Influence of the interface design on the yield force of the implant-abutment complex before and after cyclic mechanical loading.

Marc Philipp Dittmer1, Stephanie Dittmer, Lothar Borchers, Philipp Kohorst, Meike Stiesch.   

Abstract

PURPOSE: The aim of this in vitro study was to evaluate the implant-abutment assembly of Astra Tech (AST), Bego (BEG), Camlog (CAM), Friadent (FRI), Nobel Biocare (NOB) and Straumann (STR) with respect to yield force before and after cyclic fatigue, using a static overload test with a test set-up according to ISO 14801.
METHODS: Ten specimens of each type were split into two homogenous groups: one half was tested for static yield force without any further treatment (control), whereas the other one underwent one million cycles of mechanical loading with 100N as the upper load limit. For load-to-failure testing, specimens were then placed in a stainless steel jig and loaded in a universal testing machine under an angle of 30° with respect to the implant axis until failure. Load-displacement curves were analyzed and the yield forces at which non-linear behaviour set in (Fp) were recorded. Statistical analysis was performed using one-way ANOVA and t-test, respectively, with the level of significance set at 0.05.
RESULTS: Statistical analysis revealed that the type of implant-abutment connection has a significant influence on Fp (p<0.001). Furthermore, dynamic loading proved to significantly influence Fp of BEG and CAM (p<0.001).
CONCLUSION: None of the implant-abutment types tested would be expected to fail under clinically relevant forces, but the type of implant-abutment connection significantly influences the yield force Fp.
Copyright © 2011 Japan Prosthodontic Society. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21398198     DOI: 10.1016/j.jpor.2011.02.002

Source DB:  PubMed          Journal:  J Prosthodont Res        ISSN: 1883-1958            Impact factor:   4.642


  10 in total

1.  Tooth-implant-supported posterior fixed dental prostheses with zirconia frameworks: 3-year clinical result.

Authors:  Florian Beuer; Caroline Sachs; Julian Groesser; Jan-Frederik Gueth; Michael Stimmelmayr
Journal:  Clin Oral Investig       Date:  2015-09-24       Impact factor: 3.573

2.  Removal torque evaluation of three different abutment screws for single implant restorations after mechanical cyclic loading.

Authors:  T Paepoemsin; P A Reichart; P Chaijareenont; F P Strietzel; P Khongkhunthian
Journal:  Oral Implantol (Rome)       Date:  2016-11-16

3.  Implant fracture under dynamic fatigue loading: influence of embedded angle and depth of implant.

Authors:  Hiroaki Suzuki; Yoshiaki Hata; Fumihiko Watanabe
Journal:  Odontology       Date:  2015-12-23       Impact factor: 2.634

4.  Reliability evaluation of alumina-blasted/acid-etched versus laser-sintered dental implants.

Authors:  Erika O Almeida; Amilcar C Freitas Júnior; Estevam A Bonfante; Nelson R F A Silva; Paulo G Coelho
Journal:  Lasers Med Sci       Date:  2012-07-28       Impact factor: 3.161

5.  Influence of the connection design and titanium grades of the implant complex on resistance under static loading.

Authors:  Su-Jung Park; Suk-Won Lee; Richard Leesungbok; Su-Jin Ahn
Journal:  J Adv Prosthodont       Date:  2016-10-21       Impact factor: 1.904

6.  The fracture strength by a torsion test at the implant-abutment interface.

Authors:  Fumihiko Watanabe; Kazuhiko Hiroyasu; Kazuhiko Ueda
Journal:  Int J Implant Dent       Date:  2015-10-10

7.  Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.

Authors:  Caio Marques Martins; Elimário Venturin Ramos; Simone Kreve; Geraldo Alberto Pinheiro de Carvalho; Aline Batista Gonçalves Franco; Luís Guilherme Scavone de Macedo; Alecsandro de Moura Silva; Sérgio Candido Dias
Journal:  Dent Res J (Isfahan)       Date:  2019 Mar-Apr

8.  A Comparative 3D Finite Element Computational Study of Three Connections.

Authors:  Davide Farronato; Mattia Manfredini; Andrea Stevanello; Veronica Campana; Lorenzo Azzi; Marco Farronato
Journal:  Materials (Basel)       Date:  2019-09-26       Impact factor: 3.623

9.  Mechanical testing of thin-walled zirconia abutments.

Authors:  Luigi Canullo; Paulo G Coelho; Estevam A Bonfante
Journal:  J Appl Oral Sci       Date:  2013 Jan-Feb       Impact factor: 2.698

10.  In situ microradioscopy and microtomography of fatigue-loaded dental two-piece implants.

Authors:  Wolfram Wiest; Simon Zabler; Alexander Rack; Christian Fella; Andreas Balles; Katja Nelson; Rainer Schmelzeisen; Randolf Hanke
Journal:  J Synchrotron Radiat       Date:  2015-10-09       Impact factor: 2.616

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.