Literature DB >> 24273335

Performance Evaluation of Small Animal PET Scanners With Different System Designs.

Xiaoli Li1, Adam M Alessio, Thompson H Burnett, Thomas K Lewellen, Roberts Miyaoka.   

Abstract

This study evaluated the image quality metrics of small animal PET scanners based upon measured single detector module positioning performance. A semi-analytical approach was developed to study PET scanner performance in the scenario of multiple realizations. Positron range blurring, scanner system response function (SRF) and statistical noise were included in the modeling procedure. The scanner sensitivity map was included in the system matrix during maximum likelihood expectation maximization (MLEM) reconstruction. Several image quality metrics were evaluated for octagonal ring PET scanners consisting of continuous miniature crystal element (cMiCE) detector modules with varying designs. These designs included 8 mm and 15 mm thick crystal detectors using conventional readout with the photosensors on the exit surface of the crystal and a 15 mm thick crystal detector using our proposed sensor-on-the-entrance (SES) design. For the conventional readout design, the results showed that there was a tradeoff between bias and variance with crystal thickness. The 15 mm crystal detector had better detection task performance, while quantitation task performance was degraded. On the other hand, our SES detector had similar detection efficiency as the conventional design using a 15 mm thick crystal and had similar intrinsic spatial resolution as the conventional design using an 8 mm thick crystal. The end result was that by using the SES design, one could improve scanner quantitation task performance without sacrificing detection task performance.

Entities:  

Keywords:  Image quality metrics; maximum likelihood expectation maximization (MLEM) reconstruction; monolithic crystal detector; system response function

Year:  2013        PMID: 24273335      PMCID: PMC3834349          DOI: 10.1109/TNS.2013.2246797

Source DB:  PubMed          Journal:  IEEE Trans Nucl Sci        ISSN: 0018-9499            Impact factor:   1.679


  13 in total

Review 1.  Use of positron emission tomography in animal research.

Authors:  S R Cherry; S S Gambhir
Journal:  ILAR J       Date:  2001

2.  Depth of interaction decoding of a continuous crystal detector module.

Authors:  T Ling; T K Lewellen; R S Miyaoka
Journal:  Phys Med Biol       Date:  2007-03-29       Impact factor: 3.609

3.  Constrained Fourier space method for compensation of missing data in emission computed tomography.

Authors:  J S Karp; G Muehllehner; R M Lewitt
Journal:  IEEE Trans Med Imaging       Date:  1988       Impact factor: 10.048

4.  A high resolution, monolithic crystal, PET/MRI detector with DOI positioning capability.

Authors:  Xiaoli Li; Cate Lockhart; Tom K Lewellen; Robert S Miyaoka
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2008

5.  Objective assessment of image quality: effects of quantum noise and object variability.

Authors:  H H Barrett
Journal:  J Opt Soc Am A       Date:  1990-07       Impact factor: 2.129

6.  Parametric positioning of a continuous crystal PET detector with depth of interaction decoding.

Authors:  T Ling; T H Burnett; T K Lewellen; R S Miyaoka
Journal:  Phys Med Biol       Date:  2008-03-10       Impact factor: 3.609

7.  Comparison of Detector Intrinsic Spatial Resolution Characteristics for Sensor on the Entrance Surface and Conventional Readout Designs.

Authors:  Robert S Miyaoka; Xiaoli Li; Cate Lockhart; Tom K Lewellen
Journal:  IEEE Trans Nucl Sci       Date:  2010       Impact factor: 1.679

8.  New Continuous Miniature Crystal Element (cMiCE) Detector Geometries.

Authors:  Robert S Miyaoka; Xiaoli Li; Cate Lockhart; Tom K Lewellen
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2009-10-24

9.  Measurements of entrance-surface vs. conventional single-ended readout of a monolithic scintillator.

Authors:  William C J Hunter; Wendy McDougald; Jerome J Griesmer; Lingxiong Shao; Robert Zahn; Tom K Lewellen; Robert S Miyaoka
Journal:  IEEE Trans Nucl Sci       Date:  2011       Impact factor: 1.679

10.  Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluation.

Authors:  S Tong; A M Alessio; P E Kinahan
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.