| Literature DB >> 23425449 |
Dirk Van Gestel1, Corine van Vliet-Vroegindeweij, Frank Van den Heuvel, Wouter Crijns, Ann Coelmont, Bie De Ost, Andrea Holt, Emmy Lamers, Yasmyne Geussens, Sandra Nuyts, Danielle Van den Weyngaert, Tim Van den Wyngaert, Jan B Vermorken, Vincent Gregoire.
Abstract
BACKGROUND: Radiotherapy techniques have evolved rapidly over the last decade with the introduction of Intensity Modulated RadioTherapy (IMRT) in different forms. It is not clear which of the IMRT techniques is superior in the treatment of head and neck cancer patients in terms of coverage of the planning target volumes (PTVs), sparing the organs at risk (OARs), dose to the normal tissue, number of monitor units needed and delivery time.The present paper aims to compare Step and Shoot (SS) IMRT, Sliding Window (SW) IMRT, RapidArc (RA) planned with Eclipse, Elekta VMAT planned with SmartArc (SA) and helical TomoHD™ (HT).Entities:
Mesh:
Year: 2013 PMID: 23425449 PMCID: PMC3599972 DOI: 10.1186/1748-717X-8-37
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Patient characteristics
| 1 | Base of tongue R | C01 | T1N2aM0 | IV A | 2 & 3 R | 104 | 483 |
| 2 | Tonsil R | C09 | T2N2cM0 | IV A | 2 bilat & 3 R | 233 | 610 |
| 3 | Tonsil L | C09 | T3N2cM0 | IV A | bilat 1,2,3,4 | 380 | 955 |
| 4 | Base of tongue R | C01 | T3N2cM0 | IV A | 1b R & 2 bilat | 422 | 831 |
| 5 | Tonsil L | C09 | T2N1M0 | III | 2 L | 146 | 563 |
| average | 257 | 688 |
Bilat = bilateral; LN + = positive lymph node; R = right; L = left; PTVtotal = PTVtherapeutic and PTVprophylactic together; PTV volumes reported in this table are calculated by the Pinnacle treatment planning system.
Dose-volume constraints for PTVs and organs at risk
| PTV 56 | 56 Gy | | V59.9 Gy | ≥ 95% of planned absorbed dose | |
| PTV 69 | 69.12 Gy | | | ≥ 95% of planned absorbed dose | ≤ 107% of planned absorbed dose |
| PRV Spinal cord | | | D2 | | ≤ 50 Gy |
| PRV Brainstem | ≤ 55 Gy | | D2 | | ≤ 59 Gy |
| Parotid gland contralateral | | ≤ 23 Gy | Mean D, V27 | | |
| Parotid gland ipsilateral | | ≤ 27 Gy | Mean D, V27 | | |
| Submandibular gland | | ≤ 39 Gy | Mean D | | |
| Oral mucosa | | ≤ 27 Gy | Mean D, V27 | | |
| Mandible | | | V60 | | |
| Soft palate | | ≤ 27 Gy | Mean D, V27 | | |
| Constrictor muscles | | ≤ 55 Gy | Mean D, V20 | | |
| Cricopharyngeal muscle | | ≤ 55 Gy | Mean D, V20 | | |
| Base of tongue | | ≤ 55 Gy | Mean D, V20 | | |
| Larynx | | ≤ 40 Gy | Mean D, V40 | | |
| Esophagus superior | | ≤ 35 Gy | Mean D, V35 | | |
| Top of lung | | | V20 | | |
| Inner ear | Mean D, V45 |
PTV = planning target volume; PRV = planning risk volume; Vx =% volume receiving x Gy; Dx = dose to x% of the volume; D = dose; ALARA = as low as reasonable achievable.
Mean volumes of PTVs and OARs calculated by the different planning systems (in cubic centimetre)
| PTV 69 | | 257.0 | 254.2 | 251.2 |
| PTV 56 | | 431.4 | 425.3 | 414.5 |
| * PRV Spinal cord | | 64.0 | 63.4 | 61.8 |
| * PRV Brainstem | | 53.9 | 52.8 | 52.8 |
| * Parotid gland | contralateral | 27.6 | 26.9 | 26.7 |
| | ipsilateral | 26.5 | 26.0 | 25.7 |
| * Submandibular | contralateral | 7.4 | 7.1 | 7.2 |
| * Oral mucosa | | 51.0 | 49.1 | 50.1 |
| * Mandible | | 67.7 | 65.7 | 64.0 |
| * Middle pharyngeal constrictor | | 1.6 | 1.5 | |
| * Lower pharyngeal constrictor | | 6.5 | 6.4 | |
| * Cricopharyngeal muscle | | 2.3 | 2.3 | |
| * Oesophagus (cranial part) | | 5.3 | 5.1 | |
| * Supraglottic larynx | | 11.3 | 11.0 | |
| * Glottic larynx | | 5.1 | 5.0 | |
| * Top of lung | contralateral | 438.0 | 423.5 | 432.5 |
| | ipsilateral | 430.5 | 415.8 | 425.1 |
| * Brachial plexus | contralateral | 12.7 | 12.2 | |
| | ipsilateral | 11.5 | 11.0 | |
| * Inner ear | contralateral | 2.5 | 2.4 | |
| | ipsilateral | 2.2 | 2.1 | |
| * Brain - PRV brainstem | | 932.1 | 914.5 | 922.2 |
| * Skin near PTV | | 148.6 | 145.4 | |
| * Eye | contralateral | 6.8 | 6.5 | 6.6 |
| | ipsilateral | 6.5 | 6.4 | |
| * Shoulder | contralateral | 161.4 | 162.0 | 158.5 |
| | ipsilateral | 154.6 | 156.7 | 151.8 |
| * Non specified tissue | 6473.8 | 6373.3 | 6394.0 |
Bold = more than 5% difference in volume in comparison with Pinnacle; PTV = planning target volume; PRV = planning risk volume.
Figure 1Mean values of PTVs for the different IMRT planning systems.
Figure 2Dose volume histograms of mean PTVand mean PTV.
Figure 3Critical doses and volumes of the different organs at risk for the five planning systems.
Figure 4Dose volume histograms of mean values of both parotid glands and total body.
Mean beam on time, treatment time and monitor units for the different planning systems
| | n | mean | mean | mean | mean | mean | |||||
| 5 | 1.87 | 3.74 | 2.50 | 4.00 | 5.94 | ||||||
| 5 | 14.01 | | 8.17* | | 3.05 | | 6.00 | | 5.94 | | |
| 5 | 746 | 1125 | 415 | 608 | 5052 | ||||||
SE = standard error; * = the irradiation time of one of the sliding window IMRT plans, representative for all sliding window IMRT plans.