Literature DB >> 21969242

Angiographic CT with intravenous contrast agent application for monitoring of intracranial flow diverting stents.

Marc Saake1, Tobias Struffert, Philipp Goelitz, Sabine Ott, Frank Seifert, Oliver Ganslandt, Arnd Doerfler.   

Abstract

INTRODUCTION: Intracranial flow diverting devices are increasingly used to treat cerebral aneurysms. A reliable, non-invasive follow-up modality would be desirable. Our aim was to compare intra-arterial digital subtraction angiography (ia DSA) to angiographic computed tomography with intravenous contrast agent application (iv ACT) in the visualisation of flow diverting devices and aneurysm lumina.
METHODS: Follow-up monitoring by iv ACT (n = 36) and ia DSA (n = 25) in 14 patients treated with flow diverting devices for intracranial aneurysms was evaluated retrospectively. Images were evaluated by two neuroradiologists in anonymous consensus reading regarding the device deployment, wall apposition, neck coverage of the aneurysm, opacification of the vessel and device lumen, as well as the degree of aneurysm occlusion.
RESULTS: Corresponding ia DSA and iv ACT images were scored identically in all patients regarding the stent deployment, wall apposition and neck coverage, as well as the degree of aneurysm occlusion and patency status of the device and parent artery. Opacification of the parent vessel lumen and perfused parts of the aneurysm was considered slightly inferior for iv ACT in comparison with ia DSA (seven of 36 cases), without impact on diagnosis.
CONCLUSIONS: We demonstrated the feasibility and diagnostic value of iv ACT in follow-up imaging of intracranial flow diverting devices. Due to its high spatial resolution and non-invasive character, this novel technique might become a valuable imaging modality in these patients.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21969242     DOI: 10.1007/s00234-011-0965-9

Source DB:  PubMed          Journal:  Neuroradiology        ISSN: 0028-3940            Impact factor:   2.804


  18 in total

1.  Optimized intravenous Flat Detector CT for non-invasive visualization of intracranial stents: first results.

Authors:  Tobias Struffert; Stephan Kloska; Tobias Engelhorn; Yu Deuerling-Zheng; Sabine Ott; Marc Doelken; Marc Saake; Martin Köhrmann; Arnd Doerfler
Journal:  Eur Radiol       Date:  2010-08-14       Impact factor: 5.315

2.  Intra-aneurysmal thrombosis as a possible cause of delayed aneurysm rupture after flow-diversion treatment.

Authors:  Z Kulcsár; E Houdart; A Bonafé; G Parker; J Millar; A J P Goddard; S Renowden; G Gál; B Turowski; K Mitchell; F Gray; M Rodriguez; R van den Berg; A Gruber; H Desal; I Wanke; D A Rüfenacht
Journal:  AJNR Am J Neuroradiol       Date:  2010-11-11       Impact factor: 3.825

3.  Early fatal hemorrhage after endovascular cerebral aneurysm treatment with a flow diverter (SILK-Stent): do we need to rethink our concepts?

Authors:  Bernd Turowski; Stephan Macht; Zolt Kulcsár; Daniel Hänggi; Walter Stummer
Journal:  Neuroradiology       Date:  2010-03-26       Impact factor: 2.804

4.  Angiographic CT in cerebrovascular stenting.

Authors:  Götz Benndorf; Charles M Strother; Benjamin Claus; Ramin Naeini; Hesham Morsi; Richard Klucznik; Michael E Mawad
Journal:  AJNR Am J Neuroradiol       Date:  2005-08       Impact factor: 3.825

5.  Flat-detector computed tomography in the assessment of intracranial stents: comparison with multi detector CT and conventional angiography in a new animal model.

Authors:  Tobias Struffert; Sabine Ott; Edyta Adamek; Marc Schwarz; Tobias Engelhorn; Stephan Kloska; Yu Deuerling-Zheng; Arnd Doerfler
Journal:  Eur Radiol       Date:  2011-03-02       Impact factor: 5.315

Review 6.  Early experience with flow diverting endoluminal stents for the treatment of intracranial aneurysms.

Authors:  Brian P Walcott; Jared M Pisapia; Brian V Nahed; Kristopher T Kahle; Christopher S Ogilvy
Journal:  J Clin Neurosci       Date:  2011-05-14       Impact factor: 1.961

7.  Assessment of angiographic outcomes after flow diversion treatment of intracranial aneurysms: a new grading schema.

Authors:  Mudassar Kamran; Julia Yarnold; Iris Q Grunwald; James V Byrne
Journal:  Neuroradiology       Date:  2010-09-14       Impact factor: 2.804

8.  Flow-diverter stent for the endovascular treatment of intracranial aneurysms: a prospective study in 29 patients with 34 aneurysms.

Authors:  Boris Lubicz; Laurent Collignon; Gaï Raphaeli; Jean-Pierre Pruvo; Michaël Bruneau; Olivier De Witte; Xavier Leclerc
Journal:  Stroke       Date:  2010-08-26       Impact factor: 7.914

9.  Angiographic assessment of the performance of flow divertors to treat cerebral aneurysms.

Authors:  Chander Sadasivan; Baruch B Lieber; Liliana Cesar; Laszlo Miskolczi; Jaehoon Seong; Ajay K Wakhloo
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2006

10.  Angiographic CT with intravenous administration of contrast medium is a noninvasive option for follow-up after intracranial stenting.

Authors:  Jan-Hendrik Buhk; Paul Lingor; Michael Knauth
Journal:  Neuroradiology       Date:  2008-02-02       Impact factor: 2.804

View more
  7 in total

1.  Wall Apposition Is a Key Factor for Aneurysm Occlusion after Flow Diversion: A Histologic Evaluation in 41 Rabbits.

Authors:  A Rouchaud; C Ramana; W Brinjikji; Y-H Ding; D Dai; T Gunderson; J Cebral; D F Kallmes; R Kadirvel
Journal:  AJNR Am J Neuroradiol       Date:  2016-07-07       Impact factor: 3.825

2.  Angiographic CT: in vitro comparison of different carotid artery stents-does stent orientation matter?

Authors:  Michael Lettau; Martin Bendszus; Stefan Hähnel
Journal:  Neuroradiology       Date:  2013-02-15       Impact factor: 2.804

3.  Impact of a new metal artefact reduction algorithm in the noninvasive follow-up of intracranial clips, coils, and stents with flat-panel angiographic CTA: initial results.

Authors:  Marios-Nikos Psychogios; Bernhard Scholz; Christopher Rohkohl; Yiannis Kyriakou; Alexander Mohr; Peter Schramm; Dorothee Wachter; Katrin Wasser; Michael Knauth
Journal:  Neuroradiology       Date:  2013-04-16       Impact factor: 2.804

Review 4.  Flat-Panel Computed Tomography (DYNA-CT) in Neuroradiology. From High-Resolution Imaging of Implants to One-Stop-Shopping for Acute Stroke.

Authors:  A Doerfler; P Gölitz; T Engelhorn; S Kloska; T Struffert
Journal:  Clin Neuroradiol       Date:  2015-06-20       Impact factor: 3.649

5.  Angiographic C-arm CT visualization of the Woven EndoBridge cerebral aneurysm embolization device (WEB): first experience in an animal aneurysm model.

Authors:  T Struffert; S Lang; E Adamek; T Engelhorn; C M Strother; A Doerfler
Journal:  Clin Neuroradiol       Date:  2013-05-28       Impact factor: 3.649

6.  Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia.

Authors:  Thomas F Flood; Imramsjah M J van der Bom; Lara Strittmatter; Ajit S Puri; Gregory M Hendricks; Ajay K Wakhloo; Matthew J Gounis
Journal:  J Neurointerv Surg       Date:  2014-01-30       Impact factor: 5.836

7.  Metal artifact reduction for flat panel detector intravenous CT angiography in patients with intracranial metallic implants after endovascular and surgical treatment.

Authors:  Rastislav Pjontek; Belgin Önenköprülü; Bernhard Scholz; Yiannis Kyriakou; Gerrit A Schubert; Omid Nikoubashman; Ahmed Othman; Martin Wiesmann; Marc A Brockmann
Journal:  J Neurointerv Surg       Date:  2015-09-07       Impact factor: 5.836

  7 in total