Literature DB >> 29497547

Pacemaker malfunction associated with proton beam therapy: a report of two cases and review of literature-does field-to-generator distance matter?

Tomoko Ueyama1,2, Takeshi Arimura1,2, Takashi Ogino2, Naoaki Kondo2, Ryutaro Higashi1, Fumihiko Nakamura1, Soichiro Ito1, Takashi Yoshiura1, Yoshio Hishikawa2.   

Abstract

It is well known that radiotherapy causes malfunctions of cardiac implantable electronic devices such as pacemaker (PM) and implantable cardioverter-defibrillator because of incidental neutron production. Here, we report our experience with two cases of PM reset among seven patients with PM who underwent proton beam therapy (PBT) from January 2011 to April 2015 at our centre. Our experience shows PM reset can occur also with abdominal PBT. In both cases, PM reset was not detected by electrocardiogram (ECG) monitoring but was rather discovered by post-treatment programmer analysis. Our cases suggest that PM malfunction may not always be detected by ECG monitoring and emphasize the importance of daily programmer analysis.

Entities:  

Year:  2016        PMID: 29497547      PMCID: PMC5782476          DOI: 10.1093/omcr/omw049

Source DB:  PubMed          Journal:  Oxf Med Case Reports        ISSN: 2053-8855


INTRODUCTION

It is well known that ionizing radiation can damage cardiac implantable electronic devices (CIEDs) due to incidental neutron production by nuclear fragment reaction, which may result in a loss of cardiac function [1, 2]. In photon radiotherapy, it has been reported that beam energy and the locations of tumours are risk factors for device malfunction [3]. However, there have been no large clinical studies to investigate the influence of proton beam therapy (PBT) on CIED malfunction. Therefore, CIED-dependent patients with malignancies may face unexpected clinical challenges from PBT. In this context, we report our experience with two cases of pacemaker (PM) reset that occurred during PBT.

CASE REPORT

The characteristics of seven patients with PMs who underwent PBT at our centre from January 2011 to April 2015 are presented in Table 1. Among these, we experienced 2 (28.6%) PM resets during PBT.
Table 1:

Characteristics of patients with PMs who underwent PBT

SexAgeTumour locationField-to-generator distance (cm)Heart diseasePM manufacturerPM modePM malfunction
M75Prostate>50Complete A–V blockMedtronicDDDNone
M74Prostate>50Complete A-V blockBoston ScientificDDDNone
M90Lung18CHF, OMIBoston ScientificDDDRNone
M78Larynx13SSSSt. Jude MedicalDDDNone
M87Lung24SSSMedtronicAAIR⇄DDDRYes
M72Pancreas30SSSSt. Jude MedicalDDDYes
M85Lung17Complete A–V blockMedtronicDDDNone

CHF, congestive heart failure; OMI, old myocardial infarction; SSS, sick sinus syndrome.

Characteristics of patients with PMs who underwent PBT CHF, congestive heart failure; OMI, old myocardial infarction; SSS, sick sinus syndrome.

Case 1

This patient was an 87-year-old male diagnosed with primary lung carcinoma in the right lower lobe with a clinical Tumor-Node-Metastasis (TNM) stage of T2bN2M1a according to the Union for International Cancer Control (UICC); he underwent PM implantation (EnRhythmTM; Medtronic, MN, USA) in the left infraclavicular region for sick sinus syndrome (SSS) at the age of 83 years (Fig. 1).
Figure 1:

Chest radiograph of Case 1 before PBT. PM is implanted in the left infraclavicular region. The arrow indicates the primary lesion of lung tumour.

Chest radiograph of Case 1 before PBT. PM is implanted in the left infraclavicular region. The arrow indicates the primary lesion of lung tumour. The pacing mode was AAIR⇄DDDR. The pacing rate was low with an atrial sense–ventricular sense (AS–VS) of 77.2%, atrial sense–ventricular pacing (ASVP) of <0.1%, atrial pacing–ventricular sense (AP–VS) of 22.7% and atrial pacing–ventricular pacing (AP–VP) of <0.1%. Before treatment, we performed a phantom simulation using the same PM model as that implanted in our patient and measured neutron dose (Fig. 2). The measured neutron dose (154.6 mSv) was within the estimated values at simulation (158.4 mSv), and we found no abnormality in PM function. A total of 66 GyE was administered in 10 fractions at three different angles in 210-MeV proton beams (Fig. 3). The pacing leads were partially irradiated with one proton beam. We monitored the electrocardiogram (ECG) throughout irradiation during every treatment but found no abnormality. After the completion of PBT, a change in the PM mode from AAIR⇄DDDR to VVI was detected through programmer analysis on the final treatment day. Retrospectively, the date of the mode change corresponded to the time of irradiation (Treatment Day 8), although no ECG change was observed pre- or post-treatment in the recorded waveforms (Fig. 4). This patient remained asymptomatic after PM reset.
Figure 2:

Phantom simulation to measure neutron dose using a pocket dosimeter. The PM model and settings were the same as that for this patient. The PM leads were partially exposed to one beam, whereas the PM generator was not directly irradiated.

Figure 3:

PBT plan for Case 1. The patient experienced PM reset on the eighth day of 52.8 GyE of a total 66 GyE. A dose–volume histogram (DVH) is shown in the lower right. The volume of 95% of the prescribed dose (V95%) for the gross tumour volume (GTV), clinical target volume (CTV) and planning target volume (PTV) was 100, 100 and 95%, respectively. The maximum dose of the PM chamber was negligible.

Figure 4:

ECG of Case 1 on PBT Day 8 showing no change between pre- and post-treatment.

Phantom simulation to measure neutron dose using a pocket dosimeter. The PM model and settings were the same as that for this patient. The PM leads were partially exposed to one beam, whereas the PM generator was not directly irradiated. PBT plan for Case 1. The patient experienced PM reset on the eighth day of 52.8 GyE of a total 66 GyE. A dose–volume histogram (DVH) is shown in the lower right. The volume of 95% of the prescribed dose (V95%) for the gross tumour volume (GTV), clinical target volume (CTV) and planning target volume (PTV) was 100, 100 and 95%, respectively. The maximum dose of the PM chamber was negligible. ECG of Case 1 on PBT Day 8 showing no change between pre- and post-treatment.

Case 2

This patient was a 72-year-old male with pancreatic cancer (UICC TNM stage T3N0M0). We planned 50 GyE in 25 fractions from two different angles to deliver proton beams of 210 and 150 MeV (Fig. 5). This patient also had a previous history of SSS and implantation of PM (IDENTITY® ADx; St. Jude Medical, CA, USA) via the left subclavian approach 6 years ago. The PM mode was originally programmed to DDD. He was not completely dependent on pacing (AS–VS, 61%; ASVP, 1%; AP–VS, 34% and AP–VP, 4.8%). We simulated a phantom study in the same way as performed in Case 1. The measured dose of neutrons (96.4 mSv) fell well within the estimated values (103.2 mSv), and a PM reset occurred in the simulation. Accordingly, we performed programmer analysis during each step of treatment, which revealed that the reset of the device occurred after Treatment Day 13. Programmer message indicated a reset of the device. As with the first case, there was no change in the ECG pattern of this patient (Fig. 6). Because this was an emergency event, the patient was transported to a cardiac specialty hospital and returned to our centre with the PM settings recovered. After a discussion with the attending cardiologists, we continued the treatment regimen and experienced no problems with PM; we then completed the planned schedule.
Figure 5:

The PBT plan of Case 2. The patient experienced PM reset when a total of 26 GyE was delivered. DVH indicates V95% of GTV, CTV and PTV at 100, 100 and 98%, respectively.

Figure 6:

ECG of Case 2 on treatment Day 13. A reset by programmer analysis was noted. Arrows indicate atrium pacing. No ECG abnormality was recognized, although atrium pacing transiently increased after treatment, which is a common phenomenon.

The PBT plan of Case 2. The patient experienced PM reset when a total of 26 GyE was delivered. DVH indicates V95% of GTV, CTV and PTV at 100, 100 and 98%, respectively. ECG of Case 2 on treatment Day 13. A reset by programmer analysis was noted. Arrows indicate atrium pacing. No ECG abnormality was recognized, although atrium pacing transiently increased after treatment, which is a common phenomenon.

DISCUSSION

Although we confirmed the safety of PMs in these patients through phantom studies and monitored ECGs during treatment, PM reset occurred and we did not realize the mode change of PM for an entire 2 days in the first case. In the second case, we prepared for PM reset during treatment through the simulation, and one actually happened. We believe our simulation to expect the PM reset is effective. However, in both cases, PM malfunction was not detected by ECG monitoring but was rather discovered by programmer analysis. Fortunately, neither of the patients experienced a severe event as both had sustainable heart rates. Moreover, neither experienced an additional reset event during the remaining treatments, and both PMs continued functioning properly after reprogramming. There is little information about the effect of PBT on PM function. Oshiro et al. reported that minor PM malfunction occurred in 2 of 8 patients (25%) receiving high-energy PBT [4]. In addition, Gomez et al. reported that the frequency of CIED reset was ∼20% among patients receiving PBT to the thorax [5]. In our series, PM malfunction occurred in 28.6% (2/7) of the cases, which was higher than that reported by Zaremba et al. who encountered malfunctions in 10 (2.5%) PMs and 4 (6.8%) implanted cardiac defibrillators from 453 patients after radiotherapy using photons or electrons [3]. Thus, the incidence of PM malfunction in PBT may be higher than that in photon radiotherapy. In our cases, neutrons scattered by collimators or the patient's body might have accidentally collided with PM. Therefore, we emphasize the importance to recognize the risk of PM malfunction during PBT, even with devices located at a distance from the target of the irradiated field. Further data about the field-to-generator distance are needed. In the future, pencil beam scanning will possibly be able to reduce the neutron impact on PMs [4]. Zaremba et al. reported that beam energy plays a considerable role in inducing CIED malfunctions in photon radiotherapy [6]. There is a possibility that beam energy of proton might have an impact. Larger-scale study would be needed to verify that low-energy PBT might be safer in PM patients. Both our patients experienced PM malfunction that resulted in a mode change to VVI, which is the safety back-up programme to support infrequent pacing when necessary and might prevent severe PM malfunction. We believe that PBT is feasible for patients who are not completely PM dependent if adequate precautions are considered. We reported two cases of PM malfunction associated with PBT. There is a possibility that the incidence of PM malfunction in PBT is higher than that in photon radiotherapy. It is difficult to accurately predict a PM reset event associated with PBT by simulation or ECG monitoring as both might be insufficient as measures for safe PBT in patients with PMs. Our experience emphasizes the importance of daily programmer analysis after every treatment session.

CONFLICT OF INTEREST STATEMENT

None declared.

FUNDING

There were no sources of funding.

ETHICAL APPROVAL

None required.

CONSENT

Informed consent was obtained from these patients.

GUARANTOR

T.U. will act as guarantor.
  6 in total

1.  Runaway implantable defibrillator--a rare complication of radiation therapy.

Authors:  Jan Nemec
Journal:  Pacing Clin Electrophysiol       Date:  2007-05       Impact factor: 1.976

2.  Proton beam therapy interference with implanted cardiac pacemakers.

Authors:  Yoshiko Oshiro; Shinji Sugahara; Mio Noma; Masato Sato; Yuzuru Sakakibara; Takeji Sakae; Yasutaka Hayashi; Hidetsugu Nakayama; Koji Tsuboi; Nobuyoshi Fukumitsu; Ayae Kanemoto; Takayuki Hashimoto; Koichi Tokuuye
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-06-04       Impact factor: 7.038

3.  The effect of radiotherapy beam energy on modern cardiac devices: an in vitro study.

Authors:  Tomas Zaremba; Annette Ross Jakobsen; Anna Margrethe Thøgersen; Lars Oddershede; Sam Riahi
Journal:  Europace       Date:  2013-08-09       Impact factor: 5.214

4.  Life-threatening pacemaker dysfunction associated with therapeutic radiation: a case report.

Authors:  Andreas Zweng; Reinhard Schuster; Robert Hawlicek; Heinz Stefan Weber
Journal:  Angiology       Date:  2008-04-02       Impact factor: 3.619

5.  Malfunctions of implantable cardiac devices in patients receiving proton beam therapy: incidence and predictors.

Authors:  Daniel R Gomez; Falk Poenisch; Chelsea C Pinnix; Tommy Sheu; Joe Y Chang; Nada Memon; Radhe Mohan; Marc A Rozner; Anne H Dougherty
Journal:  Int J Radiat Oncol Biol Phys       Date:  2013-11-01       Impact factor: 7.038

6.  Risk of device malfunction in cancer patients with implantable cardiac device undergoing radiotherapy: a population-based cohort study.

Authors:  Tomas Zaremba; Annette Ross Jakobsen; Mette Søgaard; Anna Margrethe Thøgersen; Martin Berg Johansen; Laerke Bruun Madsen; Sam Riahi
Journal:  Pacing Clin Electrophysiol       Date:  2015-01-20       Impact factor: 1.976

  6 in total
  5 in total

Review 1.  A Review and Analysis of Managing Commonly Seen Implanted Devices for Patients Undergoing Radiation Therapy.

Authors:  Maria F Chan; Claire Young; Daphna Gelblum; Chengyu Shi; Carolanne Rincon; Elizabeth Hipp; Jingdong Li; Dongxu Wang
Journal:  Adv Radiat Oncol       Date:  2021-08-06

2.  Safety verification of carbon-ion radiotherapy for patients with cardiac implantable electronic devices (CIEDs).

Authors:  Naoko Okano; Makoto Sakai; Kei Shibuya; Kazuhisa Tsuda; Takao Kanzaki; Masato Sano; Yoshiaki Kaneko; Tatsuya Ohno
Journal:  J Radiat Res       Date:  2022-01-20       Impact factor: 2.724

3.  Spot-scanning proton therapy for targets with adjacent cardiac implantable electronic devices - Strategies for breast and head & neck cancer.

Authors:  Line Bjerregaard Stick; Peter Magnus Trock Lægdsmand; Henrik Laurits Bjerre; Morten Høyer; Kenneth Jensen; Maria Fuglsang Jensen; Mads Brix Kronborg; Birgitte Vrou Offersen; Camilla Jensenius Skovhus Kronborg
Journal:  Phys Imaging Radiat Oncol       Date:  2022-02-19

4.  Particle therapy using protons or carbon ions for cancer patients with cardiac implantable electronic devices (CIED): a retrospective multi-institutional study.

Authors:  Takayuki Hashimoto; Yusuke Demizu; Haruko Numajiri; Tomonori Isobe; Shigekazu Fukuda; Masaru Wakatsuki; Haruo Yamashita; Shigeyuki Murayama; Shigeyuki Takamatsu; Hiroyuki Katoh; Kazutoshi Murata; Ryosuke Kohno; Takeshi Arimura; Taeko Matsuura; Yoichi M Ito
Journal:  Jpn J Radiol       Date:  2021-11-15       Impact factor: 2.701

5.  Active-Scanned Protons and Carbon Ions in Cancer Treatment of Patients With Cardiac Implantable Electronic Devices: Experience of a Single Institution.

Authors:  Katharina Seidensaal; Semi Ben Harrabi; Eberhard Scholz; Malte Ellerbrock; Thomas Haberer; Fabian Weykamp; Matthias Mattke; Stefan E Welte; Klaus Herfarth; Jürgen Debus; Matthias Uhl
Journal:  Front Oncol       Date:  2019-08-22       Impact factor: 6.244

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.