Impact of number of channels on RF shimming at 3T.

Alexander S Childs1, Shaihan J Malik, Declan P O'Regan, Joseph V Hajnal.   

Abstract

OBJECT: At high-field strengths (≥3T) inhomogeneity of the radio frequency (RF) field and RF power deposition become increasingly problematic. Parallel Transmission (PTx)--the use of segmented transmission arrays with independently driven elements--affords the ability to combat both of these issues. There are a variety of existing designs for PTx coils, ranging from systems with two channels to systems with eight or more. In this work, we have investigated the impact of the number of independent channels on the achievable results for both homogeneity improvement and power reduction in vivo.
MATERIALS AND METHODS: A 3T Philips Achieva MRI system fitted with an 8-channel PTx body coil was driven so as to emulate configurations with 1, 2 4 and 8 independent channels. RF shimming was used in two different anatomies in order to assess improvements in RF homogeneity.
RESULTS: Significant homogeneity improvements were observed when increasing from 1 to 2, 2 to 4, and 4 to 8 channel configurations. Reductions in RF power requirements and local SAR were predicted for increasing numbers of channels.
CONCLUSION: Increasing the number of RF transmit channels adds extra degrees of freedom which can be used to benefit homogeneity improvement or power reduction for body imaging at 3T.

Entities:  

Mesh:

Year:  2013        PMID: 23315236      PMCID: PMC3728430          DOI: 10.1007/s10334-012-0360-5

Source DB:  PubMed          Journal:  MAGMA        ISSN: 0968-5243            Impact factor:   2.310


  21 in total

1.  Effect of RF coil excitation on field inhomogeneity at ultra high fields: a field optimized TEM resonator.

Authors:  T S Ibrahim; R Lee; B A Baertlein; A M Abduljalil; H Zhu; P M Robitaille
Journal:  Magn Reson Imaging       Date:  2001-12       Impact factor: 2.546

2.  A specific absorption rate prediction concept for parallel transmission MR.

Authors:  Ingmar Graesslin; Hanno Homann; Sven Biederer; Peter Börnert; Kay Nehrke; Peter Vernickel; Giel Mens; Paul Harvey; Ulrich Katscher
Journal:  Magn Reson Med       Date:  2012-01-09       Impact factor: 4.668

3.  DREAM--a novel approach for robust, ultrafast, multislice B₁ mapping.

Authors:  Kay Nehrke; Peter Börnert
Journal:  Magn Reson Med       Date:  2012-01-17       Impact factor: 4.668

4.  Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field.

Authors:  Vasily L Yarnykh
Journal:  Magn Reson Med       Date:  2007-01       Impact factor: 4.668

5.  Eight-channel transmit/receive body MRI coil at 3T.

Authors:  P Vernickel; P Röschmann; C Findeklee; K-M Lüdeke; Ch Leussler; J Overweg; U Katscher; I Grässlin; K Schünemann
Journal:  Magn Reson Med       Date:  2007-08       Impact factor: 4.668

6.  Standing-wave and RF penetration artifacts caused by elliptic geometry: an electrodynamic analysis of MRI.

Authors:  J G Sled; G B Pike
Journal:  IEEE Trans Med Imaging       Date:  1998-08       Impact factor: 10.048

7.  Toward individualized SAR models and in vivo validation.

Authors:  H Homann; P Börnert; H Eggers; K Nehrke; O Dössel; I Graesslin
Journal:  Magn Reson Med       Date:  2011-05-31       Impact factor: 4.668

8.  Eigenmode analysis of transmit coil array for tailored B1 mapping.

Authors:  Kay Nehrke; Peter Börnert
Journal:  Magn Reson Med       Date:  2010-03       Impact factor: 4.668

9.  Patient-individual local SAR determination: in vivo measurements and numerical validation.

Authors:  T Voigt; H Homann; U Katscher; O Doessel
Journal:  Magn Reson Med       Date:  2011-12-28       Impact factor: 4.668

10.  Slice-selective RF pulses for in vivo B1+ inhomogeneity mitigation at 7 tesla using parallel RF excitation with a 16-element coil.

Authors:  Kawin Setsompop; Vijayanand Alagappan; Borjan Gagoski; Thomas Witzel; Jonathan Polimeni; Andreas Potthast; Franz Hebrank; Ulrich Fontius; Franz Schmitt; Lawrence L Wald; Elfar Adalsteinsson
Journal:  Magn Reson Med       Date:  2008-12       Impact factor: 4.668

View more
  18 in total

1.  Comparison of simulated parallel transmit body arrays at 3 T using excitation uniformity, global SAR, local SAR, and power efficiency metrics.

Authors:  Bastien Guérin; Matthias Gebhardt; Peter Serano; Elfar Adalsteinsson; Michael Hamm; Josef Pfeuffer; Juergen Nistler; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2014-04-18       Impact factor: 4.668

Review 2.  Image-quality optimization and artifact reduction in fetal magnetic resonance imaging.

Authors:  Fedel Machado-Rivas; Camilo Jaimes; John E Kirsch; Michael S Gee
Journal:  Pediatr Radiol       Date:  2020-11-30

Review 3.  Fetal MRI at 3T-ready for routine use?

Authors:  Christian Weisstanner; Gerlinde M Gruber; Peter C Brugger; Christan Mitter; Mariana C Diogo; Gregor Kasprian; Daniela Prayer
Journal:  Br J Radiol       Date:  2016-10-21       Impact factor: 3.039

4.  Effect of parallel radiofrequency transmission on arterial input function selection in dynamic contrast-enhanced 3 Tesla pelvic MRI.

Authors:  Hatim Chafi; Saba N Elias; Huyen T Nguyen; Harry T Friel; Michael V Knopp; BeiBei Guo; Steven B Heymsfield; Guang Jia
Journal:  J Magn Reson Imaging       Date:  2015-06-09       Impact factor: 4.813

5.  An in vivo comparison of the DREAM sequence with current RF shim technology.

Authors:  Kay Nehrke; Alois M Sprinkart; Peter Börnert
Journal:  MAGMA       Date:  2014-08-10       Impact factor: 2.310

Review 6.  3 T: the good, the bad and the ugly.

Authors:  Martin John Graves
Journal:  Br J Radiol       Date:  2021-11-10       Impact factor: 3.039

7.  General design approach and practical realization of decoupling matrices for parallel transmission coils.

Authors:  Zohaib Mahmood; Patrick McDaniel; Bastien Guérin; Boris Keil; Markus Vester; Elfar Adalsteinsson; Lawrence L Wald; Luca Daniel
Journal:  Magn Reson Med       Date:  2015-07-31       Impact factor: 4.668

8.  Comparison of RF body coils for MRI at 3  T: a simulation study using parallel transmission on various anatomical targets.

Authors:  Xiaoping Wu; Xiaotong Zhang; Jinfeng Tian; Sebastian Schmitter; Brian Hanna; John Strupp; Josef Pfeuffer; Michael Hamm; Dingxin Wang; Juergen Nistler; Bin He; Thomas J Vaughan; Kamil Ugurbil; Pierre-Francois Van de Moortele
Journal:  NMR Biomed       Date:  2015-08-30       Impact factor: 4.044

9.  Hip imaging of avascular necrosis at 7 Tesla compared with 3 Tesla.

Authors:  J M Theysohn; O Kraff; N Theysohn; S Orzada; S Landgraeber; M E Ladd; T C Lauenstein
Journal:  Skeletal Radiol       Date:  2014-02-05       Impact factor: 2.199

10.  Impact of 3.0 T Cardiac MR Imaging Using Dual-Source Parallel Radiofrequency Transmission with Patient-Adaptive B1 Shimming.

Authors:  Haipeng Jia; Cuiyan Wang; Guangbin Wang; Lei Qu; Weibo Chen; Queenie Chan; Bin Zhao
Journal:  PLoS One       Date:  2013-06-18       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.