendrike1, Christina Liebing1, Rolf Hinzmann2, Cornelia Haug1, Guido Freckmann1.   

Abstract

BACKGROUND: Measurement results provided by blood glucose monitoring systems (BGMS) can be affected by various influencing factors. For some BGMS using glucose oxidase (GOx)-based test strips, one of these factors is the oxygen partial pressure (pO2) of the applied blood sample. Because assessing the potential influence of pO2 when measuring capillary blood samples is not straight-forward, we performed a proof of concept study.
METHOD: Influence of pO2 was investigated for two GOx-based BGMS (BGMS A and B). Measurement results of the GOx-based BGMS were compared with measurement results from a pO2-independent BGMS (BGMS C). A total of 119 samples from 60 subjects were measured, twice with BGMS C, then 6 times each with BGMS A and BGMS B or vice versa, and again twice with BGMS C. Immediately afterward, pO2 was determined. Linear regression analysis based on relative differences between results from BGMS A or BGMS B and results from BGMS C was performed to estimate the degree of pO2 influence.
RESULTS: The relative bias between the lowest and highest pO2 values differed by 14.3% for BGMS A, indicating a pO2 influence that might be clinically relevant, and by 9.7% for BGMS B, indicating that pO2 influence may be too small to be reliably detected because of the BGMS' imprecision.
CONCLUSIONS: This proof of concept study showed that with the procedures used, a potentially clinically relevant influence of pO2 in capillary blood samples on GOx-based BGMS could be detected. Further larger-scale studies are needed to verify this influence.

Entities:  

Keywords:  blood glucose monitoring system; glucose oxidase; partial pressure of oxygen; self-monitoring of blood glucose

Mesh:

Substances:

Year:  2019        PMID: 30841739      PMCID: PMC6835173          DOI: 10.1177/1932296819833369

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  19 in total

1.  Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design.

Authors:  R F Louie; Z Tang; D V Sutton; J H Lee; G J Kost
Journal:  Arch Pathol Lab Med       Date:  2000-02       Impact factor: 5.534

Review 2.  Electrochemical glucose biosensors.

Authors:  Joseph Wang
Journal:  Chem Rev       Date:  2007-12-23       Impact factor: 60.622

3.  Partial pressure of oxygen in capillary blood samples from the fingertip.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Stefan Pleus; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

4.  Impact of partial pressure of oxygen in blood samples on the performance of systems for self-monitoring of blood glucose.

Authors:  Christina Schmid; Annette Baumstark; Stefan Pleus; Cornelia Haug; Martina Tesar; Guido Freckmann
Journal:  Diabetes Technol Ther       Date:  2013-11-08       Impact factor: 6.118

5.  Influence of partial pressure of oxygen in blood samples on measurement performance in glucose-oxidase-based systems for self-monitoring of blood glucose.

Authors:  Annette Baumstark; Christina Schmid; Stefan Pleus; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

6.  Oxygen effects on glucose measurements with a reference analyzer and three handheld meters.

Authors:  Z Tang; R F Louie; M Payes; K C Chang; G J Kost
Journal:  Diabetes Technol Ther       Date:  2000       Impact factor: 6.118

Review 7.  [Air travel and respiratory diseases].

Authors:  Francisco García Río; Luis Borderías Clau; Ciro Casanova Macario; Bartolomé R Celli; Joan Escarrabill Sanglás; Nicolás González Mangado; Josep Roca Torrent; Fernando Uresandi Romero
Journal:  Arch Bronconeumol       Date:  2007-02       Impact factor: 4.872

8.  Chronic obstructive pulmonary disease, asthma, and risk of type 2 diabetes in women.

Authors:  Jamal S Rana; Murray A Mittleman; Javed Sheikh; Frank B Hu; JoAnn E Manson; Graham A Colditz; Frank E Speizer; R Graham Barr; Carlos A Camargo
Journal:  Diabetes Care       Date:  2004-10       Impact factor: 19.112

Review 9.  Glucose biosensors: an overview of use in clinical practice.

Authors:  Eun-Hyung Yoo; Soo-Youn Lee
Journal:  Sensors (Basel)       Date:  2010-05-04       Impact factor: 3.576

Review 10.  Chronic obstructive pulmonary disease and glucose metabolism: a bitter sweet symphony.

Authors:  Aibek E Mirrakhimov
Journal:  Cardiovasc Diabetol       Date:  2012-10-27       Impact factor: 9.951

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.