Takayuki Hashimoto1, Tomonori Isobe, Haruko Hashii, Hiroaki Kumada, Hiroshi Tada, Toshiyuki Okumura, Koji Tsuboi, Takeji Sakae, Kazutaka Aonuma, Hideyuki Sakurai.
Abstract
BACKGROUND: Although proton radiotherapy is a promising new approach for cancer patients, functional interference is a concern for patients with implantable cardioverter defibrillators (ICDs). The purpose of this study was to clarify the influence of secondary neutrons induced by proton radiotherapy on ICDs.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22284700 PMCID: PMC3283465 DOI: 10.1186/1748-717X-7-10
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Type, description and total number of ICD errors caused by the secondary neutron from 107 Gy proton beam irradiation for each of four ICDs (Marquis DR 7274, Medtronic, Minneapolis, MN)
| Type of error | Explanation | Change in frequency and pulse width | Alarm is sounded | ICD reprogramming | Number of sessions detected by the programmer | Number of errors revealed by analysis of the ICD data logs |
|---|---|---|---|---|---|---|
| Hard error | The ICDs stop generating pulses permanently. | Yes | No | Impossible | 0 | 0 |
| Soft error | 13 | 29 | ||||
| Power-on reset | The ICDs present sudden complete failure and switch over to the safety backup mode. | Yes | Yes | Necessary | 8 | 8 |
| Partial electrical reset | The ICDs show no change in mode of pacing. | No | Yes | Unnecessary | 5 | 7 |
| Minor error | The ICDs show no change in mode of pacing. | No | No | Unnecessary | 0 | 14 |
Abbreviation: ICD = implantable cardioverter defibrillator
Figure 1Schematic of the proton beam delivery system.
Figure 2View of the experimental set-up of proton irradiation to the phantom with implantable cardioverter defibrillators (ICDs) on the central axis of the proton beam. Four ICDs were placed on the contralateral side of the phantom. Proton beams entered the water phantom laterally, and an acrylic plastic board was set behind the ICDs to provide backscatter conditions. Calculated spread-out Bragg peak (SOBP) curves in water and experimental values are included.
Figure 3Results of predicted dose due to secondary neutrons in the phantom determined by the Monte-Carlo calculation.
Figure 4Results of predicted dose due to gamma-rays in the phantom determined by the Monte-Carlo calculation.