Measurement of breast tissue composition with dual energy cone-beam computed tomography: a postmortem study.

Huanjun Ding1, Justin L Ducote, Sabee Molloi.   

Abstract

PURPOSE: To investigate the feasibility of a three-material compositional measurement of water, lipid, and protein content of breast tissue with dual kVp cone-beam computed tomography (CT) for diagnostic purposes.
METHODS: Simulations were performed on a flat panel-based computed tomography system with a dual kVp technique in order to guide the selection of experimental acquisition parameters. The expected errors induced by using the proposed calibration materials were also estimated by simulation. Twenty pairs of postmortem breast samples were imaged with a flat-panel based dual kVp cone-beam CT system, followed by image-based material decomposition using calibration data obtained from a three-material phantom consisting of water, vegetable oil, and polyoxymethylene plastic. The tissue samples were then chemically decomposed into their respective water, lipid, and protein contents after imaging to allow direct comparison with data from dual energy decomposition.
RESULTS: Guided by results from simulation, the beam energies for the dual kVp cone-beam CT system were selected to be 50 and 120 kVp with the mean glandular dose divided equally between each exposure. The simulation also suggested that the use of polyoxymethylene as the calibration material for the measurement of pure protein may introduce an error of -11.0%. However, the tissue decomposition experiments, which employed a calibration phantom made out of water, oil, and polyoxymethylene, exhibited strong correlation with data from the chemical analysis. The average root-mean-square percentage error for water, lipid, and protein contents was 3.58% as compared with chemical analysis.
CONCLUSIONS: The results of this study suggest that the water, lipid, and protein contents can be accurately measured using dual kVp cone-beam CT. The tissue compositional information may improve the sensitivity and specificity for breast cancer diagnosis.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23718593      PMCID: PMC3663856          DOI: 10.1118/1.4802734

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  45 in total

1.  Normalized glandular dose (DgN) coefficients for arbitrary X-ray spectra in mammography: computer-fit values of Monte Carlo derived data.

Authors:  John M Boone
Journal:  Med Phys       Date:  2002-05       Impact factor: 4.071

2.  Measurement of breast density with dual X-ray absorptiometry: feasibility.

Authors:  John A Shepherd; Karla M Kerlikowske; Rebecca Smith-Bindman; Harry K Genant; Steve R Cummings
Journal:  Radiology       Date:  2002-05       Impact factor: 11.105

3.  Flat panel detector-based cone beam computed tomography with a circle-plus-two-arcs data acquisition orbit: preliminary phantom study.

Authors:  Ruola Ning; Xiangyang Tang; David Conover; Rongfeng Yu
Journal:  Med Phys       Date:  2003-07       Impact factor: 4.071

4.  A dual-energy subtraction technique for microcalcification imaging in digital mammography--a signal-to-noise analysis.

Authors:  Michael R Lemacks; S Cheenu Kappadath; Chris C Shaw; Xinming Liu; Gary J Whitman
Journal:  Med Phys       Date:  2002-08       Impact factor: 4.071

5.  Compression paddle tilt correction in full-field digital mammograms.

Authors:  Michiel G J Kallenberg; Nico Karssemeijer
Journal:  Phys Med Biol       Date:  2012-01-13       Impact factor: 3.609

6.  Quantification of breast density with spectral mammography based on a scanned multi-slit photon-counting detector: a feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Phys Med Biol       Date:  2012-07-06       Impact factor: 3.609

7.  Effect of compression paddle tilt correction on volumetric breast density estimation.

Authors:  Michiel G J Kallenberg; Carla H van Gils; Mariëtte Lokate; Gerard J den Heeten; Nico Karssemeijer
Journal:  Phys Med Biol       Date:  2012-07-27       Impact factor: 3.609

8.  Breast composition measurement with a cadmium-zinc-telluride based spectral computed tomography system.

Authors:  Huanjun Ding; Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

9.  Comparison of full-field digital mammography with screen-film mammography for cancer detection: results of 4,945 paired examinations.

Authors:  J M Lewin; R E Hendrick; C J D'Orsi; P K Isaacs; L J Moss; A Karellas; G A Sisney; C C Kuni; G R Cutter
Journal:  Radiology       Date:  2001-03       Impact factor: 11.105

10.  Cone-beam volume CT breast imaging: feasibility study.

Authors:  Biao Chen; Ruola Ning
Journal:  Med Phys       Date:  2002-05       Impact factor: 4.071

View more
  7 in total

1.  Breast density evaluation using spectral mammography, radiologist reader assessment, and segmentation techniques: a retrospective study based on left and right breast comparison.

Authors:  Sabee Molloi; Huanjun Ding; Stephen Feig
Journal:  Acad Radiol       Date:  2015-05-29       Impact factor: 3.173

2.  Postmortem validation of breast density using dual-energy mammography.

Authors:  Sabee Molloi; Justin L Ducote; Huanjun Ding; Stephen A Feig
Journal:  Med Phys       Date:  2014-08       Impact factor: 4.071

3.  Breast tissue decomposition with spectral distortion correction: a postmortem study.

Authors:  Huanjun Ding; Bo Zhao; Pavlo Baturin; Farnaz Behroozi; Sabee Molloi
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

4.  Breast density quantification with cone-beam CT: a post-mortem study.

Authors:  Travis Johnson; Huanjun Ding; Huy Q Le; Justin L Ducote; Sabee Molloi
Journal:  Phys Med Biol       Date:  2013-12-07       Impact factor: 3.609

5.  Breast tissue characterization with photon-counting spectral CT imaging: a postmortem breast study.

Authors:  Huanjun Ding; Michael J Klopfer; Justin L Ducote; Fumitaro Masaki; Sabee Molloi
Journal:  Radiology       Date:  2014-05-07       Impact factor: 11.105

6.  Diagnostic accuracy of cone-beam breast computed tomography: a systematic review and diagnostic meta-analysis.

Authors:  Johannes Uhlig; Annemarie Uhlig; Lorenz Biggemann; Uwe Fischer; Joachim Lotz; Susanne Wienbeck
Journal:  Eur Radiol       Date:  2018-09-25       Impact factor: 5.315

Review 7.  Dedicated breast CT: state of the art-Part I. Historical evolution and technical aspects.

Authors:  Yueqiang Zhu; Avice M O'Connell; Yue Ma; Aidi Liu; Haijie Li; Yuwei Zhang; Xiaohua Zhang; Zhaoxiang Ye
Journal:  Eur Radiol       Date:  2021-08-03       Impact factor: 7.034

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.