Yilei Shao1, Aizhu Tao, Hong Jiang, Meixiao Shen, Jianguang Zhong, Fan Lu, Jianhua Wang. 1. Bascom Palmer Eye Institute, University of Miami, Miami, FL, 33136, USA ; School of Ophthalmology and Optometry, Wenzhou Medical College, Wenzhou, Zhejiang, China ; These authors contributed equally to this work.
Abstract
We demonstrated a novel approach of imaging the anterior segment including the ciliary muscle using combined and synchronized two spectral domain optical coherence tomography devices (SD-OCT). In one SD-OCT, a Complementary Metal-Oxide-Semiconductor Transistor (CMOS) camera and an alternating reference arm was used to image the anterior segment from the cornea to the lens. Another SD-OCT for imaging the ciliary muscle was equipped with a light source with a center wavelength of 1,310 nm and a bandwidth of 75 nm. Repeated measurements were performed under relaxed and 4.00 D accommodative stimulus states in six eyes from 6 subjects. We also imaged dynamic changes in the anterior segment in one eye during accommodation. The biometry of the anterior segment and the ciliary muscle was obtained. The combined system appeared to be capable to simultaneously real-time image the biometry of the anterior segment, including the ciliary muscle, in vivo during accommodation.
We demonstrated a novel approach of imaging the anterior segment including the ciliary muscle using combined and synchronized two spectral domain optical coherence tomography devices (SD-OCT). In one SD-OCT, a Complementary Metal-Oxide-Semiconductor Transistor (CMOS) camera and an alternating reference arm was used to image the anterior segment from the cornea to the lens. Another SD-OCT for imaging the ciliary muscle was equipped with a light source with a center wavelength of 1,310 nm and a bandwidth of 75 nm. Repeated measurements were performed under relaxed and 4.00 D accommodative stimulus states in six eyes from 6 subjects. We also imaged dynamic changes in the anterior segment in one eye during accommodation. The biometry of the anterior segment and the ciliary muscle was obtained. The combined system appeared to be capable to simultaneously real-time image the biometry of the anterior segment, including the ciliary muscle, in vivo during accommodation.
Entities:
Keywords:
(170.3880) Medical and biological imaging; (170.4500) Optical coherence tomography; (170.4580) Optical diagnostics for medicine; (330.4460) Ophthalmic optics and devices; (330.7322) Visual optics, accommodation
As a muscle-induced activity, accommodation in the human eye is a highly fluctuant and dynamic
process. According to the classic accommodative theory of Helmholtz, the ciliary muscle contracts
when accommodation occurs, causing the release of the zonular fibers, allowing the crystalline lens
to increase curvature and thickness, and leading to an increase in the ocular refractive power to
change the focus on a close target [1]. Initially, the ciliary
muscle plays a role, while the crystalline lens serves as an effector in the process of
accommodation. The capacity for accommodation declines with age, evidently due to the degeneration
in both the elasticity of the crystalline lens and the activity of the ciliary muscle; this leads to
presbyopia [2-7].
However, this theory may not fully explain the development of presbyopia. Investigating the
crystalline lens response to ciliary muscle contraction may be beneficial to better understanding
the aging problem. Nevertheless, the relationship between the crystalline lens response and ciliary
muscle contraction has not been fully studied, mainly due to the technical difficulty of
simultaneously imaging the anterior segment and the ciliary muscle in real-time.Optical coherence tomography (OCT) is a non-contact, non-invasive imaging technology with a high
resolution and high scan speed that has the capability to image both the anterior segment [8-11] and the ciliary
muscle [12-14].
Swept-source OCT (SS-OCT), equipped with the high-speed reflective Fabry-Perot tunable lasers has
been demonstrated to attain long image depths without the drop off in the sensitivity and the axial
resolution versus depth [15-17]. Recently, using a custom tunable light source, SS-OCT has been demonstrated to
have a unique performance for anterior segment imaging [18].
Using spectral domain OCT (SD-OCT) with two channels, a prolonged scan depth can be obtained for
imaging the entire anterior segment [19,20]. Based on the previous work of our group [8,21] and that of others [22], the scan depth of the spectral domain OCT with a galvanometer-based switchable
reference arm appeared to be sufficient for in vivo imaging of the ocular anterior
segment through the pupil. Using a simple alternation of the reference arms, the entire anterior
segment has been imaged by overlapping [8,21] or stitching [21,22]. Compared to our previous work [8,21], the long scan depth system used in the present
study was optimized by improving the scan speed using complementary metal-oxide-semiconductor
transistor (CMOS) technology. It is worth noting that these systems mentioned above are not capable
of simultaneously imaging the anterior segment through the pupil and the ciliary muscle through the
sclera. We combined the long scan depth system with another SD-OCT equipped with a light source
centered at a wavelength of 1,310 nm to enable the assessment of the anterior segment including the
ciliary muscle at different accommodative states. The goal of the present study was to demonstrate a
novel approach of imaging the anterior segment including the ciliary muscle by combining two
spectral domain optical coherence tomography devices.
2. Methods
2.1 Two SD-OCT devices
The improved system, which operates at 840 nm, was based on our previously reported long scan
depth SD-OCT using a CCD camera [8,21]. The previous system demonstrated its suitability for imaging the entire
anterior segment of the eye during accommodation [8,21]. The main benefit found in the improved system was the
implementation of a CMOS camera to increase its scan speed for real-time imaging (Fig. 1
). Briefly, a superluminescent diode (SLD, InPhenix, IPSDD0808, Livermore, CA, USA) with a
central wavelength of 840 nm and the full-width at half bandwidths of 50 nm was used as the light
source. The power output on the cornea was set to be 1.25 mW which was well below the safe cut-off
value according to American National Standard Institute (ANSI) Z136.1. The spectrometer was composed
of a collimating lens (f = 50 mm, OZ Optics, Ottawa, Canada), an 1800 lines/mm volume holography
transmission grating, an image enlargement lens (Schneider, f = 240 mm, New York, NY), and a line
array CMOS camera (Basler Sprint spL4096-140k; Basler AG, Germany). The acquisition speed of the
camera was up to 70,000 A-lines/s, and the interference spectrum data were transferred using the
image acquisition board (PCIe-1429, National Instruments, USA). The measured scan depth was 12.55
mm; the axial resolution of the system near the zero-delay line was approximately 7.0 μm in
air; and the sensitivity was 99 dB near the zero delay line, with a 51 dB drop at the maximal
imaging depth. The driver was developed in a setup with Labview (National Instrument, Austin, TX)
running on a computer with Windows 7 (64 bit).
Fig. 1
Schematic diagram depicting the combined spectral-domain OCT systems. SLD1310:
superluminescent diode with a central wavelength of 1,310 nm, SLD840: superluminescent
diode with a central wavelength of 840 nm, FC: fiber coupler, PC: polarization controller,
CL1-4: collimating lenses, DC: dispersion compensator, L1-3: objective lenses,
M1-3: refractive mirror, GM: galvanometer mirror, NDF: neutral density filter, LCD:
liquid-crystal display, DG: diffraction grating, CMOS: complementary metal-oxide-semiconductor
transistor camera, InGaAs: indium gallium arsenide. Insert: X-Y alignment.
Schematic diagram depicting the combined spectral-domain OCT systems. SLD1310:
superluminescent diode with a central wavelength of 1,310 nm, SLD840: superluminescent
diode with a central wavelength of 840 nm, FC: fiber coupler, PC: polarization controller,
CL1-4: collimating lenses, DC: dispersion compensator, L1-3: objective lenses,
M1-3: refractive mirror, GM: galvanometer mirror, NDF: neutral density filter, LCD:
liquid-crystal display, DG: diffraction grating, CMOS: complementary metal-oxide-semiconductor
transistor camera, InGaAs: indium gallium arsenide. Insert: X-Y alignment.The beam from the SLD was split into the sample arm and the reference arm using a 50:50 fiber
coupler. In order to extend the effective image depth, the reference arm was specially designed,
which consisted of two reflective mirrors (usually one reflective mirror in the reference arm). The
two mirrors were mounted on two axial stages, respectively, which could adjust the optical path
difference (OPD), amounting to two reference arms with different OPDs. Moreover, a galvanometer
optical scanner with a silver mirror (GM in Fig. 1) was
implanted in the reference arm, which could rapidly turn the beam between the two mirrors that was
synchronized with the scanning. Thus a set of images that included the two frames was continually
obtained during image acquisition. The difference in the OPD between the two mirrors was set about
11 mm, which was used to place the zero-delay lines of the two images on the top and the bottom of
the anterior segment. Through image overlapping, the sensitivity drop was compensated, and the image
enhancement was realized as described in our previous work [8,21]. This device (anterior segment OCT, AS-OCT) was
used to image the entire anterior segment from the cornea to the crystalline lens through the
pupil.Another SD-OCT (CM-OCT) was used to image the ciliary muscle for acceptable penetration (Fig. 1). The spectrometer was custom-developed (Bioptigen, Durham,
NC), and this device has been described elsewhere for imaging the anterior segment [23,24]. The light source was
an SLD centered at a wavelength of 1,310 nm with a full-width at half maximum bandwidths of 75 nm.
The output on the eye was set to 2.6 mW which was well below the ANSI safety limit for maximum
permissible exposure of 15.4 mW at the wavelength [25]. In
addition, most of the 1310 nm beam did not reach the retina because the iris blocked the light and
the beam was directed to the limbal area. The scan depth was 3.8 mm with an axial resolution of ~8.0
μm in air. An indium gallium arsenide (InGaAs) camera (SU1024, Goodrich Sensors Unlimited
Inc, Princeton, NJ) was used, and the system could image at 7 frames per second with an acquisition
speed of 7,000 A-lines/s, corresponding to a frame of 1,000 A-lines. In real-time imaging, the
images were continually acquired, processed and displayed. An electronic shutter (JML Optical,
Rochester, NY) was implanted in the reference arm and used to insert a sync signal into OCT image
acquisition during real-time imaging. The system ran with a proprietary program for data
acquisition.
2.2 Combined sample arms and the synchronized real-time imaging
The sample arms of the two SD-OCT systems were combined and mounted on a slit-lamp microscope
(Fig. 2
). The probe of the AS-OCT was mounted on top of the slit-lamp microscope. A pair of X-Y
galvanometers served as the scanner to image the anterior segment. Cross-hair scanning at the
horizontal and the vertical meridians was used to align the scanning position of the eye by viewing
the iris images in both horizontal and vertical images. The probe of the CM-OCT was placed on a
translation stage that replaced the lamp holder of the slit-lamp. The platform of the stage had the
capacity to rotate and vertically move, allowing a horizontal galvanometer to linearly scan the
ciliary muscle at the temporal side through the sclera. The two beams were pre-aligned at the same
altitude to cross at the corneal apex. The angle between the two probes was set at 30~50 degrees by
moving the slit-lamp arms and rotating the stage platform. Both of the two beams from the two
devices were focused on the sample by the objective lenses (f = 100 mm), which gave the lateral
resolution of ~20 μm [10].
Fig. 2
Schematic diagram depicting the combination of the probes. LCD: liquid-crystal display.
Schematic diagram depicting the combination of the probes. LCD: liquid-crystal display.To synchronize both sub-systems, a synchronization signal was needed to set the start point. We
were not able to modify the proprietary software in the CM-OCT and the CM-OCT run during live view
and the recording. If we stopped the recording during the alignment, we were not able to see the
live view. Therefore, we implanted a shutter in the reference arm which provided the synchronization
signal to the CM-OCT accounting as the start point. The acquisition of the AS-OCT was controlled by
a foot switch that signaled the electronic shutter in the reference arm of the CM-OCT as described
above (Fig. 1). The shutter, powered with a 5 volt direct
current, could quickly block the light beam reflected in the reference arm. When the shutter closed
at the onset of scanning, the detector of the CM-OCT did not receive the light from the reference
arm, resulting in a blank frame without any information that served as a synchronization signal.
2.3 Accommodative module
A liquid-crystal display (LCD) screen showed a white Snellen letter “E” on a black
background as the fixation target and was placed 10 cm from the tested eye. A trial lens combined
with the LCD was used to compensate for the refractive error at close proximity and to stimulate the
accommodation during image acquisition. The LCD and the trial lens were located on a translation
stage with a dual axis for horizontal and vertical adjustments. The LCD was coupled with AS-OCT
using a hot mirror (Fig. 2) and was synchronized with the
acquisition of the OCT devices by the foot switch (Fig. 1) to
alter the boundaries of the target between a blurred or sharp picture.
2.4 Participants and experiments
We performed two experiments using the combined system. A total of six subjects (5 male and 1
female, age range: 26 to 47 yrs) were included. This protocol was approved by the institutional
review board for human research at the University of Miami. Informed consent was obtained from each
subject, and all patients were treated in accordance with the tenets of the Declaration of Helsinki.
In the first experiment, we imaged the left eye of each subject at two separate states. The
refractive error during near vision of each subject was corrected by placing a trial lens with
appropriate power into the fixation system. Subjects were asked to sit in front of the slit-lamp and
to look forward at the letter “E”. In AS-OCT, the cross-hair alignment live view was
used to align the iris image on both the horizontal and the vertical meridians so that the OCT beam
was perpendicular to the iris plane (Fig. 1, Insert). In
addition, the scan was centered on the apex as indicated by the specular reflex (the central
vertical high reflectivity line on Figs. 3
, 5 and 7;
Media 1). During
image acquisition, the operator adjusted the probe to ensure the beam from the AS-OCT pass through
the corneal apex which ensured to scan the anterior segment at the same sagittal section. Before the
experiment, the beam from the CM-OCT was pre-aligned with the beam from the AS-OCT at the same
altitude. In addition, the specular reflex (vertical high reflectivity line) ensured sagittal
scanning of the ciliary muscle (Figs. 4
, 5 and 7;
Media 2). After
adjusting fixation to ensure the perfect alignment of the iris and corneal apex, the operator
initiated the acquisition by stepping on the foot switch. Then, the equivalent accommodative
stimulus of 4.00D was added to the trial lens, and the eyes were imaged following the same protocol.
At each separate state, we repeated the exams twice. In this experiment, the ocular anterior segment
at both the horizontal and the vertical meridian, which consisted of 2,048 × 4,096 pixels
(lateral × axial) per frames with the scan speed of 17,500 A-lines per second, were imaged.
Meanwhile, the ciliary muscle at the horizontal meridian of the temporal side was imaged at the scan
speed of 7,000 A-lines per second; corresponding to the image size of 1,000 × 1,223 pixels
(lateral × axial). Both of the OCT devices obtained 4 frames at each state, which took
approximate 0.5 second.
Fig. 3
Anterior segment from the cornea through the crystalline lens obtained from a 26-year-old
subject. A: Frame 1 of the anterior segment; B: Frame 2 of the anterior segment; C: Combined image
of the anterior segment. PD: pupil diameter; CCT: central corneal thickness; ACD: anterior chamber
depth; CLT: central lens thickness. Bar = 1 mm.
Fig. 5
The entire anterior segment (A and B) and the ciliary muscle (C and D) obtained from a
26-year-old subject using the combined devices in relaxed (A and C) and accommodative (B and D)
states. The scan width of the anterior segment was set to 12 mm, with an image size of 2,048
× 4,096 pixels (lateral × axial), and to 10.3 mm for the ciliary muscle, with an image
size of 1,000 × 1,223 pixels (lateral × axial). Bar = 1 mm. Note the iris (C and D)
was flipped as the mirror image due to placement of the delay line inside the eye (bottom of the
image).
Fig. 7
Real-time display of the anterior segment (A) and the ciliary muscle (B) from a 26-year-old
subject during 4.00D accommodation (Media
1 and Media
2). The frame rates of the movies were 8.33 fps (A) and 7 fps (B),
respectively. Bar = 1 mm. Note the iris (B and Media
2) was flipped as the mirror image due to
placement of the delay line inside the eye (bottom of the image).
Fig. 4
Ciliary muscle obtained from a 26-year-old human subject. A: Enhanced image of the ciliary
muscle; Note the iris was flipped as the mirror image due to placement of the zero-delay line inside
the eye (bottom of the image). B: Semi-automatic segmentation of the boundaries of the cornea, the
sclera, and the ciliary muscle. C: Optical correction for image distortion followed Snell’s
principle; D: The calculation of the ciliary muscle thickness. CMT1-3: the thickness of the ciliary
muscle at 1 mm, 2 mm, and 3 mm posterior to the scleral spur; CMTM: the maximum thickness of the
ciliary muscle. Bar = 1 mm.
Anterior segment from the cornea through the crystalline lens obtained from a 26-year-old
subject. A: Frame 1 of the anterior segment; B: Frame 2 of the anterior segment; C: Combined image
of the anterior segment. PD: pupil diameter; CCT: central corneal thickness; ACD: anterior chamber
depth; CLT: central lens thickness. Bar = 1 mm.Ciliary muscle obtained from a 26-year-old human subject. A: Enhanced image of the ciliary
muscle; Note the iris was flipped as the mirror image due to placement of the zero-delay line inside
the eye (bottom of the image). B: Semi-automatic segmentation of the boundaries of the cornea, the
sclera, and the ciliary muscle. C: Optical correction for image distortion followed Snell’s
principle; D: The calculation of the ciliary muscle thickness. CMT1-3: the thickness of the ciliary
muscle at 1 mm, 2 mm, and 3 mm posterior to the scleral spur; CMTM: the maximum thickness of the
ciliary muscle. Bar = 1 mm.In the second experiment, the left eye from a 26-year-old subject was imaged in real-time twice.
The fixation target was blurred at first to fog the eye, and the picture was sharpened 1 second
after the onset of the scanning to stimulate the accommodation. During the experiment, the subject
was asked to keep the target as clear and sharp as possible. Sixty-two images, corresponding to
thirty-one combined images of the anterior segment at the horizontal meridian, with 1,024 ×
4,096 pixels (lateral × axial) per frame, were sequentially acquired using the AS-OCT in 3.72
seconds, with a single frame rate of 8.33 Hz. Meanwhile, the ciliary muscle at the temporal side was
imaged using the CM-OCT at a frame-rate of 7 Hz. The starting point of the CM-OCT acquisition was
one frame behind that of the AS-OCT because the first frame was uninformed. Thus, a total of 25
frames of the ciliary muscle were obtained over a total duration of 3.71 seconds.
2.5 Image process and analysis
Figures 3A-3B show
the two images and one combined image from the 26-year-old subject using the AS-OCT. The lateral
scanning width was set to 15 mm. The zero-delay lines were set at the top and bottom of the two
single frames, which were near the corneal apex and the posterior pole of the crystalline lens,
respectively, with almost half of the dim images being due to a drop in the signal-to-noise ratio
(SNR). An algorithm was developed and used to register the two images using the features in both of
the images, such as the anterior segment of the lens and the iris. Briefly, the contour of the
common portion, which included the surfaces of the iris and crystalline lens on the each frame, was
extracted using segmentation algorithm based on the gradient information and shortest-path search
and then converted to one dimensional curve. Then the transformation, which consisted of a
combination of rotation and translation between the two frames according to the detected contours of
the anterior surface of the iris and the lens, was used for the overlaying of the two frames. After
registering and overlaying the images, the entire ocular anterior segment, including the cornea,
iris, anterior chamber and crystalline lens, was clearly visualized in the combined image (Fig. 3C). Thus, the sensitivity drop was compensated for, and the
enhancement of the anterior surface was realized.The boundaries of the cornea and the crystalline lens were semi-automatically outlined using
custom-developed software. Optical correction was performed based on Snell’s principle. The
correction approach has been well-documented in the literature [26-30], including in our previous work [8,31]. The refractive index
of each medium (the refractive index of 1.387 for the cornea [32], 1.342 for the aqueous humor [33], and 1.408 for
the crystalline lens [34]) at a wavelength of 840 nm was
applied in the algorithm of correction. The central corneal thickness (CCT), the pupil diameter
(PD), the anterior chamber depth (ACD), the curvature radii of the anterior (CAL) and posterior
(CPL) lens surface, and the central lens thickness (CLT) were computed.Overlaying multiple images for averaging was used to enhance the images of the ciliary muscle
(Fig. 4). Four images were averaged. From the averaged
images, the boundaries of the ciliary muscle, which defined as the intraocular boundaries of the
sclera, the external edge of the pigmented ciliary epithelium and the internal edge of the iris,
were manually segmented using custom developed software. After image correction for the optical
distortion, the thickness of the ciliary muscle was evaluated by drawing lines through the points at
1 mm (CMT1) posterior to the scleral spur that were perpendicular to the local curvature of the
sclera and extended to pigmented ciliary epithelium or the internal edge of the iris [12,35]. The thicknesses of
the ciliary muscle at 2 mm (CMT2) and 3 mm (CMT3) posterior to the scleral spur and the maximum
thickness of the ciliary muscle (CMTM) were calculated [12,35]. It is worth noting that the iris in the
CM-OCT images (Figs. 4, 5
and 7) was shown as the flipped image. We put the OCT
delay-line inside the eye (the bottom of the image) to improve the image quality of the ciliary
muscle, resulting in the flipped iris image (as the conjugated artifact). The conjugated artifact is
common in SD-OCT, unless phase shift method is used [20,36]. The conjugate artifact is not in time-domain OCT [35,37] or swept light source
OCT [18]. The details of the conjugated artifact of the iris
and its relation to the placement of the delay line have been reported in some published work [20,38]. The same phenomenon
was shown in the attached video clips.The entire anterior segment (A and B) and the ciliary muscle (C and D) obtained from a
26-year-old subject using the combined devices in relaxed (A and C) and accommodative (B and D)
states. The scan width of the anterior segment was set to 12 mm, with an image size of 2,048
× 4,096 pixels (lateral × axial), and to 10.3 mm for the ciliary muscle, with an image
size of 1,000 × 1,223 pixels (lateral × axial). Bar = 1 mm. Note the iris (C and D)
was flipped as the mirror image due to placement of the delay line inside the eye (bottom of the
image).
3. Results
Figure 5 shows the ciliary muscle and the anterior segment
from the same subject before and after accommodation. Compared to the relaxed condition, pachynsis
contraction was noted in the anterior portion of the ciliary muscle under accommodative condition.
Although the thinning in the posterior part was not apparent in the figure, the averaged results
from the study group showed significant thinning at the posterior portion of the ciliary muscle
measured as the CMT3 (Fig. 6
). In addition, changes in the anterior segment, including a thickening of the lens, a
steepening in the lens surface, an increased shallowness in the anterior chamber depth, and the
onset of miosis, were clearly observed. Figure 6 shows
similar results between the two repeated measurements, as well as the biometric changes between the
two states.
Fig. 6
Biometry of the anterior segment at the horizontal (A) and the vertical (B) meridians, as well as
the ciliary muscle (C), within two visits during accommodation. PD: pupil diameter; ACD: anterior
chamber depth; CAL: curvature radius of the anterior surface of the lens; CPL: curvature radius of
the posterior surface of the lens; CLT: central lens thickness; CMT1-3: ciliary muscle thickness;
CMTM: maximum thickness of the ciliary muscle. P value: difference of the mean values between the
relaxed and the accommodative states. non-ACC: Relax status; ACC: Accommodative status; V1: visit 1;
V2: visit 2.
Biometry of the anterior segment at the horizontal (A) and the vertical (B) meridians, as well as
the ciliary muscle (C), within two visits during accommodation. PD: pupil diameter; ACD: anterior
chamber depth; CAL: curvature radius of the anterior surface of the lens; CPL: curvature radius of
the posterior surface of the lens; CLT: central lens thickness; CMT1-3: ciliary muscle thickness;
CMTM: maximum thickness of the ciliary muscle. P value: difference of the mean values between the
relaxed and the accommodative states. non-ACC: Relax status; ACC: Accommodative status; V1: visit 1;
V2: visit 2.Figure 7
shows the dynamic deformation of the ciliary muscle and the anterior segment in real-time.
The biometric parameters of the lens and the ciliary muscle were calculated as the mean values
before and after the accommodative stimulus onset, as shown in Table 1
, and the dynamic changes calculated as the function of time are shown in Fig. 8
. After the accommodative stimulus was given, the CMT1 and CMTM increased gradually, and
reached the stable status at approximate 2 second (Figs. 8A
and 8B, Table 1).
Meanwhile, the CAL, PD and ACD decreased with the increasing in the CLT, and reached the
accommodative status at approximate 2.5 second (Figs.
8C-8F, Table
1).
Table 1
The biometry of the ciliary muscle and the anterior segment before and after accommodation in
a 26-year-old subject (unit: mm)
Visit 1
Visit 2
0 D
4 D
0 D
4 D
Ciliary muscle
CMT1
0.406 ± 0.009
0.605 ± 0.008
0.413 ± 0.020
0.603 ± 0.021
CMT2
0.279 ± 0.012
0.309 ± 0.011
0.290 ± 0.012
0.317 ± 0.014
CMT3
0.154 ± 0.008
0.151 ± 0.009
0.154 ± 0.009
0.162 ± 0.014
Maximum thickness
0.408 ± 0.011
0.606 ± 0.012
0.416 ± 0.021
0.604 ± 0.021
Crystalline lens
Anterior surface curvature radius
12.05 ± 0.584
7.65 ± 0.464
11.72 ± 0.529
7.78 ± 0.387
Posterior surface curvature radius
5.82 ± 0.139
4.67 ± 0.313
5.53 ± 0.211
4.39 ± 0.327
Central thickness
3.62 ± 0.010
3.76 ± 0.010
3.63 ± 0.012
3.78 ± 0.006
Other biometry
Pupil diameter
4.09 ± 0.061
2.70 ± 0.102
4.09 ± 0.054
2.91 ± 0.043
Anterior chamber depth
3.56 ± 0.008
3.47 ± 0.006
3.55 ± 0.010
3.45 ± 0.016
Central cornea thickness
0.55 ± 0.004
0.55 ± 0.006
0.56 ± 0.003
0.56 ± 0.006
CMT1-3: ciliary muscle thickness at 1 mm, 2 mm, and 3 mm posterior to the scleral spur.
The mean and standard deviation of the biometry at 0 D were calculated from 7 frames before the
accommodative stimulus onset (t < 1.0 s).
The mean and standard deviation of the ciliary muscle thickness at 4 D were calculated from 13
frames after reaching the stable accommodative status (t > 2.0 s).
The mean and standard deviation of the crystalline lens and other biometry at 4 D were calculated
from 11 frames after reaching the stable accommodative status (t > 2.5 s).
Fig. 8
Dynamics of the biometry of the ciliary muscle (A and B) and the anterior segment (C-F) during
accommodation acquired at two visits (v1 and v2). Accommodative stimulus was set at 1 second after
the start point of the scanning. CMT1: ciliary muscle thickness at 1 mm posterior to the sclera spur
(A); CMTM: maximum thickness of the ciliary muscle (B); CAL: curvature radius of the anterior
surface of the lens (C); CLT: central lens thickness (D); PD: pupil diameter (E); ACD: anterior
chamber depth (F). V1: visit 1; V2: visit 2.
Real-time display of the anterior segment (A) and the ciliary muscle (B) from a 26-year-old
subject during 4.00D accommodation (Media
1 and Media
2). The frame rates of the movies were 8.33 fps (A) and 7 fps (B),
respectively. Bar = 1 mm. Note the iris (B and Media
2) was flipped as the mirror image due to
placement of the delay line inside the eye (bottom of the image).CMT1-3: ciliary muscle thickness at 1 mm, 2 mm, and 3 mm posterior to the scleral spur.The mean and standard deviation of the biometry at 0 D were calculated from 7 frames before the
accommodative stimulus onset (t < 1.0 s).The mean and standard deviation of the ciliary muscle thickness at 4 D were calculated from 13
frames after reaching the stable accommodative status (t > 2.0 s).The mean and standard deviation of the crystalline lens and other biometry at 4 D were calculated
from 11 frames after reaching the stable accommodative status (t > 2.5 s).Dynamics of the biometry of the ciliary muscle (A and B) and the anterior segment (C-F) during
accommodation acquired at two visits (v1 and v2). Accommodative stimulus was set at 1 second after
the start point of the scanning. CMT1: ciliary muscle thickness at 1 mm posterior to the sclera spur
(A); CMTM: maximum thickness of the ciliary muscle (B); CAL: curvature radius of the anterior
surface of the lens (C); CLT: central lens thickness (D); PD: pupil diameter (E); ACD: anterior
chamber depth (F). V1: visit 1; V2: visit 2.
4. Discussion
Imaging the entire anterior segment has presented a substantial challenge. Strenk et
al. developed high-resolution magnetic resonance imaging (MRI) for imaging the whole eye
during accommodation to investigate the relationship between the ciliary muscle and the crystalline
lens [39]. This technology offers direct visualization of the
ciliary muscle and the lens periphery simultaneously without preventing visualization of the iris.
This MRI has the disadvantage of a low spatial resolution (approximately 156 μm), which is
roughly equivalent to the change in the maximum thickness of the lens and the ciliary muscle
(approximate 100 to 200 μm) during accommodation. Time-consuming exams and high costs may
limit its application to real-time imaging and to research on dynamic accommodation. Ultrasound
biomicroscopy (UBM), which has good resolution (approximate 50 μm) and scan speed (more than
8 Hz), have been used to image the crystalline lens [40,41] and the ciliary muscle [42,43]. With this contact method, the configuration
of the ciliary muscle was clearly visualized and the changes in the accommodative response could be
documented. Moreover, with iridectomies in animal models, the ciliary process and the crystalline
lens equator can be visualized using goniovideography [42,44]. These features cannot be visualized using OCT
method if the iridectomy is not performed in human subjects.Imaging the entire anterior segment from the cornea to the crystalline lens is difficult for
SD-OCT due to the significant sensitivity decay with increased scan depth [8,22,45-47]. The decrease in the SNR with an
increase in the imaging depth from the zero-delay line reduces the image contrast, causing the
posterior portion of the lens to become unobservable [8,21,22]. With complicated
phase shift technology, the entire anterior segment can be imaged by removing the complex conjugated
image [36,38,48,49]. We previously imaged
the anterior segment using a custom-built dual-channel dual-focus OCT system [19] that required two spectrometers and a 50% attenuation of the reflected light
and resulted in a low-contrast image. Wang et al. extended the imaging range of an
SD-OCT using a fiber optic switch in the reference arm, and this device has been demonstrated
in vitro [50]. Using a fiber optic switch
may require an additional patch fiber in the sample arm and possibly complicates the dispersion
compensation [50]. As demonstrated in previous work by our
researchers [8,21] and
others [22], the use of a galvanometer to alter the reference
arm length appears to be a simple and easy way to avoid the necessity for a patch fiber and
adjustment of the dispersion compensation. Placing the zero delay line on the top of one image and
on the bottom of the other image improves the SNR throughout the entire image depth and avoids
possible conjugated artifacts. Therefore, image enhancement can be realized by the acquisition and
overlapping of two frames. This technique is useful to extend imaging depths and to compensate for
the sensitivity fall-off without eliminating complex ambiguity or reducing image contrast [8,21,22]. Additionally, this approach can further improve the scan depth by using more
mirrors [22]. In the present study, we also improved our
system by using a CMOS camera for a quicker scanning speed, and a similar design of overlapping
images was utilized to extend the image range and to achieve real-time imaging of dynamic
accommodation. The previous CCD-based approach had a limited scan speed and could not be used for
real-time imaging [8,21]. Combining this improved system with our existing 1,310 nm SD-OCT enabled us to image
the ocular anterior segment, including the cornea, anterior chamber, crystalline lens, and ciliary
muscle.Anterior segment OCT has been used to analyze the biometric changes of the ciliary muscle during
accommodation [12-14,35,37,51]. Sheppard and Davies observed that the ciliary muscle showed a
contractile shortening and a thickening in the anterior portion under the accommodative state [37]. Lossing combined an AS-OCT and an auto-refractor to
investigate the relationship between the contraction of the ciliary muscle and the accommodative
response [12]. In those studies, the measurements were taken
at separate states, but not dynamically, mainly due to their slow scan speeds (approximately 2,000
A-lines/s) using time domain OCTs (TD-OCT) [12,13,37,51].The custom-developed SD-OCT used in the present study improved the scanning speed for real-time
imaging. In this static experiment, image enhancement was accomplished by averaging multiple images.
The entire configuration of the ciliary muscle in non-human primate eyes has been previously
documented using UBM [42,44], which could be used for imaging dynamic changes of the ciliary muscle. To the best of
our knowledge, this is the first time that the dynamics of the ciliary muscle biometry on the human
eye have been visualized by SD-OCT during accommodation. The combination of the two SD-OCT devices
enabled the simultaneous imaging of the entire anterior segment. Due to the inaccessibility of the
OCT software in the custom CM-OCT, the use of the synchronization marker in image recording may be
an easy way to synchronize both systems. Further advances may involve data acquisition from two
spectrometers with one computer, similar to the system used in the two-channel OCT system [19].Our results on the biometry of the anterior segment obtained with this system are in agreement
with the classic theories of Helmholtz [1] and others [2,7,52]. The deformation included a reduction of the pupil size, a shallow anterior
chamber, a steep lens surface, an increased lens thickness [8,21,22], and
a thickened anterior portion of the ciliary muscle, accompanied by a thinning in the posterior
portion [12,13,37,51]. Further studies are
required to investigate the relationship of the ciliary muscle and the crystalline lens with aging.
Using UBM and goniovideography, Glasser et al. and Croft et al.
imaged the lens, ciliary processes and the ciliary muscle in living monkeys that were undertaken
surgical iridectomy during accommodation, which performed by stimulating the Edinger-Westphal (E-W)
nucleus [42,44,53]. Although these studies were experimented in non-human primate
eyes, the dynamic changes in the ciliary muscle and the crystalline lens during accommodation had
been reported. In the present study, real-time imaging on the human eye in vivo was
conducted and the dynamic accommodative responses of the ocular anterior segment, including the
iris, crystalline lens and ciliary muscle were documented. Our results are in agreement with the
previous findings obtained from non-human primate eyes. The findings in the present study were
expected as the anatomy of the human and the monkey accommodative apparatus is very similar [42,44,53]. The combined system demonstrated an enhanced ability to understand dynamic
accommodation. For instance, we found that the ciliary muscle started to contract 0.3 seconds after
the accommodative stimulus was given and reached a stable state in approximately 0.5 seconds. The
crystalline lens began to restore its shape following the contraction of the ciliary muscle, but its
response ended later than that of the ciliary muscle. There was an approximate 0.5 second delay
between the response times. This result implies that the deformation of the lens is slower than the
contraction of the ciliary muscle. The similar finding of delayed responses was found in non-human
primate eyes using UBM [44]. Early studies involving the
reaction time of accommodative response found an age-related deterioration in dynamic accommodation
[40,54,55]. The capacity to obtain real-time images of the response delay
on the human eye using the non-contact OCT method may add an alternative modality for studying
accommodation.There are some limitations in the present study. The convergence, rotation and movements of the
eye during accommodation may lead to errors in the dynamic measurements of the anterior segment
biometry. The movies of the accommodative process showed minimal ocular movements. The small
variability in the ocular biometry shown in Table 1 implies
that the measurements are not seriously impacted by motion-related artifacts. However, the changes
in the curvature radii of the lens surface showed more fluctuations, indicating that the lens shape
may be more critically deteriorated by motion-related artifacts. The achievement of a 3D scanning
pattern will require precise registration to avoid the influence of ocular movement in the future.
Our current analysis for real-time imaging was processed by semi-automatic software that was
extremely time-consuming. Automatic software for the registration, segmentation and calculation of
these processes is being developed. Although our combined system is able to image the anterior
segment including the ciliary muscle, the need to modify the reference arms for compensating
sensitivity drop-off and implementation of the shutter may complicate the imaging approach. Further
improvement with one system and alterative scanning channels [56] using the wavelength of 1 µm [18] may
simplify the approach. Imaging the muscle apex position and width is important in studying the
accommodative response. Apparently the image of the ciliary muscle obtained with OCT was not as
clear as that obtained with UBM [42,43]. When the images were averaged, the configuration of the ciliary muscle
appeared to be visualized as demonstrated in our 1st experiment, in which four images were averaged.
Considering the non-contact feature and better resolution compared to UBM, OCT method may add an
alternative tool for studying the real-time accommodative response, although OCT may not replace UBM
in imaging the ciliary muscle. The image in the video was not as good as the averaged image because
no averaging was applied in the real-time imaging. We did not overlay the images due to the
limitation of the scan speed (7 Hz). Nevertheless, our approach would be a start to OCT real-time
imaging of the ciliary muscle. In the future, high speed data acquisition with frame averaging may
provide clearer visualization of the ciliary muscle in real-time.In summary, we combined two custom-built spectral domain optical coherence tomography devices
(SD-OCT) to simultaneously image the anterior segment and the ciliary muscle during accommodation.
We also imaged dynamic changes of the anterior segment in one eye during accommodation. The combined
systems provided the capacity for simultaneous, real-time imaging of the biometry of the anterior
segment, including the ciliary muscle, in vivo during accommodation.
Authors: Maciej Wojtkowski; Rainer Leitgeb; Andrzej Kowalczyk; Tomasz Bajraszewski; Adolf F Fercher Journal: J Biomed Opt Date: 2002-07 Impact factor: 3.170
Authors: Marco Ruggeri; Stephen R Uhlhorn; Carolina De Freitas; Arthur Ho; Fabrice Manns; Jean-Marie Parel Journal: Biomed Opt Express Date: 2012-06-06 Impact factor: 3.732
Authors: Marco Ruggeri; Carolina de Freitas; Siobhan Williams; Victor M Hernandez; Florence Cabot; Nilufer Yesilirmak; Karam Alawa; Yu-Cherng Chang; Sonia H Yoo; Giovanni Gregori; Jean-Marie Parel; Fabrice Manns Journal: Biomed Opt Express Date: 2016-03-17 Impact factor: 3.732
Authors: Aikaterini I Moulakaki; Daniel Monsálvez-Romín; Alberto Domínguez-Vicent; José J Esteve-Taboada; Robert Montés-Micó Journal: Int Ophthalmol Date: 2017-10-26 Impact factor: 2.031