Brian Derbyshire1, Videshnandan V Raut.
Abstract
Historically, wire markers were attached to cemented all-plastic acetabular cups to demarcate the periphery and to measure socket wear. The wire shape was either a semi-circle passing over the pole of the cup, or a circle around the cup equator. More recently, "double-D" shaped markers were introduced with a part-circular aspect passing over the pole and a semi-circular aspect parallel to the equatorial plane. This configuration enabled cup retroversion to be distinguished from anteversion. In this study, the accuracy of radiographic measurement of cup orientation and wear was assessed for cups with "double-D" and circular markers. Each cup was attached to a measurement jig which could vary the anteversion/retroversion and internal/external rotation of the cup. A metal femoral head was fixed within the socket and radiographic images were created for all combinations of cup orientation settings. The images were measured using software with automatic edge detection, and cup orientation and zero-wear accuracies were determined for each setting. The median error for cup version measurements was similar for both types of wire marker (0.2° double-D marker, -0.24° circular marker), but measurements of the circular marker were more repeatable. The median inclination errors were 2.05° (double-D marker) and 0.23° (circular marker). The median overall "zero wear" errors were 0.19 mm (double-D marker) and 0.03 mm (circular marker). Measurements of the circular wire marker were much more repeatable.Entities:
Mesh:
Year: 2013 PMID: 23813165 PMCID: PMC3880854 DOI: 10.5301/hipint.5000038
Source DB: PubMed Journal: Hip Int ISSN: 1120-7000 Impact factor: 2.135
Fig. 1a) Measurement jig. The outer cylinder can be tilted to put the cup in ante/retroversion. The arrow shows the marker bead which was the target for the central x-ray beam. b) Inner cylinder face showing pin holes for attaching the cup at different rotational positions.
Fig. 2The thick line represents the shape of the polar aspect of the DDM wire marker image. Its thickness represents the error band (scatter) for edge detected points on the image of this wire. The dashed lines represent just some of the possible ellipses that could be fitted within this error band.
Fig. 3Radiographic images of the DDM cup set at different orientations: a) 20° anteversion, 20° internal rotation; b) 10° retroversion, 20° internal rotation; c) 20° anteversion, 20° external rotation; d) 10° retroversion, 20° external rotation.
STATISTICS (IN MM) OF THE “ZERO WEAR” MEASUREMENTS AT EACH SETTING OF THE MEASUREMENT JIG FOR CUPS WITH A “DOUBLE-D” MARKER (DDM) AND A CIRCULAR MARKER (CM)
| DDM Error | CM Error | Difference Statistics | ||||
|---|---|---|---|---|---|---|
| Setting | Mean | SD | Mean | SD | P | CI |
| 35I_10AV_0R | 0.322 | 0.163 | 0.056 | 0.029 | 0.003 | 0.115 to 0.417 |
| 35I_20AV_0R | 0.220 | 0.119 | 0.020 | 0.008 | 0.009 | 0.092 to 0.308 |
| 35I_10RV_0R | 0.270 | 0.161 | 0.017 | 0.010 | 0.012 | 0.106 to 0.399 |
| 35I_20RV_0R | 0.259 | 0.132 | 0.050 | 0.026 | 0.011 | 0.087 to 0.331 |
| 35I_10AV_10IR | 0.161 | 0.073 | 0.056 | 0.029 | 0.008 | 0.034 to.0.177 |
| 35I_20AV_10IR | 0.144 | 0.078 | 0.020 | 0.008 | 0.012 | 0.052 to 0.195 |
| 35I_10RV_10IR | 0.105 | 0.024 | 0.017 | 0.010 | 0.000 | 0.064 to 0.112 |
| 35I_20RV_10IR | 0.140 | 0.045 | 0.050 | 0.026 | 0.002 | 0.042 to 0.137 |
| 35I_10AV_20IR | 0.302 | 0.150 | 0.056 | 0.029 | 0.009 | 0.108 to 0.385 |
| 35I_20AV_20IR | 0.082 | 0.030 | 0.020 | 0.008 | 0.004 | 0.033 to 0.090 |
| 35I_10RV_20IR | 0.172 | 0.056 | 0.017 | 0.010 | 0.001 | 0.103 to 0.207 |
| 35I_20RV_20IR | 0.175 | 0.072 | 0.050 | 0.026 | 0.007 | 0.055 to 0.194 |
| 35I_10AV_10ER | 0.510 | 0.267 | 0.056 | 0.029 | 0.009 | 0.210 to 0.699 |
| 35I_20AV_10ER | 0.105 | 0.054 | 0.020 | 0.008 | 0.012 | 0.035 to 0.135 |
| 35I_10RV_10ER | 0.117 | 0.061 | 0.017 | 0.010 | 0.010 | 0.044 to 0.157 |
| 35I_20RV_10ER | 0.330 | 0.176 | 0.050 | 0.026 | 0.011 | 0.118 to 0.441 |
| 35I_10AV_20ER | 0.403 | 0.224 | 0.056 | 0.029 | 0.012 | 0.142 to 0.552 |
| 35I_20AV_20ER | 0.268 | 0.121 | 0.020 | 0.008 | 0.004 | 0.137 to 0.358 |
| 35I_10RV_20ER | 0.416 | 0.295 | 0.017 | 0.010 | 0.020 | 0.131 to 0.668 |
| 35I_20RV_20ER | 0.302 | 0.128 | 0.050 | 0.026 | 0.004 | 0.134 to 0.371 |
STATISTICS (IN DEGREES) OF THE VERSION MEASUREMENT ERRORS AT EACH SETTING OF THE MEASUREMENT JIG FOR CUPS WITH A “DOUBLE-D” MARKER (DDM) AND A CIRCULAR MARKER (CM)
| DDM Error | CM Error | Difference Statistics | ||||
|---|---|---|---|---|---|---|
| Settings | Mean | SD | Mean | SD | P | CI |
| 35I_10AV_0R | 0.18 | 0.49 | -0.25 | 0.13 | 0.142 | -0.190 to 1.040 |
| 35I_20AV_0R | 0.10 | 0.29 | -0.45 | 0.10 | 0.012 | 0.170 to 0.930 |
| 35I_10RV_0R | 0.33 | 0.71 | -0.23 | 0.05 | 0.221 | -0.325 to 1.425 |
| 35I_20RV_0R | 0.83 | 0.36 | -0.18 | 0.05 | 0.011 | 0.556 to 1.444 |
| 35I_10AV_10IR | 0.05 | 0.13 | -0.25 | 0.13 | 0.017 | 0.077 to 0.523 |
| 35I_20AV_10IR | -0.03 | 0.29 | -0.45 | 0.10 | 0.031 | 0.053 to 0.797 |
| 35I_10RV_10IR | 1.40 | 0.27 | -0.23 | 0.05 | 0.001 | 1.288 to 1.962 |
| 35I_20RV_10IR | 0.80 | 0.22 | -0.18 | 0.05 | 0.002 | 0.704 to 1.246 |
| 35I_10AV_20IR | 0.50 | 0.22 | -0.25 | 0.13 | 0.001 | 0.442 to 1.058 |
| 35I_20AV_20IR | -0.40 | 0.18 | -0.45 | 0.10 | 0.648 | -0.205 to 0.305 |
| 35I_10RV_20IR | 1.40 | 0.38 | -0.23 | 0.05 | 0.003 | 1.152 to 2.098 |
| 35I_20RV_20IR | 1.20 | 0.44 | -0.18 | 0.05 | 0.008 | 0.834 to 1.916 |
| 35I_10AV_10ER | -0.10 | 0.35 | -0.25 | 0.13 | 0.448 | -0.302 to 0.602 |
| 35I_20AV_10ER | -0.15 | 0.21 | -0.45 | 0.10 | 0.041 | 0.017 to 0.583 |
| 35I_10RV_10ER | 0.45 | 0.24 | -0.23 | 0.05 | 0.009 | 0.377 to 0.973 |
| 35I_20RV_10ER | 0.20 | 0.88 | -0.18 | 0.05 | 0.455 | -0.698 to 1.448 |
| 35I_10AV_20ER | -0.18 | 0.17 | -0.25 | 0.13 | 0.51 | -0.187 to 0.337 |
| 35I_20AV_20ER | -0.30 | 0.36 | -0.45 | 0.10 | 0.448 | -0.302 to 0.602 |
| 35I_10RV_20ER | 0.05 | 0.26 | -0.23 | 0.05 | 0.128 | -0.054 to 0.604 |
| 35I_20RV_20ER | 0.13 | 0.76 | -0.18 | 0.05 | 0.487 | -0.630 to 1.230 |
I = inclination.
AV = anteversion.
RV = retroversion.
IR/ER = internal/external rotation change about the polar axis (left sided cup).
STATISTICS (IN DEGREES) OF THE INCLINATION MEASUREMENT ERRORS AT EACH SETTING OF THE MEASUREMENT JIG FOR CUPS WITH A “DOUBLE-D” MARKER (DDM) AND A CIRCULAR MARKER (CM)
| DDM Error | CM Error | Difference Statistics | ||||
|---|---|---|---|---|---|---|
| Settings | Mean | SD | Mean | SD | P | CI |
| 35I_10AV_0R | 2.28 | 0.05 | 0.25 | 0.10 | < 0.0001 | 1.888 to 2.162 |
| 35I_20AV_0R | 2.28 | 0.17 | 0.3 | 0.08 | < 0.0001 | 1.743 to 2.207 |
| 35I_10RV_0R | 2.20 | 0.16 | 0.2 | 0.00 | < 0.0001 | 1.800 to 2.200 |
| 35I_20RV_0R | 2.13 | 0.26 | 0.2 | 0.08 | < 0.0001 | 1.588 to 2.262 |
| 35I_10AV_10IR | 2.13 | 0.17 | 0.25 | 0.10 | < 0.0001 | 1.633 to 2.117 |
| 35I_20AV_10IR | 1.90 | 0.14 | 0.3 | 0.08 | < 0.0001 | 1.400 to 1.800 |
| 35I_10RV_10IR | 1.95 | 0.13 | 0.2 | 0.00 | 0.0001 | 1.592 to 1.908 |
| 35I_20RV_10IR | 1.80 | 0.14 | 0.2 | 0.08 | < 0.0001 | 1.400 to 1.800 |
| 35I_10AV_20IR | 1.98 | 0.22 | 0.25 | 0.10 | < 0.0001 | 1.427 to 2.023 |
| 35I_20AV_20IR | 2.60 | 0.08 | 0.3 | 0.08 | < 0.0001 | 2.159 to 2.441 |
| 35I_10RV_20IR | 1.33 | 0.10 | 0.2 | 0.00 | < 0.0001 | 1.008 to 1.242 |
| 35I_20RV_20IR | 1.23 | 0.19 | 0.2 | 0.08 | < 0.0001 | 0.773 to 1.277 |
| 35I_10AV_10ER | 1.98 | 0.22 | 0.25 | 0.10 | < 0.0001 | 1.427 to 2.023 |
| 35I_20AV_10ER | 1.95 | 0.06 | 0.3 | 0.08 | < 0.0001 | 1.528 to 1.772 |
| 35I_10RV_10ER | 2.23 | 0.13 | 0.2 | 0.00 | < 0.0001 | 1.871 to 2.179 |
| 35I_20RV_10ER | 2.60 | 0.12 | 0.2 | 0.08 | < 0.0001 | 2.227 to 2.573 |
| 35I_10AV_20ER | 1.75 | 0.19 | 0.25 | 0.10 | < 0.0001 | 1.236 to 1.764 |
| 35I_20AV_20ER | 1.65 | 0.17 | 0.3 | 0.08 | < 0.0001 | 1.116 to 1.584 |
| 35I_10RV_20ER | 2.25 | 0.31 | 0.2 | 0.00 | 0.0009 | 1.670 to 2.430 |
| 35I_20RV_20ER | 2.85 | 0.19 | 0.2 | 0.08 | < 0.0001 | 2.395 to 2.905 |
I = inclination.
AV = anteversion.
RV = retroversion.
IR/ER = internal/external rotation change about the polar axis (left sided cup).