Literature DB >> 23960880

Intra-ocular lens power calculation in patients with high axial myopia before cataract surgery.

Raouf El-Nafees1, Ashraf Moawad, Hanem Kishk, Walid Gaafar.   

Abstract

PURPOSE: To evaluate the accuracy of different formulas used for IOL power calculation in patients with high axial myopia undergoing cataract surgery.
METHODS: A prospective clinical study was carried out on 53 eyes of 51 patients with an axial length from 25.5 to 31.4 mm including 21 males (41.2%) and 30 females (58.8%). Calculation of the IOL power to be implanted was done by three available IOL power formulas; Haigis formula, SRK/T formula, and Holladay I formula. The mean error (ME) was calculated from the difference between the formula predicted refractive error and the actual post operative refractive error.
RESULTS: There was no statistically significant difference between the mean error of the three formulas used in the overall performance or in the axial length subcategories. SRK/T formula caused the smallest mean error, (+0.17 D). Haigis formula showed a higher ME (+0.21 D) and Holladay formula caused a myopic postoperative refractive error (-0.20 D).
CONCLUSION: The calculation of IOL power in patients with high axial myopia using the third or the fourth generation formulas help in improvement of the accuracy of the calculation and decreasing the post operative refractive error. SRK/T formula showed the lowest mean error, however, there was not statistically significant difference between the three formulas used, neither in the overall performance, nor in axial length subcategories.

Entities:  

Keywords:  Cataract surgery; IOL power calculation; Myopia

Year:  2010        PMID: 23960880      PMCID: PMC3729345          DOI: 10.1016/j.sjopt.2010.03.006

Source DB:  PubMed          Journal:  Saudi J Ophthalmol        ISSN: 1319-4534


  15 in total

1.  Clinical results using the Holladay 2 intraocular lens power formula.

Authors:  K J Hoffer
Journal:  J Cataract Refract Surg       Date:  2000-08       Impact factor: 3.351

2.  Prediction of the effective postoperative (intraocular lens) anterior chamber depth.

Authors:  Thomas Olsen
Journal:  J Cataract Refract Surg       Date:  2006-03       Impact factor: 3.351

3.  Accuracy of intraocular lens power prediction using the Hoffer Q, Holladay 1, Holladay 2, and SRK/T formulas.

Authors:  Julio Narváez; Grenith Zimmerman; R Doyle Stulting; Daniel H Chang
Journal:  J Cataract Refract Surg       Date:  2006-12       Impact factor: 3.351

4.  Intraocular lens power calculation with an improved anterior chamber depth prediction algorithm.

Authors:  T Olsen; L Corydon; H Gimbel
Journal:  J Cataract Refract Surg       Date:  1995-05       Impact factor: 3.351

5.  Development of the SRK/T intraocular lens implant power calculation formula.

Authors:  J A Retzlaff; D R Sanders; M C Kraff
Journal:  J Cataract Refract Surg       Date:  1990-05       Impact factor: 3.351

6.  Accuracy of lens power calculations with the biconvex and meniscus intraocular lenses.

Authors:  H J Shammas
Journal:  Am J Ophthalmol       Date:  1988-11-15       Impact factor: 5.258

7.  Results in patients with high and low power intraocular lenses.

Authors:  R C Drews
Journal:  J Cataract Refract Surg       Date:  1986-03       Impact factor: 3.351

8.  The Hoffer Q formula: a comparison of theoretic and regression formulas.

Authors:  K J Hoffer
Journal:  J Cataract Refract Surg       Date:  1993-11       Impact factor: 3.351

9.  Comparison of the SRK II formula and other second generation formulas.

Authors:  D R Sanders; J Retzlaff; M C Kraff
Journal:  J Cataract Refract Surg       Date:  1988-03       Impact factor: 3.351

10.  Discrete nuclear sclerosis in young patients with myopia.

Authors:  B J Kaufman; J Sugar
Journal:  Arch Ophthalmol       Date:  1996-10
View more
  7 in total

1.  Respond to: A study regarding efficacy of various intraocular lens power calculation formulas in a subset of Indian myopic population.

Authors:  Ashish Mitra; Elesh Jain; Alok Sen; Shubhi Tripathi
Journal:  Indian J Ophthalmol       Date:  2015-03       Impact factor: 1.848

2.  Accuracy of axial length measurements from immersion B-scan ultrasonography in highly myopic eyes.

Authors:  Qing-Hua Yang; Bing Chen; Guang-Hua Peng; Zhao-Hui Li; Yi-Fei Huang
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

3.  Performance of the SRK/T formula using A-Scan ultrasound biometry after phacoemulsification in eyes with short and long axial lengths.

Authors:  Yunus Karabela; Mustafa Eliacik; Faruk Kaya
Journal:  BMC Ophthalmol       Date:  2016-07-08       Impact factor: 2.209

4.  The Comparative Study of Refractive Index Variations between Haigis, Srk/T and Hoffer-Q Formulas Used for Preoperative Biometry Calculation in Patients with the Axial Length >25 mm.

Authors:  Hasan Razmjoo; Hosein Atarzadeh; Neda Kargar; Mohammad Behfarnia; Kobra Nasrollahi; Ali Kamali
Journal:  Adv Biomed Res       Date:  2017-07-14

5.  Predictability of 6 Intraocular Lens Power Calculation Formulas in People With Very High Myopia.

Authors:  Yi-Ching Chu; Tzu-Lun Huang; Pei-Yao Chang; Wei-Ting Ho; Yung-Ray Hsu; Shu-Wen Chang; Jia-Kang Wang
Journal:  Front Med (Lausanne)       Date:  2022-04-08

6.  A Comparative Study on Accuracy of SRK-T And Haigis Formulas in IOL Power Calculation in Axial Myopes Undergoing Cataract Surgery.

Authors:  H M Amulya Padmini; K H Dhananjaya; Shivaraj B Budihal; Gajaraj T Naik
Journal:  J Pharm Bioallied Sci       Date:  2022-07-13

7.  Accuracy of Intraocular Lens Power Calculation Formulas for Highly Myopic Eyes.

Authors:  Yichi Zhang; Xiao Ying Liang; Shu Liu; Jacky W Y Lee; Srinivasan Bhaskar; Dennis S C Lam
Journal:  J Ophthalmol       Date:  2016-03-29       Impact factor: 1.909

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.