Custom high-resolution high-speed anterior segment spectral domain optical coherence tomography (OCT) was used to characterize three-dimensionally (3-D) the human crystalline lens in vivo. The system was provided with custom algorithms for denoising and segmentation of the images, as well as for fan (scanning) and optical (refraction) distortion correction, to provide fully quantitative images of the anterior and posterior crystalline lens surfaces. The method was tested on an artificial eye with known surfaces geometry and on a human lens in vitro, and demonstrated on three human lenses in vivo. Not correcting for distortion overestimated the anterior lens radius by 25% and the posterior lens radius by more than 65%. In vivo lens surfaces were fitted by biconicoids and Zernike polynomials after distortion correction. The anterior lens radii of curvature ranged from 10.27 to 14.14 mm, and the posterior lens radii of curvature ranged from 6.12 to 7.54 mm. Surface asphericities ranged from -0.04 to -1.96. The lens surfaces were well fitted by quadrics (with variation smaller than 2%, for 5-mm pupils), with low amounts of high order terms. Surface lens astigmatism was significant, with the anterior lens typically showing horizontal astigmatism ([Formula: see text] ranging from -11 to -1 µm) and the posterior lens showing vertical astigmatism ([Formula: see text] ranging from 6 to 10 µm).
Custom high-resolution high-speed anterior segment spectral domain optical coherence tomography (OCT) was used to characterize three-dimensionally (3-D) the human crystalline lens in vivo. The system was provided with custom algorithms for denoising and segmentation of the images, as well as for fan (scanning) and optical (refraction) distortion correction, to provide fully quantitative images of the anterior and posterior crystalline lens surfaces. The method was tested on an artificial eye with known surfaces geometry and on a human lens in vitro, and demonstrated on three human lenses in vivo. Not correcting for distortion overestimated the anterior lens radius by 25% and the posterior lens radius by more than 65%. In vivo lens surfaces were fitted by biconicoids and Zernike polynomials after distortion correction. The anterior lens radii of curvature ranged from 10.27 to 14.14 mm, and the posterior lens radii of curvature ranged from 6.12 to 7.54 mm. Surface asphericities ranged from -0.04 to -1.96. The lens surfaces were well fitted by quadrics (with variation smaller than 2%, for 5-mm pupils), with low amounts of high order terms. Surface lens astigmatism was significant, with the anterior lens typically showing horizontal astigmatism ([Formula: see text] ranging from -11 to -1 µm) and the posterior lens showing vertical astigmatism ([Formula: see text] ranging from 6 to 10 µm).
The optical quality of the human eye is mainly determined by the geometrical and optical
properties of two elements, cornea and crystalline lens. The cornea accounts for two-thirds of
the optical refractive power, while the crystalline lens provides approximately one third of the
total static refractive power of the eye, and it is the responsible for accommodation [1-10]. Accurate
description of the geometry of the eye’s optical components is critical for understanding
their contribution to optical quality. Although the geometrical properties of the cornea have
been widely studied due to its accessibility [1-3], accurate quantitative data of the
crystalline lens geometry are limited to in vitro studies [11-13], while
in vivo data are referred in most cases to axial properties (i.e. thickness)
and central areas (i.e. central radius of curvature) [14-21].The shape and alignment of the crystalline lens have been reported in vivo
and in vitro using different imaging techniques. Radii of curvature, tilt and
decentration of the lens have been measured using a Purkinje-imaging based method [20-22]. However,
although Purkinje-based phakometry is a rapid, systematic, and reliable method, it does not
provide the full geometry of the crystalline lens surfaces, nor a direct view of the lens. In
contrast, Scheimpflug imaging allows acquisition of cross-sectional images of the crystalline
lens, and with proper correction of the geometrical and optical distortions, the radii of
curvature and asphericity of the lens surfaces can be obtained [16-18,23,24]. The slit-lamp configuration of
Scheimpflug imaging-based systems (where the front view of the pupil appears superimposed to the
cross-sectional image) frequently limits the view of the posterior lens. Besides, commercial
Scheimpflug instruments, although providing images of the lens, are typically not corrected from
optical distortion, preventing quantitative evaluation of lens parameters. To our knowledge,
custom correction algorithms have only been applied to crystalline lens images in certain
Scheimpflug instruments in a laboratory setting, and the reported crystalline lens parameters
are restricted to a single cross-section of the lens [16,24]. In addition to optical techniques, the
anterior chamber of the eye can be imaged through non-optical imaging techniques such as
ultrasound [19,25]
and Magnetic Resonance Imaging (MRI) [26-28]. These non-optical techniques allow visualization of the
entire crystalline lens and its neighboring ocular structures. However, these are either
invasive or time-consuming imaging methods, with significantly lower resolution than optical
techniques, which impose major problems (low acquisition speed, motion artifacts or low sampling
density) which prevent quantifying the crystalline lens geometry with high accuracy.To our knowledge, the 3-D structure of the crystalline lens has not been investigated
quantitatively with any method, likely as a result of the associated problems of the imaging
techniques used to image the crystalline lens. Optical coherence tomography (OCT) is a
high-resolution noncontact advanced optical imaging technique capable of providing 3-D
evaluation of the anterior segment of the eye [29-33]. Several geometrical crystalline
lens parameters surfaces have been reported previously with OCT, such as lens thickness and
anterior and posterior lens radius of curvature in the horizontal meridian [34-36]. In
addition, OCT has been used recently to quantify the crystalline lens shape in 2-D in human and
primate lenses in vitro [37-39], and to estimate the Gradient Index (GRIN) distribution
(from the optical distortion produced by GRIN on the posterior lens surface) in the porcine lens
in 3-D and the human lens in 2-D [40-44].Imaging the crystalline lens with sOCT poses some challenges: (1) The axial imaging range is
limited by the resolution of the spectrometer, and this can be insufficient to image the entire
anterior segment (from the anterior cornea to the posterior lens) in a single acquisition [30]. (2) Motion artifacts can limit the repeatability of the
measurements, making it necessary to reduce acquisition time [45]; (3) Fan (arising from the scanning architecture) and optical (refraction)
distortions, if left uncorrected, prevent quantitative analysis of the crystalline lens images
[45-49].In this study we measured and quantified the crystalline lens surface in 3-D. Further
developments of the methods that we had previously developed and applied to achieve quantitative
corneal topography [45,46,49] were applied to the quantification of the
crystalline lens, overcoming limitations (2) and (3). Newly developed methods for registration
and merging of sOCT images at different focus allowed full anterior segment imaging, overcoming
limitation (1).Accurate measurement of the lens geometry is crucial in the understanding of crystalline lens
optical properties, and of the physical changes of the lens in accommodation and presbyopia, as
well as in the design and evaluation of accommodation-restoration solutions for presbyopia, and
to increase the predictability of intraocular lens implantation procedures.In this study, we present the capability of sOCT imaging to provide 3-D parameters of the
anterior and posterior lens surfaces (radii and asphericity) and lens thickness. To our
knowledge, this is the first report of 3-D elevation maps of the anterior and posterior lens
surfaces of the eye in vivo.
2. Material and Methods
2.1. Experimental sOCT setup
Images were collected using a custom-developed spectral OCT system (from a collaborative
effort with Copernicus University, Torun, Poland) [30].
The setup is based on a fiber-optics Michelson interferometer configuration with a
superluminescent diode (SLD) (λ0 = 840 nm, Δλ = 50 nm) as a
light source, and a spectrometer (volume diffraction grating, and a 12-bit 4096-pixel line-scan
CMOS camera) as a detector. The effective acquisition speed is 25000 A-Scans/s, which optimized
balance between speed and SNR. The axial range of the instrument is 7 mm, resulting in a
theoretical axial pixel resolution of 3.4 µm. The axial resolution predicted by the
bandwidth of the SLD laser source is 6.9 µm.
2.2. Validation experiments on samples in vitro
The accuracy of the method to retrieve the shape of the lens was tested on a physical water
cell model eye and on a donor lens.The physical model eye has been described in previous studies where it was used to assess the
accuracy of measurements of lens tilt and decentration from Scheimpflug and Purkinje imaging
[50] and of topographic measurements with a Time Domain
OCT system [46]. The water cell model consists of a PMMA
cornea and a spherical biconvex intraocular lens with known spherical radii, as measured by a
non-contact profilometer (PLu, Sensofar, Barcelona Spain), filled with saline solution. The
index of refraction of the materials (1.486 for the artificial cornea and lens and 1.336 for
the saline solution) were provided by the manufacturers. Measurements on the physical model eye
were obtained with the SLD power exposure fixed at 800 µW. Each set of 3-D measurements
was acquired in 0.72 seconds, similar to acquisition time used in in vivo
eyes, with two acquisitions at two different planes of focus. Measurements were collected on a
10 × 15 mm zone, using 50 B-Scans composed by a collection of 360 A-Scans, providing a
resolution of 0.04 mm for horizontal and 0.2 mm for vertical meridian.Experiments were also performed on one crystalline lens from a 65-year donor eye, within 24
hours post-mortem (Eye Bank Transplant Services Foundation, Barcelona, Spain). Protocols for
human tissue handling followed the guidelines of the Declaration of Helsinki, and had been
approved by the Institutional Review Boards. Crystalline lens was extracted immediately before
measurements and immersed in preservation medium (DMEM/F-12, D8437, Sigma, St. Louis, MO) at
25°C. The imaging procedure of the crystalline lens in vitro was
similar to that described by de Castro et al. [40].
Images were acquired with the lens in two orientations, first with the anterior surface facing
the OCT beam, “anterior up”, and then with the posterior surface facing the OCT
beam (“posterior up”), after carefully flipping the lens. A homogeneous lens
index of refraction of 1.413 was assumed in the calculations. Each single 3-D collection
consisted of 70 B-scans with 1668 A-scans, with a lateral range of 12x12 mm, providing a
resolution of 170x7 µm. The acquisition time of a full 3-D image was 4.5 s.
2.3. Experimental protocols for anterior segment image acquisition in vivo
Images were collected on the right eye of 3 young subjects (ages 28-33). The subjects were
considered healthy in a clinical ophthalmological examination. Refractions ranged between 0 to
−4.75 D sphere and 0 to 1 D of cylinder. All protocols for the measurements in
vivo had been approved by Institutional Review Boards. The subjects signed informed
consents after the nature of the study had been explained, in accordance to the tenets of the
Declaration of Helsinki. Measurements were performed under mydriasis (by Tropicamide 1%). The
subjects were stabilized using a bite bar. Alignment of the subject was achieved with respect
to the anterior corneal specular reflection, while the subject fixated on a reference maltese
cross target projected on a minidisplay at optical infinity.A total of 15 sets of 3-D data were collected in each eye: 5 repeated images of the cornea, 5
repeated images of the anterior part the lens, and 5 repeated images of the posterior part of
the lens. All 3-D sets of data contained also the iris.The SLD power exposure was fixed at 800 µW [45]. Focus was changed by an automatic displacement to achieve optimal imaging of the
different anterior segment structures (cornea, anterior and posterior lens). Images of each
axial region were collected in 0.72 s, in order to minimize the impact of motion artifacts.
Measurements were collected on a 10 × 15 mm zone, using 50 B-Scans composed by a
collection of 360 A-Scans, providing a resolution of 0.04 mm for horizontal and 0.2 mm for
vertical meridian.
In previous works corneal elevation maps from normal and pathological subjects were obtained
using sOCT and analyzed using dedicated algorithms for volumetric data extraction [49]. The current study incorporates improvements in the
processing algorithms (including a more sensitive automatic surface segmentation, resulting in
a significant reduction of the detection noise, and a new implementation of Delaunay
decomposition-based method for 3-D ray tracing and optical distortion correction methods), as
well as new routines specifically developed for the requirements of the current application.
The new algorithms include routines for: automatic detection of the pupil center and radius,
estimation of the pupil plane orientation, and 3-D automatic merging of anterior segment images
collected at different depths. The image processing algorithm can be summarized in nine steps.
Only dedicated new routines are described in detail:(1) Denoising: A rotational kernel transform was performed for an edge-preserving denoising,
using a mask of size 9 pixels [49]. In addition, a
wavelet low-pass filtering processing based on log-Gabor wavelet was used for 7 scales and 6
orientations.(2) Statistical thresholding: An adaptive algorithm based on a multimodal Gaussian fitting of
the histogram intensity (in a non-linear least squares sense) was performed on the entire 3-D
sets of data, which allowed identifying a noise class from the signal.(3) Volume clustering: The statistics of the noise and signal are overlapped. As a result,
small randomly distributed volumes are mixed with the large volumes representing the objects of
interest in 3-D (cornea, iris, lens surfaces, etc.). The number of pixels belonging to certain
a volume is obtained by means of the connectivity of points. The volumes of interest are
obtained by thresholding the number of elements according to the expected total number. Figure 1(a)
shows examples of the volumes clustered in each of the 3-D data set collected at
different depths: (a) cornea and iris, (b) anterior lens and iris, (c) posterior lens and
iris.
Fig. 1
Illustration of the three acquisitions of an individual data collection in subject S#1.
(a) cornea and iris; (b) anterior lens and iris; (c) posterior lens and iris.
Illustration of the three acquisitions of an individual data collection in subject S#1.
(a) cornea and iris; (b) anterior lens and iris; (c) posterior lens and iris.(4) Multilayer segmentation: A new approach for automatic segmentation was developed, based
on Canny detection in each A-scan (1-D signals). In an initial step, a Gaussian filter
(standard deviation σ = 5 pixels) was convolved with the signal to reduce spurious peaks
associated to noise detection in the CMOS line-camera. The first derivatives of the signal were
then obtained by convolving the resultant A-Scan with finite-size linear-phase separable
kernels for differentiation of the discrete data (A-Scan) [51]. The size of the kernels (3-9 pixels) controls the sensitivity of the algorithm to
noise (the lower the number of pixels used for the kernel the higher the noise sensitivity).
The result of the first derivative calculation provides both local maxima and minima of the
A-scan. Maxima are identified by subtraction of the slope sign of two consecutive points
(positive values). The number of local maxima is further reduced by thresholding the signal and
removing the peaks below a certain value (set to the average of the signal plus a variable
amount, in this case, 0.25 times the standard deviation of the signal). The maxima identified
in the A-scans are connected in the 2-D cross-sections (B-scans) or in the 3-D data sets using
the same neighborhood algorithm used for clustering, therefore allowing to segment layers.
Figure 2
shows the application of developed routine to the segmentation of the anterior and
posterior surface of the cornea. Figure 2(a) shows a
typical raw A-scan (in black), the smoothed signal after application of a Gaussian (σ =
5) filter (in blue), and the identified local maxima (obtained by the application of the first
derivative with a kernel of 9 pixels, red asterisks). Figure
2(b) illustrates the estimation of the maxima (red points) in a collection of A-scans
in the cornea, and the identified layers for the anterior and posterior cornea surfaces (in
green and yellow, respectively), following the application of the neighborhood algorithm.
Fig. 2
(a) Illustration of the algorithm for maxima detection: Original A-Scan black; Filtered
signal by Gaussian filtering, blue; Detected local maxima (red asterisks) by First
derivative 9-pixel kernel computation. The anterior and posterior corneal peaks are marked
by green and yellow asterisks, respectively. (b) Detection of maxima in corneal B-Scan (in
red), and multilayer segmentation of the anterior surface (green line) and posterior surface
(yellow line) by the neighborhood algorithm.
(a) Illustration of the algorithm for maxima detection: Original A-Scan black; Filtered
signal by Gaussian filtering, blue; Detected local maxima (red asterisks) by First
derivative 9-pixel kernel computation. The anterior and posterior corneal peaks are marked
by green and yellow asterisks, respectively. (b) Detection of maxima in corneal B-Scan (in
red), and multilayer segmentation of the anterior surface (green line) and posterior surface
(yellow line) by the neighborhood algorithm.(5) Pupil center reference: The pupil center was used as a reference across measurements
(i.e. cornea, anterior and posterior lens) and to define the optical zone (effective area
within the pupil). According to previous studies [52],
the pupil center shifts little across pupil sizes and light conditions. In a previous work
[49] we estimated the pupil center and optical zone by
fitting the outer edge of pupil in 2-D en face images (computed by adding all images of a 3-D
data set) to an ellipse. However, the current application requires 3-D position of the pupil
center (including the axial coordinate), and estimation of the pupil plane in 3-D, as this
information is used for registration and merging of the corneal and lens images. In this study,
the lateral pupil coordinates were obtained as in the previous study [49], while the axial coordinate was obtained by nonlinear least-square
fitting of the iris plane in 3-D (and evaluation of the plane at the estimated pupil lateral
coordinates). In addition, the characteristic vector of the iris plane was obtained, which
provided the tilt angle of this plane with respect to the OCT coordinate system.(6) Merging 3-D volumes: The pupil coordinates are assumed as a common reference for the
images of the anterior segment collected at different depths. Variations in the pupil diameter
were negligible across images. Images of the cornea, anterior and posterior lens were merged
using the pupil center and pupil plane orientation for registration. In a first step prior to
merging, the corneal image was inverted, as for efficiency in the focus range shift, the cornea
was acquired in the opposite side of the Fourier transform (in comparison with the crystalline
lens acquisition). The 3-D volumes of the anterior cornea/iris and posterior lens/iris are
shifted to the pupil center reference, and rotated (using a rotation matrix [53]) in order to superimpose the characteristic vectors of
the corresponding pupil plane to those of the anterior lens/iris [53]. Figure 3
(Media 1)
illustrates the process to obtain the full anterior segment 3-D image merging from the three
volumetric acquisitions. A full overlapping of the iris from the 3 sets of images can be
observed.
Fig. 3
(Media
1). Illustration of the merging of three volumetric
acquisitions to obtain a 3-D full anterior segment image.
(Media
1). Illustration of the merging of three volumetric
acquisitions to obtain a 3-D full anterior segment image.(7) Geometrical distances calculation and further denoising: The registration of the 3-D
volumetric data sets involved shift and rotation by coordinates in optical distances, which
actually differ across the different depths. Estimation of the geometrical parameters from the
full anterior segment images (still subject of fan and optical distortion) required re-sampling
of the surfaces in a rectangular equally distributed mesh of 100 x 100 points in the rectangle
circumscribed by the pupil ellipse, as well as interpolation. The interpolation method was
based on Delaunay description of the surfaces [32], and
the use of radial base cubic functions (as those functions do not require the use of a regular
grid of sampling without compromising accuracy). The interpolated coordinates are obtained from
the nodes of the triangle elements that tile the surfaces in the Delaunay description. The
procedure was performed for the four surfaces of the anterior segment: anterior-posterior
cornea and anterior-posterior lens. The optical distance can be then calculated by direct
subtraction of the coordinates of the different surfaces, while the geometrical distances were
obtained by dividing the optical distance by the corresponding group refractive index at the
illumination wavelength. The corneal refractive index was taken as 1.376, the aqueous humor
refractive index as 1.336, and the crystalline lens refractive index was obtained from the
age-dependent average refractive index expression derived by Uhlhorn et al. [37] (1.4176-1.4186 for the subjects of our study). The
sampled anterior segment elevation data from anterior and posterior surfaces of the cornea and
the lens were denoised using a fitting Zernike modal expansion (55 terms, 10th order). This
iterative method allows further rejection of spurious points in the surface due to detection
process.(8) Distortion corrections: In order to correct the interface layers for the effect of
distortion in general (fan distortion and refraction), it is required to propagate the
geometrical distances obtained from the previous point along the directional cosines of the
ray, taking into account the refraction that occurs at each interface between two media of
different refractive indices [46]. A procedure similarly
to that described in an earlier publication [49] and
practically applied to the cornea was followed, expanding it sequentially from the cornea to
the posterior surface of the crystalline lens. Correction of the fan and optical distortions
produces large geometrical changes in the anterior optical surfaces, which are qualitatively
observable in the diameter of the visible areas of the anterior, and to a larger extent, of the
posterior lens. Figure 4
(Media 2)
shows the difference in the segmented surfaces fully corrected from optical distortion
correction (in green), and the segmented surfaces corrected from fan distortion but not from
refraction effect (dividing only by its refractive index, in red).
Fig. 4
(Media
2) Illustration of the effect of optical (refraction)
correction (in green), in comparison with the correction of the surfaces by simple division
by their corresponding refractive indices in an anterior segment image (in red). From left
to right: segmented anterior and posterior corneal surfaces, and anterior and posterior lens
surfaces.
(Media
2) Illustration of the effect of optical (refraction)
correction (in green), in comparison with the correction of the surfaces by simple division
by their corresponding refractive indices in an anterior segment image (in red). From left
to right: segmented anterior and posterior corneal surfaces, and anterior and posterior lens
surfaces.(9) Surface fitting by quadrics (biconicoid, conicoid), and elevation maps: The surfaces were
fitted by quadrics in a 5-mm of diameter optical zone with respect to their apexes. The corneal
shape descriptive parameters (radii of curvature Rx and Ry and conic
constants Qx and Qy) were obtained by biconic fitting. In addition,
surfaces were also fitted by conicoids (defined by the radius of curvature R and conic
constants Q). The elevation maps for the anterior and posterior lens after fan distortion
correction only, and after full distortion correction and geometrical distances were
represented as the difference of elevation data from the reference sphere, where warm colors
represent points that are higher than the reference surface and cool colors represent points
below the reference. The difference maps between the surfaces obtained with full correction and
dividing the optical distances by the refractive indices were calculated by direct subtraction
of the elevation maps, fitted to a Zernike expansion (55 terms, 10th order) within the same
optical zone. The maps were represented using the so called “jet” color map in
Matlab notation, where warm colors depict high differences while cold colors depict low
differences. Thickness maps of the lens were obtained by direct subtraction of the anterior
elevation map from the posterior, as for the full corrected lens as for the optical distances
divided by refractive index one. The maps were represented using the so called
“jet” color map in Matlab notation, where warm colors depict a thicker areas
while cold colors depict thinner areas All maps were displayed in a square grid of 100x100
points in a 5-mm of diameter, with respect to the pupil center.
3. Results
3.1. In vitro samples
The images obtained from the physical model eye were processed using the developed image
analysis and distortion correction algorithms. Table 1
shows the nominal radii of curvature of the cornea and lens surfaces in the model
eye, along with the values before and after application of optical distortion correction
algorithms. Fan distortion correction was applied in all cases. Discrepancies in the retrieved
radii of curvature were 7.5% for the posterior cornea, 9.7% for the anterior lens and as high
as 23.5% for the posterior lens. The accuracy in the radius of curvature retrieval increased
dramatically after optical distortion correction, with discrepancies of 0.3% for the anterior
cornea, 2.0% for the posterior cornea, 1.4% for the anterior lens, and 2.8% for the posterior
lens.
Table 1
Surface radii of curvature from sphere fittings in the model eye: nominal values,
values estimated from OCT measurements before and after correction of optical
distortion.
Cornea
Lens
Nominal (mm)
OCT
Uncorrected (mm)
OCT
Corrected (mm)
Nominal (mm)
OCT Uncorrected (mm)
OCT
Corrected (mm)
Anterior
7.80
7.77
7.77
17.29
18.96
17.05
Posterior
6.48
6.96
6.35
11.04
13.64
10.74
The images obtained from the donor lens in vitro were processed using the
developed image analysis and distortion correction algorithms. Figure 5
shows the elevation maps obtained of the posterior surface of a donor lens in
vitro (direct view of the posterior lens, view of posterior lens distorted by the
anterior lens, and posterior lens corrected from optical distortion by the anterior lens).
Optical distortion correction decreases the error in the retrieved posterior lens radius of
curvature from 8.23% to 0.9%. Correction also decreased the discrepancies in the RMS surface
elevation map (referred to the best fitted sphere) from 48.3% (no optical distortion
correction) to 4.5% (after optical distortion).
Fig. 5
Quantitative elevation maps of the posterior lens surface for the in
vitro 65-year donor lens. Left panel: Measurements with the posterior surface of
the lens facing the OCT beam (“posterior up”); Middle panel: Measurement of
the posterior surface of the lens viewed through the anterior surface of the lens, and no
optical distortion correction (simple division by the index of refraction); Left panel:
Measurement of the posterior surface of the lens viewed through the anterior surface of the
lens, after application of optical distortion correction Maps are Zernike fits to the
elevation maps, relative to the best fitting sphere. R = radii of curvature of the best
fitting sphere (from fits to sphere quadrics).
Quantitative elevation maps of the posterior lens surface for the in
vitro 65-year donor lens. Left panel: Measurements with the posterior surface of
the lens facing the OCT beam (“posterior up”); Middle panel: Measurement of
the posterior surface of the lens viewed through the anterior surface of the lens, and no
optical distortion correction (simple division by the index of refraction); Left panel:
Measurement of the posterior surface of the lens viewed through the anterior surface of the
lens, after application of optical distortion correction Maps are Zernike fits to the
elevation maps, relative to the best fitting sphere. R = radii of curvature of the best
fitting sphere (from fits to sphere quadrics).
3.2. Cornea shape parameters
Corneal surface quantification is necessary to provide accurate optical distortion correction
of the lens surfaces. Table 2
summarizes the anterior and posterior corneal parameters in the 3 subjects of the
study after full distortion corrections, obtained from conicoid and biconicoid fittings of
corneal surfaces in a 5-mm diameter zone.
Table 2
Radii of curvature and asphericity (Q-value) of the anterior and posterior cornea from
biconicoid and conicoid fittings
Radius of
curvature
Anterior
Cornea
Posterior
Cornea
Biconicoid
Conicoid
Biconicoid
Conicoid
Rx
(mm)
Ry (mm)
R (mm)
Rx (mm)
Ry (mm)
R (mm)
S#1
7.40 ± 0.07
7.53 ± 0.03
7.47 ± 0.05
6.76 ± 0.13
6.59 ± 0.07
6.69 ± 0.10
S#2
7.95 ± 0.06
7.82 ± 0.10
7.87 ± 0.11
6.96 ± 0.07
6.95 ± 0.11
6.96 ± 0.08
S#3
7.47 ± 0.05
7.42 ± 0.14
7.43 ± 0.09
6.93 ± 0.06
6.70 ± 0.13
6.72 ± 0.11
Asphericity (Q-Value)
Qx
Qy
Q
Qx
Qy
Q
S#1
0.10 ± 0.01
−0.19 ± 0.05
0.01 ± 0.04
−0.03 ± 0.05
0.06 ± 0.08
−0.01 ± 0.09
S#2
0.20 ± 0.05
−0.18 ± 0.08
0.10 ± 0.03
0.03 ± 0.05
0.01 ± 0.11
0.04 ± 0.03
S#3
0.11 ± 0.06
−0.02 ± 0.09
−0.01 ± 0.07
−0.08 ± 0.04
0.01 ± 0.03
0.01 ± 0.08
3.3. Effects of optical distortion on crystalline lens shape
To quantify the effect of optical distortion from preceding surfaces on the crystalline lens
shape, we evaluated difference maps, obtained as the subtraction of the elevation map estimated
after full optical correction from the map obtained after fan distortion correction and a
simple division of the optical distances by the refractive indices, both for the anterior and
posterior lens surface. Not correcting for optical distortion produced discrepancies ranging
from 0 μm (in the center of the lens) up to 50 μm in the periphery of the
anterior lens surface, and up to 200 μm in the periphery of the posterior lens. The
difference maps (Fig. 6
) reveal the progressive overestimation of the lens curvatures in the non-corrected maps.
Although dominated by the differences in curvature, slight irregularities in the difference
maps indicate the effect of optical distortion in astigmatism and other higher order terms.
Fig. 6
Difference lens anterior (top) and posterior (bottom) elevation maps after optical
distortion correction relative to elevation maps obtained by simple division of the optical
distances by their corresponding refractive indices. Data are for 5-mm pupils.
Difference lens anterior (top) and posterior (bottom) elevation maps after optical
distortion correction relative to elevation maps obtained by simple division of the optical
distances by their corresponding refractive indices. Data are for 5-mm pupils.
3.4. Crystalline lens shape
Lens surfaces radii and asphericities (with and without optical distortion corrections) were
obtained from OCT images obtained in vivo in three subjects, from conicoid and
biconicoid fits to the retrieved surfaces (within 5-mm diameter zones). Data in Table 3
represent averages (and standard deviations) of 3 repeated measurements on each
eye. Not correcting for optical distortion produced a systematic overestimation of the lens
radii of curvature (Table 2), between 20.46 and 22.77%
for the horizontal meridian, and between 20.96 and 32.81% for the vertical meridian (in the
anterior lens), and between 63.40 and 72.70% for the horizontal meridian, and between 68.14 and
80.27% for the vertical meridian (in the posterior lens). For a conic fit, radii were
overestimated between 20.20 and 23.56% in the anterior lens, and between 69.28 and 78.72% in
the posterior lens.
Table 3
Radii of curvature from biconicoid and conicoid fits of the uncorrected and
optical distortion corrected crystalline lens surfaces
in vivo
Anterior
Lens
Posterior
Lens
Biconicoid
Conicoid
Biconicoid
Conicoid
Rx
(mm)
Ry (mm)
R (mm)
Rx (mm)
Ry (mm)
R (mm)
S#1
Uncorrected
17.36 ± 0.11
13.95 ± 0.21
15.42 ± 0.19
11.64 ± 0.01
13.51 ± 0.22
12.48 ± 0.11
Corrected
14.14 ± 0.13
11.24 ± 0.30
12.48 ± 0.20
6.74 ± 0.11
7.54 ± 0.20
7.25 ± 0.25
S#2
Uncorrected
13.83 ± 0.07
13.64 ± 0.24
13.74 ± 0.09
10.06 ± 0.26
10.66 ± 0.29
10.36 ± 0.20
Corrected
11.47 ± 0.09
10.27 ± 0.29
11.43 ± 0.04
6.10 ± 0.12
6.34 ± 0.04
6.12 ± 0.15
S#3
Uncorrected
15.60 ± 0.04
13.79 ± 0.38
14.64 ± 0.21
11.52 ± 0.12
13.43 ± 0.23
12.85 ± 0.04
Corrected
12.95 ± 0.12
11.40 ± 0.19
12.18 ± 0.06
7.04 ± 0.12
7.45 ± 0.16
7.19 ± 0.03
Lens asphericities (Table 4
) were negative in all eyes, although values differed significantly across
subjects. While optical distortion correction modified the asphericity parameter, there was not
a consistent trend toward overestimation or understimation without correction.
Table 4
Asphericity (Q-value) from biconicoid and conicoid fits of the uncorrected and optical
distortion corrected crystalline lens surfaces
Anterior
Posterior
Biconicoid
Conicoid
Biconicoid
Conicoid
Qx
Qy
Q
Qx
Qy
Q
S#1
Uncorrected
−0.40 ± 0.34
−0.33 ± 0.12
−6.02 ± 0.03
−0.09 ± 0.06
−0.49 ± 0.11
−0.67 ± 0.25
Corrected
−0.41 ± 0.15
−0.36 ± 0.20
−2.57 ± 0.31
−1.87 ± 0.93
−1.96 ± 0.76
−1.64 ± 0.33
S#2
Uncorrected
−0.33 ± 0.37
−0.42 ± 0.32
−0.29 ± 0.07
−0.45 ± 0.33
−0.24 ± 0.20
−0.66 ± 0.16
Corrected
−0.40 ± 0.16
−0.26 ± 0.29
−0.43 ± 0.21
−0.06 ± 0.02
−0.09 ± 0.07
−0.01 ± 0.00
S#3
Uncorrected
−0.35 ± 0.41
−0.23 ± 0.20
−1.21 ± 0.43
−0.40 ± 0.10
−0.25 ± 0.23
−0.84 ± 0.22
Corrected
−0.19 ± 0.23
−0.25 ± 0.32
−0.93 ± 0.40
−0.46 ± 0.22
−0.07 ± 0.08
−0.59 ± 0.07
3.5. Crystalline lens elevation
Lens surfaces were fitted by Zernike polynomial expansions (up to the 10th order) within a
5-mm optical zone. Figure 7
shows representative 2nd and 3rd order Zernike terms (astigmatism, coma and trefoil).
Data are averages of 3 repeated measurements on each eye, and the error bars represent the
standard deviation. Only fits corresponding to full distortion corrected surfaces are shown.
The most relevant term was astigmatism at 0 and 90 deg (). Not correcting optical distortion produced an underestimation
of the magnitude of astigmatism by 33% on average. Interestingly, all eyes showed vertical
astigmatism in the anterior lens (negative , i.e. steepest vertical meridian), but a horizontal astigmatism
in the posterior lens (positive , i.e. steepest vertical meridian).
Fig. 7
Representative 2nd and 3rd order Zernike terms from the Zernike fit to full distortion
corrected (a) anterior lens surface, and (b) posterior lens surface in 3 human lenses
in vivo. Data are average values of 3 repeated measurements on S#1 (red
squares), S#2 (green triangles), and S#3 (blue diamonds). Error bars are not represented
since the error is smaller than the symbol. Data are for 5-mm pupils.
Representative 2nd and 3rd order Zernike terms from the Zernike fit to full distortion
corrected (a) anterior lens surface, and (b) posterior lens surface in 3 human lenses
in vivo. Data are average values of 3 repeated measurements on S#1 (red
squares), S#2 (green triangles), and S#3 (blue diamonds). Error bars are not represented
since the error is smaller than the symbol. Data are for 5-mm pupils.Figure 8
shows elevation maps of the anterior and posterior lens surfaces in the 3 subjects of
the study, within a 5-mm diameter zone. The elevation maps represent 10th order Zernike fits of
the elevation maps relative to the best fitting sphere. The numbers below each map indicate the
radii of curvature of best fitting sphere.
Fig. 8
Quantitative anterior (top) and posterior (bottom) crystalline lens elevation maps in 3
eyes in vivo, after full distortion correction. Maps are Zernike fits to
the elevation maps, relative to the best fitting sphere. R = radii of curvature of the best
fitting sphere (from fits to sphere quadrics). Data are for 5-mm pupils.
Quantitative anterior (top) and posterior (bottom) crystalline lens elevation maps in 3
eyes in vivo, after full distortion correction. Maps are Zernike fits to
the elevation maps, relative to the best fitting sphere. R = radii of curvature of the best
fitting sphere (from fits to sphere quadrics). Data are for 5-mm pupils.Figure 9
shows an illustration of the repeatability of the lens elevation maps in subject S#3,
for both the anterior and posterior lens surfaces (5-mm optical zone). The numbers below each
map represent the radii of curvature of best fitting sphere in mm. The average anterior lens
radius of curvature was 12.02 ± 0.21 mm (1.7% of variability). The average posterior
lens radius of curvature was 7.37 ± 0.09 mm, (1.2% variability). The average RMS lens
elevation (relative to the best fitting sphere) was 2.08 ± 0.09 mm for the anterior lens
(4% variability) and 9.92 ± 0.83 mm for the posterior lens (8% variability).
Fig. 9
Three repeated anterior and posterior lens surface elevation maps in subject #3. R = radii
of curvature of the best fitting sphere (from fits to sphere quadrics). Data are for 5-mm
pupils.
Three repeated anterior and posterior lens surface elevation maps in subject #3. R = radii
of curvature of the best fitting sphere (from fits to sphere quadrics). Data are for 5-mm
pupils.Additionally, in order to test the overall repeatability, the focal length of the eye was
estimated by ray tracing through all reconstructed surfaces (cornea and crystalline lens). The
estimated focal length ranged from 25.28 to 25.37 mm across repeated measurements in subject
S#3, indicating within 1% repeatability.
3.6. Crystalline lens thickness
Figure 10
shows lens thickness maps in the three subjects, estimated using uncorrected surfaces
(obtained by simple divisions the optical distances by the average lens refractive index, and
fan distortion corrections) and fully corrected surfaces (within a 5-mm diameter zone, centered
at the pupil center). Central lens thickness was 3.18 ± 0.02mm for S#1, 3.36 ±
0.02mm for S#2, and 3.06 ± 0.04mm for S#3, and did not change with optical distortion
correction of the surfaces. However, not correcting for optical distortion overestimated
thickness by more than 15% on average in peripheral areas (2.5 mm from the center).
Fig. 10
Crystalline lens thickness maps obtained as direct subtraction from anterior to the
posterior elevation maps, for uncorrected surfaces (top) and for fully corrected surfaces
(bottom). Data are for 5-mm optical zone.
Crystalline lens thickness maps obtained as direct subtraction from anterior to the
posterior elevation maps, for uncorrected surfaces (top) and for fully corrected surfaces
(bottom). Data are for 5-mm optical zone.
4. Discussion
We have presented a new methodology for obtaining accurate shape and 3-D topographies of the
crystalline lens surfaces. To our knowledge, this is the first time that OCT has been applied to
obtain quantitative 3-D geometry of the crystalline lens in vivo. The method
has been tested using a model eye and a humandonor lens in vitro, showing a
good correspondence between nominal data of the model eye and the retrieved data after
correction, and the posterior lens surface (obtained with the posterior lens surface facing the
OCT beam) and the reconstructed posterior lens surface in the donor lens. The presented method
builds on previous methods developed to obtain quantitative corneal elevation from OCT images
[45,49],
including fan and optical distortion correction [32,33,45,46,49], as well as
image processing tools, which, along with newly developed routines for volume clustering (of
cornea, iris, and lens), and 3-D merging and registration, allowed quantitative lens surface
segmentation, correction and elevation. The lens surfaces were fitted by conicoids and
biconicoids, as well as Zernike polynomial descriptions, which allowed estimates of the anterior
and posterior lens radii of curvature, asphericity, astigmatism and high order
irregularities.Correction of optical distortion proved particularly relevant in the crystalline lens imaging,
giving the accumulated refraction changes by multiple preceding surfaces, and the ray
convergence. In keeping with theoretical predictions [46], correcting optical distortion improved accuracy in the estimates of the lens radii
of curvature by 25% and 65% on average in the anterior and posterior lens respectively.Most previous phakometric techniques were limited to the estimates of radii of curvature
either from indirect measurements of light reflections from the lens surfaces (Purkinje imaging)
or anterior segment imaging (such as Scheimpflug imaging). Our lens radii of curvature estimates
in vivo (11.90 ± 1.34 mm for the anterior lens and 6.86 ± 0.63
mm for the posterior lens) fall within the ranges reported in the literature. Classical eye
model [4] used 10 mm for the anterior lens and 6 mm for
the posterior lens, whereas more recent models [54] used
12.4 and 8.1 mm respectively. Dubbleman et al. [16]
reported an age-dependent expression for the anterior and posterior lens radius of human lenses
in vivo based on corrected Scheimpflug measurements, which applied to the
average data of our three subjects would predict 11.02-11.3 mm and 5.8-5.86 mm for the anterior
and posterior lens radius respectively. Rosales et al. [20] in a comparative study of the crystalline lens radii of curvature estimated from
Purkinje and Scheimpflug imaging on the same group of young subjects reported 10.8 mm and 11.1
mm, and 6.7 mm and 6.1 mm from each technique, for anterior and posterior lens radius of
curvature respectively. Differences with respect to the OCT-based estimates may arise from the
assumption of spherical surfaces in the Purkinje method [20], or the fit within a 3-mm optical zone in the Scheimpflug method [16], while we used a conic/biconic fitting within a 5-mm zone,
and the reported radii of curvature correspond to the apical zone. Nevertheless, the
intersubject variability (within the same age range than our three subjects) reported in
previous studies is large: a range of 9-14 mm for the anterior and 4.7 −7 mm for the
posterior lens radii in Dubbleman et al. [16]; 8.1-13.8
mm for the anterior and 5.1-7.15 mm for the posterior lens radii in Rosales et al. [20] in another study using Scheimpflug imaging, and 7.9-14.3
mm for the anterior and 5.2-8.65 mm for the posterior lens radii in Rosales et al. [21] using Purkinje imaging. The data reported in this study
appear less variable across subjects, and well within the reported ranges.To our knowledge, only Dubbleman et al. [16] reported
lens asphericity measured in vivo from Scheimpflug imaging, also providing
age-related expressions. The predicted asphericity from that study in the age range of our
studies (−4.56 and −3.04 for the anterior and posterior lens surfaces,
respectively) is significantly more negative than that obtained in the current OCT-based study
(−1.31 ± 1.12 and −0.75 ± 0.83 for the anterior and posterior lens
surfaces, respectively). However, the intersubject variability found by Dubbleman et al. [16] was extremely large (−20 to 7 for the anterior lens
and −13 to 2 for the posterior lens). It is likely that the specific fitting method used
by Dubbleman (two-stage fitting to a spherical surface followed by a fitting of the residuals to
conic), the use of 2-D data sets, and the reported interactions between radius of curvature and
asphericity in conic fits [55] play some role in the
differences.The 3-D acquisition and reconstruction of the lens surfaces with OCT allowed us to obtain 3-D
lens geometry in vivo, for, to our knowledge, the first time in the literature.
We had previously measured the 3-D geometry of crystalline lenses in vitro in
the porcine [40] and human [56] lens. The correction of optical distortions from the cornea allowed
reconstructions of lens surface topographies, very much like the conventional surface elevation
of the cornea, widespread in the clinical practice. As our previous OCT-based corneal
topographies, elevation measurements are based on a collection of regular B-scans, rather than
meridians [47,48].Unlike previous cross-sectional measurements of lens geometry which were limited to one
meridian, 3-D measurements allow to investigate lens surface astigmatism. Very interestingly,
the three subjects of our study show perpendicular orientation in the astigmatism of the
anterior versus the posterior lens surface. To date, astigmatism of the crystalline lens
in vivo is usually assessed by subtracting corneal from total aberrations
[57-59].
Compensation of the corneal astigmatism by internal astigmatism has been postulated [59], as well as changes of this balance with age [59]. Measurements of the crystalline lens elevation in
vivo will offer the possibility of investigating the sources of this compensation and
potential age-related changes. In a previous study we investigated lens surface contribution to
lens astigmatism in a porcine lens in vitro, and found that the lens gradient
index of refraction (GRIN) may also play a compensatory role [40]. Undergoing research in our laboratory explores the role of GRIN in astigmatism and
spherical aberration both in porcine and human lenses in vitro.Our OCT-based crystalline lens elevation maps show minimal amounts of trefoil, coma or other
high order aberrations. It is likely that the coma measured by direct subtraction of total minus
corneal aberrations arises from the misalignment of the ocular components, including the
off-axis position of the fovea [60,61]. Measurements of optical aberrations in isolated lenses using lateral
images of the ray tracing in primate lenses [62], or by
diffraction point interferometry [63] indicate presence
of trefoil and other high order aberrations. The fact that the lens surface topographies are
relative smooth suggests a contribution of the internal lens structure (likely line sutures) to
the reported lens optical irregularities.We provided the first report of lens surface shapes in vivo from OCT upon
correction of optical (refraction) distortion, although (as previous reports based on Purkinje
or Scheimpflug) we did not considered a GRIN in the lens. Borja et al. [41] suggested that the optimal refractive index for the correction of OCT
images was closer to the average refractive index [37]
than to the equivalent refractive index. In previous studies [41,44] we analyzed the potential effect of GRIN
in the visualization and quantification of the posterior lens surface, and developed a tool for
thru-GRIN optical distortion correction of OCT images, which we applied in
vitro in 2-D. We concluded that using a homogeneous index of refraction as opposed to
the actual lens GRIN profile did not produce significant discrepancies on the estimated radius
of curvature, although it could overestimate the posterior lens asphericity in a 6 mm pupil
diameter [41]. However, the results using a 4 mm pupil
diameter [44] showed an improvement only in the RMS or
peak to value differences. This result is consistent with the finding in a donor lens (now in
3-D) presented in the current study for validation. Estimates of the posterior lens surface
elevation and lens thickness in vivo could therefore benefit slightly by
increasing knowledge (and potential retrieval in vivo) of the lens GRIN.
5. Conclusions
Anterior segment OCT provided with full distortion correction and automatic analysis tools
allowed quantification of the human crystalline lens in vivo in 3-D. Correction
of optical distortion is critical as simple division of the OCT optical distances by the
refractive index produces an overestimation of the crystalline lens anterior and posterior radii
of curvature by 25% and 65% respectively. The retrieved lens radii of curvature from OCT agree
with phakometric data previously reported using Scheimpflug and Purkinke image along one
meridian. We found slight negative lens surface asphericities, crossed astigmatism in the
anterior and posterior lens surface, and low high order irregularities in the lens surfaces.
OCT-based lens topography is a promising tool to investigate the lens optical properties, the
contribution of the lens to the overall retinal image quality, and understanding the change of
the physical properties of the lens with accommodation and aging.
Authors: Michalina Gora; Karol Karnowski; Maciej Szkulmowski; Bartlomiej J Kaluzny; Robert Huber; Andrzej Kowalczyk; Maciej Wojtkowski Journal: Opt Express Date: 2009-08-17 Impact factor: 3.894
Authors: A de Castro; D Siedlecki; David Borja; Stephen Uhlhorn; Jean-Marie Parel; Fabrice Manns; S Marcos Journal: J Mod Opt Date: 2011-11-24 Impact factor: 1.464
Authors: Victor M Hernandez; Florence Cabot; Marco Ruggeri; Carolina de Freitas; Arthur Ho; Sonia Yoo; Jean-Marie Parel; Fabrice Manns Journal: Biomed Opt Express Date: 2015-10-21 Impact factor: 3.732