Combining automated attenuation-based tube voltage selection and iterative reconstruction: a liver phantom study.

Daniela B Husarik1, Sebastian T Schindera, Fabian Morsbach, Natalie Chuck, Burkhardt Seifert, Zsolt Szucs-Farkas, Hatem Alkadhi.   

Abstract

OBJECTIVES: To determine the value of combined automated attenuation-based tube-potential selection and iterative reconstructions (IRs) for optimising computed tomography (CT) imaging of hypodense liver lesions.
METHODS: A liver phantom containing hypodense lesions was imaged by CT with and without automated attenuation-based tube-potential selection (80, 100 and 120 kVp). Acquisitions were reconstructed with filtered back projection (FBP) and sinogram-affirmed IR. Image noise and contrast-to-noise ratio (CNR) were measured. Two readers marked lesion localisation and rated confidence, sharpness, noise and image quality on a five-point scale (1 = worst, 5 = best).
RESULTS: Image noise was lower (31-52%) and CNR higher (43-102%) on IR than on FBP images at all tube voltages. On 100-kVp and 80-kVp IR images, confidence and sharpness were higher than on 120-kVp FBP images. Scores for image quality score and noise as well as sensitivity for 100-kVp IR were similar or higher than for 120-kVp FBP and lower for 80-kVp IR. Radiation dose was reduced by 26% at 100 kVp and 56% at 80 kVp.
CONCLUSIONS: Compared with 120-kVp FBP images, the combination of automated attenuation-based tube-potential selection at 100 kVp and IR provides higher image quality and improved sensitivity for detecting hypodense liver lesions in vitro at a dose reduced by 26%. KEY POINTS: • Combining automated tube voltage selection/iterative CT reconstruction improves image quality. • Attenuation values remain stable on IR compared with FBP images. • Lesion detection was highest on 100-kVp IR images.

Entities:  

Mesh:

Year:  2013        PMID: 24154792     DOI: 10.1007/s00330-013-3049-x

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  36 in total

1.  Low-dose CT of the lung: potential value of iterative reconstructions.

Authors:  Stephan Baumueller; Anna Winklehner; Christoph Karlo; Robert Goetti; Thomas Flohr; Erich W Russi; Thomas Frauenfelder; Hatem Alkadhi
Journal:  Eur Radiol       Date:  2012-06-15       Impact factor: 5.315

2.  Low contrast agent and radiation dose protocol for hepatic dynamic CT of thin adults at 256-detector row CT: effect of low tube voltage and hybrid iterative reconstruction algorithm on image quality.

Authors:  Takeshi Nakaura; Shinichi Nakamura; Natsuki Maruyama; Yoshinori Funama; Kazuo Awai; Kazunori Harada; Shouzaburou Uemura; Yasuyuki Yamashita
Journal:  Radiology       Date:  2012-05-24       Impact factor: 11.105

Review 3.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

4.  Combined use of automatic tube potential selection with tube current modulation and iterative reconstruction technique in coronary CT angiography.

Authors:  Young Joo Suh; Young Jin Kim; Sae Rom Hong; Yoo Jin Hong; Hye-Jeong Lee; Jin Hur; Byoung Wook Choi
Journal:  Radiology       Date:  2013-10-28       Impact factor: 11.105

5.  Computed tomography of the chest with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination: preliminary observations.

Authors:  Angeliki Neroladaki; Diomidis Botsikas; Sana Boudabbous; Christoph D Becker; Xavier Montet
Journal:  Eur Radiol       Date:  2012-08-15       Impact factor: 5.315

6.  Assessment of image quality on effects of varying tube voltage and automatic tube current modulation with hybrid and pure iterative reconstruction techniques in abdominal/pelvic CT: a phantom study.

Authors:  Varut Vardhanabhuti; Robert Loader; Carl A Roobottom
Journal:  Invest Radiol       Date:  2013-03       Impact factor: 6.016

7.  Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver.

Authors:  Mark E Baker; Frank Dong; Andrew Primak; Nancy A Obuchowski; David Einstein; Namita Gandhi; Brian R Herts; Andrei Purysko; Erick Remer; Neil Vachhani; Neil Vachani
Journal:  AJR Am J Roentgenol       Date:  2012-07       Impact factor: 3.959

8.  Comparison of different body size parameters for individual dose adaptation in body CT of adults.

Authors:  Jan Menke
Journal:  Radiology       Date:  2005-08       Impact factor: 11.105

9.  Clinical impact of an adaptive statistical iterative reconstruction algorithm for detection of hypervascular liver tumours using a low tube voltage, high tube current MDCT technique.

Authors:  Daniele Marin; Kingshuk Roy Choudhury; Rajan T Gupta; Lisa M Ho; Brian C Allen; Sebastian T Schindera; James G Colsher; Ehsan Samei; Rendon C Nelson
Journal:  Eur Radiol       Date:  2013-07-06       Impact factor: 5.315

10.  80-kVp CT using Iterative Reconstruction in Image Space algorithm for the detection of hypervascular hepatocellular carcinoma: phantom and initial clinical experience.

Authors:  Saebeom Hur; Jeong Min Lee; Soo Jin Kim; Ji Hoon Park; Joon Koo Han; Byung Ihn Choi
Journal:  Korean J Radiol       Date:  2012-03-07       Impact factor: 3.500

View more
  8 in total

1.  Aortic CT angiography dose reduction: investigation of optimal noise index and iterative algorithm strength in combination with low kV.

Authors:  Osvaldo Rampado; Stefania Busso; Domenica Garabello; Ezio Marengo; Marco Valerio; Simona Capello; Simona Veglia; Ottavio Davini; Roberto Ropolo
Journal:  Radiol Med       Date:  2015-12-16       Impact factor: 3.469

2.  Optimizing radiation dose by using advanced modelled iterative reconstruction in high-pitch coronary CT angiography.

Authors:  Sonja Gordic; Lotus Desbiolles; Martin Sedlmair; Robert Manka; André Plass; Bernhard Schmidt; Daniela B Husarik; Francesco Maisano; Simon Wildermuth; Hatem Alkadhi; Sebastian Leschka
Journal:  Eur Radiol       Date:  2015-06-03       Impact factor: 5.315

3.  90-kVp low-tube-voltage CT pulmonary angiography in combination with advanced modeled iterative reconstruction algorithm: effects on radiation dose, image quality and diagnostic accuracy for the detection of pulmonary embolism.

Authors:  Doris Leithner; Tatjana Gruber-Rouh; Martin Beeres; Julian L Wichmann; Scherwin Mahmoudi; Simon S Martin; Lukas Lenga; Moritz H Albrecht; Christian Booz; Thomas J Vogl; Jan-Erik Scholtz
Journal:  Br J Radiol       Date:  2018-06-05       Impact factor: 3.039

4.  Diagnostic yield of 90-kVp low-tube-voltage carotid and intracerebral CT-angiography: effects on radiation dose, image quality and diagnostic performance for the detection of carotid stenosis.

Authors:  Doris Leithner; Julian L Wichmann; Scherwin Mahmoudi; Simon S Martin; Moritz H Albrecht; Thomas J Vogl; Jan-Erik Scholtz
Journal:  Br J Radiol       Date:  2018-03-08       Impact factor: 3.039

5.  Hypodense liver lesions in patients with hepatic steatosis: do we profit from dual-energy computed tomography?

Authors:  Johanna Nattenmüller; Waldemar Hosch; Tri-Thien Nguyen; Stephan Skornitzke; Andreas Jöres; Lars Grenacher; Hans-Ulrich Kauczor; Christof M Sommer; Wolfram Stiller
Journal:  Eur Radiol       Date:  2015-05-17       Impact factor: 5.315

6.  Assessment of image quality and low-contrast detectability in abdominal CT of obese patients: comparison of a novel integrated circuit with a conventional discrete circuit detector at different tube voltages.

Authors:  A Euler; T Heye; M Kekelidze; G Bongartz; Z Szucs-Farkas; C Sommer; B Schmidt; Sebastian T Schindera
Journal:  Eur Radiol       Date:  2014-10-15       Impact factor: 5.315

7.  Third-generation dual-source CT of the neck using automated tube voltage adaptation in combination with advanced modeled iterative reconstruction: evaluation of image quality and radiation dose.

Authors:  Jan-Erik Scholtz; Julian L Wichmann; Kristina Hüsers; Moritz H Albrecht; Martin Beeres; Ralf W Bauer; Thomas J Vogl; Boris Bodelle
Journal:  Eur Radiol       Date:  2015-11-11       Impact factor: 5.315

8.  Combined Use of Automatic Tube Voltage Selection and Current Modulation with Iterative Reconstruction for CT Evaluation of Small Hypervascular Hepatocellular Carcinomas: Effect on Lesion Conspicuity and Image Quality.

Authors:  Peijie Lv; Jie Liu; Rui Zhang; Yan Jia; Jianbo Gao
Journal:  Korean J Radiol       Date:  2015-05-13       Impact factor: 3.500

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.