Evaluation of test protocol variables for dental implant fatigue research.

Cornell K Lee1, Matthias Karl, J Robert Kelly.   

Abstract

OBJECTIVES: This work begins to explore the influence of cycling rate and environment on fatigue testing of dental implants according to the ISO protocol 14801.
METHODS: Twenty-four Straumann implants (4.1 mm x 12 mm) were tested up to five million cycles per ISO 14801: loaded at either 2 or 30 Hz in room air at 25 degrees C or normal saline at 37 degrees C (n=6 per group). Implant displacements/cycle were captured during all testing. Fracture-surface features were examined using scanning electron microscopy (n=12). Two complimentary methods were developed to estimate fatigue crack growth rates.
RESULTS: Failures (bulk fracture) were found to be bi-modally distributed, either <350,000 cycles or >1.5 million cycles at both cycling rates. Following initial crack formation, fatigue crack growth required merely 1100-4200 cycles to failure. Initial crack pop-in was statistically more likely under 2 Hz than 30 Hz (chi(2), p<0.05) but testing in air and normal saline were equivalent in terms of likelihood of fracture versus runout (chi(2), p>0.6). On a microscopic level, fatigue crack growth rates appears to be similar at 2 and 30 Hz, but may be slower in the presence of saline versus dry at 2 Hz. SIGNIFICANCE: Implant failure under fatigue conditions involved "classic" damage mechanisms. Failure appears more likely at 2 Hz than 30 Hz for reasons that remain to be elucidated. Saline may enable chemically assisted crack growth involving grain boundaries during the stage of fatigue crack growth, but did not influence likelihood of failure.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19646746     DOI: 10.1016/j.dental.2009.07.003

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  11 in total

1.  Accuracy of torque-limiting devices for mini-implant removal: an in vitro study.

Authors:  A Pauls; M Nienkemper; D Drescher
Journal:  J Orofac Orthop       Date:  2013-05-08       Impact factor: 1.938

2.  Reliability evaluation of alumina-blasted/acid-etched versus laser-sintered dental implants.

Authors:  Erika O Almeida; Amilcar C Freitas Júnior; Estevam A Bonfante; Nelson R F A Silva; Paulo G Coelho
Journal:  Lasers Med Sci       Date:  2012-07-28       Impact factor: 3.161

3.  Effect of loading frequency on cyclic fatigue lifetime of a standard-diameter implant with an internal abutment connection.

Authors:  Yuanyuan Duan; Jason A Griggs
Journal:  Dent Mater       Date:  2018-09-13       Impact factor: 5.304

4.  Stability of external and internal implant connections after a fatigue test.

Authors:  Paulo Cesar Pinheiro Feitosa; Ana Paula Barbosa de Lima; Laís Regiane Silva-Concílio; William Cunha Brandt; Ana Christina Claro Neves
Journal:  Eur J Dent       Date:  2013-07

5.  Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.

Authors:  Caio Marques Martins; Elimário Venturin Ramos; Simone Kreve; Geraldo Alberto Pinheiro de Carvalho; Aline Batista Gonçalves Franco; Luís Guilherme Scavone de Macedo; Alecsandro de Moura Silva; Sérgio Candido Dias
Journal:  Dent Res J (Isfahan)       Date:  2019 Mar-Apr

6.  New Dental Implant with 3D Shock Absorbers and Tooth-Like Mobility-Prototype Development, Finite Element Analysis (FEA), and Mechanical Testing.

Authors:  Avram Manea; Grigore Baciut; Mihaela Baciut; Dumitru Pop; Dan Sorin Comsa; Ovidiu Buiga; Veronica Trombitas; Horatiu Colosi; Ileana Mitre; Roxana Bordea; Marius Manole; Manuela Lenghel; Simion Bran; Florin Onisor
Journal:  Materials (Basel)       Date:  2019-10-21       Impact factor: 3.623

7.  Mechanical testing of thin-walled zirconia abutments.

Authors:  Luigi Canullo; Paulo G Coelho; Estevam A Bonfante
Journal:  J Appl Oral Sci       Date:  2013 Jan-Feb       Impact factor: 2.698

8.  Cast accuracy obtained from different impression techniques at different implant angulations (in vitro study).

Authors:  Enas A Elshenawy; Ahmed M Alam-Eldein; Fadel A Abd Elfatah
Journal:  Int J Implant Dent       Date:  2018-03-20

9.  A finite element analysis of novel vented dental abutment geometries for cement-retained crown restorations.

Authors:  Lucas C Rodriguez; Juliana N Saba; Clark A Meyer; Kwok-Hung Chung; Chandur Wadhwani; Danieli C Rodrigues
Journal:  Clin Exp Dent Res       Date:  2016-06-14

10.  Mechanical Assessment of Fatigue Characteristics between Single- and Multi-Directional Cyclic Loading Modes on a Dental Implant System.

Authors:  Won Hyeon Kim; Eun Sung Song; Kyung Won Ju; Dohyung Lim; Dong-Wook Han; Tae-Gon Jung; Yong-Hoon Jeong; Jong-Ho Lee; Bongju Kim
Journal:  Materials (Basel)       Date:  2020-03-27       Impact factor: 3.623

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.