Long-term wear of HIPed alumina on alumina bearings for THR under microseparation conditions.

T Stewart1, J Tipper, R Streicher, E Ingham, J Fisher.   

Abstract

The long term wear and wear debris generated in HIPed alumina on alumina bearings for hip prostheses with microseparation in vitro is compared to standard simulator conditions and ex vivo specimens. Microseparation studies were completed to five million cycles at two severity levels in attempts to rigorously evaluate the long-term tribological performance of the bearings. During the first million cycles (bedding-in) of the microseparation tests characteristic stripe wear was observed on all of the femoral heads with a matching area on the rim of the acetabular inserts. Under mild microseparation conditions an average wear rate of 0.55 mm3/million cycles was observed during the initial million cycles which reduced to a steady state level of 0.1 mm3/million cycles. Under more severe conditions an average wear rate of 4.0 mm3/million cycles was observed during bedding-in which reduced to a steady state level of 1.3 mm3/million cycles. These compare to a bedding-in wear rate of 0.11 mm3/million cycles and steady-state wear rate of 0.05 mm3/million cycles for the same material under normal simulation with no microseparation. Furthermore, under microseparation the wear mechanisms and wear debris were similar to those observed in previous alumina retrieval studies with debris ranging from 10 nm to 1 microm in size. Copyright 2001 Kluwer Academic Publishers

Entities:  

Year:  2001        PMID: 15348364     DOI: 10.1023/a:1012802308636

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  4 in total

1.  An in vivo determination of total hip arthroplasty pistoning during activity.

Authors:  A V Lombardi; T H Mallory; D A Dennis; R D Komistek; R A Fada; E J Northcut
Journal:  J Arthroplasty       Date:  2000-09       Impact factor: 4.757

2.  Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns.

Authors:  J Nevelos; E Ingham; C Doyle; R Streicher; A Nevelos; W Walter; J Fisher
Journal:  J Arthroplasty       Date:  2000-09       Impact factor: 4.757

3.  The influence of acetabular cup angle on the wear of "BIOLOX Forte" alumina ceramic bearing couples in a hip joint simulator.

Authors:  J E Nevelos; E Ingham; C Doyle; A B Nevelos; J Fisher
Journal:  J Mater Sci Mater Med       Date:  2001-02       Impact factor: 3.896

4.  Analysis of retrieved alumina ceramic components from Mittelmeier total hip prostheses.

Authors:  J E Nevelos; E Ingham; C Doyle; J Fisher; A B Nevelos
Journal:  Biomaterials       Date:  1999-10       Impact factor: 12.479

  4 in total
  21 in total

1.  Assessment of wear on the cones of modular stainless steel Exeter hip stems.

Authors:  R P S Chaplin; A J C Lee; R M Hooper
Journal:  J Mater Sci Mater Med       Date:  2004-09       Impact factor: 3.896

2.  Wear degradation of long-term in vivo exposed alumina-on-alumina hip joints: linking nanometer-scale phenomena to macroscopic joint design.

Authors:  Yasuhito Takahashi; Nobuhiko Sugano; Wenliang Zhu; Takashi Nishii; Takashi Sakai; Masaki Takao; Giuseppe Pezzotti
Journal:  J Mater Sci Mater Med       Date:  2011-11-22       Impact factor: 3.896

3.  Wear patterns of, and wear volume formulae for, cylindrically elongated acetabular cup liners.

Authors:  James Shih-Shyn Wu; Shu-Ling Hsu; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2010-04-28       Impact factor: 2.602

Review 4.  Edge loading has a paradoxical effect on wear in metal-on-polyethylene total hip arthroplasties.

Authors:  William H Harris
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

5.  High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings.

Authors:  J C Keurentjes; R M Kuipers; D J Wever; B W Schreurs
Journal:  Clin Orthop Relat Res       Date:  2008-02-26       Impact factor: 4.176

6.  Delta ceramic-on-alumina ceramic articulation in primary THA: prospective, randomized FDA-IDE study and retrieval analysis.

Authors:  Adolph V Lombardi; Keith R Berend; Brian E Seng; Ian C Clarke; Joanne B Adams
Journal:  Clin Orthop Relat Res       Date:  2010-02       Impact factor: 4.176

7.  Evaluating the accuracy of wear formulae for acetabular cup liners.

Authors:  James Shih-Shyn Wu; Shu-Ling Hsu; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2009-10-10       Impact factor: 2.602

8.  Using nonlinear finite element models to analyse stress distribution during subluxation and torque required for dislocation of newly developed total hip structure after prosthetic impingement.

Authors:  Wei-Min Chi; Chien-Chung Lin; Ying-Jui Ho; Hsiao-Che Lin; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2017-07-01       Impact factor: 2.602

9.  Squeaking and microcracks in a delta-delta ceramic coupling: pin-on-disc study.

Authors:  Kiyokazu Fukui; Ayumi Kaneuji; Tadami Matsumoto; Kazuhiro Shintani
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-03-12

Review 10.  Review on squeaking hips.

Authors:  Yadin David Levy; Selin Munir; Shane Donohoo; William Lindsay Walter
Journal:  World J Orthop       Date:  2015-11-18
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.