It is well known that radiotherapy causes malfunctions of cardiac implantable electronic devices such as pacemaker (PM) and implantable cardioverter-defibrillator because of incidental neutron production. Here, we report our experience with two cases of PM reset among seven patients with PM who underwent proton beam therapy (PBT) from January 2011 to April 2015 at our centre. Our experience shows PM reset can occur also with abdominal PBT. In both cases, PM reset was not detected by electrocardiogram (ECG) monitoring but was rather discovered by post-treatment programmer analysis. Our cases suggest that PM malfunction may not always be detected by ECG monitoring and emphasize the importance of daily programmer analysis.
It is well known that radiotherapy causes malfunctions of cardiac implantable electronic devices such as pacemaker (PM) and implantable cardioverter-defibrillator because of incidental neutron production. Here, we report our experience with two cases of PM reset among seven patients with PM who underwent proton beam therapy (PBT) from January 2011 to April 2015 at our centre. Our experience shows PM reset can occur also with abdominal PBT. In both cases, PM reset was not detected by electrocardiogram (ECG) monitoring but was rather discovered by post-treatment programmer analysis. Our cases suggest that PM malfunction may not always be detected by ECG monitoring and emphasize the importance of daily programmer analysis.
It is well known that ionizing radiation can damage cardiac implantable electronic devices
(CIEDs) due to incidental neutron production by nuclear fragment reaction, which may result
in a loss of cardiac function [1, 2]. In photon radiotherapy, it has been reported
that beam energy and the locations of tumours are risk factors for device malfunction [3]. However, there have been no large clinical
studies to investigate the influence of proton beam therapy (PBT) on CIED malfunction.
Therefore, CIED-dependent patients with malignancies may face unexpected clinical challenges
from PBT. In this context, we report our experience with two cases of pacemaker (PM) reset
that occurred during PBT.
CASE REPORT
The characteristics of seven patients with PMs who underwent PBT at our centre from January
2011 to April 2015 are presented in Table 1. Among these, we experienced 2 (28.6%) PM resets during PBT.
Table 1:
Characteristics of patients with PMs who underwent PBT
Characteristics of patients with PMs who underwent PBTCHF, congestive heart failure; OMI, old myocardial infarction; SSS, sick sinus
syndrome.
Case 1
This patient was an 87-year-old male diagnosed with primary lung carcinoma in the right
lower lobe with a clinical Tumor-Node-Metastasis (TNM) stage of T2bN2M1a according to the
Union for International Cancer Control (UICC); he underwent PM implantation
(EnRhythmTM; Medtronic, MN, USA) in the left infraclavicular region for sick
sinus syndrome (SSS) at the age of 83 years (Fig. 1).
Figure 1:
Chest radiograph of Case 1 before PBT. PM is implanted in the left
infraclavicular region. The arrow indicates the primary lesion of lung
tumour.
Chest radiograph of Case 1 before PBT. PM is implanted in the left
infraclavicular region. The arrow indicates the primary lesion of lung
tumour.The pacing mode was AAIR⇄DDDR. The pacing rate was low with an atrial
sense–ventricular sense (AS–VS) of 77.2%, atrial
sense–ventricular pacing (AS–VP) of <0.1%, atrial
pacing–ventricular sense (AP–VS) of 22.7% and atrial
pacing–ventricular pacing (AP–VP) of <0.1%.Before treatment, we performed a phantom simulation using the same PM model as that
implanted in our patient and measured neutron dose (Fig. 2). The measured neutron dose (154.6 mSv) was within the estimated
values at simulation (158.4 mSv), and we found no abnormality in PM function. A total of
66 GyE was administered in 10 fractions at three different angles in 210-MeV proton beams
(Fig. 3). The pacing leads were
partially irradiated with one proton beam. We monitored the electrocardiogram (ECG)
throughout irradiation during every treatment but found no abnormality. After the
completion of PBT, a change in the PM mode from AAIR⇄DDDR to VVI was detected
through programmer analysis on the final treatment day. Retrospectively, the date of the
mode change corresponded to the time of irradiation (Treatment Day 8), although no ECG
change was observed pre- or post-treatment in the recorded waveforms (Fig. 4). This patient remained asymptomatic after PM
reset.
Figure 2:
Phantom simulation to measure neutron dose using a pocket dosimeter. The PM
model and settings were the same as that for this patient. The PM leads were
partially exposed to one beam, whereas the PM generator was not directly
irradiated.
Figure 3:
PBT plan for Case 1. The patient experienced PM reset on the eighth day of
52.8 GyE of a total 66 GyE. A dose–volume histogram (DVH) is shown in the
lower right. The volume of 95% of the prescribed dose (V95%) for the
gross tumour volume (GTV), clinical target volume (CTV) and planning target volume
(PTV) was 100, 100 and 95%, respectively. The maximum dose of the PM chamber
was negligible.
Figure 4:
ECG of Case 1 on PBT Day 8 showing no change between pre- and
post-treatment.
Phantom simulation to measure neutron dose using a pocket dosimeter. The PM
model and settings were the same as that for this patient. The PM leads were
partially exposed to one beam, whereas the PM generator was not directly
irradiated.PBT plan for Case 1. The patient experienced PM reset on the eighth day of
52.8 GyE of a total 66 GyE. A dose–volume histogram (DVH) is shown in the
lower right. The volume of 95% of the prescribed dose (V95%) for the
gross tumour volume (GTV), clinical target volume (CTV) and planning target volume
(PTV) was 100, 100 and 95%, respectively. The maximum dose of the PM chamber
was negligible.ECG of Case 1 on PBT Day 8 showing no change between pre- and
post-treatment.
Case 2
This patient was a 72-year-old male with pancreatic cancer (UICC TNM stage T3N0M0). We
planned 50 GyE in 25 fractions from two different angles to deliver proton beams of 210
and 150 MeV (Fig. 5). This patient also
had a previous history of SSS and implantation of PM (IDENTITY® ADx; St.
Jude Medical, CA, USA) via the left subclavian approach 6 years ago. The PM mode was
originally programmed to DDD. He was not completely dependent on pacing (AS–VS,
61%; AS–VP, 1%; AP–VS, 34% and AP–VP,
4.8%). We simulated a phantom study in the same way as performed in Case 1. The
measured dose of neutrons (96.4 mSv) fell well within the estimated values (103.2 mSv),
and a PM reset occurred in the simulation. Accordingly, we performed programmer analysis
during each step of treatment, which revealed that the reset of the device occurred after
Treatment Day 13. Programmer message indicated a reset of the device. As with the first
case, there was no change in the ECG pattern of this patient (Fig. 6). Because this was an emergency event, the
patient was transported to a cardiac specialty hospital and returned to our centre with
the PM settings recovered. After a discussion with the attending cardiologists, we
continued the treatment regimen and experienced no problems with PM; we then completed the
planned schedule.
Figure 5:
The PBT plan of Case 2. The patient experienced PM reset when a total of 26
GyE was delivered. DVH indicates V95% of GTV, CTV and PTV at 100, 100 and
98%, respectively.
Figure 6:
ECG of Case 2 on treatment Day 13. A reset by programmer analysis was noted.
Arrows indicate atrium pacing. No ECG abnormality was recognized, although atrium
pacing transiently increased after treatment, which is a common
phenomenon.
The PBT plan of Case 2. The patient experienced PM reset when a total of 26
GyE was delivered. DVH indicates V95% of GTV, CTV and PTV at 100, 100 and
98%, respectively.ECG of Case 2 on treatment Day 13. A reset by programmer analysis was noted.
Arrows indicate atrium pacing. No ECG abnormality was recognized, although atrium
pacing transiently increased after treatment, which is a common
phenomenon.
DISCUSSION
Although we confirmed the safety of PMs in these patients through phantom studies and
monitored ECGs during treatment, PM reset occurred and we did not realize the mode change of
PM for an entire 2 days in the first case. In the second case, we prepared for PM reset
during treatment through the simulation, and one actually happened. We believe our
simulation to expect the PM reset is effective. However, in both cases, PM malfunction was
not detected by ECG monitoring but was rather discovered by programmer analysis.
Fortunately, neither of the patients experienced a severe event as both had sustainable
heart rates. Moreover, neither experienced an additional reset event during the remaining
treatments, and both PMs continued functioning properly after reprogramming.There is little information about the effect of PBT on PM function. Oshiro et
al. reported that minor PM malfunction occurred in 2 of 8 patients (25%)
receiving high-energy PBT [4]. In addition,
Gomez et al. reported that the frequency of CIED reset was
∼20% among patients receiving PBT to the thorax [5]. In our series, PM malfunction occurred in 28.6% (2/7) of
the cases, which was higher than that reported by Zaremba et al. who
encountered malfunctions in 10 (2.5%) PMs and 4 (6.8%) implanted cardiac
defibrillators from 453 patients after radiotherapy using photons or electrons [3]. Thus, the incidence of PM malfunction in PBT may
be higher than that in photon radiotherapy.In our cases, neutrons scattered by collimators or the patient's body might have
accidentally collided with PM. Therefore, we emphasize the importance to recognize the risk
of PM malfunction during PBT, even with devices located at a distance from the target of the
irradiated field. Further data about the field-to-generator distance are needed. In the
future, pencil beam scanning will possibly be able to reduce the neutron impact on PMs
[4]. Zaremba et al. reported
that beam energy plays a considerable role in inducing CIED malfunctions in photon
radiotherapy [6]. There is a possibility that
beam energy of proton might have an impact. Larger-scale study would be needed to verify
that low-energy PBT might be safer in PM patients.Both our patients experienced PM malfunction that resulted in a mode change to VVI, which
is the safety back-up programme to support infrequent pacing when necessary and might
prevent severe PM malfunction. We believe that PBT is feasible for patients who are not
completely PM dependent if adequate precautions are considered.We reported two cases of PM malfunction associated with PBT. There is a possibility that
the incidence of PM malfunction in PBT is higher than that in photon radiotherapy. It is
difficult to accurately predict a PM reset event associated with PBT by simulation or ECG
monitoring as both might be insufficientas measures for safe PBT in patients with PMs. Our
experience emphasizes the importance of daily programmer analysis after every treatment
session.
CONFLICT OF INTEREST STATEMENT
None declared.
FUNDING
There were no sources of funding.
ETHICAL APPROVAL
None required.
CONSENT
Informed consent was obtained from these patients.
Authors: Tomas Zaremba; Annette Ross Jakobsen; Anna Margrethe Thøgersen; Lars Oddershede; Sam Riahi Journal: Europace Date: 2013-08-09 Impact factor: 5.214
Authors: Daniel R Gomez; Falk Poenisch; Chelsea C Pinnix; Tommy Sheu; Joe Y Chang; Nada Memon; Radhe Mohan; Marc A Rozner; Anne H Dougherty Journal: Int J Radiat Oncol Biol Phys Date: 2013-11-01 Impact factor: 7.038
Authors: Katharina Seidensaal; Semi Ben Harrabi; Eberhard Scholz; Malte Ellerbrock; Thomas Haberer; Fabian Weykamp; Matthias Mattke; Stefan E Welte; Klaus Herfarth; Jürgen Debus; Matthias Uhl Journal: Front Oncol Date: 2019-08-22 Impact factor: 6.244