Reliability of the corneal thickness measurements with the Pentacam HR imaging system and ultrasound pachymetry.

László Módis1, Eszter Szalai, Gábor Németh, András Berta.   

Abstract

PURPOSE: To obtain the central corneal thickness (CCT) of healthy eyes using the rotating Scheimpflug imaging system and compare these results with conventional ultrasound pachymetry values. Interoperator reliability and repeatability of the different techniques were also determined.
METHODS: CCT was measured on the right eye of 46 healthy subjects with the Pentacam HR (Pentacam High Resolution; Oculus, Wetzlar, Germany) and contact ultrasound pachymetry (AL-2000; Tomey, Tennenlohe, Germany) by 2 independent investigators.
RESULTS: Both observers obtained significantly higher CCT readings with the Pentacam HR imaging system (572 ± 33 and 575 ± 31 μm) than with the ultrasound device (546 ± 27 and 548 ± 28 μm) (P < 0.0001, paired Student t test). Strong and significant correlation was found between the 2 pachymetry techniques regardless of the investigator (r = 0.845 and 0.831, Spearman correlation coefficient) (P < 0.0001). For the Pentacam HR values, statistically significant difference was detected between the observers (P = 0.033). For both pachymetry methods, the degree of agreement between the investigators was reflected by the high interoperator intraclass correlation coefficient value (ICC = 0.96). No significant difference was disclosed between the 3 consecutive readings of either instrument (P > 0.05, analysis of variance).
CONCLUSIONS: Significant difference in CCT values between the high-resolution Pentacam imaging system and the ultrasound pachymeter suggests that the 2 techniques cannot be used interchangeably in clinical practice.

Mesh:

Year:  2011        PMID: 21139499     DOI: 10.1097/ICO.0b013e318200096a

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  5 in total

1.  Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry.

Authors:  Mohammad Soleimani; Hassan Hashemi; Shiva Mehravaran; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Akbar Fotouhi
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

2.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

3.  Topographic typology in a consecutive series of refractive surgery candidates.

Authors:  Seyed-Farzad Mohammadi; Vahid Mohammadzadeh; Sakineh Kadivar; Amir-Houshang Beheshtnejad; Amir Hossein Norooznezhad; Seyed-Hassan Hashemi
Journal:  Int Ophthalmol       Date:  2017-07-04       Impact factor: 2.031

4.  Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound.

Authors:  J Jorge; Jl Rosado; Ja Díaz-Rey; Jm González-Méijome
Journal:  Clin Ophthalmol       Date:  2013-02-26

5.  Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy.

Authors:  Nader Nassiri; Kourosh Sheibani; Sare Safi; Saman Nassiri; Alireza Ziaei; Alireza Ziaee; Farnaz Haji; Shiva Mehravaran; Nariman Nassiri
Journal:  J Ophthalmic Vis Res       Date:  2014-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.