Correction of acetabular cup orientation measurements for X-ray beam offset.

Brian Derbyshire1.   

Abstract

It is important to achieve correct orientation of an acetabular cup in order to avoid impingement, dislocation and increased wear. Cup orientation is usually assessed from a pelvis radiograph in which the central X-ray beam has been focused on the pubic symphysis. However, the angular offset of this beam from a beam passing through the centre of the cup causes rotational artefact in the cup image. Three simple algorithms for correcting this 3D rotational artefact are proposed, and these were validated using 3D CAD techniques by simulating the X-ray focus, target point and film position, as well as cup orientation. The accuracy of the correction was assessed in different practical scenarios. Cup version could be corrected to within a few hundredths of a degree, and inclination to within 0.2 degrees , if the X-ray target position was accurately located. However, only a small error in the correction was incurred if this position was erroneously located by as much as 10mm. The error in correcting the rotation position about the axis of symmetry depended on the cup size and anteversion setting, and it varied sinusoidally according to the position of measurement. Generally, it was within any possible measurement error. For the largest cup size considered (60mm outer diameter), the error ranged between -0.4 degrees and 0.8 degrees at the highest anteversion setting (30 degrees ).

Entities:  

Mesh:

Year:  2008        PMID: 18359261     DOI: 10.1016/j.medengphy.2008.02.001

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  13 in total

1.  Assessment of inter- and intra-observer reliability in the determination of radiographic version and inclination of the cup in metal-on-metal hip resurfacing.

Authors:  Aleksi Reito; Timo Puolakka; Antti Paakkala; Jorma Pajamäki
Journal:  Int Orthop       Date:  2011-08-12       Impact factor: 3.075

2.  Central X-ray beam correction of radiographic acetabular cup measurement after THA: an experimental study.

Authors:  T Schwarz; M Weber; M Wörner; T Renkawitz; J Grifka; B Craiovan
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-10-06       Impact factor: 2.924

Review 3.  Radiographic assessment of the cup orientation after total hip arthroplasty: a literature review.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Ann Transl Med       Date:  2020-02

4.  How Reliable is the Acetabular Cup Position Assessment from Routine Radiographs?

Authors:  Jaime A Carvajal Alba; Heather K Vincent; Jagdeep S Sodhi; Loren L Latta; Hari K Parvataneni
Journal:  Iowa Orthop J       Date:  2017

5.  Three-dimensional orientation and location-dependent varying rules of radiographic angles of the acetabular cup.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Int Orthop       Date:  2018-02-17       Impact factor: 3.075

6.  The synergetic effect of pelvic rotation and X-ray offset on radiographic angles of the acetabular cup.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Guo-Fei Sun; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Med Biol Eng Comput       Date:  2019-09-09       Impact factor: 2.602

7.  Silent soft tissue pathology is common with a modern metal-on-metal hip arthroplasty.

Authors:  Henry Wynn-Jones; Rory Macnair; James Wimhurst; Nish Chirodian; Brian Derbyshire; Andoni Toms; John Cahir
Journal:  Acta Orthop       Date:  2011-04-19       Impact factor: 3.717

8.  Validation of a new 2-D technique for radiographic wear measurement of cemented, highly cross-linked polyethylene acetabular cups.

Authors:  Brian Derbyshire; Bilal Barkatali
Journal:  Med Eng Phys       Date:  2017-07-03       Impact factor: 2.242

9.  The efficacy of a "double-D-shaped" wire marker for radiographic measurement of acetabular cup orientation and wear.

Authors:  Brian Derbyshire; Videshnandan V Raut
Journal:  Hip Int       Date:  2013-05-24       Impact factor: 2.135

10.  2D measurements of cup orientation are less reliable than 3D measurements (Acta Orthop 2015; 86 (4): 485-90.).

Authors: 
Journal:  Acta Orthop       Date:  2016-03-15       Impact factor: 3.717

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.