Literature DB >> 18507546

Effects of scatter radiation on ICD and CRT function.

Suraj Kapa1, Luis Fong, Charles R Blackwell, Michael G Herman, Paula J Schomberg, David L Hayes.   

Abstract

BACKGROUND: Effects of direct radiation on implantable cardiac devices have been well studied. However, the effects of scatter radiation are not as clear. Recommendations on management of patients with implantable cardiac devices undergoing radiotherapy are based on limited studies mostly involving pacemakers. We sought to elucidate the effects of scatter radiation on implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy (CRT)-ICDs.
METHODS: We exposed 12 ICDs and eight CRT-ICDs to 400 cGy of scatter radiation from a 6-MV photon beam. Devices were programmed with nominal parameters and interrogated prior to radiation, after each fraction, upon completion of the radiation course and again 1 week later. A retrospective review of patients undergoing radiotherapy at the Mayo Clinic-Rochester between 2002 and 2007 in whom the device was outside the radiation field was also performed. There were 13 patients with devices undergoing radiotherapy during this time period, 12 of whom were interrogated prior to and after radiation.
RESULTS: Interrogation reports were reviewed for device reset or parameter changes. There was no evidence of reset or malfunction during or after radiation. Also, no episodes of device reset, inappropriate sensing or therapy, or changes in programmed parameters were found in our review of patients undergoing radiotherapy.
CONCLUSIONS: Device reset or malfunction associated with scatter radiation likely represents an unpredictable, rare occurrence. While we see no clear contraindication to radiotherapy in patients with ICDs or CRT-ICDs, precautions should be taken to avoid direct radiation exposure and to closely evaluate patient outcomes before and after the radiation course.

Entities:  

Mesh:

Year:  2008        PMID: 18507546     DOI: 10.1111/j.1540-8159.2008.01077.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  17 in total

Review 1.  [Perioperative management of patients with implanted pacemakers or cardioverter/defibrillators. Recommendations of the Austrian Society for Anaesthesiology, Resuscitation and Intensive Care Medicine, the Austrian Society for Cardiology and the Austrian Society for Surgery].

Authors:  H Gombotz; M Anelli Monti; N Leitgeb; M Nürnberg; B Strohmer
Journal:  Anaesthesist       Date:  2009-05       Impact factor: 1.041

Review 2.  Arrhythmias and Other Electrophysiology Issues in Cancer Patients Receiving Chemotherapy or Radiation.

Authors:  Federico Viganego; Robin Singh; Michael G Fradley
Journal:  Curr Cardiol Rep       Date:  2016-06       Impact factor: 2.931

3.  Experimental study of neutron-induced soft errors in modern cardiac pacemakers.

Authors:  Alexandre Trigano; Guillaume Hubert; Jannie Marfaing; Karine Castellani
Journal:  J Interv Card Electrophysiol       Date:  2011-08-24       Impact factor: 1.900

4.  Radiotherapy-induced malfunctions of cardiac implantable electronic devices in cancer patients.

Authors:  Vincenzo L Malavasi; Giuseppina De Marco; Jacopo F Imberti; Filippo Placentino; Marco Vitolo; Ercole Mazzeo; Gianfranco Cicoria; Edoardo Casali; Vincenzo Turco; Frank Lohr; Giuseppe Boriani
Journal:  Intern Emerg Med       Date:  2019-12-02       Impact factor: 3.397

5.  A randomized in vitro evaluation of transient and permanent cardiac implantable electronic device malfunctions following direct exposure up to 10 Gy.

Authors:  Maria Daniela Falco; Domenico Genovesi; Luciana Caravatta; Clelia Di Carlo; Ekaterina Bliakharskaia; Marianna Appignani; Massimiliano Faustino; Nanda Furia; Enrico Di Girolamo
Journal:  Strahlenther Onkol       Date:  2020-06-19       Impact factor: 3.621

Review 6.  [Radiation therapy in patients with cardiac pacemakers or implantable cardioverter defibrillators. Interdisciplinary safety recommendations].

Authors:  M Dorenkamp; C Stromberger; C von Heymann; W Haverkamp; P Wust; M Roser
Journal:  Strahlenther Onkol       Date:  2012-11-11       Impact factor: 3.621

7.  Safe Delivery of Endoscopic Brachytherapy in a Patient with a Dual Chamber Pacemaker.

Authors:  Chang Nancy Wang; Cynthia Yeung; Andres Enriquez; Adrian Baranchuk
Journal:  J Atr Fibrillation       Date:  2019-06-30

8.  Malfunctions of implantable cardiac devices in patients receiving proton beam therapy: incidence and predictors.

Authors:  Daniel R Gomez; Falk Poenisch; Chelsea C Pinnix; Tommy Sheu; Joe Y Chang; Nada Memon; Radhe Mohan; Marc A Rozner; Anne H Dougherty
Journal:  Int J Radiat Oncol Biol Phys       Date:  2013-11-01       Impact factor: 7.038

9.  DEGRO/DGK guideline for radiotherapy in patients with cardiac implantable electronic devices.

Authors:  Benjamin Gauter-Fleckenstein; Carsten W Israel; Marc Dorenkamp; Jürgen Dunst; Mattias Roser; Rainer Schimpf; Volker Steil; Jörg Schäfer; Ulrike Höller; Frederik Wenz
Journal:  Strahlenther Onkol       Date:  2015-03-05       Impact factor: 3.621

10.  A clinical example of extreme dose exposure for an implanted cardioverter-defibrillator : Beyond the DEGRO guidelines.

Authors:  Yoana Hristova; Janett Köhn; Stefanie Preuß; Claus Rödel; Panagiotis Balermpas
Journal:  Strahlenther Onkol       Date:  2017-05-31       Impact factor: 3.621

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.