Measurement accuracy in acetabular cup wear. Three retrospective methods compared with Roentgen stereophotogrammetry.

T Ilchmann1, B Mjöberg, H Wingstrand.   

Abstract

The accuracy of three methods (the simple and noncomputerized Scheier-Sandel and Charnley-Duo methods and the computerized Ein Bild Roentgen Analyse [EBRA] method) for retrospective wear measurements of the acetabular cup from standard pelvis radiographs was studied. Measurements on 13 hip prostheses were compared with those obtained by roentgen stereophotogrammetry analysis. The Scheier-Sandel method had the lowest accuracy and the EBRA method had the best accuracy. The Charnley-Duo method was almost as good when starting analysis 3 months after surgery and is easier to use. The EBRA method is useful for accurate measurements on a small number of patients; the Charnley-Duo method is recommended for clinical wear studies on a larger number of patients.

Entities:  

Mesh:

Year:  1995        PMID: 9273375     DOI: 10.1016/s0883-5403(05)80208-6

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  15 in total

1.  Wear patterns of, and wear volume formulae for, cylindrically elongated acetabular cup liners.

Authors:  James Shih-Shyn Wu; Shu-Ling Hsu; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2010-04-28       Impact factor: 2.602

2.  Early polyethylene wear and excessive acetabular granuloma in an uncemented HA-coated total hip arthroplasty--midterm results of a prospective study.

Authors:  Jörg Axel Karl Ohnsorge; Jack Davis; Uwe Maus; Muharrem Saklak; Markus Weisskopf; Dieter Christian Wirtz
Journal:  HSS J       Date:  2006-09

3.  Evaluating the accuracy of wear formulae for acetabular cup liners.

Authors:  James Shih-Shyn Wu; Shu-Ling Hsu; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2009-10-10       Impact factor: 2.602

4.  Clinical and radiographic results and wear performance in different generations of a cementless porous-coated acetabular cup.

Authors:  E García-Rey; E García-Cimbrelo
Journal:  Int Orthop       Date:  2007-01-31       Impact factor: 3.075

5.  The final follow-up plain radiograph is sufficient for clinical evaluation of polyethylene wear in total hip arthroplasty. A study of validity and reliability.

Authors:  Maiken Stilling; Kristian Larsen; Niels T Andersen; Kjeld Søballe; Søren Kold; Ole Rahbek
Journal:  Acta Orthop       Date:  2010-10       Impact factor: 3.717

6.  Migration pattern of cementless press fit cups in the presence of stabilizing screws in total hip arthroplasty.

Authors:  C Zilkens; S Djalali; B Bittersohl; T Kälicke; C N Kraft; R Krauspe; Marcus Jäger
Journal:  Eur J Med Res       Date:  2011-03-28       Impact factor: 2.175

7.  Effectiveness of computer-navigated minimally invasive total hip surgery compared to conventional total hip arthroplasty: design of a randomized controlled trial.

Authors:  Inge H F Reininga; Robert Wagenmakers; Inge van den Akker-Scheek; A Dennis Stant; Johan W Groothoff; Sjoerd K Bulstra; Wiebren Zijlstra; Martin Stevens
Journal:  BMC Musculoskelet Disord       Date:  2007-01-11       Impact factor: 2.362

8.  2D measurements of cup orientation are less reliable than 3D measurements.

Authors:  Kinner Davda; Niall Smyth; Justin P Cobb; Alister J Hart
Journal:  Acta Orthop       Date:  2015-02-12       Impact factor: 3.717

9.  No superior performance of hydroxyapatite-coated acetabular cups over porous-coated cups.

Authors:  Kestutis Valancius; Kjeld Søballe; Poul T Nielsen; Mogens B Laursen
Journal:  Acta Orthop       Date:  2013-10-31       Impact factor: 3.717

10.  Superior accuracy of model-based radiostereometric analysis for measurement of polyethylene wear: A phantom study.

Authors:  M Stilling; S Kold; S de Raedt; N T Andersen; O Rahbek; K Søballe
Journal:  Bone Joint Res       Date:  2012-08-01       Impact factor: 5.853

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.