| Literature DB >> 32256447 |
Brynn E Marks1, Joseph I Wolfsdorf2.
Abstract
Regular self-monitoring of blood glucose levels, and ketones when indicated, is an essential component of type 1 diabetes (T1D) management. Although fingerstick blood glucose monitoring has been the standard of care for decades, ongoing rapid technological developments have resulted in increasingly widespread use of continuous glucose monitoring (CGM). This article reviews recommendations for self-monitoring of glucose and ketones in pediatric T1D with particular emphasis on CGM and factors that impact the accuracy and real-world use of this technology.Entities:
Keywords: blood glucose self monitoring; continuous glucose monitor; diabetes management; ketone; type 1 diabetes (T1D)
Mesh:
Substances:
Year: 2020 PMID: 32256447 PMCID: PMC7089921 DOI: 10.3389/fendo.2020.00128
Source DB: PubMed Journal: Front Endocrinol (Lausanne) ISSN: 1664-2392 Impact factor: 5.555
ISPAD and ADA glycemic and A1c targets (6, 7).
| A1c target | <7.5% | <7.0% |
| Pre-meal | 90–130 mg/dL | 70–130 mg/dL |
| Postmeal | n/a | 90–180 mg/dL |
| Before bedtime | 90–150 mg/dL | 90–150 mg/dL |
ADA, American Diabetes Association; ISPAD, International Society for Pediatric and Adolescent Diabetes.
Recommendations for blood glucose monitoring.
| •Routinely before meals and snacks. |
Factors impacting the accuracy of self-monitored blood glucose.
| Improper handwashing | • Residual water on fingertips | • Traces of glucose-containing substances on fingertips | |
| Deteriorated test-strips | • Improper coding | ||
| Altitude | • High Altitude (>2,000 m) | ||
| Oxygen concentrations | • High partial pressure of oxygen | • Low partial pressure of oxygen | |
| Extreme temperatures | • Cold temperatures (<10°C) | • Rapid increase in ambient temperature | • Hot temperatures (>39°C) |
| Alternative test sites | • Testing at cold sites (e.g., forearm) | • Lag time created by decreased blood flow to alternative sites | |
| Laboratory values | • Elevated hematocrit (>45%) | • Low hematocrit (<35%) | |
| Altered tissue oxygenation | • Partial pressure of oxygen >150 mmHg | • Partial pressure of oxygen <45 mmHg | |
| Medication related | • Ascorbic acid | • Acetaminophen | |
Glucose oxidase-based devices only.
Rates of sensor glucose change as indicated by different CGM systems.
| 0–1 mg/dL/min | → | No arrow | → | → |
| 1–2 mg/dL/min | ↗ or ↘ | ↑ or ↓ | ↗ or ↘ | ↗ or ↘ |
| 2–3 mg/dL/min | ↑ or ↓ | ↑↑ or ↓↓ | ↑ or ↓ | ↑ or ↓ |
| > 3 mg/dL/min | ↑↑ or ↓↓ | ↑↑↑ or ↓↓↓ | n/a | n/a |
Comparison of currently available CGM systems.
| Non-adjunctive dosing indication | Yes | No | Yes | Yes |
| MARD | 9.0% | 9.6% | 9.4% | 8.5% |
| Alerts | Yes | Yes | No | Yes |
| Sensor warm up time | 2 h | 2 h | 1 h | ~26 h |
| Sensor wear time | 10 days | 7 days | 14 days | 90 days |
| Transmitter life | 90 days | 1 year, rechargeable | 14 days | 1 year, rechargeable |
| Share/follow app | Yes | Yes | Yes | Yes |
| Interfering substances | No acetaminophen interference (up to 4 g/day) | Acetaminophen | Vitamin C, Aspirin | Mannitol, Tetracycline, Aspirin |
| Unique features | T:slim integration for Basal IQ and Control IQ | Sugar IQ decision support | No alerts for highs/lows unless scanned; | Implanted CGM; |
Non-adjunctive dosing refers to the ability to make treatment decisions for both hypo- and hyperglycemia based on a CGM sensor glucose value without a confirmatory fingerstick BG measurement.
MARD or mean absolute relative difference is the average of the absolute differences between reference blood glucose measurements and glucose measurements obtained by CGM.
Pediatric and adult insulin dose adjustments for trend arrows for Dexcom G5 and G6 and Freestyle Libre [adapted from Laffel et al. (33), Aleppo et al. (34), and Kudva et al. (35)].
| ↑↑ or ↓↓ | <25 | ±4.0 units | <25 | ±4.5 units | ||
| ↑ or ↓ | <25 | ±3.0 units | <25 | ±3.5 units | <25 | ±3.5 units |
| ↗ or ↘ | <25 | ±2.0 units | <25 | ±2.5 units | <25 | ±2.5 units |
| → | All CF | No adjustment | All CF | No adjustment | All CF | No adjustment |
CF, correction factor.
Comparisons of the accuracy requirements for blood glucose meters, as dictated by the ISO 15197:2013, and iCGM, as required by the US FDA.
| Overall Accuracy | ≥95% within ± 15 mg/dL for glucose concentrations <100 mg/dL | >87% within ± 20% |
| <70 mg/dL | n/a | >85% within ± 15 mg/dL |
| 70–180/dL | n/a | >70% within ± 15% |
| >180 mg/dL | n/a | >80% within ± 15% |
| ≥99% within consensus error grid zones A and B | ≤1% of glucose rates of change >1 mg/dL/min if true rate of change is < −2 mg/dl/min | |
Adapted from Freckmann et al. (.
International consensus for clinical CGM targets.
| ≥250 | ≥13.9 | <5% | 72 min |
| >180 | ≥10.0 | <25% | <6 h |
| 70–180 | 3.9–10.0 | >70% | 16 h, 48 min |
| <70 | <3.9 | <4% | <58 min |
| <54 | <3.0 | <1% | <15 min |
If <25 years old with an A1c goal of 7.5% may use a target of 60%.
Figure 1An example of how CGM metrics are displayed in a standardized ambulatory glucose profile (AGP). (A) Average glucose, (B) Time very low (<54 mg/dL), (C) Time low (<70 mg/dL), (D) Time in target range (70–180 mg/dL), (E) Time high (>180 mg/dL), (F) Time very high (> 250 mg/dL), (G) Coefficient of variation, (H) Percentage of time CGM is active. In the top figure, the hatched area shows the target range; composite data for 14-days are shown as median (orange line), interquartile range (blue shaded area), 10 and 90th percentiles (green dashed line). Individual days are shown below.
Benefits and limitations of continuous glucose monitoring.
| • Awareness of glycemic values and trends | • Body image concerns |
When to measure urine or blood ketones.
| •When BG is unexpectedly high; e.g., fasting ≥250 mg/dL or unexpectedly ≥300 mg/dL for more than 2–3 h |
Measurement of ketosis with urine and blood ketone monitoring.
| Negative | <5 mg/dL | <0.3 mmol/L |
| Trace | 5 mg/dL | 0.3–0.5 mmol/L |
| Small | 15 mg/dL | 0.6–0.9 mmol/L |
| Moderate | 40 mg/dL | 1.0–1.5 mmol/L |
| Large | 80–160 mg/dL | ≥1.6 mmol/L |