Literature DB >> 16002221

Radiotherapy to patients with artificial cardiac pacemakers.

S Sundar1, R P Symonds, C Deehan.   

Abstract

BACKGROUND: The in vitro studies show that the modern cardiac pacemakers utilising the complementary metal-oxide semi-conductor (CMOS) circuitry can be adversely affected by therapeutic radiation. However, the published clinical data are sparse regarding the safety of radiotherapy delivery to patients with artificial pacemakers. Despite the potential risk of life threatening complications, there are no national guidelines and most radiotherapy departments have no formal clinical risk management strategy in place. A literature review was performed to assess the risks involved in irradiating patients with pacemakers and to identify strategies, which minimise the risk of pacemaker malfunction. Recommendations for radiotherapy departments are made.
CONCLUSION: Modern multi-programmable pacemakers are very sensitive to therapeutic megavoltage irradiation. There is no safe radiation threshold for megavoltage radiation. The low energy kilovoltage X-rays used for radiotherapy simulation cause no pacemaker malfunction. Megavoltage radiation can be safely delivered to patients with cardiac pacemakers provided direct irradiation of pacemakers is avoided, adequate monitoring is done during and after irradiation, and the dose to the pacemaker generator is kept below 2 Gy. Close liaison with cardiologists and a pacemaker clinic is essential and radiotherapy departments should have protocols in place to identify and care for cancer patients with pacemakers.

Entities:  

Mesh:

Year:  2005        PMID: 16002221     DOI: 10.1016/j.ctrv.2005.05.002

Source DB:  PubMed          Journal:  Cancer Treat Rev        ISSN: 0305-7372            Impact factor:   12.111


  15 in total

Review 1.  [Perioperative management of patients with implanted pacemakers or cardioverter/defibrillators. Recommendations of the Austrian Society for Anaesthesiology, Resuscitation and Intensive Care Medicine, the Austrian Society for Cardiology and the Austrian Society for Surgery].

Authors:  H Gombotz; M Anelli Monti; N Leitgeb; M Nürnberg; B Strohmer
Journal:  Anaesthesist       Date:  2009-05       Impact factor: 1.041

2.  Management of patients with implanted cardiac devices during radiotherapy: results of a Spanish survey in radiation oncology departments.

Authors:  S Sabater; A Montero; T López Fernández; J J González Ferrer; M Arenas
Journal:  Clin Transl Oncol       Date:  2018-05-23       Impact factor: 3.405

Review 3.  [Radiation therapy in patients with cardiac pacemakers or implantable cardioverter defibrillators. Interdisciplinary safety recommendations].

Authors:  M Dorenkamp; C Stromberger; C von Heymann; W Haverkamp; P Wust; M Roser
Journal:  Strahlenther Onkol       Date:  2012-11-11       Impact factor: 3.621

4.  Effect of X-ray dose rates higher than 8 Gy/min on the functioning of cardiac implantable electronic devices.

Authors:  Kazuhiko Nakamura; Takahiro Aoyama; Naoki Kaneda; Masashi Otsuji; Yoshitaka Minami; Ami Sakuragi; Masaru Nakamura
Journal:  J Radiat Res       Date:  2020-05-22       Impact factor: 2.724

5.  Malfunctions of implantable cardiac devices in patients receiving proton beam therapy: incidence and predictors.

Authors:  Daniel R Gomez; Falk Poenisch; Chelsea C Pinnix; Tommy Sheu; Joe Y Chang; Nada Memon; Radhe Mohan; Marc A Rozner; Anne H Dougherty
Journal:  Int J Radiat Oncol Biol Phys       Date:  2013-11-01       Impact factor: 7.038

6.  Radiotherapy in patients with cardiac implantable electronic devices: clinical and dosimetric aspects.

Authors:  Giulia Riva; Ombretta Alessandro; Ruggero Spoto; Annamaria Ferrari; Cristina Garibaldi; Federica Cattani; Rosa Luraschi; Elena Rondi; Nicola Colombo; Fulvio Lorenzo Francesco Giovenzana; Carlo Maria Cipolla; Mikolaj Winnicki; Martina Persiani; Fabiana Castelluccia; Massimo Sarra Fiore; Roberto Orecchia; Barbara Alicja Jereczek-Fossa
Journal:  Med Oncol       Date:  2018-04-17       Impact factor: 3.064

7.  DEGRO/DGK guideline for radiotherapy in patients with cardiac implantable electronic devices.

Authors:  Benjamin Gauter-Fleckenstein; Carsten W Israel; Marc Dorenkamp; Jürgen Dunst; Mattias Roser; Rainer Schimpf; Volker Steil; Jörg Schäfer; Ulrike Höller; Frederik Wenz
Journal:  Strahlenther Onkol       Date:  2015-03-05       Impact factor: 3.621

Review 8.  Management of radiation therapy patients with cardiac defibrillator or pacemaker.

Authors:  Francesca Salerno; Sara Gomellini; Cristina Caruso; Raffaele Barbara; Daniela Musio; Tamara Coppi; Mario Cardinale; Vincenzo Tombolini; Ugo de Paula
Journal:  Radiol Med       Date:  2015-12-26       Impact factor: 3.469

9.  PAIDEIA: pacemaker and implanted cardioverter defibrillator management in radiation therapy-a survey by the Young Group of the Italian Association of Radiotherapy and Clinical Oncology (AIRO).

Authors:  Anna Rita Alitto; Silvia Chiesa; Pierfrancesco Franco; Michele Fiore; Lorenza Marino; Paolo Borghetti; Isacco Desideri; Daniela Greto; Alba Fiorentino
Journal:  Radiol Med       Date:  2019-12-12       Impact factor: 3.469

10.  Implantable cardiac pacemaker failure by cumulative dose effects of flattening filter free beams.

Authors:  Kazuhiko Nakamura; Takahiro Aoyama; Naoki Kaneda; Masashi Otsuji; Yoshitaka Minami; Ami Sakuragi; Masaru Nakamura
Journal:  J Radiat Res       Date:  2021-07-10       Impact factor: 2.724

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.