| Literature DB >> 24927727 |
Jin Ho Song1, Ji-Young Jung2, Hyung-Wook Park3, Gi Woong Lee4, Soo-Min Chae4, Chul Seung Kay2, Seok Hyun Son5.
Abstract
The aim of this study was to compare lateral photon-electron (LPE), helical tomotherapy (HT), and volumetric-modulated arc therapy (VMAT) plans for total scalp irradiation. We selected a single adult model case and compared the dosimetric results for the three plans. All plans mainly used 6-MV photon beams, and the prescription dose was 60 Gy in 30 fractions. First, we compared the LPE, HT and VMAT plans, with all plans including a 1-cm bolus. We also compared HT plans with and without the bolus. The conformity indices for LPE, HT and VMAT were 1.73, 1.35 and 1.49, respectively. The HT plan showed the best conformity and the LPE plan showed the worst. However, the plans had similar homogeneity indexes. The dose to the hippocampus was the highest in the VMAT plan, with a mean of 6.7 Gy, compared with 3.5 Gy in the LPE plan and 4.8 Gy in the HT plan. The doses to the optical structures were all within the clinically acceptable range. The beam-on time and monitor units were highest in the HT plan. The HT plans with and without a bolus showed similar target coverage and organ-at-risk (OAR) sparing. The HT plan showed the best target coverage and conformity, with low doses to the brain and hippocampus. This plan also had the advantage of not necessarily requiring a bolus. Although the VMAT plan showed better conformity than the LPE plan and acceptable OAR sparing, the dose to the hippocampus should be considered when high doses are prescribed.Entities:
Keywords: hippocampus; total scalp irradiation; lateral photon–electron technique; helical tomotherapy; volumetric-modulated arc therapy
Mesh:
Year: 2014 PMID: 24927727 PMCID: PMC4497384 DOI: 10.1093/jrr/rru049
Source DB: PubMed Journal: J Radiat Res ISSN: 0449-3060 Impact factor: 2.724
Fig. 1.(a) The photon and electron field in the conventional plan. The fields are matched on the skin. The arrowheads indicate the matching line. (b) The positions of the three centers in the volumetric-modulated arc therapy plan.
Fig. 2.Axial, coronal and sagittal images of the isodose distribution of each plan with a bolus. (LPE = lateral photon–electron plan, HT = helical tomotherapy plan, VMAT = volumetric-modulated arc therapy plan.)
Fig. 3.Dose–volume histograms for the scalp, brain and hippocampus. (Red line = the conventional lateral photon–electron plan, blue line = helical tomotherapy, green line = volumetric-modulated arc therapy.)
Dosimetric comparisons of target coverage in the LPE, HT and VMAT plans
| LPE | HT | VMAT | HT/LPE | VMAT/LPE | VMAT/HT | ||
|---|---|---|---|---|---|---|---|
| CI | 1.73 | 1.35 | 1.49 | −22.0 | −13.9 | 10.4 | |
| HI | 1.15 | 1.15 | 1.17 | 0 | 1.7 | 1.7 | |
| CN | 0.57 | 0.74 | 0.66 | 29.8 | 15.8 | −10.8 | |
| COIN | 0.56 | 0.74 | 0.66 | 32.1 | 17.9 | −10.8 |
LPE = lateral photon–electron plan, HT = helical tomotherapy, VMAT = volumetric-modulated arc therapy, CI = conformity index, HI = homogeneity index, CN = conformation number, COIN = conformation index.
Dosimetric analyses of the mean, maximum and minimum doses to nine OARs in the LPE, HT and VMAT plans
| OARs | Mean (Gy) | Max (Gy) | Min (Gy) | ||||||
|---|---|---|---|---|---|---|---|---|---|
| LPE | HT | VMAT | LPE | HT | VMAT | LPE | HT | VMAT | |
| Brain | 16.8 | 9.8 | 14.6 | 64.0 | 57.4 | 53.1 | 0.9 | 1.0 | 0.9 |
| Brain stem | 2.5 | 3.6 | 6.2 | 4.0 | 5.7 | 8.6 | 1.3 | 1.5 | 2.7 |
| Hippocampus | 3.5 | 4.8 | 6.7 | 4.6 | 6.5 | 9.8 | 1.5 | 4.7 | 4.4 |
| Lt eyeball | 1.6 | 5.2 | 6.2 | 1.8 | 18.7 | 13.3 | 1.4 | 1.8 | 3.9 |
| Rt eyeball | 2.0 | 4.5 | 5.9 | 6.2 | 17.3 | 12.2 | 1.2 | 1.7 | 3.8 |
| Lt lens | 1.6 | 3.0 | 6.0 | 1.8 | 3.6 | 6.2 | 1.4 | 2.4 | 5.5 |
| Rt lens | 1.7 | 2.4 | 5.5 | 2.0 | 3.0 | 6.4 | 1.5 | 2.0 | 4.5 |
| Lt optic nerve | 1.9 | 3.5 | 5.9 | 2.3 | 5.0 | 6.2 | 1.7 | 2.8 | 5.4 |
| Rt optic nerve | 1.8 | 2.9 | 6.4 | 2.1 | 3.6 | 6.6 | 1.6 | 2.4 | 5.9 |
| V10Gy (%) | V20Gy (%) | V30Gy (%) | |||||||
| Brain | 39.2 | 25.0 | 48.9 | 29.4 | 12.5 | 22.6 | 22.9 | 7.4 | 11.3 |
| V40Gy (%) | V50Gy (%) | ||||||||
| Brain | 17.4 | 3.0 | 2.9 | 11.8 | 0.3 | 0.1 | |||
OARs = organs at risk, LPE = lateral photon–electron plan, HT = helical tomotherapy, VMAT = volumetric-modulated arc therapy, Lt = left side, Rt = right side, V10Gy = volume receiving at least 10 Gy, V20Gy = volume receiving at least 20 Gy, V30Gy = volume receiving at least 30 Gy, V40Gy = volume receiving at least 40 Gy, V50Gy = volume receiving at least 50 Gy.
Dosimetric comparisons of the QIs for nine OARs between the LPE, HT and VMAT plans
| OARs | QI values | ||
|---|---|---|---|
| HT/LPE | VMAT/LPE | VMAT/HT | |
| Brain | 0.58 | 0.86 | 1.48 |
| Brain stem | 1.44 | 2.48 | 1.72 |
| Hippocampus | 1.37 | 1.91 | 1.39 |
| Lt eyeball | 3.25 | 3.87 | 1.19 |
| Rt Eyeball | 2.25 | 2.96 | 1.31 |
| Lt lens | 1.87 | 3.44 | 2.00 |
| Rt Lens | 1.41 | 3.23 | 2.29 |
| Lt optic nerve | 2.17 | 2.69 | 1.24 |
| Rt optic nerve | 1.74 | 3.14 | 1.83 |
| mean ± SD | 1.78 ± 0.74 | 2.73 ± 0.90 | 1.60 ± 0.38 |
OARs = organs at risk, QI = quality index, LPE = lateral photon–electron plan, HT = helical tomotherapy, VMAT = volumetric-modulated arc therapy, Lt = left side, Rt = right side, SD = standard deviation.
Beam-on times and MUs in a single fraction in the LPE, HT and VMAT plans
| LPE | HT | VMAT | |
|---|---|---|---|
| Beam-on time (min) | 1.5 | 14.7 | 11.3 |
| MUs | 616 | 13 088 | 1 791 |
LPE = lateral photon–electron plan, HT = helical tomotherapy, VMAT = volumetric-modulated arc therapy, MUs = Monitor Units.
Dosimetric comparisons of target coverage for HT plans with and without a bolus
| HT with a bolus | HT without a bolus | HT without a bolus/HT with a bolus | ||
|---|---|---|---|---|
| CI | 1.35 | 1.26 | ΔCI (%) | −6.7 |
| HI | 1.15 | 1.14 | ΔHI (%) | −0.8 |
| CN | 0.74 | 0.79 | ΔCN (%) | 6.7 |
| COIN | 0.74 | 0.79 | ΔCOIN (%) | 6.7 |
HT = helical tomotherapy, CI = conformity index, HI = homogeneity index, CN = conformation number, COIN = conformation index.
Dosimetric comparisons of the mean, maximum and minimum doses to nine OARs for HT plans with and without a bolus
| OARs | Mean (Gy) | Max (Gy) | Min (Gy) | QI value | |||
|---|---|---|---|---|---|---|---|
| With a bolus | Without a bolus | With a bolus | Without a bolus | With a bolus | Without a bolus | ||
| Brain | 9.8 | 10.3 | 57.4 | 59.5 | 1.0 | 0.9 | 1.05 |
| Brain stem | 3.6 | 3.3 | 5.7 | 4.4 | 1.5 | 1.4 | 0.92 |
| Hippocampus | 4.8 | 4.7 | 6.5 | 6.4 | 4.7 | 3.5 | 0.98 |
| Lt eyeball | 5.2 | 4.8 | 18.7 | 20.9 | 1.8 | 2.1 | 0.92 |
| Rt eyeball | 4.5 | 6.0 | 17.3 | 19.7 | 1.7 | 1.7 | 1.33 |
| Lt lens | 3.0 | 3.7 | 3.6 | 4.5 | 2.4 | 3.1 | 1.23 |
| Rt lens | 2.4 | 2.7 | 3.0 | 3.4 | 2.0 | 2.2 | 1.12 |
| Lt optic nerve | 3.5 | 3.8 | 5.0 | 5.8 | 2.8 | 3.3 | 1.08 |
| Rt optic nerve | 2.9 | 3.2 | 3.6 | 3.9 | 2.4 | 2.7 | 1.10 |
| V10Gy (%) | V20Gy (%) | V30Gy (%) | |||||
| Brain | 25.0 | 25.4 | 12.5 | 13.9 | 7.4 | 8.8 | |
| V40Gy (%) | V50Gy (%) | ||||||
| Brain | 3.0 | 4.4 | 0.3 | 1.1 | |||
OARs = organs at risk, QI = quality index, Lt = left side, Rt = right side, V10Gy = volume receiving at least 10 Gy, V20Gy = volume receiving at least 20 Gy, V30Gy = volume receiving at least 30 Gy, V40Gy = volume receiving at least 40 Gy, V50Gy = volume receiving at least 50 Gy.
Fig. 4.Comparisons of dose–volume histograms for the brain: helical tomotherapy plans with and without a bolus. (Red line = with a bolus; blue line = without a bolus.)