Effect of X-ray dose rates higher than 8 Gy/min on the functioning of cardiac implantable electronic devices.

Kazuhiko Nakamura1, Takahiro Aoyama2, Naoki Kaneda1, Masashi Otsuji3, Yoshitaka Minami1, Ami Sakuragi1, Masaru Nakamura1.   

Abstract

Direct irradiation may cause malfunctioning of cardiac implantable electronic devices (CIEDs). Therefore, a treatment plan that does not involve direct irradiation of CIEDs should be formulated. However, CIEDs may be directly exposed to radiation because of the sudden intrafractional movement of the patient. The probability of CIED malfunction reportedly depends on the dose rate; however, reports are only limited to dose rates ≤8 Gy/min. The purpose of this study was to investigate the effect of X-ray dose rates >8 Gy/min on CIED function. Four CIEDs were placed at the center of the radiation field and irradiated using 6 MV X-ray with flattening filter free (6 MV FFF) and 10 MV X-ray with flattening filter free (10 MV FFF). The dose rate was 4-14 Gy/min for the 6 MV FFF and 4-24 Gy/min for 10 MV FFF beams. CIED operation was evaluated with an electrocardiogram during each irradiation. Three CIEDs malfunctioned in the 6 MV FFF condition, and all four CIEDs malfunctioned in the 10 MV FFF condition, when the dose rate was >8 Gy/min. Pacing inhibition was the malfunction observed in all four CIEDs. Malfunction occurred simultaneously along with irradiation and simultaneously returned to normal function on stopping the irradiation. An X-ray dose rate >8 Gy/min caused a temporary malfunction due to interference. Therefore, clinicians should be aware of the risk of malfunction and manage patient movement when an X-ray dose rate >8 Gy/min is used for patients with CIEDs.
© The Author(s) 2020. Published by Oxford University Press on behalf of The Japanese Radiation Research Society and Japanese Society for Radiation Oncology.

Entities:  

Keywords:  cardiac implantable electronic devices; dose rate; malfunction; radiotherapy

Mesh:

Year:  2020        PMID: 32253430      PMCID: PMC7299257          DOI: 10.1093/jrr/rraa016

Source DB:  PubMed          Journal:  J Radiat Res        ISSN: 0449-3060            Impact factor:   2.724


  23 in total

Review 1.  Radiotherapy to patients with artificial cardiac pacemakers.

Authors:  S Sundar; R P Symonds; C Deehan
Journal:  Cancer Treat Rev       Date:  2005-07-06       Impact factor: 12.111

2.  Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators.

Authors:  Coen W Hurkmans; Egon Scheepers; Bob G F Springorum; Hans Uiterwaal
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-09-01       Impact factor: 7.038

3.  Radiation-induced effects in multiprogrammable pacemakers and implantable defibrillators.

Authors:  F Rodriguez; A Filimonov; A Henning; C Coughlin; M Greenberg
Journal:  Pacing Clin Electrophysiol       Date:  1991-12       Impact factor: 1.976

4.  Does high-power computed tomography scanning equipment affect the operation of pacemakers?

Authors:  Satoshi Yamaji; Shinobu Imai; Fumio Saito; Hiroshi Yagi; Toshio Kushiro; Takahisa Uchiyama
Journal:  Circ J       Date:  2006-02       Impact factor: 2.993

Review 5.  Radiotherapy-induced pacemaker and implantable cardioverter defibrillator malfunction.

Authors:  Fernando Tondato; Daniel W Ng; Komandoor Srivathsan; Gregory T Altemose; Michele Y Halyard; Luis R Scott
Journal:  Expert Rev Med Devices       Date:  2009-05       Impact factor: 3.166

Review 6.  Current status and future perspective of flattening filter free photon beams.

Authors:  Dietmar Georg; Tommy Knöös; Brendan McClean
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

7.  DEGRO/DGK guideline for radiotherapy in patients with cardiac implantable electronic devices.

Authors:  Benjamin Gauter-Fleckenstein; Carsten W Israel; Marc Dorenkamp; Jürgen Dunst; Mattias Roser; Rainer Schimpf; Volker Steil; Jörg Schäfer; Ulrike Höller; Frederik Wenz
Journal:  Strahlenther Onkol       Date:  2015-03-05       Impact factor: 3.621

8.  Radiotherapy-Induced Malfunction in Contemporary Cardiovascular Implantable Electronic Devices: Clinical Incidence and Predictors.

Authors:  Jonathan D Grant; Garrett L Jensen; Chad Tang; Julianne M Pollard; Stephen F Kry; Sunil Krishnan; Anne H Dougherty; Daniel R Gomez; Marc A Rozner
Journal:  JAMA Oncol       Date:  2015-08       Impact factor: 31.777

Review 9.  Radiotherapy in patients with cardiac pacemakers.

Authors:  A Last
Journal:  Br J Radiol       Date:  1998-01       Impact factor: 3.039

10.  Management of radiation oncology patients with a pacemaker or ICD: a new comprehensive practical guideline in The Netherlands. Dutch Society of Radiotherapy and Oncology (NVRO).

Authors:  Coen W Hurkmans; Joost L Knegjens; Bing S Oei; Ad J J Maas; G J Uiterwaal; Arnoud J van der Borden; Marleen M J Ploegmakers; Lieselot van Erven
Journal:  Radiat Oncol       Date:  2012-11-24       Impact factor: 3.481

View more
  3 in total

Review 1.  A Review and Analysis of Managing Commonly Seen Implanted Devices for Patients Undergoing Radiation Therapy.

Authors:  Maria F Chan; Claire Young; Daphna Gelblum; Chengyu Shi; Carolanne Rincon; Elizabeth Hipp; Jingdong Li; Dongxu Wang
Journal:  Adv Radiat Oncol       Date:  2021-08-06

2.  Implantable cardiac pacemaker failure by cumulative dose effects of flattening filter free beams.

Authors:  Kazuhiko Nakamura; Takahiro Aoyama; Naoki Kaneda; Masashi Otsuji; Yoshitaka Minami; Ami Sakuragi; Masaru Nakamura
Journal:  J Radiat Res       Date:  2021-07-10       Impact factor: 2.724

3.  Local dose rate effects in implantable cardioverter-defibrillators with flattening filter free and flattened photon radiation.

Authors:  Benjamin Gauter-Fleckenstein; Erol Tülümen; Boris Rudic; Martin Borggrefe; Martin Polednik; Jens Fleckenstein
Journal:  Strahlenther Onkol       Date:  2022-03-10       Impact factor: 4.033

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.