Malfunctions of implantable cardiac devices in patients receiving proton beam therapy: incidence and predictors.

Daniel R Gomez1, Falk Poenisch, Chelsea C Pinnix, Tommy Sheu, Joe Y Chang, Nada Memon, Radhe Mohan, Marc A Rozner, Anne H Dougherty.   

Abstract

PURPOSE: Photon therapy has been reported to induce resets of implanted cardiac devices, but the clinical sequelae of treating patients with such devices with proton beam therapy (PBT) are not well known. We reviewed the incidence of device malfunctions among patients undergoing PBT. METHODS AND MATERIALS: From March 2009 through July 2012, 42 patients with implanted cardiac implantable electronic devices (CIED; 28 pacemakers and 14 cardioverter-defibrillators) underwent 42 courses of PBT for thoracic (23, 55%), prostate (15, 36%), liver (3, 7%), or base of skull (1, 2%) tumors at a single institution. The median prescribed dose was 74 Gy (relative biological effectiveness; range 46.8-87.5 Gy), and the median distance from the treatment field to the CIED was 10 cm (range 0.8-40 cm). Maximum proton and neutron doses were estimated for each treatment course. All CIEDs were checked before radiation delivery and monitored throughout treatment.
RESULTS: Median estimated peak proton and neutron doses to the CIED in all patients were 0.8 Gy (range 0.13-21 Gy) and 346 Sv (range 11-1100 mSv). Six CIED malfunctions occurred in 5 patients (2 pacemakers and 3 defibrillators). Five of these malfunctions were CIED resets, and 1 patient with a defibrillator (in a patient with a liver tumor) had an elective replacement indicator after therapy that was not influenced by radiation. The mean distance from the proton beam to the CIED among devices that reset was 7.0 cm (range 0.9-8 cm), and the mean maximum neutron dose was 655 mSv (range 330-1100 mSv). All resets occurred in patients receiving thoracic PBT and were corrected without clinical incident. The generator for the defibrillator with the elective replacement indicator message was replaced uneventfully after treatment.
CONCLUSIONS: The incidence of CIED resets was about 20% among patients receiving PBT to the thorax. We recommend that PBT be avoided in pacing-dependent patients and that patients with any type of CIED receiving thoracic PBT be followed closely.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 24074931      PMCID: PMC3931127          DOI: 10.1016/j.ijrobp.2013.07.010

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  16 in total

Review 1.  Radiotherapy to patients with artificial cardiac pacemakers.

Authors:  S Sundar; R P Symonds; C Deehan
Journal:  Cancer Treat Rev       Date:  2005-07-06       Impact factor: 12.111

2.  Proton beam therapy interference with implanted cardiac pacemakers.

Authors:  Yoshiko Oshiro; Shinji Sugahara; Mio Noma; Masato Sato; Yuzuru Sakakibara; Takeji Sakae; Yasutaka Hayashi; Hidetsugu Nakayama; Koji Tsuboi; Nobuyoshi Fukumitsu; Ayae Kanemoto; Takayuki Hashimoto; Koichi Tokuuye
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-06-04       Impact factor: 7.038

3.  Measurement of neutron dose equivalent and its dependence on beam configuration for a passive scattering proton delivery system.

Authors:  Xin Wang; Narayan Sahoo; Ronald X Zhu; John R Zullo; Michael T Gillin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-01-25       Impact factor: 7.038

4.  Radiation therapy in patients with implanted cardiac pacemakers and implantable cardioverter defibrillators: a prospective survey in Japan.

Authors:  Toshinori Soejima; Eisaku Yoden; Yasumasa NIshimura; Seiji Ono; Akihiro Yoshida; Haruyuki Fukuda; Noboru Fukuhara; Ryohei Sasaki; Kayoko Tsujino; Yoshiki Norihisa
Journal:  J Radiat Res       Date:  2011       Impact factor: 2.724

Review 5.  The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the perioperative management of patients with implantable defibrillators, pacemakers and arrhythmia monitors: facilities and patient management this document was developed as a joint project with the American Society of Anesthesiologists (ASA), and in collaboration with the American Heart Association (AHA), and the Society of Thoracic Surgeons (STS).

Authors:  George H Crossley; Jeanne E Poole; Marc A Rozner; Samuel J Asirvatham; Alan Cheng; Mina K Chung; T Bruce Ferguson; John D Gallagher; Michael R Gold; Robert H Hoyt; Samuel Irefin; Fred M Kusumoto; Liza Prudente Moorman; Annemarie Thompson
Journal:  Heart Rhythm       Date:  2011-07       Impact factor: 6.343

6.  Management of radiation oncology patients with implanted cardiac pacemakers: report of AAPM Task Group No. 34. American Association of Physicists in Medicine.

Authors:  J R Marbach; M R Sontag; J Van Dyk; A B Wolbarst
Journal:  Med Phys       Date:  1994-01       Impact factor: 4.071

Review 7.  Effect of radiation therapy on the latest generation of pacemakers and implantable cardioverter defibrillators: A systematic review.

Authors:  F Hudson; D Coulshed; E D'Souza; C Baker
Journal:  J Med Imaging Radiat Oncol       Date:  2010-02       Impact factor: 1.735

8.  Effects of scatter radiation on ICD and CRT function.

Authors:  Suraj Kapa; Luis Fong; Charles R Blackwell; Michael G Herman; Paula J Schomberg; David L Hayes
Journal:  Pacing Clin Electrophysiol       Date:  2008-06       Impact factor: 1.976

Review 9.  Treatment of patients with cardiac pacemakers and implantable cardioverter-defibrillators during radiotherapy.

Authors:  Amy N Solan; Merrill J Solan; Greg Bednarz; Michael B Goodkin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-07-01       Impact factor: 7.038

10.  Influence of secondary neutrons induced by proton radiotherapy for cancer patients with implantable cardioverter defibrillators.

Authors:  Takayuki Hashimoto; Tomonori Isobe; Haruko Hashii; Hiroaki Kumada; Hiroshi Tada; Toshiyuki Okumura; Koji Tsuboi; Takeji Sakae; Kazutaka Aonuma; Hideyuki Sakurai
Journal:  Radiat Oncol       Date:  2012-01-29       Impact factor: 3.481

View more
  14 in total

Review 1.  Arrhythmias and Other Electrophysiology Issues in Cancer Patients Receiving Chemotherapy or Radiation.

Authors:  Federico Viganego; Robin Singh; Michael G Fradley
Journal:  Curr Cardiol Rep       Date:  2016-06       Impact factor: 2.931

2.  Radiotherapy-induced malfunctions of cardiac implantable electronic devices in cancer patients.

Authors:  Vincenzo L Malavasi; Giuseppina De Marco; Jacopo F Imberti; Filippo Placentino; Marco Vitolo; Ercole Mazzeo; Gianfranco Cicoria; Edoardo Casali; Vincenzo Turco; Frank Lohr; Giuseppe Boriani
Journal:  Intern Emerg Med       Date:  2019-12-02       Impact factor: 3.397

3.  Effect of X-ray dose rates higher than 8 Gy/min on the functioning of cardiac implantable electronic devices.

Authors:  Kazuhiko Nakamura; Takahiro Aoyama; Naoki Kaneda; Masashi Otsuji; Yoshitaka Minami; Ami Sakuragi; Masaru Nakamura
Journal:  J Radiat Res       Date:  2020-05-22       Impact factor: 2.724

4.  Prospective exploratory analysis of cardiac biomarkers and electrocardiogram abnormalities in patients receiving thoracic radiation therapy with high-dose heart exposure.

Authors:  Daniel R Gomez; Syed Wamique Yusuf; Mark F Munsell; James W Welsh; Zhongxing Liao; Steven H Lin; Hubert Y Pan; Joe Y Chang; Ritsuko Komaki; James D Cox; Mary Frances McAleer; David R Grosshans
Journal:  J Thorac Oncol       Date:  2014-10       Impact factor: 15.609

5.  DEGRO/DGK guideline for radiotherapy in patients with cardiac implantable electronic devices.

Authors:  Benjamin Gauter-Fleckenstein; Carsten W Israel; Marc Dorenkamp; Jürgen Dunst; Mattias Roser; Rainer Schimpf; Volker Steil; Jörg Schäfer; Ulrike Höller; Frederik Wenz
Journal:  Strahlenther Onkol       Date:  2015-03-05       Impact factor: 3.621

Review 6.  A Review and Analysis of Managing Commonly Seen Implanted Devices for Patients Undergoing Radiation Therapy.

Authors:  Maria F Chan; Claire Young; Daphna Gelblum; Chengyu Shi; Carolanne Rincon; Elizabeth Hipp; Jingdong Li; Dongxu Wang
Journal:  Adv Radiat Oncol       Date:  2021-08-06

7.  Pacemaker malfunction associated with proton beam therapy: a report of two cases and review of literature-does field-to-generator distance matter?

Authors:  Tomoko Ueyama; Takeshi Arimura; Takashi Ogino; Naoaki Kondo; Ryutaro Higashi; Fumihiko Nakamura; Soichiro Ito; Takashi Yoshiura; Yoshio Hishikawa
Journal:  Oxf Med Case Reports       Date:  2016-08-29

8.  Assessment of Radiation-Induced Malfunction in Cardiac Implantable Electronic Devices.

Authors:  Amin Zagzoog; Matt Wronski; David H Birnie; Cynthia Yeung; Adrian Baranchuk; Jeffrey S Healey; Mehrdad Golian; Usama Boles; Aldo G Carrizo; Suzette Turner; Ahmed Hassan; Elsayed Ali; Sharath K Kumar; Steve Russell; Mohammed Shurrab; Eugene Crystal
Journal:  CJC Open       Date:  2021-07-11

9.  Safety verification of carbon-ion radiotherapy for patients with cardiac implantable electronic devices (CIEDs).

Authors:  Naoko Okano; Makoto Sakai; Kei Shibuya; Kazuhisa Tsuda; Takao Kanzaki; Masato Sano; Yoshiaki Kaneko; Tatsuya Ohno
Journal:  J Radiat Res       Date:  2022-01-20       Impact factor: 2.724

10.  Active-Scanned Protons and Carbon Ions in Cancer Treatment of Patients With Cardiac Implantable Electronic Devices: Experience of a Single Institution.

Authors:  Katharina Seidensaal; Semi Ben Harrabi; Eberhard Scholz; Malte Ellerbrock; Thomas Haberer; Fabian Weykamp; Matthias Mattke; Stefan E Welte; Klaus Herfarth; Jürgen Debus; Matthias Uhl
Journal:  Front Oncol       Date:  2019-08-22       Impact factor: 6.244

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.