Normalized glandular dose (DgN) coefficients for arbitrary X-ray spectra in mammography: computer-fit values of Monte Carlo derived data.

John M Boone1.   

Abstract

Normalized glandular dose (DgN) values have been reported by several investigators for specific spectra, however for unconventional or unanticipated x-ray spectra considered for use in mammography, practical methods are not available for DgN computation. In this study, the previously validated SIERRA Monte Carlo code was used to compute the normalized glandular dose coefficients for monoenergetic energies from 8 keV to 50 keV. The overall mammography geometry used was a 65 cm source to image distance, a 1.2 cm air gap between the breast and the detector, and breast thicknesses ranging from 2 to 9 cm. A 4 mm layer of skin was also modeled, and semicircular breast radii of 8.5 cm and 10.0 cm were studied. Breast compositions of 0% glandular, 50% glandular, and 100% glandular were evaluated. The Monte Carlo derived DgN results demonstrated coefficients of variation less than 0.3%. The monoenergetic DgN values, DgN(E), were computer fit using commercial software and the best fit equations are reported. All fits resulted in r2 values of 0.9999 or better. The computer fit equations, along with easy to use spectral modeling routines, are available electronically on the web.

Entities:  

Mesh:

Year:  2002        PMID: 12033583     DOI: 10.1118/1.1472499

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  60 in total

1.  Dedicated breast CT: radiation dose for circle-plus-line trajectory.

Authors:  Srinivasan Vedantham; Linxi Shi; Andrew Karellas; Frederic Noo
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

2.  First application of computed radiology to mammography with synchrotron radiation.

Authors:  E Quai; R Longo; F Zanconati; G Jaconelli; M Tonutti; A Abrami; F Arfelli; D Dreossi; G Tromba; M A Cova
Journal:  Radiol Med       Date:  2012-06-28       Impact factor: 3.469

3.  Dual-energy contrast-enhanced digital mammography: initial clinical results.

Authors:  Clarisse Dromain; Fabienne Thibault; Serge Muller; Françoise Rimareix; Suzette Delaloge; Anne Tardivon; Corinne Balleyguier
Journal:  Eur Radiol       Date:  2010-09-14       Impact factor: 5.315

4.  Dosimetric characterization of a dedicated breast computed tomography clinical prototype.

Authors:  Ioannis Sechopoulos; Steve Si Jia Feng; Carl J D'Orsi
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

5.  Quantification of breast density with spectral mammography based on a scanned multi-slit photon-counting detector: a feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Phys Med Biol       Date:  2012-07-06       Impact factor: 3.609

6.  Breast dose in mammography is about 30% lower when realistic heterogeneous glandular distributions are considered.

Authors:  Andrew M Hernandez; J Anthony Seibert; John M Boone
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

7.  Optimization of contrast-enhanced spectral mammography depending on clinical indication.

Authors:  Clarisse Dromain; Sandra Canale; Sylvie Saab-Puong; Ann-Katherine Carton; Serge Muller; Eva Maria Fallenberg
Journal:  J Med Imaging (Bellingham)       Date:  2014-12-30

Review 8.  Breast cancer imaging: a perspective for the next decade.

Authors:  Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

9.  Breast tissue characterization with photon-counting spectral CT imaging: a postmortem breast study.

Authors:  Huanjun Ding; Michael J Klopfer; Justin L Ducote; Fumitaro Masaki; Sabee Molloi
Journal:  Radiology       Date:  2014-05-07       Impact factor: 11.105

10.  Quantitative contrast-enhanced spectral mammography based on photon-counting detectors: A feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Med Phys       Date:  2017-06-28       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.